LIQUICHEK PEDIATRIC CONTROL, LEVEL 1; LIQUICHEK PEDIATRIC CONTOL, LEVEL 2; LIQUICHEK PEDIATRIC CONTROL, BILEVEL MINIPAK

K130963 · Bio-Rad Laboratories, Inc. · JJY · May 6, 2013 · Clinical Chemistry

Device Facts

Record IDK130963
Device NameLIQUICHEK PEDIATRIC CONTROL, LEVEL 1; LIQUICHEK PEDIATRIC CONTOL, LEVEL 2; LIQUICHEK PEDIATRIC CONTROL, BILEVEL MINIPAK
ApplicantBio-Rad Laboratories, Inc.
Product CodeJJY · Clinical Chemistry
Decision DateMay 6, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1
AttributesPediatric

Indications for Use

Liquichek Pediatric Control is intended for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the analytes listed in this package insert.

Device Story

Liquichek Pediatric Control is a bovine serum-based quality control material containing added chemicals, therapeutic drugs, preservatives, and stabilizers. It is used in clinical laboratories to monitor the precision of testing procedures for analytes including bilirubin (direct/total), caffeine, calcium, chloride, glucose, magnesium, potassium, sodium, and theophylline. The control is provided in liquid form. Laboratory personnel use the product to verify the performance of testing instrumentation and reagents by comparing results against provided mean values and ±3SD ranges. It serves as a guide for laboratories to establish their own acceptable performance ranges, helping ensure the accuracy and reliability of patient test results.

Clinical Evidence

No clinical data. Bench testing only. Stability studies (real-time and accelerated) were conducted to establish shelf-life, thawed unopened stability (45 days at 2-8°C), and thawed opened stability (14 days at 2-8°C).

Technological Characteristics

Bovine serum matrix with added chemicals, therapeutic drugs, preservatives, and stabilizers. Liquid form. Storage at -20 to -50°C. Analyte list includes bilirubin, caffeine, calcium, chloride, glucose, magnesium, potassium, sodium, and theophylline.

Indications for Use

Indicated for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for specific analytes in a clinical laboratory setting.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION ## TRIAGE-QUICK REVIEW DECISION SUMMARY 510(k) #: k130963 This 510(k) was reviewed under OIR’s Triage-Quick Review Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA. The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant’s 510(k) summary for a summary of the information that supports this SE determination.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...