K130940 · Kwang Yang Motor Co., Ltd. · INI · Oct 7, 2013 · Physical Medicine
Device Facts
Record ID
K130940
Device Name
KYMCO EQ20C SCOOTER
Applicant
Kwang Yang Motor Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Oct 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
The EQ 20C is a motorized four-wheeled scooter for indoor/outdoor mobility. It features an un-foldable platform, adjustable tiller, two lead-acid batteries, motor/electromagnetic brake assembly, and seat/backrest. The user operates the device via a tiller handlebar and a thumb-operated potentiometer throttle control lever to manage forward/reverse motion. Releasing the throttle automatically engages the electromagnetic brake to slow the scooter to a stop. The device is intended for use by disabled or elderly individuals who require a seated transportation solution. It provides mobility benefits by allowing independent navigation in various environments.
Clinical Evidence
Bench testing only. The device complied with ANSI/RESNA WC Vol. 1 and 2, ISO 7176 series (stability, energy consumption, braking, etc.), ISO 10993 (biocompatibility), ISO 14971 (risk management), and electromagnetic compatibility standards (EN 61000 series, CISPR 11).
Technological Characteristics
Motorized four-wheeled scooter; PU foam/nylon fabric seat; lead-acid batteries; electromagnetic brake; PG S-Drive 45A controller; 250W motor. Complies with ANSI/RESNA WC, ISO 7176, ISO 10993, and EN 61000 standards. Standalone operation.
Indications for Use
Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility assistance.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
K1BO940
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Applicant
Name: KWANG YANG MOTOR CO., LTD.
Address: No. 35, Wan Hsing Street, San Min Dist., 803, Kaohsiung, Taiwan
Registration number: 3003851898
Contact person: Yen, Wen Hsi
Phone: +886-7-3822526
Fax: +886-7-3825834
e-mail: yen@mail.kymco.com
#### Date prepared: Aug. 15, 2013
#### Device
Trade name: KYMCO
Model: EQ 20C scooter
Common name: Electrical scooter
Classification name: Motorized three-wheeled vehicle
Medical specialty (Panel): Physical Medicine Device
Regulation number: 890.3800
Product Code: INI
Classification: Class II
#### Predicate devices
Trade name: KYMCO
Model: EQ 30
510(k) number: K072630
Manufacture: KWANG YANG MOTOR CO., LTD.
Regulation number: 890.3800
Product Code: INI
Classification: Class II
#### Intend use of device
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
#### Device description:
The EQ 20C scooter consists of an un-foldable platform which connects the two front
**OCT 07 2013**
{1}------------------------------------------------
wheels and two rear wheels, an adjustable tiller, two lead acid batteries with an on-board charger, a motor/electromagnetic brake assembly, a electric motor controller and a seat /backrest set.
The patient uses the tiller handle/handlebar for steering and a thumb operated potentiometer throttle control lever located at the top of the tiller to engage and disengage the scooter motion in both the forward and reverse directions. When the throttle control lever is released, the electromagnetic brake will be actuated and the scooter is slow to stop.
#### Summary of non-clinical testing
The EQ 20C scooter complied with the requirements of ANSI/RESNA WC Vol. 1 Sec. 7 and Sec 8, ANSI/RESNA WC Vol. 2 Sec. 21, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 7176-16, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 14971, EN 61000-4-2, EN 61000-4-3, EN 61000-4-8, and CISPR 11.
#### Statement of substantial equivalence
The EQ 20C scooter is substantially equivalent to the EQ 30 (K072630). They have same intended use of a motor driven, indoor and outdoor transportation vehicle to provide mobility to a disabled or elderly person limited to a seated position.
The design and technological characteristics of this device is basically similar to the predicate device chosen. Both are un-foldable scooters and have the same user interface. While there are minor differences between the devices, do not alter the intended use function and use of the device. Moreover, the non-clinical tests and the predicate comparisons demonstrate that these differences in their technological characteristics do not raise any questions as to the safety and effectiveness.
#### Conclusion
In accordance with the Federal Food, Drug and Cosmelic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, KWANG YANG MOTOR CO., LTD. concludes that, EQ 20C scooter is substantially equivalent to predicate devices as described herein.
{2}------------------------------------------------
| | The substantial equivalence comparison of the EQ 20C and EQ 30 (K072630) | |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| | EQ 20C | EQ30 (K072630) |
| ended use | They are motor driven, indoor and outdoor transportation vehicle with the intended use to provide<br>mobility to a disabled or elderly person limited to a seated position. | |
| Maximum load capacity | 127 kg (280 lbs) | 150 kg (330 lbs) |
| Overall dimension | | |
| Height | 921 mm (36.3") | 1250 mm (49.2") |
| Length | 1080 mm (43.5") | 1240 mm (48.8") |
| Width | 520 mm (20.5") | 560 mm (22") |
| Seat Overall dimension | | |
| Height | 550 mm (21.7") (from ground) | 756 mm (29.8") |
| Width | 427 mm (16.8") | 455 mm (17.9") |
| Depth | 350 mm (13.8") | 400 mm (15.7") |
| Seat / Seat cushion / Backrest | | |
| Material | PU foam covered by nylon fabric cloth | |
| Motor | | |
| Rated power | 250 W | 350 W |
| Input voltage | | DC 24V |
| Amount | | 1 Pc |
| Brake system | | Intelligent regenerative electromagnetic brake and hand brake |
| Controller | PG S-Drive, 45A | KYSA 70 |
| Differential mechanism | | Differential rate: 19.7 : 1 |
| Rear wheel drive | | Sealed transaxle direct drive |
| (Continuous) | The substantial equivalence comparison of the EQ 20C and EQ 30 (K072630) | |
| | EQ 20C | EQ30 (K072630) |
| Free-wheel mode | | Yes |
| Braking distance | Forward:1.0 m (39.4") at max speed | Forward: 1.24 m (48.8") at max speed |
| Net weight with battery | 55.5 kg (122 lbs) | 84.4 kg(185.7 lbs) |
| Slope grade ability | 8 degree | 10 degree |
| Per-charge distance | Up to 21 km (13 miles) | Up to 37 km (23.1 miles) |
| Maximum speed | Up to 6.5 km/h (4.1 mph) , variable | Up to 8 km/hr (4.7 mph), variable |
| Speed mode | | Single mode, Variable |
| Turning radius | 1200 mm (47.2") | 1565 mm (61.6") |
| Maximum curb height | 50 mm (2") | 80 mm (3.14") |
| Suspension | | Front: No, Rear: Yes |
| Head/Tail lights<br>Signal light | | Head/tail lights and signal lights of the EQ 20C, and EQ 30 scooters have the same functions, but<br>only the styles are different. |
| Warning light | | Yes |
| Horn | | Yes |
| Anti-tip wheels | | Yes |
| Operation mode | | Thumb operated potentiometer throttle control lever |
| Armrest | | Yes, Foldable |
| Tiller foldable | | Yes |
| Headrest | No | Yes |
| Height adjustable | No | Yes |
| | | |
| Continuous) | The substantial equivalence comparison of the EQ 20C and EQ 30 (K072630) | |
| | EQ 20C | EQ30 (K072630) |
| Battery | | |
| Type | | Lead-acid |
| Rated power | 22 Ah | 31 Ah |
| Output voltage | | 24 VDC |
| Amount | | DC 12V x 2 Pcs |
| Battery level indicator | | Yes |
| Charger | | |
| Type | | on-board, Automatic Type |
| Input power requirement | | 115-230 VAC |
| Output voltage / current | | DC 24V/5A |
| Front wheels | | |
| Diameter | 70 mm/62.5 mm - 100 mm(2.8"/2.5"-4") | 250 mm x 125 mm(10" x 5") |
| type | | non-pneumatic tires |
| Material | | Polyurethane (PU) |
| Amount | | 2 |
| Rear wheels | | |
| Diameter | 70 mm/62.5 mm - 100 mm(2.8"/2.5"-4") | 250 mm x 125 mm(10" x 5") |
| type | | non-pneumatic tires |
| Material | | Polyurethane (PU) |
| Amount | | 2 |
:
,
{3}------------------------------------------------
{4}------------------------------------------------
1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -
. . . . .
1
-
:
•
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WQ66-G609 Silver Spring, MD 20993-0002
October 7, 2013
Kwang Yang Motor Co., Ltd. c/o Mrs. Junnata Chang 16F-2(16A), No. 462, Sec. 2. ChongDe Rd. Beitun Dist. Taichung CHINA (Taiwan) 406
Re: K 130940
Trade/Device Name: Kymco EO 20C scooter Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: Class II Product Code: INI Dated: August 15, 2013 Received: August 22, 2013
Dear Junnata Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device arrated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Junnata Chang
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Victor Krauthamer -S
Victor Krauthamer, Ph.D. Acting Director Division of Neurological. and Physical Medicine Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use
510(k) Number (if known): K130940
Device Name: KYMCO EQ 20C SCOOTER
Indications For Use:
The device is a motor drive, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disable or elderly person limited to a seated position.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use ___ × (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Victor Krauthamer -S 2013.10.04 19:07:38 -04'00'
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.