1417WCA Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
Device Story
1417WCA is a wired/wireless digital X-ray flat panel detector; captures radiographic images of human anatomy. Device consists of a Cesium Iodide scintillator coupled to an a-Si TFT sensor. Integrates with existing radiographic X-ray generators and console PC systems. Captures X-ray signals, converts to RAW files, and transmits via IEEE 802.11a/g/n wireless or wired connection to console software. Console software converts RAW files to DICOM 3.0 format for viewing, diagnosis, and analysis by clinicians. Used in clinical settings (upright stands, tables, universal stands) by trained operators. Provides high-speed digital imaging, replacing traditional film/screen systems to improve workflow and diagnostic efficiency.
Clinical Evidence
No clinical trials; clinical evidence based on comparative review of clinical images. Images from 1417WCA and 1417PCA were evaluated by a licensed US radiologist to confirm diagnostic equivalence across similar age groups and anatomical structures. Non-clinical bench testing included MTF, DQE, and NPS analysis per IEC 62220-1, showing identical performance to the predicate. Safety and EMC testing conducted per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Amorphous Silicon (a-Si) TFT detector with Cesium Iodide scintillator. 14 x 17 inch imaging area; 127 µm pixel pitch; 3.9 lp/mm resolution; 14-bit A/D conversion. Connectivity: Wired/Wireless (IEEE 802.11a/g/n). Dimensions: 460 x 417 x 15.9 mm. Weight: 3.9 kg. Standards: IEC 60601-1, IEC 60601-1-2, FCC Part 15.
Indications for Use
Indicated for digital imaging in general radiographic systems for human anatomy. Intended to replace film/screen-based systems for general diagnostic procedures. Contraindicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K130935
Page 1 of 5
Rev. 3
### 510(k) Summary
# JUL 0 2 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date 510K summary prepared: March 26, 2013
Submitter's Name, address, telephone number, a contact person:
| Submitter's Name : | Rayence Co., Ltd. |
|-------------------------|-------------------------------------------|
| Submitter's Address: | 1F, 2F, 3F, #402, 14, Samsung 1-ro 1-gil, |
| | Hwaseong-si, Gyeonggi-do, Korea |
| Submitter's Telephone: | +82-31-8015-6459 |
| Contact person: | Mr. Kee Dock Kim / Manager |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net) |
| (U.S. Designated agent) | |
| Address: | 12946 Kimberley Ln, Houston, TX 77079 |
| Telephone: | +713-467-2607 |
| Fax: | +713-464-8880 |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
| Trade/proprietary name: | 1417WCA |
|-------------------------|-----------------------------------------|
| Common Name: | Digital Flat Panel X-ray Detector |
| Classification Name : | 21CFR892.1680 / Stationary x-ray system |
| Product Code: | MQB |
1
{1}------------------------------------------------
#### Predicate Device :
| Manufacturer | : Rayence Co., Ltd. |
|---------------|-------------------------------------------|
| Device | : 1417PCA |
| 510(k) Number | : K122919 (Decision Date - JAN. 31. 2013) |
#### Device Description :
1417WCA is a wired/wireless digital X-ray flat panel detector that can acquire radiographic images of human anatomy when used with existing radiographic x-ray systems. The wireless LAN((IEEE 802.11a/g/n) communication signals images captured to the system and improves the user operability through high-speed processing. This X-ray imaging detector consists of a scintillator directly coupled to an a-Si TFT sensor. 1417WCA is designed specifically to be integrated with a console PC system and X-Ray generator to digitalize x-ray images into RAW files. The RAW files can be made to DICOM compatible image files which can be viewed by console SW for a radiographic image diagnosis and analysis.
#### Indication for use :
1417WCA Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
#### Summary of the technological characteristics of the device compared to the predicate device:
The 1417WCA SSXI detector described in this 510(k) has the same indications for use and similar technical characteristics as its predicate device, 1417PCA flat panel detector, of Rayence Co., Ltd. Table 1 summarizes the technological characteristics of the 1417WCA and 1417PCA, the predicate device.
{2}------------------------------------------------
| Characteristic | Proposed<br>Rayence Co.,Ltd.<br>1417WCA | Predicate<br>Rayence Co.,Ltd.<br>1417PCA |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | - | -K122919 |
| Intended Use | 1417WCA Digital Flat Panel X-Ray Detector is indicated for<br>digital imaging solution designed<br>for general radiographic system<br>for human anatomy. It is intended<br>to replace film or screen based<br>radiographic systems in all<br>general purpose diagnostic<br>procedures. Not to be used for<br>mammography. | 1417PCA Digital Flat Panel X-Ray Detector is indicated for<br>digital imaging solution designed<br>for general radiographic system<br>for human anatomy. It is intended<br>to replace film or screen based<br>radiographic systems in all<br>general purpose diagnostic<br>procedures. Not to be used for<br>mammography. |
| Detector Type | Amorphous Silicon, TFT | Amorphous Silicon, TFT |
| Scintillator | Cesium Iodide | Cesium Iodide |
| Imaging Area | 14 x 17 inches | 14 x 17 inches |
| Total Pixel<br>Number | 3328 x 2816 pixels | 3328 x 2816 pixels |
| Pixel pitch | 127 µm | 127 µm |
| Resolution | 3.9 lp/mm | 3.9lp/mm |
| A/D conversion | 14 bit | 14 bit |
| Preview Image | 2~3 seconds (wired)<br>/ 3~5 seconds (wireless) | 2~3 seconds |
| Data output | RAW<br>*The RAW files are<br>convertible into DICOM 3.0<br>by console S/W | RAW<br>*The RAW files are<br>convertible into DICOM 3.0<br>by console S/W |
| Dimensions | 460 x 417 x 15.9 mm | 460 x 417 x 15.9 mm |
Table 1: Comparison of 1417WCA and 1417PCA
3
:
{3}------------------------------------------------
| Weight | 3.9 kg (incl. battery pack) | 3.6 kg |
|-------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Application | Wireless portable system<br>Available with upright stand,<br>table, universal stand | Portable system<br>Available with upright stand,<br>table, universal stand |
| Feature | Image: X-ray detector with handle on top | Image: X-ray detector with handle on top and crosshair in the center |
Table 2: Size Comparison of 1417WCA and 1417PCA
| Item | Unit | 1417WCA | 1417PCA |
|--------------------------------------------|--------|-------------|-------------|
| Pixel Pitch | μm | 127 x 127 | 127 x 127 |
| Total Pixel Number | pixels | 3328 x 2816 | 3328 x 2816 |
| Effective Pixel Area | mm | 415 x 350 | 415 x 350 |
| Effective Pixel Number(light<br>sensitive) | pixels | 3268x 2756 | 3268x 2756 |
| Fill factor | % | 61.03 | 61.03 |
| Weight | Kg | 3.9 Kg | 3.6 Kg |
#### Summary of Performance Testing:
The wireless/wired 1417WCA flat panel detector is a modified version of 1417PCA (K122919), FDA cleared predicate device from Rayence. Indications for use, material, form factor, performance, and safety characteristics between 1417WCA and 1417PCA are the same. The non-clinical test report and clinical consideration report were prepared and submitted to FDA separately to demonstrate the substantial equivalency between two different detectors. The non-clinical test report contains the MTF, DQE and NPS test results of 1417WCA and 1417PCA by using the identical test equipment and same analysis method described by IEC 62220-1 The comparisoin of the MTF for 1417WCA and 1417PCA detector demonstated that the MTF of the 1417WCA detector performed almost same with 1417PCA. Therefore the overall resolution performance and shapness of 1417WCA is almost same with 1417PCA. The DQE represents the ability to visualize object details of a certain size and contrast. 1417WCA demonstrated almost same DQE
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows handwritten text on a white background. The text includes the identifier "K130935" at the top, followed by "Page 5 of 5" in a cursive style. Below the page number, the text "Rev.3" is present, indicating a revision number.
performance with 1417PCA at various spatial frequencies and provides almost same Signal-toNoise Ratio (SNR) transfer from the input to the output of a detector as a function of frequency. At the lowest spatial frequency, the DQE test for 1417WCA and 1417 PCA resulted 74%, respectively. The NPS test for 1417WCA and 1417 PCA exhibited almost identical performance between the two devices. Therefore, the image quality of 1417WCA is substantially equivalent to 1417PCA at the same patient exposure setting.
To further demonstrate the substantial equivalency of two devices, clinical images are obtained from both devcies and reviewd by a licensed US radiologist to render an expert opinion. Both the test (1417WCA) and control group (1417PCA) are evaluated according to similar age group and anatomical structures were compared in accordance with the test protocol of diagnostic radiography evaluation procedure.
Based on the non-clinical and clinical consideration test and the outcome of a comparative review by an expert for both devices, we can claim the substantial equivalency between 1417WCA and its predicate device, 1417PCA in terms of image quality.
#### Safety, EMC and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1:1988 + A1:1991 + A2:1995 (Medical electrical equipment Part 1: General Requirements for Safety) was performed, and EMC testing were conducted in accordance with standard IEC60601-1-2:2007 (CISPR 11:2009/A1: 2010), EN60601-1-2:2007 +A1:2010 (Medical electrical equipment - Part 1-2: General Requirements for safety - Collateral Standard : Electromagnetic Compatibility Requirements and tests). The equipment also complies with the standard FCC Rule part(s) 47CFR PART 15.107(B) / 47CFR PART 15.109(G) CLASSB. All test results were satisfactory.
#### Conclusions :
In accordance with the Federal Food. Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Rayence Co., Ltd, concludes that 1417WCA is safe and effective and substantially equivalent in comparison with 1417PCA, the predicate device as described herein.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a partial view of a document or logo. On the left, there's a stylized graphic with curved lines, possibly part of a circular design. To the right, the word "DEPARTMENT" is printed in a clear, sans-serif font. A horizontal line is present beneath the word, suggesting it might be a heading or part of a formal document.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20003-0002
July 2. 2013
Ravence Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 12946 Kimberley Lane HOUSTON TX 77079
Re: K130935
Trade/Device Name: Digital Flat Panel X-ray Detector / 1417WCA Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: April 04, 2013 Received: April 04, 2013
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
#### Page 2 - Mr. Kim
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Janning M. Menge
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if known): K130935
Device Name: Digital Flat Panel X-ray Detector / 1417 WCA
Indications for Use:
1417WCA Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Stamford M. Weiss
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) K 130935
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.