EASY DAY (METHAFILCON A) 1-DAY SOFT (HYDROPHILIC) CONTACT LENS
K130931 · Jiangsu Horien Contact Lens Co., Ltd. · LPL · Aug 30, 2013 · Ophthalmic
Device Facts
Record ID
K130931
Device Name
EASY DAY (METHAFILCON A) 1-DAY SOFT (HYDROPHILIC) CONTACT LENS
Applicant
Jiangsu Horien Contact Lens Co., Ltd.
Product Code
LPL · Ophthalmic
Decision Date
Aug 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
EASY DAY (methafilcon A) 1-Day Soft ( Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia) in not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 0.75 diopters that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from 0 ~ - 12.00D. The lens is indicated for single use daily disposable wear and patients are instructed to discard the lens after each removal.
Device Story
EASY DAY is a single-vision aspherical soft contact lens made of methafilcon A (HEMA/MAA copolymer). It is tinted blue for handling visibility and supplied sterile in buffered saline. Designed for daily wear to correct myopia in non-aphakic patients, the lens is discarded after each daily use. The device is prescribed by eye care professionals. It functions by resting on the cornea to refract light, compensating for refractive ametropia. Benefits include vision correction and convenience of a daily disposable modality.
Clinical Evidence
Clinical study involved at least 30 evaluable subjects' eyes over 1 month. Results indicated nearly 100% vision correction and high patient satisfaction with lens wear and care. No adverse safety issues reported.
Technological Characteristics
Material: methafilcon A (HEMA and MAA copolymer cross-linked with EGDMA). Diameter: 14.20 ±0.2mm. Power: 0 to -12.00 D. Water content: 53-57%. Oxygen permeability: 19.7 Dk units. Refractive index: 1.409. Light transmittance: >92%. Sterilization: Steam.
Indications for Use
Indicated for daily wear correction of myopia in non-aphakic, non-diseased eyes with astigmatism up to 0.75D. Spherical powers 0 to -12.00D. Single-use daily disposable.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC, ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens (K000384)
Submission Summary (Full Text)
{0}------------------------------------------------
K13093
## 510 (k) Summary SUMMARY OF SAFETY AND EFFECTIVENESS FOR EASY DAY (methafilcon A )1-Day Soft (hydrophilic) Contact Lens for Daily Wear
Prepared Date: Aug. 20, 2013
| Prepared Date: | Aug. 20, 2013 |
|-----------------|-----------------------------------------------------------|
| | <b>Submitter Information:</b> |
| Company: | Jiangsu Horien Contact Lens Co., Ltd. |
| | No.88 Danfu RD, Situ Town, Danyang City,Jiangsu Province, |
| | China.,212331 |
| Contact Person: | Jinhua Hu |
| | Manager of Registration Division |
| Phone: | 86-511-86809314 |
| FAX: | 86-511-86206366 |
| | AUG 3 0 2013 |
| Identification of Device: | |
|---------------------------|-------------------------------------------------------------------|
| Classification Name: | Lens, Contact, (Disposable) |
| Trade Name: | EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact<br>Lens |
| Common or usual Name: | Soft (hydrophilic) Contact lens (daily wear) |
| FDA Classification: | Class II |
#### Predicate Device:
FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC, ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens cleared via K000384 COOPERVISION, INC.
#### Indications for Use
EASY DAY (methafilcon A) 1-Day Soft ( Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia) in not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 0.75 diopters that does not interfere with visual acuity. The lens may be prescribed in spherical powers ranging from 0 ~ - 12.00D. The lens is indicated for single use daily disposable wear and patients are instructed to discard the lens after each removal.
#### Description of Device
EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) contact lens is available as a single vision aspherical lens. The lens material, methafilcon A is a copolymer of 2-hydroxyethyl
{1}------------------------------------------------
methacrylate (HEMA) and methacrylic acid (MAA) cross-linked with ethylene glycol dimethacrylate (EGDMA) . The lens with visible tint is tinted blue using reactive blue 4K to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing buffered saline solution.
#### Summary of Clinical Study:
The EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lenses for Daily Wear were tested at least 30 evaluable subjects' eyes separately for 1 month. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.
#### Nonclinical Studies:
A series of nonclinical performance tests were performed to demonstrate the safety and effectiveness of the EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear, and establish substantial equivalence to predicate lenses - FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC, ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens (K000384). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification [510(k)] Guidance Document for Class II Daily Wear Contact Lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:
- a) Technological characteristics studies
EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear designs in the following parameter ranges:
Diameter range:14.20 ±0.2mm
Power range: -0.0 D to -12.00 D
Center Thickness: 0.06~0.100mm varies with power
#### Lenses have the following properties:
Refractive index: 1.409 ±0.005
Light transmittance: >92%
Water content: 53 to 57 %
Oxygen permeability (edged corrected): 19.7±20% { 10++1(cm2/sec)(mlO2/ml-mmHg) }
There characterizations of EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lenses are equivalent and comparable to those of predicate lenses.
#### b) Biocompatibility
"In accordance with the May 1994 FDA guideline titled Premarket Notification [510(k)] Guidance Document for Class II Daily Wear Contact Lenses, toxicity studies have been conducted on the model: EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens
{2}------------------------------------------------
for Daily Wear. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated. The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study.
#### c) Microbiology
Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group 4. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.
#### d) Leachability
Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.
#### Substantial equivalence Statement:
Testing performed on the EASY DAY (methafilcon A) I-Day Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices- FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC. ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens (K000384), when used in accordance with the instructions for use. It is due to the facts that the risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.
In conclusion, it is Jiangsu Horien's EASY DAY (methafilcon A )1-Day Soft (hydrophilic) Contact Lens conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the EASY DAY (methafilcon A) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear, with the same established safety profile and effectiveness as the predicate device-- FREQUENCY 55, FREQUENCY 55 ASPHERIC, ENCORE, ENCORE TORIC, ONEVUE, COOPERFLEX (methafilcon A ) Soft (hydrophilic) Contact Lens (K000384).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem in a circular fashion.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
August 30, 2013
Jiangsu Horien Contact Lens Co., Ltd. % Ms. Jennifer Reich Harvest Consulting Corp 2904 N. Boldt Drive, Flagstaff, AZ 86001
Re: K130931
> Trade/Device Name: EASY DAY(methafilcon A ) 1-Day Soft (Hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact lens (daily wear) Regulatory Class: Class II Product Code: LPL Dated: July 29, 2013 Received: July 19, 2013
Dear Ms. Reich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Ms. Jennifer Reich
You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kesia Y. Alexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number: K130931
Device Name: EASY DAY (methafilcon A) I-Day Soft ( Hydrophilic) Contact Lens
### Indications for Use:
EASY DAY (methafilcon A) I-Day Soft ( Hydrophilic) Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia) in not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 0.75 diopters that does not interfere with visual acuity. The lensmay be prescribed in spherical powers ranging from 0 ~ - 12.00D.
The lens is indicated for single use daily disposable wear and patients are instructed to discard the lens after each removal.
Prescription Use______________________________________________________________________________________________________________________________________________________________ X (part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Leonid Livshitz · -2 2013.08.23 : '00'04- 17:19:55 (Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices
510(k) Number:
page 1 of _ l ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.