K130927 · Cardinal Health200, LLC · FRG · Jun 18, 2013 · General Hospital
Device Facts
Record ID
K130927
Device Name
DURABLUE STERILIZATION WRAP
Applicant
Cardinal Health200, LLC
Product Code
FRG · General Hospital
Decision Date
Jun 18, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
DuraBlue™ Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider using: Pre-vacuum steam at 270°F/132°C for 4 minutes . 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% . relative humidity for 60 minutes . Advanced Sterilization Products (ASP) STERRAD® 100S system. . Advanced Sterilization Products (ASP) STERRAD® NX system, Standard and Advanced Cycles . Advanced Sterilization Products (ASP) STERRAD® 100NX system, Standard, Flex, Express, and . DUO Cycles . Lumen, Non Lumen, and Flexible Cycles in the Amsco® V-PRO™ 1, Amsco® V-PRO™ 1 Plus . and Amsco® V-PRO™ maX Low Temperature Sterilization Systems The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used.
Device Story
DuraBlue™ Sterilization Wrap consists of double-layer 100% polypropylene SMS fabric; used by healthcare providers to wrap medical devices prior to sterilization. Device acts as a barrier allowing sterilant penetration while maintaining sterility post-process. Six models (CH100-CH600) vary by basis weight to accommodate different load weights. Used in clinical settings; manual operation by sterile processing staff. Wraps are applied using simultaneous double-wrapping technique; packs are opened aseptically. Benefits include protection of sterilized instruments from contamination during storage and transport.
Clinical Evidence
Bench testing only. Sterility validation studies performed for pre-vacuum steam, EO, STERRAD (100S, NX, 100NX), and V-PRO systems. Validation confirmed maintenance of sterility for specified load weights and lumen configurations (diameters/lengths) per model. No clinical data.
Technological Characteristics
Material: 100% polypropylene spunbond-meltblown-spunbond (SMS) fabric. Form factor: Double-layer wrap in six basis-weight models (CH100-CH600). Sterilization compatibility: Pre-vacuum steam, 100% EO, and low-temperature hydrogen peroxide gas plasma (STERRAD/V-PRO).
Indications for Use
Indicated for use by healthcare providers to enclose medical devices for sterilization via pre-vacuum steam, 100% ethylene oxide, or specific low-temperature hydrogen peroxide gas plasma systems (STERRAD and V-PRO). Intended to maintain sterility of enclosed devices until use.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
K130927
Image /page/0/Picture/1 description: The image contains the logo for Cardinal Health. The logo consists of a stylized graphic above the company name. The graphic appears to be three curved lines that converge at a point, resembling a bird in flight or a stylized leaf. The text "CardinalHealth" is written in a simple, sans-serif font.
1430 Waukegan Road Waukegan, IL 60085
www.cardinalhealth.com
# 510(k) SUMMARY DuraBlue™ Sterilization Wrap
Manufacturer:
Cardinal Health 200, LLC 1430 Waukegan Road Waukegan, IL 60085
1430 Waukegan Road Waukegan, IL 60085
Requlatory Affairs Contact:
Telephone Number:
Fax Number:
Date summary Prepared:
Trade Name:
Classification:
Classification Name:
Common Name:
Product Code:
Predicate Device:
(847) 887-3323
(847) 887-2461
Lavenia Ford
June 14, 2013
DuraBlue™ Sterilization Wrap
Class II per 21 CFR § 880.6850
Sterilization Wrap
Sterilization Wrap
FRG
K123857 – DuraBlue™ Sterilization Wrap – Johnson & Johnson STERRAD® NX and STERRAD® 100NX
K123289 - DuraBlue™ Sterilization Wrap - Pre-Vacuum Steam (4 min/270°F) & 100% Ethylene Oxide (EO)
K120658 – DuraBlue™ Sterilization Wrap – STERRAD® 100S and Arreoooo - Barablao - StormEx. Pro TM 1 Plus, and Amsco® V-PRO™ MAX Low Temperature Sterilization Systems
JUN 1 8 2013
{1}------------------------------------------------
## Description:
DuraBlue™ Sterilization Wraps are double layer sterilization wraps made from 100% polypropylene spunbond-meltblown-spunbond (SMS) fabric. They are intended to enclose another medical device that is to be sterilized by a health care provider using:
- Pre-vacuum steam at 270°F/132°C for 4 minutes .
- 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-. 80% relative humidity for 60 minutes
- Advanced Sterilization Products (ASP) STERRAD® 100S System .
- Advanced Sterilization Products (ASP) STERRAD® NX System, Standard and Advanced ● Cycles
- Advanced Sterilization Products (ASP) STERRAD® 100NX, Standard, Flex, Express, and . DUO cycles
- 2 @ Cyrist
Lumen, Non Lumen, and Flexible Cycles by the STERIS Amsco® V-PRO™ 1, Amsco® VPRO™ 1 Plus and Amsco® V-PRO™ MAX Low Temperature Sterilization Systems .
This wrap design allows for use of the simultaneous double-wrapping technique and also allows for a sterilized pack to be opened aseptically.
The proposed device is identical to the device cleared in the predicate submissions. The only modification to the predicate devices is the clarification to the sterility maintenance information provided in the Indications for Use portion of the Instructions for Use labeling. This change is being made in order to comply with 21CFR880.6850 which states "A sterilization wrap (pack, sterilization wrapper, bag, or accessories) is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used."
This submission covers six different models of DuraBlue™ Sterilization Wrap. Each model is made from material of a different basis weight, though all models utilize the same material technology.
### Indications for Use
DuraBlue™ Sterilization Wrap is intended to enclose another medical device that is to be sterilized by a health care provider using:
- Pre-vacuum steam at 270°F/132°C for 4 minutes .
- 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-. 80% relative humidity for 60 minutes
- Advanced Sterilization Products (ASP) STERRAD® 100S System .
- Advanced Sterilization Products (ASP) STERRAD® NX System, Standard and Advanced . Cvcles
- Advanced Sterilization Products (ASP) STERRAD® 100NX, Standard, Flex, Express, and . DUO cvcles
- Lumen, Non Lumen, and Flexible Cycles by the STERIS Amsco® V-PRO™ 1, Amsco® VPRO™ 1 Plus and Amsco® V-PRO™ MAX Low Temperature Sterilization Systems .
The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used.
For pre-vacuum steam sterilization, the wrap has been validated for dry times of 20 minutes for Y Models CH100 and CH200 and 30 minutes for Models CH300, CH500, and CH600. Models
{2}------------------------------------------------
CH400, CH500, and CH600 have been validated for pre-vacuum steam sterilization of lumens 3 mm in diameter or larger and 400 mm in length or less.
For EO sterilization, the wrap has been validated for an aeration time of 8 hours at 55°C. Models CH400, CH500, and CH600 have been validated for sterilization of two lumens of 3 mm diameter or larger and 400 mm in length or less for EO sterilization.
All models of DuraBlue™ Sterilization Wrap have been validated for Advanced Sterilization Products (ASP) STERRAD® 100S sterlization of lumens 2.5 mm in diameter or larger and 250 mm in length or less.
All models of DuraBlue™ Sterilization Wrap have been validated for use with the Advanced Sterilization Products (ASP) STERRAD® NX and STERRAD® 100NX in Table 1.
All models of DuraBlue™ Sterilization Wrap have been validated for use with the STERIS Amsco® V-PRO™ cycles in Table 2. The DuraBlue™ Sterilization Wrap was validated to be effectively aerated during the pre-programmed STERIS Amsco® V-PRO™ sterilization cycles.
{3}------------------------------------------------
# Table 1 – Validated Advanced Sterilization Products STERRAD® NX and STERRAD® 100NX Sterilization Cycles
,
:
| Advanced<br>Sterilization<br>Products<br>(ASP)<br>STERRAD®<br>System and<br>Cycle | Maximum<br>Recommended<br>Chamber Load | Intended Load |
|-----------------------------------------------------------------------------------|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASP<br>STERRAD® NX<br>Standard Cycle | 10.7 lbs | Reusable metal and non-metal medical devices, including up to 10<br>lumens of the following lumen dimensions per chamber load:<br>• an inside diameter of 1 mm or larger and a length of 150<br>mm or shorter of single-channel stainless steel lumens<br>• an inside diameter of 2 mm or larger and a length of 400<br>mm or shorter of single-channel stainless steel lumens |
| ASP<br>STERRAD® NX<br>Advanced Cycle | 10.7 lbs | Reusable metal and non-metal medical devices, including up to 10<br>lumens of the following lumen dimensions per chamber load:<br>• an inside diameter of 1 mm or larger and a length of 500<br>mm or shorter of single-channel stainless steel lumens<br>OR<br>One single-channel Flexible Endoscope with or without a silicone<br>mat and no additional load. The flexible endoscope may contain:<br>• a single-channel Teflon®/Polyethylene lumen with an inside<br>diameter of 1 mm or larger and a length of 850 mm or<br>shorter |
| ASP<br>STERRAD® 100NX<br>Standard Cycle | 21.4 lbs | Reusable metal and non-metal medical devices, including up to 10<br>lumens of the following lumen dimensions per chamber load:<br>• an inside diameter of 0.7 mm or larger and a length of 500<br>mm or shorter of single-channel stainless steel lumens (A<br>maximum of five lumens per tray per sterilization cycle) |
| ASP<br>STERRAD® 100NX<br>Flex Cycle | 12.2 lbs | One or two single-channel Flexible Endoscope with or without a<br>silicone mat and no additional load. The flexible endoscope may<br>contain:<br>• a single-channel Teflon®/Polyethylene lumen with an inside<br>diameter of 1 mm or larger and a length of 850 mm or<br>shorter (A maximum of two flexible endoscopes, one per<br>tray per sterilization cycle) |
| ASP<br>STERRAD® 100NX<br>Express Cycle | 10.7 lbs | Non-lumened reusable metal and non-metal medical devices<br>requiring surface sterilization, or sterilization of mated stainless steel<br>and titanium surfaces, and rigid or semi-rigid endoscopes without<br>lumens |
| ASP<br>STERRAD® 100NX<br>Duo Cycle | 13.2 lbs | One or two single-channel Flexible Endoscope with or without a<br>silicone mat and no additional load. The flexible endoscope may<br>contain:<br>• a single-channel Teflon®/Polyethylene lumen with an inside<br>diameter of 1 mm or larger and a length of 875 mm or<br>shorter |
.
1000 Children
.
{4}------------------------------------------------
| Table 2 – Validated STERIS Amsco® V-PRO™ cycles | | |
|-------------------------------------------------|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Amsco® V-<br>PRO® Cycle | Maximum<br>Recommended<br>Chamber Load | Intended Load |
| Lumen Cycle | 19.65 lbs | Reusable metal and non-metal medical devices, including up to 20<br>lumens of the following dimensions per chamber load:<br>an inside diameter of 1 mm or larger and a length of 125 mm or shorter an inside diameter of 2 mm or larger and a length of 250 mm or shorter an inside diameter of 3 mm or larger and a length of 400 mm or shorter |
| Non Lumen<br>Cycle | 19.65 lbs | Non-lumened reusable metal and non-metal medical devices |
| Flexible Cycle | 24 lbs | Single or dual lumen surgical flexible endoscopes and bronchoscopes in<br>either of two load configurations:<br>1. Two trays, each containing a flexible endoscope with a light cord (if<br>not integral to endoscope) and mat with no additional load<br>2. One tray containing a flexible endoscope with a light cord (if not<br>integral to endoscope) and mat and an additional tray containing<br>non-lumened medical devices<br>The flexible endoscope(s) may contain either:<br>a single lumen with an inside diameter of 1 mm or larger<br>and a length of 1050 mm or shorter two lumens, with one lumen having an inside diameter of 1<br>mm or larger and a length of 998 mm or shorter and the<br>other lumen having an inside diameter of 1 mm or larger<br>and a length of 850 mm or shorter |
and the control control control control controlled
100 - 100 - 100 - 100 -
# Table 2 – Validated STERIS Amsco® V-PRO™ cycles
{5}------------------------------------------------
Table 3 - Wrap Model Recommendations1
| | | Maximum Recommended Wrapped Package Content<br>Weight2 | | | |
|-----------------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------|-----------------------------------------------------------------|----------------------------------------------------------------------------|----------------------------|
| Sterilization<br>Wrap Model | Intended Load | Pre-Vacuum<br>Steam and EO | Advanced<br>Sterilization<br>Products (ASP)<br>STERRAD®<br>100S | Advanced<br>Sterilization<br>Products<br>(ASP)<br>STERRAD®<br>NX and 100NX | STERIS<br>Amsco®<br>V-PRO™ |
| CH100 | Very light weight package<br>(for example: towel packs or<br>batteries). | 3 lbs | 3 lbs | 10.7 lbs | 3 lbs |
| CH200 | Light weight package<br>(for example: telescope with light<br>cord). | 6 lbs | 6 lbs | 10.7 lbs | 6.5 lbs |
| CH300 | Light to moderate weight package<br>(for example: general use medical<br>instruments) | 9 lbs | 9.7 lbs | 10.7 lbs | 9 lbs |
| CH400 | Moderate to heavy weight package<br>(for example: general use medical<br>instruments) | 13 lbs | 9.7 lbs | 10.7 lbs | 9.1 lbs |
| CH500 | Heavy weight package<br>(for example: general use medical<br>instruments) | 17 lbs | 9.7 lbs | 10.7 lbs | 9.1 lbs |
| CH600 | Very heavy weight package<br>(for example: general use medical<br>instruments) | 25 lbs | 9.7 lbs | 10.7 lbs | 9.1 lbs |
The following loads were used in the pre-vacuum steam Sterility Validation Studies:
· CH100: 16 huck towels (17 in. x 29 in.).
• CH200: 2 huck towels (17 in. x 29 in.), 3 fluid-resistant drapes (108 in. x 72 in.).
· CH300: 16 huck towels (17 in. x 29 in.), 1 fluid-resistant table cover (90 in. x 60 in.), 5 lbs of metal mass.
· CH400: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 8 Ibs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH500: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 12 Ibs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH600: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 20 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
{6}------------------------------------------------
The following loads were used in the EO Sterility Validation Studies.
- · CH100: 16 huck towels (17 in. x 29 in.).
· CH200: 2 huck towels (17 in. x 29 in.), 2 fluid-resistant drapes (108 in. x 88 in.), 2.5 lbs of metal mass.
· CH300: 16 huck towels (17 in. x 29 in.), 1 fluid-resistant table cover (90 in. x 60 in.), 5 lbs of metal mass.
· CH400: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm I. D x 400 mm), and 7.5 Ibs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH500: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm 1. D x 400 mm), and 11.5 Ibs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH600: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm 1. D x 400 mm), and 19.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
The following loads were used in the Advanced Sterilization Products (ASP) STERRAD® 100S Sterility Validation Studies:
· CH100: Metal Instruments
· CH200 - CH600: 15 in. x 10 in. x 1.2 in. tray containing metal instruments.
The following loads were used in the Advanced Sterilization Products (ASP) STERRAD® NX and STERRAD® 100NX Sterility Validation Studies:
· CH100 - CH600: 23 in. x 11 in. x 4 in. tray containing metal instruments.
The following loads were used in the STERIS Amsco® V-PRO™ sterility Validation Study:
- · CH100: Metal Instruments
- · CH200 CH600: 17 in. x 10 in. x 3.5 in. tray containing metal instruments
Note: The loads used in the Sterility Validation Study corresponded to the maximum wrapped package content weights in Table 3.
Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irregularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use.
2 It is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight and size of individual. content types that were validated for the DuraBlue™ Sterilization Wraps.
### Substantial Equivalence
The proposed DuraBlue " Sterilization Wrap is substantially equivalent to the predicate devices. Both devices:
- Have the same intended use
- Have the same material composition
- Have the same physical and chemical properties
- Have the same configurations/dimensions
- Demonstrate maintenance of package sterility
- Are indicated for the same sterilization parameters
- Are indicated for the same Maximum Wrapped Package Content Weight
{7}------------------------------------------------
# Summary of Testing
DuraBlue™ Sterilization Wrap performance has been tested in accordance with the applicable requirements recommended in the FDA's Guidance Document Premarket Notification 510(k) Submissions for Medical Sterilization Packaging System in Health Care Facilities; Draft Guidance for Industry and FDA (March 7, 2002) in the predicate Premarket Notifications. Risk analysis showed that the modification to the information provided in the labeling does not present and increase or change in risk of illness or injury associated with the use of this device. The only modification to the labeling is the clarification to the sterility maintenance information provided in the Indications for Use portion of the Instructions for Use ' labeling.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine and health. The caduceus is depicted as a staff with a snake winding around it, and it is accompanied by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## June 18, 2013
Cardinal Health Ms. Lavenia Ford Manager, Regulatory Affairs 1430 Waukegan Road WAUKEGAN, IL 60085
Re: K130927
Trade/Device Name: DuraBlue™ Sterilization Wrap Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: May 16, 2013 Received: May 20, 2013
## Dear Ms. Ford:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Ms. Ford
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Susan Runner DDS, MA Mary S. Runner--S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Appendix 2 – Indications for Use
:
:
.
. . . . . .
:
:
.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of three curved lines above the company name. The company name is written in a bold, sans-serif font.
## Indications for Use
510(k) Number (if known):
Device Name: DuraBlue™ Sterilization Wrap
## Indications for Use
DuraBlue™ Sterilization Wrap is intended to be used to enclose another medical device that is to be sterilized by a health care provider using:
- Pre-vacuum steam at 270°F/132°C for 4 minutes .
- 100% ethylene oxide (EO) with a concentration of 725-735 mg/L at 131°F/55°C and 40%-80% . relative humidity for 60 minutes
- . Advanced Sterilization Products (ASP) STERRAD® 100S system.
- . Advanced Sterilization Products (ASP) STERRAD® NX system, Standard and Advanced Cycles
- Advanced Sterilization Products (ASP) STERRAD® 100NX system, Standard, Flex, Express, and . DUO Cycles
- Lumen, Non Lumen, and Flexible Cycles in the Amsco® V-PRO™ 1, Amsco® V-PRO™ 1 Plus . and Amsco® V-PRO™ maX Low Temperature Sterilization Systems
The wrap is intended to allow sterilization of the enclosed medical device(s) and to maintain sterility of the enclosed device(s) until used.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Devices Evaluation (ODE)
Susan Runner DOS,MA Mary S.
Runner -S
Division Sign-Off Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K130927 510(k)
{12}------------------------------------------------
For pre-vacuum steam sterilization, the wrap has been validated for dry times of 20 minutes for Models CH100 and CH200, and for 30 minutes for Models CH300, CH400, CH500 and CH600. Models CH400, CH500 and CH600 have been validated for pre-vacuum steam sterilization of two lumens 3 mm in diameter or larger and 400 mm in length or less.
For EO sterilization, the wrap has been validated for an aeration time of 8 hours at 55 °C. Models CH400, CH500 and CH600 have been validated for EO sterilization of two lumens of 3 mm diameter or larger and 400 mm in length or less.
All models of DuraBlue™ Sterilization Wrap have been validated for Advanced Sterilization Products (ASP) STERRAD® 100S sterilization of lumens 2.5 mm in diameter or larger and 250 mm in length or less.
All models of DuraBlue™ Sterilization Wrap have been validated for use with the Advanced Sterilization Products (ASP) STERRAD® NX and STERRAD® 100NX cycles in Table 1.
All models of DuraBlue™ Sterilization Wrap have been validated for use with the STERIS Amsco® V-PRO® cycles in Table 2. The DuraBlue™ Sterilization Wrap was validated to be effectively aerated during the pre-programmed STERIS Amsco® V-PRO® sterilization cycles.
{13}------------------------------------------------
Table 1: Validated Advanced Sterilization Products (ASP) STERRAD® NX and STERRAD® 100NX Cycles
-
....
.
.
- -- · -
·
| Advanced<br>Sterilization<br>Products<br>(ASP)<br>STERRAD®<br>System and<br>Cycle | Maximum<br>Recommended<br>Chamber Load | Intended Load |
|-----------------------------------------------------------------------------------|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASP<br>STERRAD® NX<br>Standard Cycle | 10.7 lbs | Reusable metal and non-metal medical devices, including up to<br>10 lumens of the following lumen dimensions per chamber load:<br>• an inside diameter of 1 mm or larger and a length of 150<br>mm or shorter of single-channel stainless steel lumens<br>• an inside diameter of 2 mm or larger and a length of 400<br>mm or shorter of single-channel stainless steel lumens |
| ASP<br>STERRAD® NX<br>Advanced Cycle | 10.7 lbs | Reusable metal and non-metal medical devices, including up to<br>10 lumens of the following lumen dimensions per chamber load:<br>• an inside diameter of 1 mm or larger and a length of 500<br>mm or shorter of single-channel stainless steel lumens<br>OR<br>One single-channel Flexible Endoscope with or without a silicone<br>mat and no additional load. The flexible endoscope may contain:<br>• a single-channel Teflon®/Polyethylene lumen with an<br>inside diameter of 1 mm or larger and a length of 850 mm<br>or shorter |
| ASP<br>STERRAD®<br>100NX<br>Standard Cycle | 21.4 lbs | Reusable metal and non-metal medical devices, including up to<br>10 lumens of the following lumen dimensions per chamber load:<br>• an inside diameter of 0.7 mm or larger and a length of 500<br>mm or shorter of single-channel stainless steel lumens (A<br>maximum of five lumens per tray per sterilization cycle) |
| ASP<br>STERRAD®<br>100NX Flex<br>Cycle | 12.2 lbs | One or two single-channe! Flexible Endoscope with or without a<br>silicone mat and no additional load. The flexible endoscope may<br>contain:<br>• a single-channel Teflon®/Polyethylene lumen with an<br>inside diameter of 1 mm or larger and a length of 850 mm<br>or shorter (A maximum of two flexible endoscopes, one<br>per tray per sterilization cycle) |
| ASP<br>STERRAD®<br>100NX Express<br>Cycle | 10.7 lbs | Non-lumened reusable metal and non-metal medical devices<br>requiring surface sterilization, or sterilization of mated stainless<br>steel and titanium surfaces, and rigid or semi-rigid endoscopes<br>without lumens |
| ASP<br>STERRAD®<br>100NX Duo Cycle | 13.2 lbs | One or two single-channel Flexible Endoscope with or without a<br>silicone mat and no additional load. The flexible endoscope may<br>contain:<br>• a single-channel Teflon®/Polyethylene lumen with an<br>inside diameter of 1 mm or larger and a length of 875 mm<br>or shorter |
{14}------------------------------------------------
The following loads were used in the EO Sterility Validation Studies:
· CH100: 16 huck towels (17 in. x 29 in.).
· CH200: 2 huck towels (17 in. x 29 in.), 2 fluid-resistant drapes (108 in. x 88 in.), 2.5 lbs of metal mass.
· CH300: 16 huck towels (17 in. x 29 in.), 1 fluid-resistant table cover (90 in. x 60 in.), 5 lbs of metal mass.
· CH400: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm ID x 400 mm) and 7.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. trav.
· CH500: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm ID x 400 mm) and 11.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
· CH600: 4 stacked tray liners (20 in. x 25 in.), 2 lumens (3 mm ID x 400 mm) and 19.5 lbs of metal mass in a 23 in. x 11 in. x 3.5 in. tray.
The following loads were used in the Advanced Sterilization Products (ASP) STERRAD® 100S Sterility Validation Studies:
· CH100: Metal instruments.
· CH200 - CH600: 15 in. x 10 in. x 1.2 in. tray containing metal instruments.
The following loads were used in the Advanced Sterilization Products (ASP) STERRAD® NX and STERRAD® 100NX Sterility Validation Studies:
· CH100 - CH600: 23 in. x 11 in. x 4 in. tray containing metal instruments.
The following loads were used in the STERIS Amsco® V-PRO® Sterility Validation Studies: · CH100: Metal instruments.
· CH200 - CH600: 17 in. x 10 in. x 3.5 in. tray containing metal instruments.
Note: The loads used in each Sterility Validation Study corresponded to the maximum wrapped package content weights in Table 3.
'Individual results may differ due to factors such as variations in handling practices, wrapping techniques, and folding methods. Results may also differ due to the use of irreqularly shaped contents, which may put added stress on the wrap. Each healthcare facility should determine for itself which wrap model is most appropriate for each intended use.
2 It is recommended to not exceed the maximum wrapped package content weights indicated for each wrap model. Furthermore, it is recommended to not exceed the number, weight and size of individual content types that were validated for the DuraBlue™ Sterilization Wraps.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.