CORRECT PLUS 1

K130920 · Pentron Clinical · ELW · Jul 2, 2013 · Dental

Device Facts

Record IDK130920
Device NameCORRECT PLUS 1
ApplicantPentron Clinical
Product CodeELW · Dental
Decision DateJul 2, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

Correct Plus 1 is a vinyl polysiloxane impression material which is to be used for taking impression of inlays, crowns & bridges, orthodontic appliances, precision attachments and veneers.

Device Story

Correct Plus 1 is a vinyl polysiloxane (VPS) dental impression material used by dental professionals to capture accurate negative imprints of hard (teeth) and soft oral tissues. The material is placed into the mouth via a dental impression tray; it exhibits thixotropic properties and hydrophilic behavior to ensure detail capture in the presence of oral fluids. Upon setting, it forms an elastic solid that is removed from the mouth to provide a stable negative mold, which is subsequently used to create a cast of the dentition. Available in multiple viscosities (light, medium, heavy/putty) and set times, it supports various clinical procedures including two-step heavy-wash or putty-wash techniques and monophase impressions. The device benefits patients by enabling the fabrication of precise dental restorations and appliances.

Clinical Evidence

No clinical testing was conducted. Substantial equivalence is supported by biocompatibility testing (ISO 10993 MEM Elution, Kligman Maximization, and Oral Irritation tests) and extensive bench testing comparing physical and performance characteristics (e.g., work time, set time, viscosity, dimensional change, detail reproduction, and mechanical strength) against the predicate device.

Technological Characteristics

Vinyl polysiloxane (VPS) impression material. Thixotropic, hydrophilic formulation. Available in light, medium, and heavy (putty) viscosities. Supplied in auto-mix cartridges. Shelf life: 3 years. No electronic components or software.

Indications for Use

Indicated for taking impressions of inlays, crowns, bridges, orthodontic appliances, precision attachments, and veneers in patients requiring dental restorative or orthodontic procedures.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130920 Image /page/0/Picture/1 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a swoosh above the letters. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are printed in a smaller, sans-serif font. The logo appears to be black and white. JUL 0 2 2013 #### 510(k) Summary #### Submitter: Sybron Dental Specialties, Inc. 1717 West Collins Avenue Orange, California 92867 (714) 516-7602 - Phone (714) 516-7488 - Facsimile Wendy Garman - Contact Person Date Summary Prepared: June 2013 - Trade name Correct Plus 1 이 - Common name Impression Material 0 - Classification name Material, Impression (21 CFR 872.3660, Product Code ELW) o #### Devices for Which Substantial Eguivalence is Claimed: Correct Plus, Class II, K001218, Product Code ELW, Pentron Clinical 0 #### Summary #### Device Description Correct Plus 1 impression material (Fast Set) are versatile impression materials designed to accommodate impressions of inlays, crowns and bridges, orthodontic appliances, precision attachments and veneers. The hydrophilic nature of these materials provides outstanding detail in the presence of fluids and exceptional tear strength ensures your detailed impression will remain intact upon removal. Correct Plus 1 impression material provides multiple packaging options and various set times to accommodate all of your impression needs. For optimal patient acceptance, Correct Plus 1 impression material is available in unflavored and Berry. #### Indications of Use Correct Plus 1 is a vinyl polysiloxane impression material which is to be used for taking impression of inlays, crowns & bridges, orthodontic appliances, precision attachments and veneers. . HAP West Collins Avenue, Orange, CA 92867 800-537-7824 714-516-7400 15 of 17 {1}------------------------------------------------ | Description<br>Information | Correct Plus 1 | Correct Plus<br>K001218 | | |----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company Name | Pentron Clinical | Pentron Clinical | | | Intended Use | Correct Plus 1 is a vinyl polysiloxane<br>impression material which is to be<br>used for taking impression of inlays,<br>crowns & bridges, orthodontic<br>appliances, precision attachments<br>and veneers. | Correct Plus is a vinyl polysiloxane<br>impression material which is to be<br>used for taking impression of<br>inlays, crowns & bridges,<br>orthodontic appliances, precision<br>attachments and veneers. | | | Description of<br>Material | Vinly Polysiloxane | Vinyl Polysiloxane | | | Mode of Use | Thick n' Thin fast set:<br>A thixotropic light viscosity wash<br>impression material that is used in<br>two-step heavy-wash or putty-wash<br>procedures to capture accurate<br>subgingival details. | Hydrophilic Light Body:<br>A thixotropic light viscosity wash<br>impression material that is used in<br>two-step heavy-wash or putty-<br>wash procedures to capture<br>accurate subgingival details. | | | | Universal fast set:<br>A medium viscosity material that<br>produces a highly detailed,<br>extremely accurate monophase<br>impression. It can also be used as an<br>alginate substitute impression<br>material. | Hydrophilic Universal:<br>A medium viscosity material that<br>produces a highly detailed,<br>extremely accurate monophase<br>impression. It can also be used as<br>an alginate substitute impression<br>material. | | | | Auto-mix Putty fast set:<br>A heavy viscosity material that<br>provides the hydraulic drive of a<br>hand-mix putty in a convenient auto-<br>mix cartridge. It can be used in all<br>applications putties are used. | Putty:'<br>A heavy viscosity material that<br>provides the hydraulic drive of a<br>hand-mix putty in a convenient<br>auto-mix cartridge. It can be used<br>in all applications putties are used. | | | | Principles of<br>Operation | Correct Plus 1 is a dental impression<br>that takes imprints of hard (teeth)<br>and/or soft tissues. Correct Plus 1<br>captures a part or all of a person's<br>dentition and surrounding structures<br>of oral cavity. The dental impression | Correct Plus is a dental impression<br>that takes imprints of hard (teeth)<br>and/or soft tissues. Correct Plus<br>captures a part or all of a person's<br>dentition and surrounding<br>structures of oral cavity. The | | | | Description<br>Information | Correct Plus 1 | | | | forms an imprint (i.e. a 'negative'<br>mold) of teeth and soft tissues,<br>which can then be used to make a<br>cast of the dentition. An impression<br>is made by placing a viscous,<br>thixotropic Impression material into<br>the mouth via a dental Impression<br>tray. The material, then sets to<br>become an elastic solid, and, when<br>removed from the mouth, provides a<br>detailed and stable negative of<br>teeth. | dental impression forms an imprint<br>(i.e. a 'negative' mold) of teeth and<br>soft tissues, which can then be<br>used to make a cast of the<br>dentition. An impression is made<br>by placing a viscous, thixotropic<br>impression material into the<br>mouth via a dental impression<br>tray. The material, then sets to<br>become an elastic solid, and, when<br>removed from the mouth, provides<br>a detailed and stable negative of<br>teeth. | | Shelf- Life | 3 Years | 3 Years | | ### Summary of Technological Characteristics Compared to Predicate K130920, Correct Plus 1 16 of 17 . ' {2}------------------------------------------------ #### Non-Clinical Performance Data Biocompatibility study was completed, which demonstrates that the material is safe for its intended use. Correct Plus 1 was tested through the following tests: ISO L929 MEM Elution Test, ISO Kligman Maximization Test and ISO Oral Irritation Test. The 510(k) submission also includes data from bench testing used to evaluate performance characteristics of Correct Plus 1 as compare to the predicate device, Correct Plus currently marketed by Pentron Clinical. The characteristics evaluated include Work Time, Oral Set Time, Catalyst Viscosity, Base Viscosity, Out Gassing, Hardness, Mixed Consistency, Dimensional Change, Detail Reproduction, Compatibility with Gypsum, Elastic Recovery, Strain in Compression, Tensile Strength, Tensile Elongation, Tear Strength and Contact Angle. #### Clinical Testing Clinical testing has not been conducted on this product. #### Conclusion Based upon the biocompatibility tests and bench testing, the clinical performance of Correct Plus 1 is substantially equivalent to the predicate device. 17 of 17 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### July 2, 2013 Sybron Dental Specialties, Incorporated C/O Ms. Wendy Garman 1717 West Collins Avenue ORANGE CA 92867 Re: K130920 Trade/Device Name: Correct Plus 1 Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: April 1, 2013 Received: April 3, 2013 #### Dear Ms. Garman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. # Kwame O. Ulmer -S Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K130920 Device Name: Correct Plus 1 Indications For Use: Correct Plus 1 is a vinyl polysiloxane impression material which is to be used for taking impression of inlays, crowns & bridges, orthodontic appliances, precision attachments and veneers. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sheena A. Green -S 2013.06.28 16:56:48 -04'00' for M. Susan Runner, DDS, MA (Division Sign-Off) (Division Sign-on) Division of Anesthesiology, General Hospital Division of Antalia.org/w/index. Pentron Clinical - Correct Plus 1 - 510(k) Submis Seguion 004 - Page 2 of 2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...