K130897 · Medcomp ( Medical Components) · LJS · Oct 22, 2013 · General Hospital
Device Facts
Record ID
K130897
Device Name
MEDCOMP VASCU-PICC WITH VALVE
Applicant
Medcomp ( Medical Components)
Product Code
LJS · General Hospital
Decision Date
Oct 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Peripherally Inserted Central Vein Access Catheters with valve technology are designed for Long or Short-term peripheral access to the central venous system for intravenous therapy and blood sampling, and allows for central venous pressure monitoring.
Device Story
Vascu-PICC with Valve Technology is a peripherally inserted central venous catheter (3F/4F single lumen; 4F/5F double lumen). Device features bi-directional valve integrated into luer-lock connector; valve remains closed until flow is induced in either direction. Catheter includes depth markings and tapered diameter near hub for kink resistance and venotomy bleeding control. Used by clinicians for central venous access; facilitates IV therapy, blood sampling, and pressure monitoring. Sterile, single-use device.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench/performance testing conducted in accordance with international standards and internal engineering protocols. Biocompatibility testing confirms materials meet ISO 10993 requirements.
Technological Characteristics
Peripherally inserted central venous catheter; 3F/4F single lumen or 4F/5F double lumen configurations. Features integrated bi-directional valve in luer-lock connector. Materials previously cleared in Medcomp devices; compliant with ISO 10993. Includes depth markings and kink-resistant hub design. Sterile, single-use.
Indications for Use
Indicated for patients requiring long or short-term peripheral access to the central venous system for intravenous therapy, blood sampling, and central venous pressure monitoring.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
Medcomp, Vascu-PICC and Midline Catheters (K121094)
Bard Access Systems, Inc., PowerPICC SOLO (K072230)
Submission Summary (Full Text)
{0}------------------------------------------------
# K120897
Image /page/0/Picture/1 description: The image shows a logo with the text "medcomp." in a stylized font. Above the text is an abstract geometric shape, primarily black and white. The shape appears to be a stylized letter or symbol, possibly representing the company's initials or a related concept.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax: 215-256-1787
www.medcompnet.com
| Section 5 | | 510(k) SUMMARY | Traditional 510K |
|-----------|-----------------------------------|------------------------------------------------------------------------------------------------------------|------------------|
| A | Submitter Information: | | |
| | Submitter: | MEDCOMP®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201, x2271<br>Fax: (215) 256-9191 | |
| | Contact: | Jessica Leo<br>Regulatory Associate | |
| | Date Prepared: | March 28, 2013 | OCT 22 2013 |
| B. | Trade Name:<br>Technology | Medcomp® Vascu-PICC® with Valve | |
| | Common Name: | Catheter, Intravascular, Therapeutic,<br>Long-Term | |
| | Classification Name: | Long Term Intravascular Catheter (80 LJS) | |
| | Regulation Name:<br>intravascular | Percutaneous, implanted, long-term | |
| | C.F.R. Section: | catheter<br>880.5970 | |
| | Class: | II | |
| | | | |
| C. | Predicate Devices: | K121094 Medcomp, Vascu-PICC® and Midline<br>Catheters, class II 880.5970 | |
#### D. Device Description:
The Vascu-PICC® with Valve Technology is available in various configurations; they are a 3F & 4F single lumen and 4F & 5F double lumen. The catheter lumen terminates through an extension to a female luer-lock connector. Assembled within each huer is a bi-directional valve that can control fluid flow in two directions. The valve is normally closed but opens when flow is induced in either direction. Each extension is marked with the lumen gauge size and "Valved catheter". The transition between lumen and extension is housed within a molded hub. The hub is marked with catheter French size. The outside diameter of the lumen increases gradually near the hub to aid in kink resistance and to provide a mechanical obstruction to bleeding from the venotomy. The lumen is marked with depth marks every centimeter.
K072230 Bard Access Systems, Inc., PowerPICC SOLO™, class II 880.5970
The catheter is packaged sterile in a variety of tray configurations with the necessary accessories to facilitate catheter insertion.
ど。 Indications for Use:
Medcomp Vascu-PICC with Valve Technology Summary
Page 1 of 2
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with a geometric design above the word "medcomp". The geometric design features a combination of thick, black lines forming an abstract shape, with a white circle incorporated into the design. The word "medcomp" is written in a lowercase, italicized font below the geometric shape.
F.
1499 Delp Drive
Hurleysville, PA 19438
Tel: 215-256-4201
Fax: 215-256-1787
www.medcompnet.com
The Peripherally Inserted Central Vein Access Catheters with valve technology are designed for Long or Short-term peripheral access to the central venous system for intravenous therapy and blood sampling, and allows for central venous pressure monitoring.
### Comparison to Predicate Devices:
The Vascu-PICC® with valve technology catheter is substantially equivalent to the predicate devices in terms of intended use, anatomical location, basic design, materials, performance, labeling, manufacturing process and method of sterilization.
The difference between the Vascu-PICC® with valve technology and the predicate devices (K121094) is the addition of the valve technology to the luer.
#### G. Bench / Performance Data:
Performance testing of the proposed device was conducted in accordance with applicable international standards and FDA guidance documents. Performance standards for pressure injection have not been established by FDA under section 514 of the Federal Food, Drug and Cosmetic Act, Testing is based upon internal engineering testing methods.
The results of these tests in conjunction with the substantial equivalence claims effectively demonstrate the proposed devices are equivalent to the predicate devices.
#### ਜ਼ੁ Biocompatibility:
Testing for all materials used for the Vascu-PICC® with valve technology has been submitted in previously cleared Medcomp devices. All biocompatibility testing demonstrates the materials used meet the requirements of ISO 10993.
#### L Technological Characteristics:
Technological similarities between the proposed device and predicate devices remain the same.
#### J. Summary of Substantial Equivalence:
The proposed device meets the performance criteria of design verification as specified by ISO standards, guidance documents and internal test protocols. The proposed device has the same intended use, operation and function as the predicates. There are no differences that raise new issues of safety and effectiveness. The proposed device is substantially equivalent to the legally marketed predicate device.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## October 22, 2013
MEDCOMP (Medical Components) Jessica Leo Regulatory Associate 1499 Delp Drive Harleysville, PA 19438
Re: K130897
Trade/Device Name: Vascu-PICC with Valve Technology Regulation Number: 21 CFR 880.5970 Regulation Name: Catheter, Intravascular, Therapeutic, Long-Term Regulatory Class: II Product Code: LJS Dated: September 3, 2013 Received: September 10, 2013
Dear Ms. Leo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include reguirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Leo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
MarySBanner-S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
K130897 510(k) Number (if known):
Vascu-PICC® with Valve Technology Device Name:
Indications for Use:
The Peripherally Inserted Central Vein Access Catheters with valve technology are designed for Long or Short-term peripheral access to the central venous system for intravenous therapy and blood sampling, and allows for central venous pressure monitoring.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/4/Picture/10 description: The image shows the FDA logo. The logo is a stylized version of the letters "FDA". The letters are outlined in black. The logo is simple and recognizable.
Richard C. Chapman 2013.10.22 14:06:05 -04'00'
Page _1_ of _1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.