The Medcomp Split Cath® Rg is intended for use in attaining long term vascular access for Hemodialysis and Apheresis in the adult patient.
Device Story
Split Cath® Rg is a long-term, 14 French dual-lumen catheter for hemodialysis and apheresis; provides dedicated arterial/venous access lumens. Catheter features thermo-sensitive polyurethane that softens at body temperature to reduce vessel trauma; venous lumen is 1.0" longer than arterial to prevent recirculation. Device is designed for retrograde insertion; packaged as separate components (lumen, attachable hub, extension assembly) for assembly by clinical personnel post-insertion. Hub assembly includes a stainless steel cannula, compression ring, and hub clamps to ensure secure connection. Used in hospital settings by clinicians. Consistent flows up to 400ml/min. Output is vascular access for extracorporeal blood treatment; facilitates hemodialysis/apheresis procedures, benefiting patients requiring chronic blood access.
Clinical Evidence
Bench testing only. Testing included air leakage, liquid leakage, force at break, recirculation, flow vs. pressure, and priming volume. Additional mechanical hemolysis testing and MRI safety evaluation were performed due to design modifications.
Indicated for long-term vascular access for hemodialysis and apheresis in adult patients. May be inserted percutaneously, primarily in the internal jugular vein, with alternate sites including subclavian and femoral veins.
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image is a black and white logo. The logo features a stylized depiction of a person in motion, possibly running or dancing, composed of three curved lines. The figure is enclosed within a circle, and the text "ARIZONA MINISTRIES" is written around the circumference of the circle. There is also a small dot inside the circle.
#### DEPARTMENT OF HEALTH & IIUMAN SERVICES
Public Health Service
Food and Dree Administration 10903 Now Hamoshing Avenue Doeument Caniral Center - WO66-G60 Suver Spring, MD 20991-00002
April 28. 2014
MEDCOMP® Timothy Holvick Regulatory Associate 1499 Delp Drive Harleysville, PA 19438
Re: K130889
> Trade/Device Name: Split Cath® Rg Regulation Number: 21 CFRS 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: March 17. 2014 Received: March 18, 2014
Dear Timothy Holwick,
We have reviewed your Section 510(k) nremurket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnaciment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval upplication (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially cquivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976. the enacunent date of the Medical Device Amendmens. Please note: If you purchase your device components in bulk (i.c., unfinished) and further process (c.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
{1}------------------------------------------------
Page 2 - Timothy Holwick
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Herbert P. Lerner-S
for
Beniamin R. Fisher. Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo and contact information for Medcomp. The address is 1499 Delp Drive, Harleysville, PA 19438. The phone number is 215-256-4201, the fax number is 215-256-1787, and the website is www.medcompnet.com.
# Indications for Use
510(k) Number (if known): K130889
Split Cath® Ry Device Name:
Indications for Use:
The Medcomp Split Cath® Ry is indicated for use in attaining Long-Term vascular access for Hemodialysis und Apheresis in the adult patient.
It may be inserted perculuneously and is primarily placed in the internal jugular vein. Altemate insertion sites include the subclavian vein and femoral vein.
Prescription Use ਮ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Usc (2) CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Herbert P. Lerner -S 2014.04.28 16:20:38 -04'00' Page _1_ of
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for a company called "medcomp". The logo is a stylized letter "M" that is made up of geometric shapes. The letter "M" is in gray, and the background is white. The word "medcomp" is written in lowercase letters below the logo.
edCOMP
1499 Delp Drive Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787
www.medcompnet.com
Section 5 510(k) SUMMARY Submitter Information: Submitter: MEDCOMP@ 1499 Delp Drive Harleysville, PA 19438 (215) 256-4201 Telephone (215) 256-9191 Fax Timothy Holwick. Regulatory Associate Date Prepared: March 25, 2013 Device Name: Split Cathe Ra
Traditional 510K
| Device Name: | Split-Cath® Rg |
|-----------------------|-------------------------------------|
| Common Name: | Catheter, Hemodialysis, Implanted |
| Classification Name: | Blood Access device and accessories |
| C.F.R. Section: | 876.5540 |
| Classification Panel: | Gastroenterology and Urology |
| Class: | III, MSD |
#### Predicate Devices:
Contact:
K121848, Split Cath® III concurrence date September 21, 2012. Class III CFR 8876.5540 K 022678 Split Cath IV (Split Stream), concurrence date February 24, 2003 Class III CFR \$876.5540
## Device Description:
The Split Cath® Rg Long Term Catheter provides two dedicated (arterial/venous) access lumens. Each lumen is connected through an extension line with female luer connectors. The transition between lumen and extension is housed within a molded hub. To prevent recirculation, the venous end of the lumen is approximately 1.0" longer than the arterial. The catheter is capable of providing consistent flows up to 400ml/min.
The catheter lumen is composed of a soft, thermo-sensitive, polyurethane material that is rigid upon insertion and once it reaches body temperature becomes soft to reduce vessel trauma.
The dialysis extensions are color coded with a white luer, red sleeve, and red clamp for the arterial lumen. The easy identification of the venous lumen, there is a white luer, blue sleeve, and blue clamp.
The proposed device is 14 French in size and comes in the following lumen lengths: 24.28. 32, 36, and 40cm. The catheter is packaged in separate components as they are assembled during retrograde insertion. After the lumen is inserted, it is attached to the Attachable Hub and Extension Assembly using the Lumen Clamp.
## Intended Used:
The Medcomp Split Cath® Rg is intended for use in attaining long term vascular access for Hemodialysis and Apheresis in the adult patient.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows an abstract design with geometric shapes. The design features a combination of curved and straight lines, creating a sense of depth and dimension. The shapes are arranged in a way that suggests a sense of movement and flow. The overall composition is visually striking and intriguing.
eCOMH
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Indications for Use:
The Medcomp Split Cath® Rg is indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis in the adult patient.
If may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion sites include the subclavian vein and femoral vein.
Comparison to Predicate Devices:
The Split Cath® Rg is substantially equivalent to the predicate devices in terms of intended use, materials, anatomical location, basic design, performance, labeling, manufacturing process and method of sterilization.
| ATTRIBUTE | Split Cath RG (Proposed Device) | Split Cath III K121848<br>(Predicate Device) |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INDICATIONS<br>FOR USE | The Medcomp Split Cath Rg is<br>indicated for use in attaining Long-Term<br>vascular access for Hemodialysis and<br>Apheresis. | The Medcomp Split Cath III is<br>indicated for use in attaining long<br>term vascular access for<br>Hemodialysis and Apheresis in the<br>adult patient. |
| TARGET<br>POPULATION | Adult | Adult |
| INSERTION SITE | Internal jugular vein<br>Subclavian vein | Internal jugular vein<br>Subclavian vein<br>Femoral vein<br>Translumbar |
| WHERE USED | Hospital | Hospital |
| STERILIZATION<br>METHOD | 100% Ethylene Oxide | 100% Ethylene Oxide |
| MATERIAL /<br>ADDITIVES | Lumen - Polyurethane: ChronoFlex<br>Cuff - Polyester<br>Hub/Suture Wing and Luers: Isoplast<br>Extension: Polyurethane<br>Clamps - Acetal<br>Extension Sleeve - Silicone<br>Cannula and Pins - Stainless Steel | Lumen - Polyurethane: ChronoFlex<br>Cuff - Polyester<br>Hub and Suture Wing -<br>Polyurethane<br>Extension - Polyurethane<br>Clamps - Acetal<br>Luers - Isoplast<br>Extension Sleeve - Silicone |
| DESIGN<br>SPECIFICATION | LUMEN: 14F<br>LUMEN LENGTHS: 24, 28, 32, 36, and<br>40cm | LUMEN: 14F<br>LUMEN LENGTHS: 20, 22, 24, 28,<br>32, 36, 40, and 55cm |
## Performance Standards:
Performance standards have not been established by the FDA under section 514 of the Federal Food, Drug and Cosmetic Act.
Biocompatibility:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows an abstract design with geometric shapes. The design features a combination of curved and angular elements, creating a sense of depth and dimension. The shapes are filled with a gradient of gray, adding to the visual complexity of the image. The overall composition is visually striking and evokes a sense of modern art.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax. 215-256-1787
www.medcompnet.com
Results for all biocompatibility testing demonstrate the materials used meet the requirements of ISO 10993.
## Nonclinical Testing:
Performance Bench Testing:
- 0 Air Leakage
- ହ Liquid Leakage
- 0 Force at Break
- 电 Recirculation
- 国 Flow vs. Pressure
- Priming Volume 0
In addition to the above testing, mechanical hemolysis testing was performed on the proposed device due to the change in blood flow path when compared to the predicate device. Furthermore, the proposed device was evaluated for MRI safety due to the presence of the stainless steel cannula.
#### Technological Characteristics:
The proposed device is intended for retrograde insertion whereas the predicate device is intended for antegrade insertion. To facilitate this retrograde insertion. the proposed device is packaged in a disassembled form to allow for insertion personnel to assemble the hub assembly once the lumen has been inserted and tunneled back to the insertion site.
In order to facilitate proper connection between the lumen and hub assembly. the hub assembly is outfitted with a stainless steel cannula that is not present in the predicate device. The sure fit of this cannula to lumen, followed by application of the compression ring and hub clamps, is essential to device integrity following the insertion and assembly procedure.
Once inserted and assembled, the principles of operation with regard to hemodialysis are the same as the predicate devices. There are no new questions raised regarding the safety or effectiveness of the device.
#### Summary of Substantial Equivalence:
In this submission, testing is provided to address the difference in technological characteristics between the proposed and predicate devices. Nonclinical testing and discussion of previously submitted information on the predicate devices establish that there are no new issues of safety or effectiveness for the proposed device.
The proposed device meets the performance criteria of design verification as specified by ISO standards, guidance documents and test protocols. The proposed device has the same intended use, operation and function as the predicates. There are no differences that raise new issues of safety and effectiveness. The proposed device is substantially equivalent to the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.