K130864 · Aat Alber Antriebstechnik GmbH · ILK · Feb 6, 2014 · Physical Medicine
Device Facts
Record ID
K130864
Device Name
C-MAX+ (PLUS)
Applicant
Aat Alber Antriebstechnik GmbH
Product Code
ILK · Physical Medicine
Decision Date
Feb 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Powered patient stairway chair lift; Intended for use in mitigating mobility impairment caused by injury or other disease by moving a person up and down a stairway.
Device Story
Powered patient stairway chair lift; transports patients up/down stairs. Input: user-controlled speed/direction via handle-mounted controls. Operation: battery-powered stair-climbing mechanism; attendant-operated; features single-step or fluent climbing modes; automatic brakes for safety. Output: physical movement of patient between levels. Used in interior/exterior settings; operated by care attendant. Benefits: reduces physical exertion for attendant; provides safe mobility for patients with ambulatory impairments. Device is modular, breaking into 3 parts for transport.
Clinical Evidence
Bench testing only. Testing performed per EN 12182 (general requirements), ISO 7176-9 (climatic), ISO 7176-14 (power/control systems), ISO 7176-16 (flammability), ISO 7176-21 (EMC), ISO 7176-23 (stair-climbing safety), and ISO 10993-5 (in vitro cytotoxicity).
Technological Characteristics
Reinforced rigid metal profile frame; battery-powered (24V DC motor); attendant-operated control system with adjustable speed and climbing direction; automatic braking system. Dimensions: 1490-1775mm height, 527mm width. Weight: 79.3kg. Max load: 300kg. Biocompatibility per ISO 10993-5.
Indications for Use
Indicated for disabled seated persons with ambulatory impairments, including spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, and rheumatism, to move between levels.
Regulatory Classification
Identification
Powered patient stairway chair lifts: A powered patient stairway chair lift is a motorized lift equipped with a seat and permanently mounted in one location that is intended for use in mitigating mobility impairment caused by injury or other disease by moving a person up and down a stairway. All other powered patient transport: A powered patient transport is a motorized device intended for use in mitigating mobility impairment caused by injury or other disease by moving a person from one location or level to another, such as up and down flights of stairs (e.g., attendant-operated portable stair-climbing chairs). This generic type of device does not include motorized three-wheeled vehicles or wheelchairs.
Special Controls
(i) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of American Society of Mechanical Engineers (ASME) A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate that the safety controls are adequate to prevent a free fall of the chair in the event of a device failure;
(ii) Appropriate analysis and nonclinical testing must demonstrate the ability of the device, including armrests, to withstand the rated load with an appropriate factor of safety;
(iii) Appropriate restraints must be provided to prevent the user from falling from the device (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”);
(iv) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized editions of AAMI/ANSI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Safety—Collateral Standard: Electromagnetic Compatibility—Requirements and Tests,” and ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must validate electromagnetic compatibility and electrical safety; and
(v) Appropriate analysis and nonclinical testing must demonstrate the resistance of the device upholstery to ignition.
{0}------------------------------------------------
FFB - 6 2014
Image /page/0/Picture/1 description: The image shows a logo with the letters "AAT" in bold, white font. The letters are positioned at the bottom of a black square with a rounded corner at the top right. The background is white, providing a strong contrast to the black square and white letters.
## k130864
# 510(k) SUMMARY
February 4, 2014
Office of Device Evaluation U.S. Food & Drug Administration
In accordance with Section 807.92 (c), is hereby made the 510(k)-summary for our device "c-max+ (plus)"P, that we plan to introduce into interstate commercial distribution.
| Applicant / 510(k)-owner: | AAT Alber Antriebstechnik GmbH |
|---------------------------|--------------------------------|
| | Ehestetter Weg 11 |
| | D-72458 Albstadt-Ebingen |
| | Phone: Tel. +49.7431.1295-0 |
| | Fax +49.7431.1295-35 |
| | Email: info@aat-online.de |
#### Organization Number: 239600
Contact Person: Mrs. Stefanie D. Bankston BEO MedConsulting Berlin GmbH 3001 Ferndale Dr. League City TX 77573 Phone: 713-483 46 17 email: s.bankston@beoberlin.com
#### Devices Name:
- Proprietary name: c-max+ (plus) a.
- Common Name: Powered patient stairway chair lift b.
- Classification Name: 21 CFR 890.5150 Powered patient transport C.
- Device Class: II ರ.
- Classification Panel: Physical Medicine e.
- f. Product Code: ILK
#### ldentification of the legally marketed device to which we claim equivalence:
The c-max+ (plus) is substantial equivalent in intended use, design and function to the c-max by AAT Alber Antriebstechnik GmbH (K103122).
#### Device Description:
With the c-max+ people can transport a patient safely and comfortably up- and downstairs. The maximum load is 300kg. The device is safe due to automatic brakes and a rigid frame. It is suitable for almost all kinds of stairs, winding stairs excluded. The climbing system requires low maintenance and care.
On stairs as well as on level ground C-Max+ proves its versatility. Removable arm rests make easy transferring from one chair to another possible. Due to its compact dimensions and foldable foot rest, the C-Max+ is easy to manoeuvre even on very narrow stair cases. For small, compact apartments with narrow doorways, the C-Max+ represents the ideal solution
Unsere Hausanschrift
AAT Alber Antriebstechnik GmbH · Ehestetter Weg 11 · D-72458 Albstadt-Ebinger Tel. +49.7431.1295-0 · Fax +49.7431.1295-35 · info@aat-online.de · www.aat-online.de Postfach 100 560 · D-72426 Albstadt-Ebingen
Amtsgericht Stuttgart HRB 1006 . Sitz: Albstadt . Ust ID Nr. DE 173669635 Geschäftsführer: Thomas Alber, Markus Alber
Unsere Bankverbindungen Sparkasse Zollernalb · Konto 31712831 · BLZ 65351260 Swift Code: SOLADES1BAL IBAN: DE06 6535 1260 0031 7128 31
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters "AAT" in white against a black background. The logo is a square with rounded corners. The letters "AAT" are stacked vertically, with "AA" on top and "T" on the bottom. The letters are in a bold, sans-serif font. The logo is simple and modern.
Scientific Concept: At the heart of the matter lies the patented climbing mechanism, which virtually climbs stairs all by itself. It also makes the C-Max+ particularly versatile and safe.
Function: Once the patient has seated and fastened the seat belt, the care attendant needs little physical power to handle the C-Max+. The C-Max+ is individually adjustable. Wheels make the transport on the floor easy. The battery-powered stair-climbing mechanism is usercontrolled with adjustable speed and climbing-direction. The attendant can also choose a single-step-mode alternatively to fluent climbing. A safety-brake ensures safe stops during the stair-climbing.
Image /page/1/Picture/3 description: The image is a composite of six different views of a stair climbing wheelchair. The top left image shows a person sitting in the wheelchair on a flat surface. The top middle image shows a person sitting in the wheelchair on a set of stairs. The top right image shows a close-up of the wheelchair's handle and controls. The bottom three images show close-ups of the wheelchair's wheels and suspension system as it navigates stairs.
C-Max+ ascends:
Unsere Hausanschrift
AAT Alber Antriebstechnik GmbH · Ehestetter Weg 11 · D-72458 Albstadt-Ebingen Tel. +49.7431.1295-0 · Fax +49.7431.1295-35 · info@aat-online.de · www.aat-online.de Postfach 100 560 · D-72426 Albstadt-Ebingen
Unsere Bankverbindungen Sparkasse Zollernalb · Konto 31712831 · BLZ 65351260 Swift Code: SOLADES1BAL IBAN: DE06 6535 1260 0031 7128 31
Amtsgericht Stuttgart HRB 1006 · Sitz: Albstadt · Ust ID Nr. DE 173669635 Geschäftsführer: Thomas Alber, Markus Alber
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters "AAT" in white against a black background. The background is a square with a rounded corner at the top right. The letters are stacked vertically, with "A" on top, followed by another "A", and then "T" at the bottom.
## C-Max+ descends:
Image /page/2/Picture/2 description: The image contains three panels showing different parts of a machine. The first panel shows a wheel with an arrow pointing to the left. The second panel shows a handle with a control panel. The third panel shows a wheel with a support leg.
Image /page/2/Picture/3 description: This image shows the features of a device, with labels and numbers pointing to different parts. The device has an adjustable handle, an up/down switch, speed control, and an on/off switch. Other features include a quick pin, a port for a magnetic charger system, a helix cable, and a seat belt system. The image also shows the locking pin for the foot rest, the seat unit, the release lever, the foot belt, and the foot rest.
#### Intended use:
Powered patient stairway chair lift; Intended for use in mitigating mobility impairment caused by injury or other disease by moving a person up and down a stairway.
Unsere Hausanschrift
AAT Alber Antriebstechnik GmbH · Ehestetter Weg 11 · D-72458 Albstadt-Ebingen Tel. +49.7431.1295-0 · Fax +49.7431.1295-35 · info@aat-online.de · www.aat-online.de Postfach 100 560 · D-72426 Albstadt-Ebingen
Unsere Bankverbindungen Sparkasse Zollernalb · Konto 31712831 · BLZ 65351260 Swift Code: SOLADES1BAL IBAN: DE06 6535 1260 0031 7128 31
3
Amtsgericht Stuttgart HRB 1006 · Sitz: Albstadt · Ust ID Nr. DE 173669635 Geschäftsführer: Thomas Alber, Markus Alber
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters "AAT" in bold, white font. The letters are positioned in the lower-left corner of a black square. The upper-right corner of the square is rounded, giving it a modern and stylized appearance.
#### Indication for use:
Support for disabled seated persons i.e. with ambulatory impairments, including people with spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc. to move from one level to another.
# Performance Standards:
Non-clinical tests were performed to determine substantial equivalence. The tests were performed according to recognized standards for powered patient stairway chair lifts. These are the applicable standards:
- To demonstrate the device's general safety, we performed non-clinical tests . according to EN 12182: Technical aids for disabled persons-general requirements and test methods.
- . To demonstrate the device's safe function under different standardized climatic conditions, we performed non-clinical tests according to ISO 7176-9: Wheelchairs -Part 9: Climatic tests for electric wheelchairs.
- . To demonstrate the device's electrical safety, we performed non-clinical tests according to ISO 7176-14: Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters -Requirements and test methods
- To demonstrate the device's safety concerning flammability, we performed non-. clinical tests according to ISO 7176-16: Wheelchairs - Part 16: Resistance to ignition of upholstered parts -- Requirements and test methods
- . To demonstrate the device's electromagnetic-compatibility, we performed non-clinical tests according to ISQ 7176-21: Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
- To demonstrate the device's mechanical and functional safety, we performed non-. clinical tests according to ISO 7176-23: Wheelchairs - Part 23: Requirements and test methods for attendant-operated stair-climbing devices
- . To demonstrate the device's biocompatibility, we performed non-clinical tests according to ISO 10993-5: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
Unsere Hausanschrift
AAT Alber Antriebstechnik GmbH · Ehestetter Weg 11 · D-72458 Albstadt-Ebingen Tel. +49.7431.1295-0 · Fax +49.7431.1295-35 · info@aat-online.de · www.aat-online.de Postfach 100 560 · D-72426 Albstadt-Ebingen
Unsere Bankverbindungen Sparkasse Zollernalb · Konto 31712831 · BLZ 65351260 Swift Code: SOLADES1BAL IBAN: DE06 6535 1260 0031 7128 31
Amtsgericht Stuttgart HRB 1006 - Sitz: Albstadt · Ust ID Nr. DE 173669635 Geschäftsführer: Thomas Alber, Markus Alber
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters "AAT" in bold, white font against a black square. The top-right corner of the square is rounded. The letters are stacked vertically, with "AA" on top and "T" below. The logo has a simple, modern design.
## Comparison to legally marketed device (Substantial Equivalence):
The c-max+ (plus) is essentially equivalent in intended use, design and function to the c-max by AAT Alber Antriebstechnik GmbH (K103122).
| | c-max+ (plus) | c-max (K103122)<br>Predicate device |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for use | Same but the use on winding<br>stairs is excluded | support for disabled seated<br>persons e.g. with ambulatory<br>impairments, including people<br>with spinal cord injury, spina<br>bifida, cerebral palsy, multiple<br>sclerosis, muscular dystrophy,<br>polio, rheumatism, etc. to move<br>from one level to another. |
| Manual wheelchair<br>accommodation | same | Integrated seating-unit. No<br>wheelchair adaption |
| Accessories /<br>Options | Head rest, seat belts, different<br>footrests; power charger for the<br>car, magnetic charging system | Head rest, seat belts, different<br>footrests; power charger for the<br>car |
| Stairway<br>Configuration | Straight and U-turn curves.<br>Up to 8" max. step-height.<br>Stairs must be in good condition.<br>Interior/exterior. Most stairway<br>material types | Curves, spiral turns.<br>Up to 9" max. step-height.<br>Stairs must be in good condition.<br>Interior/exterior. Most stairway<br>material types |
| Level of operator<br>difficulty | same | Requires some coordination.<br>Depends on rider weight, number<br>of stairs and operator ability.<br>Automatic braking. |
| Portability for<br>Travel | Breaks down into 3 parts,<br>heaviest part is 110 lbs. | Breaks down into 4 parts,<br>heaviest part is 37 lbs. |
| | Statement: due to its ability to<br>carry a higher load the frame<br>design is more rigid. | |
| Performance<br>Max. Load | 300 kg | 140 / 160 kg |
| steps | Approx. 300 | Approx. 375 |
| speed | 3-8 steps / minute<br>continually adjustable | 8 - 23 steps / minute<br>continually adjustable |
| | Statement: due to its ability to<br>carry a higher load, it is slower<br>than the p.d. | |
| Frame/chassis | Reinforced rigid metal profile | Rigid metal profile |
| Technical data | | |
| Height | 1490 - 1775 mm | 1090 - 1400 mm |
| Width | 527 mm | 440 - 485 mm |
| Total weight | 79.3 kg | 33,4 kg |
| Power supply | | |
| | | rechargeable |
| battery charger | same | (input) 90-240 V AC<br>(output) 24 V DC |
| motor | same | 24V / 275 W / DC |
### The Chart below summarizes the similarities and differences:
Unsere Hausanschrift
AAT Alber Antriebstechnik GmbH · Ehestetter Weg 11 · D-72458 Albstadt-Ebingen Tel. +49.7431.1295-0 · Fax +49.7431.1295-35 · info@aat-online.de · www.aat-online.de Postfach 100 560 · D-72426 Albstadt-Ebingen
Unsere Bankverbindungen Sparkasse Zollernalb · Konto 31712831 · BLZ 65351260 Swift Code: SOLADES1BAL IBAN: DE06 6535 1260 0031 7128 31
5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "AAT" in bold, white font against a black background. The background is a square with the upper right corner rounded off. The letters are stacked vertically, with "AA" on top and "T" below. The logo is simple and modern, with a focus on the acronym.
Analysis on the differences:
- . The main difference between the subject device C-Max+ und the predicate device C-Max is its max. load. The C-Max+ can carry twice as much as the predicate device. Due to this higher performance, the C-max+ (plus) has a reinforced structure that makes him stronger and slightly heavier and slower.
- Due to its ability to carry a higher load, it is slower than the predicate device. The . speed difference doesn't cause any safety problems.
- . The fact that the c-max+ total weight is way higher than the c-max's one is only due to the possibility of being able to carry a higher max. Ioad. The frame is more rigid. The weight difference doesn't cause any safety problems.
#### Quality Assurance and Manufacturing Controls:
AAT Alber Antriebstechnik GmbH operates to an established and certified quality management system according to ISO 9001, and ISO 13485.
#### Conclusion:
The subject device C-Max+ is as safe, as effective, and performs as well as the predicate device but with a higher max. load. The performed non-clinical tests demonstrate the safety of the subject device's performance, electricity, electromagnetic-compatibility, mechanical strength and durability, flammability and biological/toxicological aspects.
Sincerely.
## AAT Alber Antriebstechnik GmbH
**Michael** Vent
p.p. Stefanie D. Bankston Official Correspondent for AAT Alber Antriebstechnik GmbH
SD
Unsere Hausanschrift
AAT Alber Antriebstechnik GmbH . Ehestetter Weg 11 . D-72458 Albstadt-Ebingen Tel. +49.7431.1295-0 · Fax +49.7431.1295-35 · info@aat-online.de · www.aat-online.de Postfach 100 560 - D-72426 Albstadt-Ebingen
Unsere Bankverbindungen Sparkasse Zollernalb · Konto 31712831 · BLZ 65351260 Swift Code: SOLADES1BAL IBAN: DE06 6535 1260 0031 7128 31
Amtsgericht Stuttgart HRB 1006 · Sitz: Albstadt · Ust ID Nr. DE 173669635 Geschäftsführer: Thomas Alber, Markus Alber
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human services.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2014
AAT Alber Antriebstechnik GmbH c/o Stefanie D. Bankston BEO MedConsulting Berlin GmbH Texas Office 3001 Ferndale Dr. League City. TX 77573
Re: K130864
Trade/Device Name: c-max+ (plus) Regulation Number: 21 CFR 890.5150 Regulation Name: Powered patient transport Regulatory Class: Class II Product Code: ILK Dated: December 3. 2013 Received: December 27, 2013
# Dear Ms. Bankston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{7}------------------------------------------------
Page 2 - Stefanie D. Bankston
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Joyce M. Whang -S
for Carlos Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known) K130864
Device Name c-max+ (plus)
#### Indications for Use (Describe)
The product c-max+ (plus) offers motorized stair-climbing support for disabled seated persons e.g. with ambulatory impairments, including people with spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc. to move from one level to another.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FO ﻟﻠﻘﺎﻧﻮﻥ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤ Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.