CAVITRON PROPHY-JET AIR POLISHING PROPHYLAXIS SYSTEM
Applicant
Dentsply Intl., Inc.
Product Code
ELC · Dental
Decision Date
Jul 23, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System is designed for use in general prophylaxis and periodontal treatments and other areas of operative dentistry. It can also be used for: - Stain removal, - Prophylaxis of orthodontic patients, - Preparing tooth surfaces prior to bonding and sealant procedures, and - Implant debridement
Device Story
Air polishing unit delivering air/water/powder mixture to tooth enamel via insert tip; polishes without direct contact. Uses water-soluble prophy powders (sodium bicarbonate or aluminum trihydroxide). Features: Tap-On™ wireless foot pedal, prophy mode auto cycles, illuminated diagnostic display, automated purge function, detachable sterilizable handpiece, swivel cable with lavage control. Used in dental clinics by professionals. Connects to external air/water lines and AC power. Output: controlled spray for cleaning/polishing. Benefits: efficient removal of stains/deposits, preparation for dental procedures, and safe implant maintenance.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including stain removal, enamel abrasion, functional verification, and electrical safety/EMC testing (IEC 60601-1, IEC 60601-1-2).
Technological Characteristics
Air polishing system; utilizes water-soluble powders. Features wireless/wired foot pedal (two-position: rinse/polish), detachable autoclavable handpiece, and lavage control. Connectivity: external air/water lines, AC power. Software: embedded firmware for auto-cycles, purge function, and diagnostic display. Standards: IEC 60601-1, IEC 60601-1-2, ISO 14971 (risk management).
Indications for Use
Indicated for general prophylaxis, periodontal treatments, and operative dentistry, including stain removal, orthodontic patient prophylaxis, tooth surface preparation for bonding/sealants, and implant debridement.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
Predicate Devices
Dentsply International Cavitron® SPS Ultrasonic Scaler and Prophy-Jet® (K970342)
Dentsply International Cavitron® RF Ultrasonic Scaler with Sterimate Handpiece (K052334)
EMS Electro Medical Systems, S.A. EMS Air-Flow Master Piezon (K110173)
{0}------------------------------------------------
K) 30862
## SECTION 5- 510(k) SUMMARY
Cavitron® Prophy-Jet® Air Polishing Prophylaxis System
Image /page/0/Picture/4 description: The image shows the word "DENTSPLY" in a bold, sans-serif font. The letters are all capitalized and black. The word appears to be a logo or brand name. The font is slightly distressed, giving it a textured appearance.
### DENTSPLY Internationa
World Headquarters Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405-0872 1800) 877-0020 Fax (717) 849-4343 ww.dentsply.com
- Submitter Information: 1.
DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405
JUL 2 3 2013
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-849-4229 |
| Fax Number: | 717-849-4343 |
March 19, 2013 Date Prepared:
#### 2. Device Name:
- Proprietary Name: Cavitron® Prophy-Jet® Air Polishing Prophylaxis System .
- Classification Name: Ultrasonic scaler .
- CFR Number: 872.4850 o
- Device Class: II o
- Product Code: ELC o
#### 3. Predicate Device:
The proposed Cavitron® Prophy-Jet® Air Polishing Prophylaxis System is substantially equivalent to the following legally marketed devices:
- Dentsply International Cavitron® SPS Ultrasonic Scaler and Prophy-Jet®, . K970342, product code ELC.
- Ultrasonic Scaler with Sterimate International Cavitron® RF Dentsply . Handpiece, K052334, product code ELC.
- EMS Electro Medical Systems, S.A. EMS Air-Flow Master Piezon, K110173, . Product Codes ELC, EFB, EJR
- Satelec Air-N-Go, K110379, Product Code EFB .
The following accessories used with the proposed Cavitron® Prophy-Jet® Air Polishing Prophylaxis System were previously cleared by FDA:
- Cavitron® PROPHY-JET® Sodium Bicarbonate Prophy Powder (K970342) .
- Cavitron® JET-Fresh® Aluminum Trihydroxide Prophy Powder (K014188) .
- Cavitron® JET-Mate™ detachable sterilizable handpiece (K023697) .
- Cavitron® Jet Air Polishing Insert (K041141) .
{1}------------------------------------------------
#### 4. Description of Device:
The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System is an air polishing unit that delivers an air/water/powder mixture at the air polishing insert tip that polishes the tooth enamel without direct contact of the device. The system is used with water-soluble prophy powders such as Cavitron® PROPHY-JET® Sodium Bicarbonate Prophy Powder (K970342) or Cavitron® JET-Fresh® Aluminum Trihydroxide Prophy Powder (K014188) (for patients with low sodium requirements). Updated design features on the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System include a Tap-On™ Wireless Foot Pedal, prophy mode auto cycles, an illuminated diagnostic display, automated purge function, Cavitron® JET-Mate™ detachable sterilizable handpiece (K023697), and a swivel handpiece cable with more precise lavage control. Upon installation, the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System unit is connected to external air and water lines, as well as an AC power cord.
#### Intended Use: 5.
The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System is designed for use in general prophylaxis and periodontal treatments and other areas of operative dentistry. It can also be used for:
- Stain removal, .
- Prophylaxis of orthodontic patients, .
- Preparing tooth surfaces prior to bonding and sealant procedures, and .
- . Implant debridement
#### Identification of Risk Analysis Method: 6.
Risk analysis was performed on the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System utilizing an FMEA process based on ISO 14971. The results of the risk analysis performed on the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System concluded that all device design controls and process controls will be able to mitigate known potential failures and effects. In addition, performance testing and electrical safety testing were performed to mitigate other potential risks.
#### Description of Safety and Substantial Equivalence: 7.
#### Technological Characteristics: 7.1
The technological characteristics of the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System are very similar to the predicate devices; Cavitron® SPS Ultrasonic Scaler and Prophy Jet (K970342) and Cavitron® RF Ultrasonic Scaler with Sterimate Handpiece (K052334). All devices operate by the use of a foot pedal. The foot pedal in the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System and the predicate Cavitron® RF Ultrasonic Scaler with Sterimate Handpiece (K052334) can be connected to the system with a wire or by wireless means. The foot pedals of the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System and the predicate Cavitron® SPS Ultrasonic Scaler and Prophy Jet (K970342) have a two position operation. The first position is for rinsing, and second position is for air polishing. The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System and the predicate Cavitron® SPS Ultrasonic Scaler and
{2}------------------------------------------------
Prophy Jet (K970342) both utilize the same water-soluble air polishing prophy powders and air polishing insert. The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System and the predicate Cavitron® RF Ultrasonic Scaler with Sterimate Handpiece (K052334) both have a handpiece which can be removed to autoclave. Both the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System and the predicate Cavitron® SPS Ultrasonic Scaler and Prophy Jet (K970342) have a selector pointer on the top of the powder bowl which determines the amount of prophy powder which can be delivered to the patient. Both systems require the same input air and water pressure to function correctly. The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System and the predicate Cavitron® RF Ultrasonic Scaler with Sterimate Handpiece (K052334) have lavage control on the handpiece to select a water flow rate for system operation,. All lavage controls deliver the same range of water flow to the patient.
The proposed Cavitron® Prophy-Jet® Air Polishing Prophylaxis system has added indications for stain removal, prophylaxis of orthodontic patients, preparing tooth surfaces prior to bonding and sealant procedures, and implant debridement. Although these indications are not included in the cleared indications of the air polishing predicate Cavitron® SPS Ultrasonic Scaler and Prophy Jet (K970342), they are cleared indications for newer air polishing devices. Specifically, the Satelec Air-N-Go dental prophylaxis handpiece (cleared under premarket notification K110379) and the EMS Medical Systems Air-Flow Master Piezon (cleared under premarket notification K110173) include indications for the removal of, "soft deposits and areas of discoloration", the preparation of teeth for, "conventional dental procedures such as the placement of composite fillings", "plaque and stain removal for orthodontic patients", and "cleaning of implant fixtures" and abutments. There are also multiple literature references that make a case for the safety of cleaning implants using an air polishing device.
#### Non-Clinical Performance Data: 7.2
Performance testing focused on verification of design, stain removal, enamel abrasion, function and safety of the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System. Below is a summary of the testing performed:
- IEC 60601-1-2 Medical Electrical Equipment PART 1-2: General . Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility (2007)
- IEC 60601-1 Medical Electrical Equipment Part 1:General requirements . for basic safety and essential performance (2005)
- FDA -- Guidelines for the Content of Premarket Submissions for Software . Contained in Medical Devices - Software Validation
- Internal specification and testing for handpiece operation, including lavage . control
- Internal specification and testing for autocycle operation .
- Internal specification and testing for indicator lights and user interface . functionality
- Internal specification and testing for radio frequency distance for foot . pedal
- Internal specification and third party testing for stain removal and abrasion �
{3}------------------------------------------------
The results of these performance tests support the substantial equivalence of the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System ..
#### 7.3 Clinical Performance Data:
No human clinical data has been provided to support substantial equivalence.
#### Conclusion as to Substantial Equivalence: 7.4
The similarities in design, function, safety and intended use of the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System with the legally marketed predicate devices support substantial equivalence. The performance and safety data support the substantial equivalence of the Cavitron® Prophy-Jet® Air Polishing Prophylaxis System.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image contains the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## July 23, 2013
DENTSPLY International, Incorporated C/O Ms. Helen Lewis Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405-0872
Re: K130862
Trade/Device Name: Cavitron® Prophy-Jet® Air Polishing Prophylaxis System Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: June 19, 2013 Received: June 20, 2013
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Kwame Ulmer. M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# SECTION 4. INDICATIONS FOR USE STATEMENT
KI 30862 510(k) Number (if known):
Device Name: Cavitron® Prophy-Jet® Air Polishing Prophylaxis System
## Indications for Use:
The Cavitron® Prophy-Jet® Air Polishing Prophylaxis System is designed for use in general prophylaxis and periodontal treatments and other areas of operative dentistry. It can also be used for:
- . Stain removal,
- Prophylaxis of orthodontic patients, .
- Preparing tooth surfaces prior to bonding and sealant procedures, and ●
- Implant debridement .
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Susan Runner, DDS, MA
(Division Sign-Off) (Division Sign-off)
Division of Anesthestology, General Hospital Division of Antonious of the Devices
510(k) Number: K130862
Premarket Notification
Cavitron® Prophy-Ject® Air Polishing Prophylaxis System
DENTSPLY International
000010
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.