INTEL-GE CARE INNOVATIONS CONNECT RCM

K130821 · Intel-Ge Care Innovations, LLC · DRG · Aug 9, 2013 · Cardiovascular

Device Facts

Record IDK130821
Device NameINTEL-GE CARE INNOVATIONS CONNECT RCM
ApplicantIntel-Ge Care Innovations, LLC
Product CodeDRG · Cardiovascular
Decision DateAug 9, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.2910
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Intel-GE Care Innovations Connect RCM is intended to collect vital sign measurements from physiological measurement devices intended for use in the home. Patients can review the stored vital sign measurement information and receive educational and motivational content from caregivers. Patients can also engage in videoconferences with caregivers and answer the caregivers' questions by participating in surveys. Care Innovations Connect RCM is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. Care Innovations Connect RCM is contraindicated for patients requiring direct medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered. Care Innovations Connect RCM will be available for over the counter use.

Device Story

Connect RCM is a software application for home-based remote patient monitoring. It interfaces with commercially available physiological measurement devices (e.g., blood pressure monitors, weight scales, glucose meters, pulse oximeters) via a patient's COTS personal computing device. The system collects and transmits vital sign data to a clinician database; supports manual data entry; provides educational/motivational content; and facilitates videoconferencing and surveys between patients and caregivers. The software runs on a web server and utilizes a 'Device Connector' application on the patient's local platform. It is not interpretive, does not provide real-time data, and is not for diagnosis. Clinicians access the data via a database system to support patient management. The device benefits patients by enabling remote monitoring and communication with caregivers without requiring in-person visits for non-time-critical care.

Clinical Evidence

No clinical data. The device relies on bench testing and conformity to recognized consensus standards (AAMI/ANSI/IEC ES 60601-1) to demonstrate safety and efficacy.

Technological Characteristics

Software application for COTS platforms; interfaces with third-party physiological measurement devices via standard protocols. Connectivity: Networked/Cloud-based. Safety standards: AAMI/ANSI/IEC ES 60601-1. No hardware components.

Indications for Use

Indicated for patients in the home setting willing and capable of managing device use. Contraindicated for patients requiring direct medical supervision or emergency intervention.

Regulatory Classification

Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130821 pg 1 of 4 # 510(k) Summary As required by 21 CFR §807.92(c) #### Submitter | 510(k) Owner: | Intel-GE Care Innovations, LLC | |-----------------|----------------------------------------------| | Address: | 3721 Douglas Blvd, #100, Roseville, CA 95661 | | Telephone: | 916.847.7794 | | Contact Person: | Maureen Glynn | | Date Prepared: | March 20th 2013 | ### Device Information **AUG 09 2013** | Trade Name: | Intel-GE Care InnovationsTM Connect RCM | |----------------------|---------------------------------------------------------------------------------------------------------| | Common Name: | Remote Patient Monitoring System | | Classification Name: | Transmitters and Receivers, Physiological Signal,<br>Radiofrequency (21 CFR 870.2910, Product Code DRG) | Substantial Equivalence of intended use is claimed to the following Over the Counter devices: - 1. Intel-GE Care Innovations Guide (K130290) - Wellaho Personalized Outpatient Management System (K123671) 2. - 3. Vignet Telehealth Manager (K113446) ## Device Description Care Innovations™ Connect RCM is a software application for use with measurement devices commercially available for home use. The software executes on a web server and is accessed via a browser from the patient's COTS personal computing device. A small validated software application known as Device Connector runs on patients' home COTS platforms. Off the Shelf (OTS) software is also used with the internally developed software to provide functionality such as: setting & receiving email and text-based notifications, creating & editing calendar entries, playing Learn More videos, and holding a video conference with a clinician. Care Innovations™ Connect RCM is a software application for use with measurement devices commercially available for home use. Connect RCM provides the same client capabilities of collecting and transmitting patient data to the clinician database system as the predicate devices, and uses the existing clinician database system in the Intel-GE Care Innovations 10 Guide (K130290). No changes were required to the existing clinician database system to support the new client software. {1}------------------------------------------------ Patients can also enter measurement data manually entered data is stored in the backend clinician database as well as the Personal Health Data Record. It is flagged as manually entered data. ## Indications for Use Intel-GE Care Innovations Connect RCM is intended to collect vital sign measurements from physiological measurement devices intended for use in the home. Patients can review the stored vital sign measurement information and receive educational and motivational content from caregivers. Patients can also engage in videoconferences with caregivers and answer the caregivers' questions by participating in surveys. Care Innovations Connect RCM is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. Care Innovations Connect RCM is contraindicated for patients requiring direct medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered. Care Innovations Connect RCM will be available for over the counter use. ### Technological Characteristics Connect RCM is substantially equivalent to the technology predicate device, the Intel-GE Care Innovations Guide (K130290), in terms of data collection software functionality. communication method of patient's COTS platform with backend server, types of third-party physiological measurement devices which can be interfaced to the patient's COTS platform. implementation method of collecting data from third-party devices, third party device driver software, connectivity, communication protocol, and display method. #### Table 1: Peripherals Comparison between the intended use predicate devices and Connect RCM | Physiological<br>Parameter | Vignet Telehealth<br>Manager<br>(K113446) | Intel-GE Care<br>Innovations Guide<br>(K103290) | Intel-GE Care<br>Innovations Connect<br>RCM<br>(K130821) | |----------------------------|-------------------------------------------|-------------------------------------------------|----------------------------------------------------------| | Blood<br>Pressure | A&D UA-767-<br>PBT<br><br>A&D UA-851- | A&D UA-767PC<br>(K982481) | A&D UA-767PC<br>(K982481) | {2}------------------------------------------------ | | PBT | | | |---------------------------|------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------| | Weight | A&D UC-<br>321PBT | A&D UC-321PBT<br>(exempt) | A&D UC-321PBT<br>(exempt) | | | Omron Weight<br>Scale TD-3250C | A&D UC-321PL<br>(exempt) | A&D UC-321PL<br>(exempt) | | Blood<br>Glucose<br>Level | Entra Health<br>Systems MGH-<br>BT1 (K081703) | Bayer Diagnostics<br>Ascensia Breeze2<br>(K062347) | Bayer Diagnostics<br>Ascensia Breeze2<br>(K062347) | | | Fora Care, Inc.<br>Taidoc<br>Technology TD-<br>3250C | Bayer Diagnostics<br>Ascensia Contour<br>Blood Glucose<br>Monitoring System<br>(K062058) | Bayer Diagnostics<br>Ascensia Contour<br>Blood Glucose<br>Monitoring System<br>(K062058) | | | | LifeScan OneTouch<br>Ultra Family of Blood<br>Glucose Monitoring<br>Systems (K043197) | LifeScan OneTouch<br>Ultra Family of Blood<br>Glucose Monitoring<br>Systems (K043197) | | | | LifeScan OneTouch<br>Ultra 2 of Blood<br>Glucose Monitoring<br>Systems (K053529) | LifeScan OneTouch<br>Ultra 2 of Blood<br>Glucose Monitoring<br>Systems (K053529) | | Oxygen<br>Saturation | | Nonin 4100 Pulse<br>Oximeter (K043359) | Nonin 4100 Pulse<br>Oximeter (K043359) | | | | Onyx® II 9560<br>(K081285) | Onyx® II 9560<br>(K081285) | 。 {3}------------------------------------------------ | 510(k) Notification Submission - Abbreviated<br>Intel-GE Care Innovations™ Connect RCM | K130821<br>pg 4 of 4 | |----------------------------------------------------------------------------------------|----------------------| |----------------------------------------------------------------------------------------|----------------------| | Parameter | Intel-GE Care<br>Innovations™ Health<br>Guide (K130290) | Connect RCM<br>(K130821) | |--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | Software Safety Standard | AAMI/ANSI/IEC ES 60601-<br>1 Medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance | AAMI/ANSI/IEC ES 60601-<br>1 Medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance | | Hardware Safety Standard | UL 60950-1:2007<br>Information Technology<br>Equipment - Safety - Part 1:<br>General Requirements | N/A | #### Table 2: Main difference between Connect RCM meeting minimum specifications and the technology predicate device ## Safety and Efficacy Connect RCM does not rely on an assessment of clinical performance data. The device will conform to FDA 's recognized consensus standards and relies on its conformity to demonstrate the safety and efficacy. The device has the same functionality as the predicate device, the Intel-GE Care Innovations Guide (K130290), and introduces no new questions concerning the safety or efficacy and is, therefore, substantially equivalent to the predicate device. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its body and wings. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 9, 2013 Intel-GE Care Innovations, LLC c/o Ms. Maureen Glynn 3721 Douglas Blvd. #100 Roseville, CA 95661 Re: K130821 Trade/Device Name: Intel-GE Care Innovations Connect RCM Regulation Number: 21 CFR 870.2910 Regulation Name: Transmitters and Receivers, Physiological Signal, Radiofrequency Regulatory Class: Class II (two) Product Code: DRG Dated: July 8, 2013 Received: July 11, 2013 Dear Ms. Glynn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Ms. Maureen Glynn Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Owen P. Faris -S for Bram D. Zuckerman Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### 510(k) Notification Submission - Abbreviated Intel-GE Care Innovations TM LLC Intel-GE Care Innovations Connect RCM K130821 pg 1 of 1 ## Indications for Use: 510(k) Number: Device Name: Intel-GE Care Innovations™ Connect RCM Indications for Use: Intel-GE Care Innovations Connect RCM is intended to collect vital sign measurements from physiological measurement devices intended for use in the home. Patients can review the stored vital sign measurement information and receive educational and motivational content from caregivers. Patients can also engage in videoconferences with caregivers and answer the caregivers' questions by participating in surveys. Care Innovations Connect RCM is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required. Care Innovations Connect RCM is contraindicated for patients requiring direct medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered. Care Innovations Connect RCM will be available for over the counter use. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Digitally signed by Owen P. Faris -S Date: 2013.08.09 16:14:50-04'00'
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...