ISSDE is indicated for use with InterX Transcutaneous Electrical Nerve Stimulator, previously cleared by FDA (K042912). The ISSDE provides the conductive interface between the InterX stimulation device and the patient's skin. The ISSDE is not made with natural rubber latex, is reusable (single patient only), and is provided packaged and labeled as a sterile device, for single patient use. As a sterile device, the ISSDE can be applied to patients in a sterile post-operative healthcare setting. NRG only guarantees sterility for the first application, however, ISSDE can be repositioned and reapplied (single patient only) as a non-sterile device for subsequent applications. Refer to the labeling for reuse instructions.
Device Story
ISSDE is a flexible, sterile laminate electrode designed to interface with the InterX Transcutaneous Electrical Nerve Stimulator. It consists of a top cover, reinforcement film, dual conductive members with printed silver, biocompatible conductive hydrogel, and a 2-conductor cable. The device is applied to the patient's skin to deliver electrical stimulation. It is intended for single-patient, multiple-application use; provided sterile for the first application in post-operative settings and can be repositioned for subsequent applications. The electrode does not contain active electronics, software, or firmware. It facilitates the transmission of electrical signals from the stimulator to the patient, potentially aiding in pain management or therapeutic stimulation as determined by the clinician using the InterX system.
Clinical Evidence
Bench testing only. Performance evaluation included impedance testing at 1000 Hz, energy distribution uniformity, and biocompatibility (cytotoxicity, delayed-type hypersensitivity, and primary skin irritation per ISO 10993). Sterilization validation was performed per ANSI/AAMI/ISO 11137. Results confirmed impedance and energy distribution are substantially equivalent to predicate devices.
Technological Characteristics
Flexible laminate electrode; materials include white polyethylene foam top cover, silver-printed conductive members, biocompatible hydrogel, and 24 AWG tinned copper 2-conductor cable. Connectivity via touch-proof keyed connector. Sterilization via gamma radiation. No active electronics, software, or firmware.
Indications for Use
Indicated for use with InterX Transcutaneous Electrical Nerve Stimulator as a conductive interface between the device and patient skin. For single-patient use; latex-free; provided sterile for initial application in post-operative settings.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
Predicate Devices
Superior StarBurst TENS/NMES/FES Stimulating Electrodes (K083350)
Sterile Post Operative Electrodes Self-Adhering (K900997)
Reusable and Self-Adhering TENS.NMES.FES Stimulating Electrodes (Classic # 2404) (K900519)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for NRG. The logo is in black and white and features the letters "NRG" in a bold, sans-serif font. The letters are enclosed in a rounded rectangle. There is a small dot to the right of the letter G.
# Section 5 - 510(k) Summary
K1308.16
JUN 2 1 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.87(h) and 21 CFR 807.92.
G R O U P
Submitter Information: Neuro Resource Group, Inc. 1100 Jupiter Road, Suite 190 Plano, TX 75074 Establishment # 3004786509
Official Contact: Dave Turner, CEO Neuro Resource Group, Inc. 1100 Jupiter Road, Suite 190 Plano. TX 75074 Phone: 972-665-1810 X 223 Fax: 972-665-1814
# Date Summary Prepared: 06/20/13
Product Name & Classification: Trade Name: InterX Sterile Self-Adhesive Dual Electrode (ISSDE) Classification Name: Cutaneous Electrodes Classification Regulation: 21CFR 882.1320 Classification Code: GXY Classification Panel: Neurology
#### Predicate Device Information:
- 1. Device Name: Superior StarBurst TENS/NMES/FES Stimulating Electrodes Manufacturer: Covidien, LP 510(k) Reference: K083350
- 2. Device Name: Sterile Post Operative Electrodes Self-Adhering Manufacturer: Covidien 510(k) Reference: K900997 (510(k) holder: Katecho, Inc.)
- 3. Device Name: Reusable and Self-Adhering TENS.NMES.FES Stimulating Electrodes (Classic # 2404) Manufacturer: Covidien 510(k) Reference: K900519 (510(k) holder: Classic Medical Products, Inc.)
#### Product Description:
The Proposed ISSDE is a flexible, sterile laminate composed of a top cover, a reinforcement film, dual conductive members with printed silver, biocompatible conductive hydrogel, a release liner which protects the hydrogel before use and during storage and a 2-conductor cable secured between the reinforcement film and the conductive members. The electrodes do not contain active electronics, software or firmware. The cable connects the dual-electrode to the Inter X electrical stimulation device. A single finished dual-electrode is placed into a protective pouch
{1}------------------------------------------------
which is then sealed, gamma sterilized and boxed for shipment. The Proposed ISSDE is designed for single-patient/multiple application use.
### Intended Use:
ISSDE is indicated for use with InterX Transcutaneous Electrical Nerve Stimulator, previously cleared by FDA (K042912). The ISSDE provides the conductive interface between the InterX stimulation device and the patient's skin. The ISSDE is not made with natural rubber latex, is reusable (single patient only), and is provided packaged and labeled as a sterile device, for single patient use. As a sterile device, the ISSDE can be applied to patients in a sterile post-operative healthcare setting. NRG only guarantees sterility for the first application, however, ISSDE can be repositioned and reapplied (single patient only) as a non-sterile device for subsequent applications. Refer to the labeling for reuse instructions.
### Indications for Use:
ISSDE is indicated for use with InterX Transcutaneous Electrical Nerve Stimulator. The ISSDE provides the conductive interface between the InterX stimulation device and the patient's skin. The ISSDE is not made with natural rubber latex, is reusable (single patient only), and is provided packaged and labeled as a sterile device, for single patient use.
#### Technological Characteristics:
The Proposed ISSDE exhibits identical technological characteristics as compared to the following predicates:
- Covidien's Superior StarBurst TENS/NMES/FES Stimulating Electrodes (K083350) -
- Covidien's Sterile Post Operative Electrodes Self-Adhering (K900997) "
- Covidien's Reusable and Self-Adhering TENS.NMES.FES Stimulating Electrodes . (Classic # 2404) (K900519)
The electrode is composed of a top cover, a reinforcement film, dual conductive members with printed silver, biocompatible conductive hydrogel, a release liner which protects the hydrogel before use and during storage and a 2-conductor cable secured between the reinforcement film and the conductive members.
# Bench Testing:
The InterX Sterile Self-Adhesive Dual Electrode (ISSDE) and the predicate devices were tested per protocol in accordance with the FDA's "Draft Guidance for Industry and Staff: Class II Special Controls Guidance Document: Cutaneous Electrode (April 5, 2010)". ISSDE gel electrodes were sterilized to a very high radiation level (greater than 40 kGy) to create a worst case scenario for gel degradation. Electrical performance of the ISSDE was tested for uniform energy distribution along the length of each electrode. Comparison of energy distribution of the ISSDE electrodes were made to FDA cleared predicate gel electrode devices. Another performance factor of particular interest was impedance, which was measured across the total surface area of each ISSDE gel electrode. The results from the tests performed provide evidence that design verification per 21CFR-820.30(g) was met.
ISSDE complies with the applicable requirements of the following standards:
| No. | Standard Reference & Revision | Title | Deviations |
|-----|-------------------------------|---------------------------------------------------------------------|------------|
| 1 | ISO 15223-1(Second Edition) | Medical Devices – Symbols to be<br>used with medical device labels, | None |
{2}------------------------------------------------
| | | labeling, and information to be | |
|---|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| | | supplied - Part 1: General | |
| | | requirements | |
| 2 | ISO 10993-1:1997 | Biological evaluation of medical<br>devices - Part 1: Evaluation &<br>Testing | None |
| 3 | ISO 10993-5:1999 | Biological evaluation of medical<br>devices - Part 5: Tests for In Vitro<br>Cytotoxicity | None |
| 4 | ISO 10993-10:1996 | Biological evaluation of medical<br>devices - Part 10: Tests for Irritation<br>and Delayed-type Hypersensitivity | None |
| 5 | ISO 11607-1:2006 (R) 2010 | Packaging for terminally sterilized<br>medical device- Part 1:<br>Requirements for materials, sterile<br>barrier systems and packaging<br>systems | None |
| 6 | ANSI/AAMI/ISO 11137-1:2006/(R)<br>2010 | Sterilization of health care products<br>- Radiation -Part 1: Requirements<br>for development, validation, and<br>routine control of a sterilization<br>process for medical devices | None |
| 7 | ANSI/AAMI/ISO 11137-2:2012 | Sterilization of health care products<br>- Radiation -Part 2: Establishing the<br>sterilization dose | None |
Guidance Document: The recommendations of the FDA's "Draft Guidance for Industry and Staff: Class II Special Controls Guidance Document: Cutaneous Electrode (April 5, 2010)" were taken into account in preparing this 510(k) submission.
### Performance Data:
Design validation of the ISSDE was confirmed and documented. Tests to measure impedance and variation of energy distribution, which determine safety and efficacy, were conducted to compare the ISSDE to three predicate devices.
Values of current were calculated from testing each electrode. Impedance values were tabulated for all gel electrodes using samples per electrode type. Impedance values for the ISSDE are confirmed to be substantially equivalent to predicate devices
Variation of energy distribution was also determined to be substantially equivalent, which confirms avoidance of hot spots.
The results of the impedance and variation of energy distribution tests conducted confirm that the ISSDE design is substantially equivalent to three FDA cleared predicate devices.
Performance data for the Proposed ISSDE is compared to that of the predicate devices mentioned above. Results from the biocompatibility study, physical and functional performance evaluation demonstrate that the Proposed ISSDE is substantially equivalent to the legally marketed devices.
{3}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
# Substantial Equivalence Table:
:
:
ה
:
:
| Substantial<br>Equivalence<br>Item | InterX Sterile<br>Self-Adhesive<br>Dual Electrode<br>(ISSDE)<br>(New Device) | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 |
|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterile | YES | NO | YES | NO |
| Intended Use | Sterile, post-operative,<br>reusable, self-adhesive,<br>stimulating electrode (single<br>patient) | Non-sterile, reusable, self-<br>adhesive, stimulating<br>electrode (single patient) | Sterile, post-operative, self-<br>adhesive, stimulating<br>electrode (single patient) | Non-sterile, reusable, self-<br>adhesive, stimulating and<br>recording electrode<br>(single patient) |
| Indications for<br>Use | ISSDE is<br>indicated for use<br>with InterX<br>Transcutaneous<br>Electrical Nerve<br>Stimulator. The<br>ISSDE provides<br>the conductive<br>interface<br>between the<br>InterX<br>stimulation<br>device and the<br>patient's skin.<br>The ISSDE is<br>not made with<br>natural rubber<br>latex, is reusable<br>(single patient<br>only), and is<br>provided<br>packaged and<br>labeled as a<br>sterile device,<br>for single patient<br>use. | The Superior<br>Starburst<br>Reusable Self-<br>adhering<br>TENS/NMES/F<br>ES Stimulating<br>Electrodes are<br>indicated for use<br>with<br>transcutaneous<br>electrical<br>stimulation<br>devices as non-<br>sterile, Latex<br>free, reusable<br>device for single<br>patient use only.<br>The electrodes<br>provide the<br>conductive<br>interface<br>between the<br>stimulation<br>device and the<br>patient's skin. | The Sterile Post<br>Op Self-<br>adhering TENS<br>Stimulating<br>Electrodes are<br>indicated for use<br>with<br>transcutaneous<br>electrical<br>stimulation<br>devices as<br>sterile, Latex<br>free for single<br>patient use only.<br>The electrodes<br>provide the<br>conductive<br>interface<br>between the<br>stimulation<br>device and the<br>patient's skin. | The Classic<br>Reusable Self-<br>adhering<br>TENS/NMES/F<br>ES Stimulating<br>Electrodes are<br>indicated for use<br>with<br>transcutaneous<br>electrical<br>stimulation<br>devices as non-<br>sterile, Latex<br>free, reusable<br>device for single<br>patient use only.<br>The electrodes<br>provide the<br>conductive<br>interface<br>between the<br>stimulation<br>device and the<br>patient's skin. |
| Hydrogel<br>Volume<br>Resistivity | 1500 ohm-cm<br>(max) | 2500 ohm-cm<br>(max) | 2500 ohm-cm<br>(max) | 2500 ohm-cm<br>(max) |
| Electrode Area<br>(surface area in<br>contact with<br>skin) | 67.58 cm2 | 50.23 cm2 | 57.64 cm2 | 47.47 cm2 |
| Substantial<br>Equivalence<br>Item | InterX Sterile<br>Self-Adhesive<br>Dual Electrode<br>(ISSDE)<br>(New Device) | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 |
| Hydrogel<br>Biocompatibility<br>ISO 10993-1 | Cytotoxicity<br>Test - Passed<br>Delayed Contact<br>Hypersensitivity<br>Test - Passed<br>Primary Skin<br>Irritation Test -<br>Passed | Cytotoxicity<br>Test - Passed<br>Delayed Contact<br>Hypersensitivity<br>Test - Passed<br>Primary Skin<br>Irritation Test -<br>Passed | Cytotoxicity<br>Test - Passed<br>Delayed Contact<br>Hypersensitivity<br>Test - Passed<br>Primary Skin<br>Irritation Test -<br>Passed | Cytotoxicity<br>Test - Passed<br>Delayed Contact<br>Hypersensitivity<br>Test - Passed<br>Primary Skin<br>Irritation Test -<br>Passed |
| | Primary Skin<br>Irritation Test -<br>Passed | Primary Skin<br>Irritation Test -<br>Passed | Primary Skin<br>Irritation Test -<br>Passed | Primary Skin<br>Irritation Test -<br>Passed |
| Hydrogel<br>Stainless Steel<br>Adhesion (180°<br>peel) | 136 grams<br>minimum | 136 grams<br>minimum | 180 grams<br>minimum | 136 grams<br>minimum |
| Conductive<br>Electrode<br>Element | Carbon film<br>with silver flood<br>coated over<br>entire surface | Carbon film<br>with silver flood<br>coated gradient<br>pattern | Carbon film<br>with silver flood<br>coated over<br>entire surface | Carbon film<br>with silver flood<br>coated over<br>entire surface |
| Electrode Top<br>Cover | Non-Conductive<br>White<br>Polyethylene<br>Foam, 31.25<br>mils thick | Non-Conductive<br>White Tricot | Non-Conductive<br>White<br>Polyethylene<br>Foam, 31.25<br>mils thick | Non-Conductive<br>White<br>Polyethylene<br>Foam, 31.25<br>mils thick |
| Sterilization<br>Method | Gamma<br>Sterilized | N/A | Gamma<br>Sterilized | N/A |
| Storage Liner | Polyester with<br>silicone release | Polyester with<br>silicone release | Silicone coated<br>bulk paper | Polyester with<br>silicone release |
| Electrode Lead<br>Wire | 2-conductor<br>cable, 24 AWG<br>tinned copper<br>conductors with<br>PVC jacket | Single conductor<br>24 AWG tinned<br>copper with<br>PVC jacket | Single conductor<br>24 AWG tinned<br>copper with<br>PVC jacket | Single conductor<br>24 AWG tinned<br>copper with<br>PVC jacket |
| Electrode<br>Packaging | Sealed Pouch | Sealed Pouch | Sealed Pouch | Sealed Pouch |
| Number per<br>Package | 1 | 4 | 2 | 4 |
| Electrode<br>Permanence | Disposable | Disposable | Disposable | Disposable |
| Patient<br>Preparation | Clean Dry Skin | Clean Dry Skin | Clean Dry Skin | Clean Dry Skin |
| Electrode<br>Placement | On Patient Skin<br>In Intended<br>Treatment Area | On Patient Skin<br>In Intended<br>Treatment Area | On Patient Skin<br>In Intended<br>Treatment Area | On Patient Skin<br>In Intended<br>Treatment Area |
| Latex Content | Not made with<br>natural rubber<br>latex | Latex Free | Latex Free | Latex Free |
| Substantial<br>Equivalence<br>Item | InterX Sterile<br>Self-Adhesive<br>Dual Electrode<br>(ISSDE)<br>(New Device) | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 |
| Electrode<br>Impedance @<br>1000 Hz | Less than 50<br>ohms | Less than 50<br>ohms | Less than 50<br>ohms | Less than 50<br>ohms |
| Wire<br>Attachment<br>Integrity (force<br>to remove lead<br>wire from<br>electrode) | More than 25<br>lbs. force | More than 10<br>lbs. force | More than 10<br>lbs. force | More than 15<br>lbs. force |
| Connector<br>Disengagement<br>Force | More than 2 lbs.<br>force | More than 2 lbs.<br>force | More than 1.75<br>lbs. force | More than 2 lbs.<br>force |
| Connector<br>Design | Overmolded<br>keyed connector<br>with 2-pole 1.0<br>mm sockets,<br>touch-proof | Overmolded<br>.080" single pole<br>socket, touch-<br>proof | .080" single pole<br>socket with heat<br>shrink covering,<br>touch-proof | Overmolded<br>.080" single pole<br>socket, touch-<br>proof |
| 21 CFR 898<br>Compatible? | Yes | Yes | Yes | Yes |
| Lead Wire to<br>Electrode<br>Attachment<br>Method | Cyanoacrylate<br>adhesive<br>between wire<br>insulation and<br>silver film,<br>tinned copper<br>wire in contact | Cyanoacrylate<br>adhesive<br>between wire<br>insulation and<br>silver film,<br>tinned copper<br>wire in contact | Cyanoacrylate<br>adhesive<br>between wire<br>insulation and<br>silver film,<br>tinned copper<br>wire in contact | Cyanoacrylate<br>adhesive<br>between wire<br>insulation and<br>silver film,<br>tinned copper<br>wire in contact |
.
·
·
.
{4}------------------------------------------------
.
.
:
:
·
,
{5}------------------------------------------------
# Conclusions:
The Proposed ISSDE is similar in intended use, functional design, principles of operation, materials, packaging and other technological characteristics to the predicate devices. A fter analyzing the performance and testing, it is concluded that the ISSDE is as safe and effective as the predicate devices and do not raise any new issues of safety and effectiveness.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 21, 2013
Neuro Resource Group, Inc. c/o David Turner, CEO Neuro Resource Group 1100 Jupiter Road, Ste. 190 Plano, TX 75074
Re: K130816
Trade/Device Name: InterX Sterile Self-Adhesive Dual Electrode (ISSDE) Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: March 13, 2013 Received: March 25, 2013
Dear Mr. Turner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class 11 (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{7}------------------------------------------------
# Page 2 - David Turner
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
hup://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
hup://www.fda.gov/MedicalDevices/Safetv/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Joyce M. Whang -S
Victor Krauthamer, Ph.D. for Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130816
InterX Sterile Self-Adhesive Dual Electrode (ISSDE) Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
ISSDE is indicated for use with InterX Transcutaneous Electrical Nerve Stimulator. The ISSDE provides the conductive interface between the InterX stimulation device and the patient's skin. The ISSDE is not made with natural rubber latex, is reusable (single patient only), and is provided packaged and labeled as a sterile device, for single patient use.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M. Whang -S
(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number K130816
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.