K130760 · Beijing Jinhengwei Technology Development Co., Ltd. · GEI · Sep 16, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130760
Device Name
ELECTROSURGICAL GENERATOR
Applicant
Beijing Jinhengwei Technology Development Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Electrosurgical Generator (HV-300A, HV-400) is a non-sterile, reusable multi-purpose electrosurgical generator that is designed to perform monopolar and bipolar functions in the operating arena.
Device Story
Electrosurgical generator (models HV-400, HV-300A) provides high-frequency current for monopolar and bipolar surgical procedures. Device comprises main unit with front panel (touchscreen, operation keys, indicators, accessory connectors), back panel (volume control, foot switch interface, power input), and internal control/display boards. Operated by clinicians in operating rooms; utilizes foot switch and touchscreen interface to control output parameters. Delivers HF energy to surgical accessories to perform cutting and coagulation. Benefits include versatile surgical energy delivery for various procedures. Device is non-sterile and reusable.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-2-2 (high-frequency surgical equipment) verified design specifications and substantial equivalence.
Technological Characteristics
HF electrosurgical generator; monopolar and bipolar output; 110-220V AC power; TFT touchscreen and foot switch interface. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2 standards.
Indications for Use
Indicated for use as a multi-purpose electrosurgical generator to perform monopolar and bipolar surgical functions in an operating room setting.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
K130760
#### Section 3 510 Summary
page 1 of (4)
# Section 3 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number:
- 1. Date of Submission: Mar. 15, 2013
SEP 16 2013
2. Sponsor Identification
> Beijing Jinhengwei Technology Development Co.,Ltd. East side on 2nd floor of south building, No.631 xingong village, Nanyuan area, Fengtai District, Beijing, China 100069
Establishment Registration Number: Not yet registered
Contact Person: Mr. Ligang Mao Position: General Manager Tel: +86-10-67916414 Fax: +86-10-67916414 Email: youyou1975@vip.sina.com
- Submission Correspondent 3.
Mr. Ligang Mao Beijing Jinhengwei Technology Development Co.,Ltd. East side on 2nd floor of south building,No.631 xingong village, Nanyuan area, Fengtai District, Beijing, China 100069 T: +86-10-67916414 F: +86-10-67916414 E: youyou1975@vip.sina.com
{1}------------------------------------------------
Section 3 510 Summary
KIK30760
Page 2 of (4)
- 4. Proposed Device Identification
Proposed Device Name: Electrosurgical Generator Proposed Device Model: HV-400, HV-300A Device Common Name: Electrosurgical Unit and Accessories
Regulatory Information: Classification Name:Electrosurgical Unit and Accessories: Classification: II; Product Code: GEI; Regulation Number: 21 CFR 878.4400; Review Panel: General& Plastic Surgery;
Intended Use Statement:
The Electrosurgical Generator (HV-300A, HV-400) is a non-sterile, reusable multi-purpose electrosurgical generator that is designed to perform monopolar and bipolar functions in the operating arena.
- Predicate Device Identification న.
510(k) Number: K072041 Predicate Device Name: Bovie IDS-400 Manufacturer: Bovie Medical
510(k) Number: K944602 Predicate Device Name: Force FX Manufacturer: Valleylab Inc.
- Device Description 6.
The proposed device (HV-400 and HV-300A) is an electrosurgical generator with, monopolar and bipolar, high frequency current isolation outputs.
The proposed device is independent equipment, which include generator and accessories. The main components are front panel ( power switch, touchable screen, operation key, indicator, connector for accessories), back panel (volume controller, foot switch interface, power cable outlet and fuse) and inner component (display panel, main board, control board, speaker etc.). The accessories include power cable, III-2
{2}------------------------------------------------
Section 3 510 Summary
K130760
#### Page 3 of (4)
cable and foot switch).
... . . . . .
The two models of proposed device share appearance, function, accessories, the only difference is the output parameter
- 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2005 Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic compatibility - requirements and tests
IEC 60601-2-2:2009 Medical electrical equipment Part 2-2: Particular requirements for safety of high frequency surgical equipment.
- 8. Substantially Equivalent (SE) Conclusion
The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
| Item | Proposed Device(s) | Predicate Device<br>(K072041) | Predicate Device<br>(K944602) |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|-------------------------------|
| Product Code | GEI | Same | Same |
| Regulation<br>Number | 878.4400 | Same | Same |
| Intended Use | The Electrosurgical Generator<br>(HV-300A, HV-400) is a<br>non-sterile, reusable multi-purpose<br>electrosurgical generator that is<br>designed to perform monopolar and<br>bipolar functions in the operating<br>arena. | Same | Similar |
| Type | HF energy | Same | Same |
| Output waveform | Monopolar and bipolar | Same | Same |
Table 3-1 Comparison of Technology Characteristics
111-3
{3}------------------------------------------------
. .
: . Section 3 510 Summary
K130760
page 4 of (4)
| Operation<br>system | Touchable TFT screen and<br>foot switch | Same | Same |
|---------------------|-----------------------------------------|------|------|
| Working voltage | 110-220V AC±10% | Same | Same |
| Safety | IEC 60601-1 | Same | Same |
| EMC | IEC 60601-1-2 | Same | Same |
| Performance | IEC 60601-2-2 | Same | Same |
| Features | LCD touchable screen,<br>foot switch | Same | Same |
The proposed device(s), Electrosurgical Generator (HV-300A, are determined to be Substantially Equivalent (SE) to the predicate device(s), Bovie IDS-400 and Force FX, in respect of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal for the Department of Health and Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three stylized lines forming the body and wings.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
Beijing Jinhengwei Technology Development Corporation, Ltd. Mr. Ligang Mao General Manager East side on 2nd Floor of south building, No.631 xingong village Nanyuan area, Fengtai District, Beijing, China 100069
September 16, 2013
### Re: K130760
Trade/Device Name: Electrosurgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 23, 2013 Received: August 1, 2013
Dear Mr. Mao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device fabeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Ligang Mao
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default,htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toil-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default,htm.
Sincerely yours,
Neil R Ogden .. 2013.09.26 14:29:58 -04'00' for Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### K130760
# Section II Indications for Use
510(k) Number: Device Name:
.
Indications for Use:
The proposed device is intended for use is:
The Electrosurgical Generator (HV-300A, HV-400) is a non-sterile, reusable multi-purpose electrosurgical generator that is designed to perform monopolar and bipolar functions in the operating arena.
& PRESCRIPTION USE (Part 21 CFR 801 Subpart D)
# O OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Long H. Chen-A
Page 1 of 1
(Division Sign-off)for MXM Division of Surgical Devices 510 (k) Number K130760
11-11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.