← Product Code [LKF](/productcode/LKF) · K130758

# KRONNER SIDE-KICK UTERINE MANIPULATOR HOLDER (K130758)

_Kronner Prototypes, Inc. · LKF · Dec 11, 2013 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K130758

## Device Facts

- **Applicant:** Kronner Prototypes, Inc.
- **Product Code:** [LKF](/productcode/LKF.md)
- **Decision Date:** Dec 11, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.4530
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The Kronner Side-Kick is intended to assist the surgical staff in mounting, positioning and holding a uterine manipulator during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment. Caution: Federal law restricts this device to sale by or on the order of a physician.

## Device Story

Kronner Side-Kick is a mechanical holder for uterine/vaginal manipulators used in laparoscopic/open gynecological surgery. Device mounts to standard OR tables; holds manipulators (e.g., RUMI, V-Care, Arch Koh) in position to reduce staff fatigue and free personnel. System consists of an arm assembly, gas supply module, and foot pedal. Gas supply module (connected to wall/portable nitrogen or compressed air) controls joint locking/unlocking. Foot pedal releases joints for positioning; releasing pedal locks joints. Device remains external to patient. Operated by trained OR personnel. Benefits include stable positioning of surgical instruments during procedures.

## Clinical Evidence

Bench testing only. Factory testing confirmed compatibility with Arch Koh, RUMI, and V-Care manipulators, ability to move the uterus through required surgical ranges, holding strength, and ease of assembly in an OR environment.

## Technological Characteristics

Materials: Aluminum and stainless steel. Energy source: Pneumatic (nitrogen or compressed air). Connectivity: None. Form factor: Articulated arm assembly with gas supply module and foot pedal. Sterilization: Not specified. Software: None.

## Regulatory Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

## Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

## Predicate Devices

- Coopers Uterine Positioning System ([K071405](/device/K071405.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## DEC 1 1 2013

#### K130758

# 510(K) SUMMARY

as required by 807.92

Dec. 6, 2013 Date:

Kronner Prototypes, Inc. Submitter: 1443 Upper Cleveland Rapids Road Roseburg, OR 97471 Tel: (541- 672-2543

ER Number: 3018984

Contact:

Richard F. Kronner M.D. President Telephone: (541) 672-2543 Fax: (541) 672-1074 E-mail: kronner@rosenet.net ﺗ ﺘﺴﺒﻪ ﻧ

Subject Device:

Kronner Side-Kick Uterine Manipulator Holder

Common Name:

Trade Name:

Holder, Positioner, Arm, Instrument Guide

Classification Name:

no known industry name for this device

Device Class:

Unclassified (LKF)

Kronner Prototypes, Inc. Kronner Side-Kick uterine manipulator Holder

{1}------------------------------------------------

### Predicate Device: Intuitive Surgical (K071405) marketed as: Coopers Uterine Positioning System by Coopersurgical K071405

#### Device Description:

The Kronner Side-Kick uterine manipulator holder is used to mount, position and hold

in position uterine or vaginal manipulators used in laparoscopic surgical procedures. It is made mostly from aluminum and stainless steel and is mounted to standard operating room tables and locked in position allowing surgical devices to be securely held in position for long periods of time. Use of the Sidek-Kick frees operating room staff for other activities and reduces fatigue associated with manually maintaining a device in position during surgical procedures.

The manipulator is inserted into the patient using standard technique, then the Side-Kick is joined to the manipulator.

With proper adapters the Side-Kick may be used with various manipulators, such as the RUMI, the V-Care and the Arch Koh,

A gas supply module, which can be attached to the operating table rail, transfers nitrogen or compressed air from the wall or a portable tank to the Side-Kick holder through two flexible lines. When the switch is turned on the Side-Kick joints are locked. When the switch is turned off the Side-Kick joints are unlocked.

Pressing a foot pedal releases all the Side-Kick joints for position changes. When the foot pedal is released the joints lock. The foot pedal is connected to the gas supply module through two lines covered by a protective hose.

A branched gas line connects the two luers of the Arm assembly to the gas supply module. A single gas line connects the gas supply module to the main pivot.

The Kronner Side-Kick remains external to the patient's body at all times.

#### Intended Use:

The Kronner Side-Kick is intended to assist the surgical staff in mounting, positioning and holding a uterine manipulator during gynecological laparoscopic surgical procedures. It is intended for use by trained operating room personnel in an operating room environment.

"Caution: Federal law restricts this device to sale by or on the order of a physician.

{2}------------------------------------------------

#### Comparison to predicate device:

Based on the comparison of design, technology, materials, manufacturing, performance, specifications, and method of use, the Kronner Side-Kick is substantially equivalent to the previously identified pre-amendment and 510(k) cleared predicate device.

#### Technological Characteristics:

The technological characteristics of the subject device are the equivalent to the predicate devices.

#### Performance data:

Design analysis and testing has been conducted to confirm that basic functional characteristics of the subject device is substantially equivalent to the predicate device cited, and that design output meets the design input requirements.

#### Conclusion:

Based upon available technical information, intended use and performance information provided in this pre-market notification, the Kronner Side-Kick described herein is substantially equivalent to current legally marketed pre-amendment and 510(k) cleared predicate devices.

### Kronner Sidekick Factory Testing

Kronner Sidekick and Sidekick uterine manipulator adapters are factory tested so they meet the following acceptance criteria:

1. That Arch Koh, RUMI, and V care manipulators can be attached to Sidekick adapters satisfactorally and can be used, as required, for the intended use .

2. That Arch Koh, RUMI, and V care manipulators will be able to move the uterus through a range required for the surgery.

3. That Arch Koh, RUMI, and V care manipulators will be held with the strength required for holding the uterus during the surgery.

4. That all Sidekick components will perform as required.

5. That all Sidekick components can be easily assembled in an operating room environment.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all capital letters. The text is in a bold, sans-serif font and is centered horizontally. The words are arranged on a single line and the text appears to be extracted from a document or a heading.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Kronner Prototypes, Inc. Richard F. Kronner, M.D. 1443 Upper Cleveland Rapids Road Roseburg, Oregon 97471

December 11, 2013

Re: K130758

Trade/Device Name: Kronner Side-Kick Uterine Manipulator Holder Regulatory Class: Unclassified Product Code: LKF Dated: October 23, 2013 Received: November 01, 2013

Dear Dr. Kronner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Richard F. Kronner, M.D.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Joshua C. Nipper -S

Binita Ashar, MD, MBA, FACS For Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## K130758

## Statement of Indications for Use

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: Kronner Sidekick Uterine Manipulator Holder

Indications For Use:

නියෝගයේ පොහොත් පැවැති ප්‍රධාන අවස්ථානයේ ප්‍රධාන බවට පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පිහිටි පි

For holding uterine manipulators during laparoscopic or open surgery on female pelvic organs.

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

#### CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)

| Prescription Use<br>(Per 21 CFR 801.109) | XX | OR | Over-The-Counter Use |
|------------------------------------------|----|----|----------------------|
|------------------------------------------|----|----|----------------------|

| Long H. Chen - A             | for BSA                  |
|------------------------------|--------------------------|
| (Division Sign-off)          | (Optional Format 1-2-96) |
| Division of Surgical Devices |                          |
| 510(k) Number: K130758       |                          |

5-1

---

**Source:** [https://fda-staging.innolitics.com/device/K130758](https://fda-staging.innolitics.com/device/K130758)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
