ENDOSCOPIC LINEAR CUTTING STAPLERS WITH SINGLE USE LOADING UNITS
K130738 · Nanjing Maidixin Medical Device Co., Ltd. · GDW · Dec 13, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130738
Device Name
ENDOSCOPIC LINEAR CUTTING STAPLERS WITH SINGLE USE LOADING UNITS
Applicant
Nanjing Maidixin Medical Device Co., Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Dec 13, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Endoscopic Linear Cutting Staple Use Loading Units have applications in general, gynecologic, pediatric and thoracic surgery for resection, and creation of anastomoses. They may be used for transection of liver substance, hepatic vasculature and biliary structures.
Device Story
Endoscopic linear cutting stapler with single-use loading units (SULU); manual operation; used in general, gynecologic, pediatric, and thoracic surgery. Device places four or six rows of titanium staples while simultaneously dividing tissue along a central line. Available in straight and articulating configurations. Reloadable up to 10 times per procedure. Includes safety mechanism (green button) to prevent misfiring. Used by surgeons in clinical settings to facilitate tissue resection and anastomosis creation; benefits include standardized staple placement and efficient tissue division.
Clinical Evidence
No clinical data. Bench testing only, including physical performance, endotoxin, package integrity, and shelf-life testing.
Technological Characteristics
Manual endoscopic linear stapler; titanium staples (ASTM F 67-06); stainless steel body; straight and articulating SULU configurations; closed staple heights 0.75-2.0 mm; irradiation sterilization (SAL 10^-6).
Indications for Use
Indicated for patients undergoing general, gynecologic, pediatric, or thoracic surgery requiring tissue resection, transection, or creation of anastomoses, including liver substance, hepatic vasculature, and biliary structures.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Endoscopic Linear Cutting Staplers with Single Use Loading Units (K120179)
Submission Summary (Full Text)
{0}------------------------------------------------
Section III 510(k) Summary
Project #: M0032012Aa
## DEC 1 3 2013
Proposed Device Identification 4.
> Proposed Device Name: Endoscopic Linear Cutting Staplers with Single Use Loading Units Device Common Name: Stapler
Regulatory Information: Classification Name: Staple, Implantable; Classification: II; Product Code: GDW: Regulation Number: 21 CFR 878.4750; Review Panel: General & Plastic Surgery;
### Intended Use Statement:
The Endoscopic Linear Cutting Staple Use Loading Units have applications in general, gynecologic, pediatric and thoracic surgery for resection, and creation of anastomoses. They may be used for transection of liver substance, hepatic vasculature and biliary structures.
#### Predicate Device Identification 5.
510(k) Number:K 120179
Product Name: Endoscopic Linear Cutting Staplers with Single Use Loading Units Manufacturer: Reach Surgical, Inc.
#### 6. Device Description
The proposed devices, Endoscopic Linear Cutting Staplers with Single Use Loading Units (SULU), are sterilized and disposable surgical instruments, which have applications in general, gynecologic, pediatric and thoracic surgery for resection, transection of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and billiary structures.
It places four or six rows of titanium staples and simultaneously divides the tissue from central line. The Endoscopic Linear Cutter Staplers can be adapted for all of the SULU sizes available, and they may be reloaded and fired no more than 10 times in a single procedure.
The SULU has two configurations: (1) Straight SULU and (2) Articulating SULU, each of them has various specifications.
- Non-Clinical Test Conclusion 7.
111-2
{1}------------------------------------------------
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. These tests include: Physical Performance Testing, Endotoxin Testing, Package Integrity Testing, and Shelf Life Testing.
- 8. Substantially Equivalent (SE) Conclusion
The following table compares the proposed device to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
| Item | Proposed Device | Predicate Device |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| Product Code | GDW | Same |
| Regulation Number | 21 CFR 878.4750 | Same |
| Class | II | Same |
| Intended Use | The Endoscopic Linear Cutting Staplers with Single<br>Use Loading Units have applications in general,<br>gynecologic, pediatric and thoracic surgery for<br>resection, transection, and creation of anastomoses.<br>They may be used for transection and resection of<br>liver substance, hepatic vasculature and biliary<br>structures. | Same |
| Cutting Mechanism | Linear | Same |
| Operation Principle | Manual | Same |
| Safety Mechanism | Green button for preventing from mis-firing. | Same |
| Suture Length | 46, 61 mm | Similar |
| Closed Staples Height | 0.75, 1.0, 1.5, 2.0 mm | Same |
| Closed Staples Form | Image: staple form | Same |
| Staple Materiel | Unalloyed Titanium conforms to ASTM F 67-06 | Same |
| Stapler Materials | Stainless Steel | Same |
| Sterilization | Irradiation Sterilized, SAL: 10-6 | Similar |
.Table 111-1 Comparison of Technology Characteristics
The proposed device, Endoscopic Linear Cutting Staplers with Single Use Loading Units, is determined to be Substantially Equivalent (SE) to the predicate device, Endoscopic Linear Cutting Staplers with Single Use Loading Units (K120179), in respect of safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing or supporting another, representing health and human services.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Nanjing Maidixin Medical Device Company, LTD % Ms. Diana Hong Mid-Link Consulting Company LTD P.O. Box 120-119 Shanghai, China 200120
December 13, 2013
Re: K130738
Trade/Device Name: Endoscopic Linear Cutting Staplers with Single Use Loading Units Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: November 7, 2013 Received: November 12, 2013
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Diana Hong
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
For
Sincerely vours.
Joshua
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
## Section II Indications for Use
510(k) Number: K130738 Device Name: Endoscopic Linear Cutting Staplers with Single Use Loading Units
Indications for Use:
The Endoscopic Linear Cutting Staple Use Loading Units have applications in general, gynecologic, pediatric and thoracic surgery for resection, and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures.
ZPRESCRIPTION USE (Part 21 CFR 801 Subpart D)
OR
OOVER-THE-COUNTER USE (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
# David Krause SS
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K130738
II-I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.