← Product Code [HRS](/productcode/HRS) · K130731

# STABILIZ FIXATION SYSTEM - SPL LOCKING SCREW (K130731)

_Stabiliz Orthopaedics, LLC · HRS · Jul 19, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K130731

## Device Facts

- **Applicant:** Stabiliz Orthopaedics, LLC
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Jul 19, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Stabiliz Fixation System with SPL Locking Screw is intended for fixation of fractures, osteotomies and nonunions of the humerus, radius, ulna, distal tibia, and fibula.

## Device Story

Stabiliz Fixation System comprises metallic bone fixation hardware including straight LCP plates, locking screws, cortical screws, and cancellous screws. Key feature is SPL Locking Screw, which incorporates resorbable threaded polymer collar. System used by orthopedic surgeons in clinical settings for stabilization of bone fractures, osteotomies, and nonunions. Hardware provides mechanical fixation to bone segments to facilitate healing. Device is purely mechanical; no software or electronic components.

## Clinical Evidence

Bench testing only. Testing included in vitro degradation, biomechanical loading, corrosion, and biocompatibility studies to demonstrate substantial equivalence to predicate devices.

## Technological Characteristics

Metallic bone fixation system including plates and screws. SPL Locking Screw features a resorbable threaded polymer collar. Mechanical fixation principle. No energy source, connectivity, or software.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Stabiliz Fixation System ([K120651](/device/K120651.md))
- Zimmer® MotionLoc Screws ([K123918](/device/K123918.md))
- Zimmer® MotionLoc Screws ([K101696](/device/K101696.md))
- Synthes 3.5mm Broad LC-DCP Plate ([K020872](/device/K020872.md))
- Biomet Orthopedics Lactosorb® Tibial L-15 Screw and Washer ([K033233](/device/K033233.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K130731 JUL 1 9 2013 Page 1 of 2

# 510(k) Summary

| Device Trade Name: | Stabiliz Fixation System with SPL Locking Screw                                                                                                                                |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:      | Stabiliz Orthopaedics, LLC<br>665 Stockton Drive<br>Exton, PA 19341                                                                                                            |
| Contact:           | Douglas L. Cerynik, MD<br>President & CEO<br>Stabiliz Orthopaedics, LLC<br>665 Stockton Drive<br>Exton, PA 19341<br>Office: (610) 458-8555<br>cerynik@stabilizorthopaedics.com |
| Prepared by:       | Musculoskeletal Clinical Regulatory Advisers, LLC<br>1331 H Street NW, 12th Floor<br>Washington, DC 20005<br>Phone: 202.552.5800<br>Fax: 202.552.5798                          |
| Date Prepared:     | July 18, 2013                                                                                                                                                                  |
| Classifications:   | 21 CFR 888.3040: Smooth or threaded metallic bone fixation<br>fasteners                                                                                                        |
|                    | and                                                                                                                                                                            |
|                    | 21 CFR 888.3030: Single/multiple component metallic bone<br>fixation appliances and accessories                                                                                |
| Class:             | II                                                                                                                                                                             |
| Product Codes:     | HWC and HRS                                                                                                                                                                    |

# Indications for Use:

The Stabiliz Fixation System with SPL Locking Screw is intended for fixation of fractures, osteotomies and nonunions of the humerus, radius, ulna, distal tibia, and fibula.

{1}------------------------------------------------

# Device Description:

The Stabiliz Fixation System includes standard straight LCP plates, metal locking screws, cortical screws, fully-threaded cancellous screws, and partially-threaded cancellous screws in various sizes and geometries. The SPL Locking Screw is a metal locking screw with a resorbable threaded polymer collar.

## Predicate Devices:

Data provided in the 510(k) support the substantial equivalence of the Stabiliz Fixation System to the following predicate devices: Stabiliz Fixation System (K120651); Zimmer® MotionLoc Screws (K123918 and K101696); Synthes 3.5mm Broad LC-DCP Plate (K020872); and Biomet Orthopedics Lactosorb® Tibial L-15 Screw and Washer (K033233).

### Substantial Equivalence:

The components of the Stabiliz Fixation System are substantially equivalent to the identified predicates with respect to indications for use, geometry, available sizes, materials, methods of fixation to bone, and performance.

### Preclinical Testing:

The non-clinical tests performed by the company include in vitro degradation testing, biomechanical loading study, corrosion testing, and biocompatibility testing. The results demonstrate that the Stabiliz Fixation System is substantially equivalent to legally marketed predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a symbol consisting of three stylized human figures.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

# July 19, 2013

Stabiliz Orthopaedics, LLC % Douglas L. Cerynik,M.D. President & CEO 665 Stockton Drive Exton, Pennsylvania 19341

Re: K130731

Trade/Device Name: Stabiliz Fixation System with SPL Locking Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: June 18, 2013 Received: June 19, 2013

Dear Dr. Cerynik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

## Page 2 - Douglas L. Cerynik, M.D

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Erin I. Keith

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number: K130731

Device Name: Stabiliz Fixation System with SPL Locking Screws

The Stabiliz Fixation System with SPL Locking Screw is intended for fixation of fractures, osteotomies and nonunions of the humerus, radius, ulna, distal tibia, and fibula.

V Prescription Use (Part 29 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (29 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Lori A. Wiggins

Division of Orthopedic Devices

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K130731](https://fda-staging.innolitics.com/device/K130731)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).