K130716 · Bioretec, Ltd. · MAI · Aug 28, 2013 · Orthopedic
Device Facts
Record ID
K130716
Device Name
BIORETEC ACTIVASCREW INTERFERENCE
Applicant
Bioretec, Ltd.
Product Code
MAI · Orthopedic
Decision Date
Aug 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ActivaScrew™ Interference is indicated for fixation of tissue, including ligament or tendon to bone, or a bone-tendon to bone in the presence of appropriate immobilization / controlled mobilization. Interference fixation is intended for surgeries of knee, shoulder, elbow, ankle, foot and hand/wrist.
Device Story
ActivaScrew™ Interference is a fully threaded, bioabsorbable interference screw used for orthopedic soft tissue-to-bone or bone-to-bone fixation. Constructed from poly(L-lactide-co-glycolide) (PLGA), the device provides mechanical fixation during surgical procedures in the knee, shoulder, elbow, ankle, foot, and hand/wrist. It is intended for use by surgeons in clinical settings. The screw functions by providing interference fixation, securing tissue in place while the patient undergoes appropriate immobilization or controlled mobilization. The device is designed to degrade over time, eliminating the need for secondary removal. Clinical benefit is derived from its mechanical stability during the healing process and its bioabsorbable nature.
Clinical Evidence
Bench testing only. Performance was verified through comparative non-clinical testing against predicate devices, including pull-out testing, shear strength testing, inherent viscosity, and dimensional stability testing. Results confirmed the device's mechanical properties and degradation profile are sufficient for its intended use and substantially similar to the predicate devices.
Technological Characteristics
Constructed of bioabsorbable poly(L-lactide-co-glycolide) (PLGA). Fully threaded screw design with diameters of 4-10 mm and lengths of 10-33 mm. Mechanical fixation principle. No software or electronic components.
Indications for Use
Indicated for fixation of tissue (ligament or tendon to bone) or bone-tendon to bone in knee, shoulder, elbow, ankle, foot, and hand/wrist surgeries, requiring immobilization or controlled mobilization.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for bioretec. The logo consists of the word "bioretec" in lowercase letters, with a stylized graphic above it. The graphic is an abstract design, possibly representing a stylized leaf or a flowing shape, rendered in black and white.
#### G 510(k) SUMMARY ·
For the Bioretec ActivaScrew™ Interference
### MANUFACTURER .
Bioretec Ltd. Hermiankatu 22, Modulight Building FI-33720 Tampere FINLAND
Contact person: Ms. Minna Räsänen Quality and Regulatory Affairs Manager Direct: +358 20 778 9509 Mobile: +358 40 868 1606 Fax: +358 3 317 0225 E-mail: Minna.Rasanen@bioretec.com
Date prepared: March 13", 2013
### DEVICE NAME
Trade Name: Bioretec ActivaScrew™ Interference
Common Name: Biodegradable interference screw
AUG 2 8 2013
## DEVICE CLASSIFICATION AND PRODUCT CODE
Product Code: MAI
Regulation Number: 21 CFR 888.3030
Device Classification Name: Single/multiple component metallic bone fixation appliances and accessories
Classification Panel: Orthopedic
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for bioretec. The logo consists of a stylized graphic above the word "bioretec". The graphic is a black abstract shape that resembles a stylized leaf or wing. The word "bioretec" is written in a lowercase sans-serif font.
### PREDICATE DEVICES
- Inion Hexalon™ Biodegradable ACL/PCL Screw (K071464) 1.
- Arthrex Bio-Tenodesis™ Screw from Tenodesis™ Screw Family (K010525, K011007, 2. K020043, K041356, K051726)
- 3. Bioretec ActivaScrew™ (K062980, K072848, K081392)
### DEVICE DESCRIPTION AND PRINCIPLES OF OPERATION
The ActivaScrew™ Interference is indicated for fixation of tissue, including ligament or tendon to bone, or a bone-tendon to bone in the presence of appropriate immobilization / controlled mobilization. Interference fixation is intended for surgeries of knee, shoulder, elbow, ankle, foot and hand/wrist.
The product family consists of fully threaded ActivaScrew™ Interference, diameters 4 - 10 mm and lengths 10 - 33 mm.
The ActivaScrew™ Interference is constructed of bioabsorbable poly(L-lactide-co-glycolide) (PLGA).
#### SUBSTANTIAL EQUIVALENCE TO MARKETED PRODUCTS
The ActivaScrew™ Interference bioabsorbable screw is substantially equivalent to biodegradable screws, intended for similar indications, which have been cleared through a 510(k) pre-market notification pathway.
The Bioretec ActivaScrew™ Interference has the same intended use and principles of operation, and very similar design characteristic as the predicate devices Inion Hexalon™ Biodegradable ACL/PCL Screw (K071464) and Arthrex Bio-Tenodesis™ Screw from Tenodesis™ Screw Family (K010525, K011007, K020043, K041356, K051726). The material of the ActivaScrew™ Interference is the same as the material of the previously cleared ActivaScrew™ (K062980, K072848, K081392).
The performance of the ActivaScrew™ Interference was compared to the predicate devices by pull-out testing, shear strength testing, inherent viscosity and dimensional stability testing.
These test results also verify that the performance of ActivaScrew™ Interference is sufficient for its indications, and that the device is substantially similar to its predicates with regard to its mechanical properties, and long term degradation properties. This performance testing does not raise any new questions of safety or effectiveness of the ActivaScrew™ Interference and demonstrates that the ActivaScrew™ Interference performs mechanically in a manner
{2}------------------------------------------------
K130716 (3/3)
Image /page/2/Picture/1 description: The image contains the word "bioretec" in lowercase letters, with a stylized graphic above it. The graphic appears to be an abstract design, possibly representing a stylized leaf or wing-like shape. The word "bioretec" is in a simple, sans-serif font and is positioned directly below the graphic.
substantially similar to the tested predicate device. Thus, non-clinical tests and in vitro testing determined that the ActivaScrew™ Interference has substantially similar performance as compared to its predicate devices.
:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 28, 2013
Biorectec Limited Ms. Mari Ruotsalainen Quality and Regulatory Affairs Manager Hermiankatu 22, Modulight Building FI-33720 Tampere Finland
Re: K130716
Trade/Device Name: Bioretec ActivaScrew 131 Interference Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: July 5, 2013 Received: July 5, 2013
Dear Ms. Ruotsalainen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of r ne general ochariting practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrs/default.htm.
Sincercly yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedie Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Bioretec. The logo consists of a stylized, abstract symbol above the word "bioretec" in lowercase letters. The symbol is black and appears to be composed of curved, overlapping shapes, possibly representing biological forms or processes.
#### F Indications for Use Statement
Submitter: Bioretec Ltd. K130716 510(k) Number: ActivaScrew™ Interference Device Name:
Indications for Use:
The ActivaScrew™ Interference is indicated for fixation of tissue, including ligament or tendon to bone, or a bone-tendon to bone in the presence of appropriate immobilization / controlled mobilization. Interference fixation is intended for surgeries of knee, shoulder, elbow, ankle, foot and hand/wrist.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
Division of Orthopedic Devices
Page 2_of_1
Traditional 510(k) Date: March 13th, 2013
F-1 Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.