← Product Code [JWH](/productcode/JWH) · K130708

# SMITH & NEPHEW PATIENT MATCHED CUTTING BLOCKS (K130708)

_Smith & Nephew, Inc. · JWH · Dec 11, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K130708

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Dec 11, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Indications for Use

Smith & Nephew's Patient Matched Cutting Blocks are intended to be used as patientspecific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. knee systems and their cleared indications for use: - Genesis 2 Knee System . - Legion Knee System o - O Journey BCS Knee System - . Journey II Knee System The Patient Matched Cutting Blocks are intended for single use only.

## Device Story

Device consists of patient-specific surgical cutting blocks designed from patient imaging data (MRI, X-ray). Software applications process imaging data to create custom blocks for intra-operative use. Surgeon uses blocks to assist in positioning total knee replacement components and guiding bone marking prior to resection. Used in clinical settings by orthopedic surgeons. Output is physical instrumentation that improves accuracy of resection depth, coronal alignment, rotation, and tibial posterior slope. Benefits include improved surgical precision compared to conventional instrumentation.

## Clinical Evidence

No clinical data. Bench testing included a cadaveric assessment comparing patient-matched cutting blocks to conventional instrumentation. Results demonstrated equivalent performance in resection depth, coronal alignment (varus/valgus), rotation, and tibial posterior slope within clinically acceptable parameters.

## Technological Characteristics

Patient-specific surgical instrumentation designed via software from patient imaging (MRI, X-ray). Single-use. Software validation performed per FDA guidance (2002).

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Patient Matched Cutting Blocks ([K082358](/device/K082358.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# > We are smith&nephew

| Submitted by:                                | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116                                                                                                                   |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Summary:                             | January 10, 2014                                                                                                                                                                                                    |
| Contact Person and Address:                  | Bradley Heil<br>Regulatory Affairs Specialist<br>T (901) 399-6339<br>F (901) 566-7831                                                                                                                               |
| Name of Device:                              | Smith & Nephew, Inc. Patient Matched Cutting Block                                                                                                                                                                  |
| Common Name:                                 | Knee Prosthesis                                                                                                                                                                                                     |
| Device Classification Name and<br>Reference: | CFR 888.3560 Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis<br>21 CFR 888.3565 Knee joint patellofemorotibial<br>metal/polymer porous-coated uncemented prosthesis |
| Device Class:                                | Class II                                                                                                                                                                                                            |
| Panel Code:                                  | Orthopaedics/87                                                                                                                                                                                                     |
| Product Code:                                | OOG, JWH, MBH                                                                                                                                                                                                       |

### Device Description

The Patient Matched Cutting Blocks were previously cleared for market via premarket notification K082358. Subject of this premarket notification are modifications to the software components used to design and manufacture the Patient Matched Cutting Blocks. The Patient Matched Cutting Blocks are designed and manufactured from patient imaging data (MRJ, X-Ray).

### Technological Characteristics

Software verification and validation testing was completed in line with FDA's guidance document entitled, "General Principles of Software Validation; Final Guidance for Industry and FDA Staff," dated January 11, 2002. A review of this testing has demonstrated that there are no new issues related to the safety and effectiveness of the subject device and the software will perform as intended. Clinical data was not needed to support the safety and effectiveness of the subject devices.

### Intended Use

Smith & Nephew's Patient Matched Cutting Blocks are intended to be used as patientspecific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. knee systems and their cleared indications for use:

{1}------------------------------------------------

- Genesis 2 Knee System .
- Legion Knee System o
- O Journey BCS Knee System
- . Journey II Knee System

The Patient Matched Cutting Blocks are intended for single use only.

#### Performance Data

Smith & Nephew conducted a cadaveric assessment of the subject devices. The results of this study show that the patient matched cutting blocks designed using the case processing applications perform equivalent to conventional instrumentation or within clinically acceptable parameters with respect to resection depth, coronal alignment (varus/valgus), rotation and tibial posterior slope.

#### Substantial Equivalence Information

The subject devices are identical in function, intended use, indications for use, and material composition, and very similar in overall design to the Patient Matched Cutting Blocks cleared via premarket notification K082358.

### Table 1: Substantially equivalent predicates to the modified Patient Matched Cutting Blocks

| Manufacturer            | Description                    | Submission<br>Number | Clearance Date |
|-------------------------|--------------------------------|----------------------|----------------|
| Smith & Nephew,<br>Inc. | Patient Matched Cutting Blocks | K082358              | 11/25/2008     |

#### Conclusion

As previously noted, this Traditional 510(k) Premarket Notification is being submitted to request clearance for the modified Patient Matched Cutting Blocks. Based on the similarities to the predicate components, a cadaveric assessment, and a review of the software validation testing performed, the devices are substantially equivalent to above predicate Patient Matched Cutting Blocks.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and appear to be abstract representations of human faces. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 28, 2104

Smith & Nephew, Inc. Mr. Bradley Heil Regulatory Affairs Specialist 1450 Brooks Road Memphis, Tennessee 38116

Re: K130708

Trade/Device Name: Smith & Nephew Patient Matched Cutting Blocks Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis. Regulatory Class: Class II Product Code: JWH, MBH, OOG Dated: October 29, 2013 Received: October 30, 2013

Dear Mr. Heil:

This letter corrects our substantially equivalent letter of December 11, 2013.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. Bradley Heil

comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Vincent J. Devlin -S

- for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

# 510(k) Number (if known): K130708

## Device Name: Smith & Nephew, Inc. Patient Matched Cutting Blocks

## Indications for Use:

Smith & Nephew's Patient Matched Cutting Blocks are intended to be used as patient-specific surgical instrumentation to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

The Patient Matched Cutting Blocks are intended for use with the following existing Smith & Nephew, Inc. knee systems and their cleared indications for use:

- · Genesis II Knee System
- · Legion Knee System
- · Journey BCS Knee System
- · Journey II Knee System

The Patient Matched Cutting Blocks are intended for single use only.

Prescription Use X (Part 21 CFR 801 Subpart AND/OR D)

Over-The-Counter Use

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.

Division of Orthopedic Devices

Page _ 1 of _ 1

---

**Source:** [https://fda-staging.innolitics.com/device/K130708](https://fda-staging.innolitics.com/device/K130708)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).