K130704 · Philips Digital Mammography Sweden AB · MUE · Dec 13, 2013 · Radiology
Device Facts
Record ID
K130704
Device Name
MICRODOSE SI
Applicant
Philips Digital Mammography Sweden AB
Product Code
MUE · Radiology
Decision Date
Dec 13, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1715
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K130704 · Dec 13, 2013
MICRODOSE SI
Philips Digital Mammography Sweden AB
Clinical screening population dataset (routine clinical mammography images)
The sponsor used a retrospective clinical screening dataset to validate the breast density measurement software by assessing agreement between the device's density score and radiologist-assigned BI-RADS scores, as well as evaluating correlations with age and breast volume.
Breast density; Screening population; BI-RADS correlation; Retrospective clinical data
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical screening population dataset analysis; Retrospective analysis of clinical screening data
Clinical screening population
Not applicable for this study
Agreement between MicroDose Density score and BI-RADS breast composition score; correlation of volumetric glandularity with age, breast volume, and BI-RADS score.
Indications for Use
The MicroDose SI (model L50) is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MicroDose SI (model L50) is intended to be used in the same clinical applications as traditional film/screen systems.
Device Story
MicroDose SI (model L50) is a full-field digital mammography (FFDM) system. It utilizes photon-counting crystalline silicon strip detectors to acquire X-ray images. The system discriminates between high and low energy photons based on an energy threshold to perform spectral imaging. This spectral data allows for the calculation of volumetric breast glandularity and breast thickness at the pixel level. The device integrates these measurements to calculate total breast volume and total fibroglandular volume, producing a breast density score that correlates to BI-RADS categories. Output is provided as a DICOM Structured Report sent to PACS and as private attributes in the image header. Used in clinical settings by radiologists to aid in the assessment of breast tissue composition. The device provides adjunctive information to assist clinical decision-making regarding breast density.
Clinical Evidence
Clinical validation included a study on a screening population dataset. Endpoints included: agreement between MicroDose Density score and MQSA-qualified radiologist BI-RADS scores; correlation of measurements across CC/MLO views and laterality; correlation of volumetric glandularity with age and breast volume; and correlation of volumetric glandularity with BI-RADS breast composition scores. Results confirmed expected physiological trends (e.g., decrease in glandularity with age/volume, increase with BI-RADS score). Bench testing included comparisons to breast tissue equivalent phantoms.
Technological Characteristics
Photon-counting crystalline silicon strip detectors; spectral imaging via energy threshold discrimination; 24 x 26 cm imaging area. Connectivity via DICOM Structured Report to PACS. Software-based volumetric analysis of breast density. No specific material standards or software architecture class (e.g., ML/DL) explicitly detailed beyond algorithmic calculation of glandularity.
Indications for Use
Indicated for generating mammographic images for screening and diagnosis of breast cancer in patients requiring mammography.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "PHILIPS" in a bold, outlined font. The letters are large and spaced closely together. The font style gives the word a textured appearance, as if it were stamped or printed with a slightly uneven distribution of ink.
DEC 1 3 2013
K130704
Page 1 of 4
# 510(k) Summary (21 CFR 807.92)
### A. SUBMITTER INFORMATION
| Submitter's name | Philips Digital Mammography Sweden AB |
|-----------------------------------|---------------------------------------|
| Address | Smidesvägen 5 |
| | SE-171 41 Solna |
| | Sweden |
| Establishment registration number | 3009307584 |
| Contact person | Mr. Gustav Lins |
| Telephone | +46-8-623 52 00 |
| FAX | +46-8-623 52 01 |
| Title | Manager Regulatory Affairs |
| Date of the summary preparation | 2013-10-09 |
### B. DEVICE IDENTIFICATION
| Device trade name | MicroDose SI (model L50) |
|-----------------------------|----------------------------------------------|
| Device common name | Full-Field Digital Mammography X-ray System |
| Classification name | Full-Field Digital Mammography X-ray System |
| Classification product code | MUE, LLZ (secondary) |
| Device class | II |
| Regulation code | 21 CFR 892.1715, 21 CFR 892.2050 (secondary) |
### C. DEVICE DESCRIPTION
This application is to add functionality to the previously cleared MicroDose SI, Model L50, (K123995) that allows for breast density measurements.
The L50 submission described that the detector electronics have been modified to do spectral imaging through the functionality to discriminate between high and low energy photons based on an energy threshold. The first application of this technology is Spectral Breast Density Measurement where breast glandularity and thickness is measured simultaneously by the spectral difference in X-ray attenuation between adipose and fibroglandular tissue. The MicroDose'Sl can thereby measure volumetric glandularity and breast thickness for each pixel in the image. Total breast volume is calculated by integrating the breast thickness over the breast region. taking pixel size and beam geometry into account. Fibroglandular thickness is calculated as the product of breast thickness and volumetric glandularity for each pixel. Total fibroglandular volume is then calculated by integrating the fibroglandular thickness over the breast region. The total breast volumetric glandularity is then given as the ratio of fibroglandular to breast volume. The correlation between the MicroDose Density Score and the radiologist's BI-RADS score has The following Information is sent as part of the content of a DICOM Structured Report, SR, to selected destinations (for example a PACS). Breast analysis information is also included as private attributes in the DICOM header of the acquired images.
- Volumetric glandularity (%) �
- Glandular volume (cm3) ●
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "PHILIPS" in a bold, outlined font. The letters are large and evenly spaced, filling most of the frame. The font has a slightly distressed or textured appearance, giving it a vintage or worn look.
- Breast volume (cm³) .
- Breast density score (I, III, III, IV). The density score correlates to the BI-RADS categories. .
### D. indications for use
The MicroDose SI (model L50) is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MicroDose SI (model L50) is intended to be used in the same clinical applications as traditional film/screen systems.
### E. SUBSTANTIAL EQUIVALANCE
The subject device is identical to the MicroDose SI (model L50) in regards to FFDM characteristics. Furthermore, with regards to the applications described in this submission, it is substantially equivalent to the software predicates Volpara and Quantra.
| | MicroDose SI (model<br>L50) | MicroDose SI (model<br>L50) | Volpara<br>K102556 | Quantra<br>K120472 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject device | K123995<br>Predicate | Predicate | Predicate |
| Manufacturer | Philips Digital<br>Mammography<br>Sweden AB | Philips Digital<br>Mammography<br>Sweden AB | Matakina Technology Limited | Hologic, Inc. |
| Indication For<br>Use | The MicroDose SI<br>(model L50) is<br>indicated for<br>generating<br>mammographic<br>images that can be<br>used for screening<br>and diagnosis of<br>breast cancer. | The MicroDose SI<br>(model L50) is<br>indicated for<br>generating<br>mammographic<br>images that can be<br>used for screening<br>and diagnosis of<br>breast cancer. | Volpara is a software application<br>intended for use with digital<br>mammography systems. Volpara<br>calculates volumentric breast<br>density as a ratio of<br>fibroglandular tissue and total<br>breast volume estimates.<br>Volpara provides these numerical<br>values for each image to aid<br>radiologists in the assessment of<br>breast tissue composition.<br>Volpara produces adjunctive<br>information. It is not an<br>interpretive or diagnostic aid.<br>Volpara is a software application<br>which runs on Windows or Linux-<br>based computers. | QuantraTM is a software<br>application intended for use with<br>images acquired using digital<br>breast X-ray systems. Quantra<br>calculates volumetric breast<br>density as a ratio of<br>fibroglandular tissue area and<br>total breast area estimates. It<br>segregates breast density into BI-<br>RADS-like breast composition<br>categories, which may be useful<br>in the reporting of consistent<br>breast composition values for<br>each image, breast and subject,<br>to aid radiologists in the<br>assessment of breast tissue<br>composition. Quantra produces<br>adjunctive information, it is not<br>an interpretive or diagnostic aid.<br>Quantra runs on a Windows<br>platform. |
| Device<br>Classification<br>Name | Full field digital,<br>System, X-ray,<br>Mammographic | Full field digital,<br>System, X-ray,<br>Mammographic | System, Image Processing,<br>Radiological | System, Image Processing,<br>Radiological |
| Device Class | II | II | II | II |
| Detector<br>technology | Photon counting<br>Crystalline silicon<br>strip detectors<br>Rejection of<br>electronic noise<br>24 x 26 cm imaging<br>area | Photon counting<br>Crystalline silicon<br>strip detectors<br>Rejection of<br>electronic noise<br>24 x 26 cm imaging<br>area | N/A | N/A |
| Image<br>Evaluation<br>Source | Digital<br>Mammography<br>Images | Digital Mammography Images | Digital Mammography Images | Digital Mammography Images |
| Anatomical<br>Assessment<br>Area | Breast | Breast | Breast | Breast |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "PHILIPS" in a bold, outlined font. The letters are uppercase and have a slightly distressed or textured appearance. The word is horizontally oriented and centered in the image.
| | MicroDose Si (model L50)<br>Subject device | MicroDose Si (model L50)<br>K123995<br>Predicate | Volpara<br>K102556<br>Predicate | Quantra<br>K120472<br>Predicate |
|-------------------------------------------|------------------------------------------------------------------------------------------------|--------------------------------------------------|---------------------------------|-----------------------------------------|
| Assessment<br>output | Glandular volume (cm³)<br>(volume of fibroglandular tissue) | N/A | Volume of fibroglandular tissue | Volume of fibroglandular tissue; |
| | Volumetric glandularity (%)<br>(volumetric breast density) | | Volumetric breast density | Volumetric breast density |
| | Breast volume (cm³) | | Breast volume | Area breast density measure |
| | Breast density score (I, II, III, IV). The density score correlates to the Bi-RADS categories. | | | BI-RADS like breast composition measure |
| Image storage<br>and report<br>generation | DICOM Structured report to send to PACS | | Output to console | DICOM Structured report to send to PACS |
### E. PERFORMANCE/BENCH TESTING
The MicroDose SI L50 Breast Density Measurement application has been developed in accordance with Philips Digital Mammography Sweden AB's product development process. It has been verified and validated in accordance with the company's design control process. The design control process is divided into several levels of verification, unit, integration and system level. The results from the performance testing have been included in the application.
The bench testing and clinical validation of the MicroDose SI included:
- Comparisons of MicroDose Breast Density Measurement parameters to industry standard . breast tissue equivalent phantoms;
- . Validation that the Breast Density Measurement application can be easily incorporated into the clinical workflow of the radiologists;
- . Based on a substantial clinical screening population dataset:
- The agreement between the MicroDose Density score and the BI-RADS breast 0 composition score given by MQSA qualified radiologists was evaluated for a screening population;
- Evaluation of the correlation between MicroDose Breast Density Measurements in o CC and MLO views and left and right laterality to confirm consistent results;
- Evaluation of the correlation between volumetric glandularity and age to confirm o the expected decrease with age;
- o Evaluation of the correlation between volumetric glandularity and breast volume to confirm the expected decrease with volume;
- Evaluation of the correlation between volumetric glandularity and BI-RADS breast O composition score to confirm the expected increase with BI-RADS score.
-
{3}------------------------------------------------
K130704
page 4 of 4
The verification and validation activities confirm that the established acceptance criteria were met.
## G. CONCLUSION
The breast density measurement functionality that is the subject of this application relies on the technology of a FFDM device which has already been cleared by the FDA (MicroDose SI, K123995). Breast density assessments have previously been introduced by Quantra (K120472) and Volpara (K102556). These devices provide information that is very similar to that generated by the MicroDose SI (Model L50), and are intended for the same purpose, i.e. to generate adjunctive information regarding breast density. Like the breast density functionality of the predicate devices Volpara and Quantra, the MicroDose Breast Density Measurement aids the clinician in the assessment of breast tissue composition. The Breast Density Measurement functionality is well characterized by the design control process and the physical laboratory tests show results that are comparative to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling a bird or abstract human figure, with three curved lines representing arms or wings. The symbol is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". The text is arranged around the circumference of the circle, with the symbol placed centrally to the right.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
December 13, 2013
PHILIPS DIGITAL MAMMOGRAPHY SWEDEN AB GUSTAV LINS MANAGER REGULATORY AFFAIRS SMIDESVAGEN 5, SOLNA, SW SE-171 41 SWEDEN
Re: K130704
Trade/Device Name: MicroDose SI Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE, LLZ Dated: October 29, 2013 Received: October 31, 2013
Dear Mr. Lins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2-Mr. Lins
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR-Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/dcfault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.ida.gov/MedicalDevices/Resources/orYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K130704
## Device Name
MicroDose SI (Model L50)
## Indications for Use (Describe)
The MicroDose SI (model L50) is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer.
The MicroDose SI (model L50) is intended to be used in the same clinical applications as traditional film/screen systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
[] Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY |
|------------------|
|------------------|
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | Michael D. O'Hara |
|------------------------------------------------------------------------------|-------------------|
|------------------------------------------------------------------------------|-------------------|
| FORM FDA 3881 (9/13) | Page 1 of 2 |
|----------------------|-------------|
|----------------------|-------------|
ORM FUA 3681 (8/13)
178 Publisher Ser sam 1 100 : 441-4741 141-4741 EF
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.