← Product Code [MBP](/productcode/MBP) · K130703

# VIVORTE BVF (K130703)

_Vivorte, Inc. · MBP · Sep 12, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K130703

## Device Facts

- **Applicant:** Vivorte, Inc.
- **Product Code:** [MBP](/productcode/MBP.md)
- **Decision Date:** Sep 12, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Vivorté BVF™ is a moldable, self-setting, gradually resorbable, calcium phosphate bone void filler with shaped particles of human bone that contain osteoinductive demineralized bone matrix (DBM). Vivorté BVF™ is indicated for use to fill bony voids or defects of the skeletal system (i.e., extremities, pelvis) that may be surgically created or osseous defects created from traumatic injury to the bone and only for bony voids or defects that are not intrinsic to the stability of the bony structure. Vivorté BVF™ may be manually applied to the bony defect or applied to the defect through a cannula. Vivorté BVF™ is gradually resorbed and remodeled by the body as new bone formation occurs during the healing process.

## Device Story

Vivorté BVF is a moldable, self-setting, resorbable bone void filler composed of calcium phosphate and shaped human bone particles containing demineralized bone matrix (DBM). Device is prepared intraoperatively by mixing components; applied manually or via cannula to skeletal bony voids or defects (extremities, pelvis) not intrinsic to structural stability. Material isothermically hardens in vivo into a carbonated apatite (hydroxyapatite) composite. Provides osteoconductive scaffold and osteoinductive DBM to support bone remodeling and healing. Intended for use by surgeons in clinical settings. Benefits include filling defects with a material that resorbs and remodels as new bone forms.

## Clinical Evidence

No clinical data. Substantial equivalence established via chemical, physical, and mechanical bench testing, biocompatibility/toxicity testing (ISO 10993, ASTM, USP), and animal testing.

## Technological Characteristics

Moldable, self-setting calcium phosphate bone void filler. Composed of carbonated apatite (hydroxyapatite) and human bone particles with DBM. Isothermic hardening in vivo. Mechanical properties exceed human cancellous bone. Sterile, single-use, prescription device.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- ETEX EquivaBone® Osteoinductive Bone Graft Substitute ([K090855](/device/K090855.md))
- Synthes® Norian® SRS® Bone Void Filler ([K011897](/device/K011897.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) Summary

510(k) SPONSOR: Vivorté, Inc. 1044 E. Chestnut Street Louisville, KY 40204 Contact: Robert Burden Email: rburden@vivorte.com Telephone: (502) 693 - 2432 FAX: (502) 714 - 7235

**SEP** 12 2013

Vivorté BVFTM TRADE NAME:

Calcium phosphate bone void filler; bone void filler COMMON NAME:

| Product      | Regulation and<br>Description                                              | Product Code | Device                                                               | Device Class |
|--------------|----------------------------------------------------------------------------|--------------|----------------------------------------------------------------------|--------------|
| Vivorté BVF™ | 21 CFR 888.3045<br>- Resorbable<br>calcium salt bone<br>void filler device | MBP          | Filler, bone void,<br>osteoinductive<br>(w/o human<br>growth factor) | II           |
|              |                                                                            | MQV          | Filler, bone void,<br>calcium compound                               | II           |

ETEX EquivaBone® Osteoinductive Bone Graft Substitute PREDICATE DEVICES: (K090855)

Synthes® Norian® SRS® Bone Void Filler (K011897)

#### DEVICE DESCRIPTION:

Vivorté BVF™ is a moldable, self-setting, gradually resorbable, calcium phosphate bone void filler with shaped particles of human bone that contain osteoinductive demineralized bone matrix (DBM). Vivorté BVF™ is indicated for use to fill bony voids or defects of the skeletal system (i.e., extremities, pelvis) that may be surgically created or osseous defects created from traumatic injury to the bone and only for bony voids or defects that are not intrinsic to the stability of the bony structure. Vivorté BVF™ may be manually applied to the bony defect or applied to the defect through a cannula. Vivorté BVF™ isothermically hardens in vivo to form a composite of a carbonated apatite (hydroxyapatite) and shaped particles of human bone that contain DBM. Vivorté BVF™ has a compressive and bending strength that is greater than that of human cancellous bone. The carbonated apatite (hydroxyapatite) which closely resembles the mineral phase of bone provides an osteoconductive scaffold and the shaped particles of human bone contain osteoinductive demineralized bone matrix (DBM). The composite device is gradually

{1}------------------------------------------------

resorbed and remodeled by the body as new bone formation occurs during the healing process. Vivorté BVF™ is provided in various kit sizes corresponding to the amount of bone void filler produced when the components of the kit are mixed together.

Each lot of allograft human bone in the Vivorté BVF™ is assayed for osteoinductive potential using an athymic mouse or rat model. The results of the osteoinductive potential testing may or may not be indicative of the osteoinductivity of Vivorté BVF™ in humans.

## INTENDED USE AND INDICATIONS FOR USE:

Vivorté BVF™ is a moldable, self-setting, resorbable, calcium phosphate bone void filler with shaped particles of human bone that contain osteoinductive demineralized bone matrix (DBM). Vivorté BVF™ is indicated for use to fill bony voids or defects of the skeletal system (i.e., extremities, pelvis) that may be surgically created or osseous defects created from traumatic injury to the bone and only for bony voids or defects that are not intrinsic to the stability of the bony structure. Vivorté BVF™ may be manually applied to the bony defect or applied to the defect through a cannula. Vivorté BVF™ is resorbed and remodeled by the body as new bone formation occurs during the healing process.

# BASIS FOR SUBSTANTIAL EQUIVALENCE:

Vivorté BVF™ is chemically and physically substantially equivalent to the ETEX EquivaBone® Osteoinductive Bone Graft Substitute device. Both of these devices are synthetic calcium phosphate based bone void fillers containing osteoinductive demineralized bone matrix. The Vivorté BVF™ and the ETEX EquivaBone® Osteoinductive Bone Graft Substitute devices are both intraoperatively prepared by mixing similar components together to produce self-hardening calcium phosphate bone void fillers that, when implanted in a bony defect, are resorbed and remodeled by the body as new bone formation occurs during the healing process. Both devices have comparable mixing, handling/working, and setting times and, when fully hardened, are composed primarily of apatites that have similar chemical, physical, and mechanical characteristics and properties. The Vivorté BVF™ device and the ETEX EquivaBone® Osteoinductive Bone Graft Substitute device have similar indications for use.

Vivorté BVFTM is also chemically and physically substantially equivalent to the Synthes® Norian® SRS® Bone Void Filler. Both the Vivorté BVF™ and the Synthes® Norian® SRS® Bone Void Filler are composed of synthetic calcium phosphate materials that are virtually identical; however, the Synthes® Norian® SRS® Bone Void Filler does not contain any demineralized bone matrix. Both the Vivorté BVF™ and the Synthes® Norian® SRS® Bone Void Filler devices are intraoperatively prepared by mixing similar components together to produce self-hardening calcium phosphate bone void fillers that, when implanted in a bony defect, are resorbed and remodeled by the body as new bone formation occurs during the healing process. Both devices have comparable mixing, handling/working, and setting times; and both of these devices, when fully hardened, are composed primarily of hydroxyapatite that have similar chemical, physical, and mechanical characteristics and properties. The indications for use of the Vivorté BVF™ and the Synthes® Norian® SRS® Bone Void Filler are similar with the

{2}------------------------------------------------

exception that the Synthes® Norian® SRS® Bone Void Filler does not contain DBM and therefore is not labeled as osteoinductive.

The Vivorté BVF™ and the ETEX and Synthes® predicate devices are provided packaged sterile and are intended as prescription use only single use devices.

#### Non-Clinical Testing

The following testing was performed to demonstrate substantial equivalency of Vivorté BVFTM to the predicate devices.

- . Chemical, physical and mechanical testing
- . Biocompatibility and toxicity (ISO 10993; ASTM; USP)
- Animal testing .

#### Clinical Testing

Clinical testing was not necessary to determine substantial equivalence between Vivorté BVFTM and the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized symbol resembling a human figure with outstretched arms, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular fashion around the symbol. The symbol is composed of three curved lines that form the shape of a person, and the text is written in a simple, sans-serif font.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MID 20993-0002

September 12. 2013

Vivorté. Incorporated % Peoples & Associates Stephen J. Peoples, VMD. MS 5010 Lodge Pole Lanc Fort Wayne, Indiana 46814

Re: K130703

Trade/Device Name: Vivorté I3VFTM Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MBP, MQV Dated: June 26, 2013 Received: July 10. 2013

Dear Dr. Peoples:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrenties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class 111 (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

#### Page 2 -Stephen J. Peoples. VMD. MS

forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indications for Use Statement

510(k) Number: K130703

Device Name: Vivorté BVFTM

Indications for Use:

Vivorté BVFTM is a moldable, self-setting, gradually resorbable, calcium phosphate bone void filler with shaped particles of human bone that contain osteoinductive demineralized bone matrix (DBM). Vivorté BVF™ is indicated for use to fill bony voids or defects of the skeletal system (i.e., extremities, pelvis) that may be surgically created or osseous defects created from traumatic injury to the bone and only for bony voids or defects that are not intrinsic to the stability of the bony structure. Vivorté BVF™ may be manually applied to the bony defect or applied to the defect through a cannula. Vivorté BVF™ is gradually resorbed and remodeled by the body as new bone formation occurs during the healing process.

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Laurence D. Coyne -S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K130703

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K130703](https://fda-staging.innolitics.com/device/K130703)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
