K130687 · Medical Components, Inc. · DYB · Aug 15, 2013 · Cardiovascular
Device Facts
Record ID
K130687
Device Name
2F & 3F VASCU-SHEATH TEARAWAY INTRODUCER
Applicant
Medical Components, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Aug 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The 2F and 3F Vascu-Sheath Tearaway Introducer is intended for percutaneous venous access by modified Seldinger Technique in neonates, infants, and children.
Device Story
The Medcomp 2F and 3F Vascu-Sheath Tearaway Introducer facilitates percutaneous venous access for catheter placement. The device is used with a .010 inch diameter guidewire placed via needle or IV catheter. The introducer sheath and dilator are inserted over the guidewire; the guidewire is then removed. The dilator is withdrawn, and a peripheral, mid-line, or central venous catheter is threaded through the sheath into the vein. Once the catheter is positioned, the sheath is split and withdrawn, leaving the catheter in place. The device is intended for use in neonates, infants, and children. It is packaged sterile in pouches or trays with accessories to assist in catheter insertion.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through in-vitro bench testing, including leakage, force at break, simulated use, guidewire passage, and accelerated aging, alongside biocompatibility testing per ISO 10993.
Technological Characteristics
Catheter introducer system consisting of a sheath and dilator. Materials tested for biocompatibility per ISO 10993. Design verification performed per ISO 14971:2000 risk management protocols. Sterile packaging. No energy source or software components.
Indications for Use
Indicated for percutaneous venous access via modified Seldinger Technique in neonates, infants, and children to facilitate insertion of peripheral, mid-line, or central venous catheters.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a stylized letter "C" shape in black and white. Below the logo is the text "medCOMP." in a serif font.
1499 Delp Drive Harleysville, PA 19438 Tel: 215-256-4201
Fax: 215-256-1787
www.medcompnet.com
| Section 5 | | 510(k) SUMMARY | Traditional 510K |
|-----------|------------------------|-----------------------------------------------------------------------------------------------------|------------------|
| A. | Submitter Information: | | |
| | Submitter: | MEDCOMP®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201<br>Fax: (215) 256-9191 | AUG 15 2013 |
| | Contact: | Jessica Leo<br>Regulatory Associate | |
| | Date Prepared: | March 11, 2013 | |
| B. | Trade Name: | Medcomp® Vascu-Sheath Tearaway Introducer | |
| | Common Name: | Introducer, Catheter | |
| | Classification Name: | Catheter Introducer (DYB) | |
| | Regulation Name: | Cardiovascular Devices | |
| | C.F.R. Section: | 870.1340 | |
| | Class: | II | |
| C. | Predicate Device: | K053092 Medcomp, Vascu-Sheath II | |
K13068
#### D. Device Description:
The Medcomp 2F and 3F Vascu-Sheath Tearaway Introducer is intended for percutaneous venous access by modified Seldinger Technique. The purpose of the device is to facilitate the insertion of a peripheral, mid-line, or central venous catheter. The introducer is inserted over a .010 inch diameter guidewire that has been placed into the target vessel using a needle or IV catheter. Upon placement of the introducer, the guidewire is removed. When the catheter is ready for insertion, the dilator of the introducer is withdrawn from the sheath and the catheter is threaded into the vein through the sheath. When the catheter has been placed, the sheath is simultaneously split and withdrawn leaving the catheter in place.
The Vascu-Sheath Tearaway Introducer is packaged sterile in either a pouch or a variety of tray configurations with additional accessories to facilitate catheter insertion.
#### Indications for Use: E.
The Vascu-Sheath Tearaway Introducer is intended for percutaneous venous access by modified Seldinger Technique in neonates, infants, and children.
#### F. Comparison to Predicate Devices:
The technological characteristics of the Vascu-Sbeath Tearaway Introducer are substantially equivalent to the predicate device in terms of intended use. design, performance, placement and method of sterilization.
Medcomp Vascu-Sheath Tearaway Introducer Summary Page 1 of 3
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image is a logo for medCOMP. The logo features a stylized, geometric shape in black and white. The shape is reminiscent of a stylized letter "C" or a partially enclosed square. Below the geometric shape, the word "medCOMP." is written in a lowercase, italicized font. The text is positioned directly beneath the open part of the geometric shape.
G.
1499 Delp Drive
Harleysville, PA 19438
Tel: 215-256-4201
Fax: 215-256-1787
www.medcompnet.com
### Bench / Performance Data:
Performance data for the Vascu-Sheath Tearaway Introducer demonstrates that this device is substantially equivalent to the legally marketed device.
Performance testing of the proposed devices was conducted in accordance with applicable international standards and FDA guidance documents along with unrecognized ISO standards and internal engineering protocols.
The results of these tests in conjunction with the substantial equivalence claims effectively demonstrate the proposed devices are equivalent to the predicate devices.
The following tests were performed:
Freedom from Liquid Leakage Force at Break Simulated Use Guidewire Passage Accelerated Aging
#### ਜ਼. Biocompatibility:
Testing for all materials used has been tested in accordance with ISO 10993. All biocompatibility testing demonstrates the materials used meet the requirements of ISO 10993.
#### I. Technological Characteristics:
Technological similarities between the proposed devices and predicate devices remain the same:
#### .J. Safety and Performance Tests
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for this device. However, design verification testing was performed according to protocols based on the recommendations/requirements of applicable FDA guidance and FDA recognized international standards. Verification testing, determined to be applicable to the safety and efficacy of the device, was shown to meet predetermined acceptance criteria listed therein.
Risk management, including a failure modes and effects analysis (FMEA), of the subject device was conducted in accordance with an internal protocol based on ISO 14971:2000, Medical Devices - Risk Management for Medical Devices. The analysis did not identify any new types of safety or efficacy questions for the subject Over the Needle Tearaway with Flared Hub device.
#### K. Summary of Substantial Equivalence:
The proposed device meets the performance criteria of design verification as specified by ISO standards, guidance documents and internal test protocols. The proposed device has the same intended use, operation and function as the predicate. There are no differences that raise new issues of safety and effectiveness. The proposed devices are substantially equivalent to the legally marketed predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the text "medCOMP." below it. The logo above the text is a stylized geometric shape, primarily black, with white space creating a sense of depth. The shape appears to be an abstract representation of a letter or symbol. The text "medCOMP." is in a serif font and is smaller in size compared to the logo.
.
1499 Delp Drive Harleysville, PA 19438
Tel: 215-256-4201
Fax: 215-256-1787
www.medcompuet.com
Clinical studies were not deemed necessary since in-vitro testing was sufficient to demonstrate safety and effectiveness by way of comparison to the legally marketed predicate device.
Medcomp Vascu-Sheath Tearaway Introducer Summary
Page 3 of 3
.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 15, 2013
Medical Components Inc. C/O Jessica Leo Regulatory Associate 1499 Delp Drive Harletsville, PA 19438
Re: K130687
Trade/Device Name: 2F and 3F Vascu-Sheath Tearaway Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Introducer, Catheter Regulatory Class: Class II Product Code: DYB Dated: July 3, 2013 Received: July 3, 2013
Dear Ms. Leo,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Jessica Leo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): K130687
Medcomp® Vascu-Sheath Tearaway Introducer Device Name:
Indications for Use:
The 2F and 3F Vascu-Sheath Tearaway Introducer is intended for percutaneous venous access by modified Seldinger Technique in neonates, infants, and children.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Bram D. Zuckerman -S 2013.08.15 16:24:12 -04'00'
Page _1_ of _1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.