PEDIATRIC FOLDING WHEELCHAIR

K130644 · Columbia Medical Mfg, LLC · LBE · Oct 16, 2013 · Physical Medicine

Device Facts

Record IDK130644
Device NamePEDIATRIC FOLDING WHEELCHAIR
ApplicantColumbia Medical Mfg, LLC
Product CodeLBE · Physical Medicine
Decision DateOct 16, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 890.3850
Device ClassClass 1
AttributesTherapeutic, Pediatric

Indications for Use

The Columbia Medical Pediatric Folding Wheelchair is a manual wheelchair designed to provide mobility aid for children and adolescents who have limited or no ability to ambulate and require a device to aid in seated mobility in the home and community.

Device Story

Manual pediatric wheelchair; provides mobility for children/adolescents unable to ambulate. Constructed of 6061 T6 aluminum tubing frame, steel components, 13.5" rear wheels, 6" front casters. Features adjustable seat/back angles to support skeletal/muscular characteristics. Designed for indoor/outdoor use on firm terrain. Foldable for transport/storage without tools. Operated by user or caregiver. Provides seated mobility; enhances patient independence and comfort.

Clinical Evidence

Bench testing only. Evaluated for front, rear, and sideway stability, dimensional conformance, fatigue, curb drop, and impact per RESNA standards. Biocompatibility testing performed on patient-contact materials (fabric seat/back, footrests).

Technological Characteristics

Frame: 1" diameter 6061 T6 aluminum tubing (powder coated), welded/bolted. Components: Steel fasteners, ABS footplates. Seating: WeatherMax FR fabric (flame retardant per CSFM T19, CPAI-84, FMVSS 302, ASTM E84 Class A). Manual operation; no energy source. Foldable form factor.

Indications for Use

Indicated for children and adolescents with limited or no ability to ambulate requiring seated mobility aid in home and community settings.

Regulatory Classification

Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130644 Image /page/0/Picture/1 description: The image shows the logo for Columbia Medical. The logo features the word "Columbia" in a stylized font, with the word "MEDICAL" in smaller letters underneath. Below that, the text "Solutions for Daily Living" is written in a smaller, different font. Section 7-1 ### 510(k) Summary-B Columbia Medical's Pediatric Folding Wheelchair 8/30/2013 #### 1) General Information | Submitter Name: | Columbia Medical, LLC | |----------------------|-------------------------------------------------| | Address: | 11724 Willake St.<br>Santa Fe Springs, CA 90670 | | Telephone: | (562) 282-0244 | | Fax: | (310) 305-1718 | | Contact Person: | Gary Werschmidt | | Registration Number: | 3005328749 | #### 2) Device | Device Trade Name: | Pediatric Folding Wheelchair | |-----------------------------|------------------------------| | Common/Generic Name: | Pediatric Wheelchair | | Device Classification Name: | Mechanical Wheelchair | | Registration Number: | 3005328749 | | Product Code: | LBE | | Device Classification: | Class 1 | | Regulatory Number: | 890.3850 | | K Number: | K130644 | ### 3) Identification of Legally Marketed Predicated Devices | Manufacturer Name: | Convaid Products, Inc | |--------------------|-----------------------| | Name: | Cruiser CP4M | | K Number: | K810676 | | Date Cleared: | 03/26/1981 | ### 4) Description of the device: This pediatric wheelchair, also commonly referred to as an adaptive stroller, is constructed of the same typical components of most manual wheelchairs. The frame consists of powder coated round tubular 6061 T6 aluminum which is 1" in diameter, and is welded and bolted. The device comes standard with 13.5" rear wheels and 6" steering front casters. The back and seating area is made of a fabric material. The wheelchair is designed for everyday indoor and outdoor use in firm terrain. The wheelchair is designed to allow for adjustments of the orientation of the seat and back angles, using tools, to provide seating and back angles to support the skeletal and muscular characteristics of the individual for comfort, control and safety during sitting and mobility. {1}------------------------------------------------ Section 7-2 Image /page/1/Picture/1 description: The image shows the logo for Columbia Medical. The word "Columbia" is written in a stylized font, with the word "MEDICAL" written in smaller letters underneath. Below the logo, the phrase "Solutions for Daily Living" is written in a simple font. ### 5) Indication For Use: The Columbia Medical Pediatric Folding Wheelchair is a manual wheelchair designed to provide mobility aid for children and adolescents who have limited or no ability to ambulate and require a device to aid in seated mobility in the home and community. #### 6) Technological Characteristics and Substantial Equivalence Device Description: The product consists of an aluminum tubing frame, steel components, rear wheels and smaller front pivoting casters for steering and turning. The product is designed to be lightweight, user adaptable, everyday wheelchair for both indoor and outdoor use. The Convaid Cruiser is made of similar tubing, aluminum and steel components, and is lightweight, user adaptable, and designed for both indoor and outdoor use. Similar to the Convaid Cruiser, the Columbia Medical Pediatric Folding Wheelchair frame is welded and botted and secured with fasteners to allow the size of the chair to be modified as needed to accommodate change or growth of the individual. The medical equipment provider can make many of the changes without additional parts. This wheelchair is designed to be folded for stowage and storage without the use of tools. Although the folding methodology for the Columbia Medical Pediatric Folding Wheelchair and the Convaid Cruiser is dissimilar, the result is a folded frame designed for easier transportability or storage. ### 7) Non-Clinical Testing Columbia Medical's Pediatric Folding Wheelchair was evaluated for front, rear and sideway stability tests, dimensional conformance, fatigue, curb drop, and impact tests, which are RESNA standards for the predicate device. Testing was performed to ensure that the Columbia Medical's device was equivalent to the predicate device. Additionally the base material in the fabric, WeatherMax FR, has flame retardant applications to meet CSFM T19, CPAI-84, FMVSS 302 and AST -E84 Class A. #### Previous Submissions for this Device 8) This device has not previously been submitted for FDA approval. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Columbia Medical. The logo features the word "Columbia" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font underneath. Below the logo is the phrase "Solutions for Daily Living" in a smaller, italicized font. The logo is simple and clean, and the text is easy to read. Section 7-3 #### 9) Patient contact with surfaces The surfaces of the wheelchair which might come in contact with the user are the fabric seat/back and the footrests while the user is seated in the wheelchair (less than 24 hour duration). The fabric seat and back are made of WeatherMax FR, in the family of Cordura, which is the similar fabric used by the predicate device fabric seat and back. The footplates are made of ABS which is similar to the material used for the footplates on the predicate device. Biocompatibility test reports for these materials are provided with this submission indicating materials are non-hazardous. #### 10) Safety The Columbia Medical Pediatric Folding Wheelchair is substantially equivalent to the predicated device listed in the 510(k). The construction and the design of this wheelchair do not raise any new issues of safety and effectiveness. #### 11) Conclusion: The Pediatric Folding Wheelchair shares similar materials, performance features and construction technology with the predicate device as well as a number of similar devices already legally marketed within the United States. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract image of an eagle with its wings spread. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hamnshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002 #### October 16, 2013 Columbia Medical Manufacturing, LLC c/o Gary Werschmidt. Chief Executive Officer 11724 Willake St. Santa Fe Springs, CA 90670 Re: K130644 Trade/Device Name: Pediatric folding wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: LBE Dated: September 4, 2013 Received: September 6, 2013 Dear Mr. Werschmidt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ Page 2 - Mr. Gary Werschmidt comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Joyce M. Whang -S - for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K130644 Device Name: Pediatric Folding Wheelchair Indications For Use: The Columbia Medical Pediatric Folding Wheelchair is a manually propelled wheelchair designed to provide mobility aid for children and adolescents who have limited or no ability to ambulate and require a device to aid in seated mobility in the home and community. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use ___x_ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) # Joyce M. Whang -S Page 1 of __1___
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...