The Barry™ Channel RFA Endoscopic Catheter is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).
Device Story
Barrx™ Channel RFA Endoscopic Catheter is a sterile, single-use bipolar electrosurgical device; delivers radiofrequency (RF) energy to GI tissue via copper electrode. Used exclusively with Flex RFA Energy Generator (K092487). Physician introduces catheter through endoscope working channel; positions electrode against target tissue by deflecting endoscope; activates energy via generator front panel or foot-pedal. Generator identifies catheter via unique plug ID; automatically sets power/energy density. Output allows coagulation of bleeding/non-bleeding sites. Benefits include targeted tissue treatment for various GI pathologies.
Clinical Evidence
Bench testing only. Testing included catheter length, shaft revolution, endoscope articulation, push/retraction/deflection/torque force, stability, bend radius, electrical insulation, apposition, simulated use, esophageal damage, and tensile force. Biocompatibility, sterilization, packaging, shelf life, and EMC/electrical safety testing also performed.
Technological Characteristics
Sterile, single-use bipolar electrosurgical catheter. Copper electrode. Energy source: Flex RFA Energy Generator. Dimensions: Models TTS-1100 and TTS-1500. Connectivity: Proprietary plug-based identification for generator communication. Sterilization: Not specified.
Indications for Use
Indicated for coagulation of bleeding and non-bleeding GI tract sites, including esophagus. Conditions include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, GAVE, and Radiation Proctitis.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
# 510(k) Summary
## SUBMITTER'S NAME, ADDRESS, TELEPHONE NUMBER, CONTACT PERSON AND DATE PREPARED
Covidien IIc 15 Hampshire Street Mansfield, MA 02048 Phone: (408) 328-7342 Facsimile: (408) 328-7342 (same as phone#) Contact: Richelle Hover Date Prepared: March 7, 2013
JUL 2 6 2013
#### NAME OF SUBJECT DEVICE AND NAME
Barrx™ Channel RFA Endoscopic Catheter
#### ESTABLISHMENT REGISTRATION NUMBER/OWNER OPERATOR NUMBER
Establishment Registration Number: 3004904811 Owner/Operator Number: 1282497
Legal Manufacturer: Covidien, Ilc 15 Hampshire Street Mansfield, MA 02048
Manufacturing Facility: Covidien, Formerly BÂRRX Medical, Inc. 540 Oakmead Parkway Sunnyvale, CA 94085
## COMMON OR USUAL NAME
Electrosurgical Coagulation Catheter
#### REGULATION DESCRIPTION
Classification: Class II, 21 CFR 876.4300 Product Code: GEI, KNS
## PREDICATE DEVICES
HALO® Catheter cleared under 510(k) K112454 HALO90 Catheter cleared under 510(k) K093008
## DEVICE DESCRIPTION
The subject device, the Barrx™ Channel RFA Endoscopic Catheter is a sterile single-use bipolar device that delivers radiofrequency (RF) energy to the treatment tissue within the gastrointestinal tract through a copper electrode. The catheter is used exclusively with the HALOFEN Energy Generator (cleared under 510(k) K092487).
{1}------------------------------------------------
#### INDICATION FOR USE
The Barry " Channel RFA Endoscopic Catheter is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).
#### TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMAPRED TO PREDICATE DEVICE
The Barry " Channel RFA Endoscopic Catheter, the HALO" catheter, and the HALO" catheter have similar technological characteristics. All catheters are single-use, have similar construction, similar principles of operation, control mechanism, similar materials, same packaging, and the same energy type.
The main difference between the Barrx™ Channel RFA Endoscopic Catheter and the HALO® and HALO® catheters is the channel through which the catheter is delivered. The HALO® and HALO® catheters are delivered to the target tissue by mounting the device to the endoscope, whereas the Barrx™ Channel RFA Endoscopic Catheter is delivered through an endoscope. The Barrx™ Channel RFA Endoscopic Catheter is compatible with the Flex RFA Energy Generator only (K092487). The HALO'ER Energy Generator will be brand marketed as Flex RFA Energy Generator. No software change has been made to the Flex RFA Energy Generator in order to accommodate the Barrx™ Channel RFA Endoscopic Catheter.
#### PRINCIPLES OF OPERATION
The Barry" Channel RFA Endoscopic Catheter follows the same principles of operation as compared to the predicates HALO® (K112454) and HALO® catheters (K093008). The Barrx™ Channel RFA Endoscopic Catheter is connected to the Flex RFA Energy Generator using an output cable. Once connected, the Generator will recognize the catheter based on a unique identification in the plug and set the appropriate power density and energy density range.
The Barry " Channel RFA Endoscopic Catheter is introduced into the esophagus through the working channel of an endoscope. Once the targeted treatment area is identified, the catheter electrode is positioned against the tissue by deflecting the endoscope. The energy activation is performed by depressing either a front panel switch on the generator or the foot-pedal. After the energy is delivered, the coagulation effect can be verified.
## SUMMARY OF TESTING PERFORMED
Verification and validation activity for the Barry " Channel RFA Endoscopic Catheter consisted of bench testing, biocompatibility testing, sterilization, packaging validation, shelf life testing, EMC and electrical safety testing, and user validation. Bench testing for the Barry" Channel RFA Endoscopic Catheter consisted of the following tests: (1) catheter length testing (2) shaft revolution testing (3) endoscope articulation testing (4) push force testing (5) retraction force testing (6) electrode deflection force testing (7) torque force and stability testing (8) bend radius testing (9) electrical insulation testing (10) apposition testing (11) simulated use testing (12) esophageal damage test (13) tensile force test.
{2}------------------------------------------------
#### CONCLUSION
Covidien, Ilc considers the Barry™ Channel RFA Endoscopic Catheter to be substantially equivalent to legally marketed predicates: HALO® Catheter cleared under 510(k) K112454 and HALO® catheters cleared under 510(k) K093008.
The test results and compliance with applicable standards provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its indications for use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human services.
#### Public Health Service
Food and Drug Administration 10903 New Hamoshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## July 26, 2013
Covidien, LLC % Richelle Hover Regulatory Affairs Specialist 15 Hampshire Street Mansfield, MA 02048
Re: K130623
Trade/Device Name: Barrx™ Channel RFA Endoscopic Catheter; Models TTS-1100 and TTS-1500 Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: KNS, GEI Dated: June 20, 2013 Received: June 21, 2013
Dear Richelle Hover,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/7 description: The image shows the name "Benjamin" in large, bold letters. The letters "R" and "A" are stylized with a decorative pattern. The last name "Fisher" is also present, followed by the letter "S".
Benjamin R. Fisher, Ph.D. · Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use Statement
510(k) Number (if known): K130623
#### Device Name:
- Barry™ Channel RFA Endoscopic Catheter; Models TTS-1100 and TTS-1500 .
#### Indications for Use:
The Barry™ Channel RFA Endoscopic Catheter is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, Angiodysplasia, Gastric Antral Vascular Ectasia (GAVE) and Radiation Proctitis (RP).
Prescription Use X (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher -S 2013.07.26 14:52:22 -04'00'
Barry™ Channel RFA Endoscopic Catheter - Traditional 510(k) Covidien, Ilc
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K130623 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.