BONE FRAGMENT AND FIXATION PLATES, SCREWS AND WASHERS
K130614 · Laura Cattabriga · HRS · Jan 17, 2014 · Orthopedic
Device Facts
Record ID
K130614
Device Name
BONE FRAGMENT AND FIXATION PLATES, SCREWS AND WASHERS
Applicant
Laura Cattabriga
Product Code
HRS · Orthopedic
Decision Date
Jan 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Core Orthopaedics bones plates and screws system are intended to be used for internal fixation of bone fractures, fusions or osteotomies in the ankle, foot, hand, wrist, clavicle, scapula, pelvis, long bone (such as humerus, ulna, radius, femur, tibia, and fibula, and small bones (such as metacarpals, metatarsals, phalanges). The washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of large bone and bone fragments.
Device Story
System consists of stainless steel bone plates, screws, and washers of various sizes and shapes. Used for internal fixation of fractures, fusions, or osteotomies in various skeletal sites. Washers function with screws to distribute load and prevent cortical breakthrough. Distributed non-sterile; sterilized by end-user via steam sterilization. Used in clinical/surgical settings by orthopedic surgeons.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence established through design, material, and dimensional comparison to predicate devices.
Technological Characteristics
Stainless steel bone fixation plates, screws, and washers. Non-sterile, steam-sterilized by end-user per AAMI/ANSI/ISO 11737 guidelines to SAL 10-6. Mechanical fixation device.
Indications for Use
Indicated for internal fixation of bone fractures, fusions, or osteotomies in ankle, foot, hand, wrist, clavicle, scapula, pelvis, long bones (humerus, ulna, radius, femur, tibia, fibula), and small bones (metacarpals, metatarsals, phalanges). Washers indicated to prevent screw head breakthrough by distributing load.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Internal Fixation Systems, Inc. Bone Plates Screws and Washers (K110086)
{0}------------------------------------------------
K130614
# JAN 17 2014
## 510(k) Summary
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Bone Fragment Fixation Plates and Screws device is provided below.
Device Common Name: Plate, Fixation, Bone Screw, Fixation, Bone Washer
Device Proprietary Name: Bone Fragment Fixation Plates, Screws and Washers
- Submitter: Laura Cattabriga 7430 Center Bay Drive N. Bay Village. Florida 33141
Contact:
Laura Cattabriga 7430 Center Bay Drive N. Bay Village, Florida 33141 email: Laurac@core-orthopaedics.com
#### Classification
21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener Regulation: (HWC)
> 21 CFR 888.3030: Single / multiple component metallic bone fixation appliances and accessories (HRS. HTN)
Panel: Orthopedics
HWC, HRS, HTN Product Code:
Indication for Use: The Core Orthopaedics bones plates and screws system are intended to be used for internal fixation of bone fractures, fusions or osteotomies in the ankle, foot, hand, wrist, clavicle. scapula. pelvis. long bone (such as humerus. ulna. radius. femur. tibia. and small bones (such as metacarpals, metatarsals, phalanges). The washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of large bone and bone fragments.
Device Description: The bone fragment fixation plates and washers of various sizes, consists of various shapes as well as screws are intended to treat fractures and osteotomies of various bones. including the clavicle, acetabulum, pelvis, scapula, humerus, radius, ulna, femur, tibia, phalanges, carpals, metacarpals, tarsals, metatarsals and fibula. The washers are to function with
{1}------------------------------------------------
screws to prevent the screw head from breaking through the cortex of the bone by providing additional surface area during screw placement. The plates, screws, and washers are a one-piece device made of stainless steel.
Performance Data: The Bone Fragment Fixation Plates Screws and Washers are substantially equivalent to the predicate devices with respect to design and material and dimensional comparison.
### Substantial Equivalence:
- Internal Fixation Systems. Inc. Bone Plates Screws and Washers . (KII0086)
- Synthes Spherical Washers (K052483) .
Technological Characteristics Comparison: The Bone Fragments Fixations Plates and Screws are equivalent to the predicate devices with respect to design. material and dimensional comparison.
Sterilization Information: The Bone Fragment Fixations Plates and Screws will be distribute non-sterile. The devices are sterilized by the end user per the AAMI Guidelines "Good Hospital Practice: Steam Sterilization and Sterility Assurance" and ANSI/AAMI/ISO 11737 guidelines to achieve the Sterility Assurance Level (SAL) of 10-6.
Conclusion: There are no significant differences between the Bone Fragment Fixations Plates and Screws and the other implants as listed in the Substantially Equivalence Devices. The Bone Fragment Fixations Plates and Screws and the predicate devices have similar design attributes, materials, and intended use thus is substantially equivalent.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the logo is an abstract symbol that represents human services.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
January 17, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Laura H. Cattabriga Chief Operating Officer / Chief Financial Officer 7430 Center Bay Drive Miami, Florida 33143
Re: K130614 Trade/Device Name: Bone Fragment Fixation Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: December 17, 2013 Received: December 18, 2013
Dear Ms. Cattabriga:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Ms. Laura H. Cattabriga
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on vour responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Mark Nijhelkerson -S
- Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Premarket Notification
Bone Fragment Fixation
Plates, Screws and Washers
# Indications for Use Statement
510(k) Number (if known): K130614
#### Device Name: Bone Fragment Fixation Plates and Screws
Indications For Use:
י
The Core Orthopaedics bones plates and screws system are intended to be used for internal fixation of bone fractures, fusions or osteotomies in the ankle, foot, hand, wrist, clavicle, scapula, pelvis, long bone (such as humerus, ulna, radius, femur, tibia, and fibula, and small bones (such as metacarpals, metatarsals, phalanges). The washers are intended to prevent a screw head from breaking through the cortex of the bone by distributing the forces/load over a large area when used for fracture fixation of large bone and bone fragments.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth L. Frank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.