Prismatik's Universal Paste Stains and Glaze are intended to be used in dental applications for coloration and finishing of glass ceramic and zirconia-based restorations.
Device Story
Universal Paste Stains and Glaze are silicate sintered glass ceramic materials used for dental restoration finishing. The system includes 17 stain shades, a fluorescent paste glaze, and a thinning liquid. The device is used by dental professionals to apply color and esthetic finishing coats to glass ceramic and zirconia-based restorations. The fluorescent glaze provides esthetic properties under UV light. The liquid is used to adjust paste consistency and clean brushes. The device is applied to the surface of restorations, which are then sintered to achieve the final finish, simulating natural tooth dentine.
Clinical Evidence
Bench testing only. Biocompatibility testing performed including cytotoxicity (DIN EN ISO 10993-5) with negative results. Chemical solubility testing performed according to DIN EN ISO 6872, showing composition and solubility similar to predicate.
Technological Characteristics
Silicate sintered glass ceramic porcelain powder. Includes stain pastes, fluorescent paste glaze, and thinning liquid. Non-sterile. Biocompatible per ISO 10993-5. Solubility per DIN EN ISO 6872. Requires machining and sintering.
Indications for Use
Indicated for coloration and finishing of glass ceramic and zirconia-based dental restorations.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
K130604
# 005-510 (k) Summary-807.92(c)
This 510 (K) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
### A. SUBMITTER INFORMATION
| Company Name: | Prismatik Dentalcraft, Inc. |
|-------------------------------|-------------------------------------------------------------------------|
| Company Address: | 2212 Dupont Dr., Suite IJK,<br>Irvine, CA 92612 |
| Company Phone: | 949-225-1269 |
| Company FAX: | 949-553-0924 |
| Facility Registration Number: | 3005477956 |
| Primary Contact Person: | Armin Zehtabchi, (949) 225-1234<br>Senior RA/QA, 510(K) Project Manager |
| Secondary Contact Person: | Marilyn Pourazar, (949) 225-1269<br>Senior Director, RA/QA |
| Date Summary Prepared: | June 21, 2013 |
### DEVICE IDENTIFICATION B.
·
| Trade/Proprietary Name: | Universal Paste Stains and Glaze |
|-------------------------|-----------------------------------|
| 21 CFR Reference: | 21 CFR 872.6660 |
| 21 CFR Common Name: | Porcelain powder for clinical use |
| Classification: | Class II |
| Product Code: | EIH |
| Panel: | Dental |
### IDENTIFICATION OF PREDICATE DEVICE C.
Trade/Proprietary Name: 3M's Bellus Shading Kit-K090718
{1}------------------------------------------------
#### DEVICE DESCRIPTION D.
Prismatik's Universal Paste Stains and Glaze are based on Silicate Sintered Glass Ceramic that is classified as Porcelain powder for clinical use (21 C.F.R. § 872.6660) and are available in a variety of colors. They include stain pastes, a glazing paste, and a liquid which can be used to thin the pastes. The pastes serve solely for the color staining and glazing of the surfaces of restorations.
The Universal Paste Stains and Glaze contains 17 stain shades, Fluorescent Paste Glaze, and Stain and Glaze Liquid, which are all silicate glass based.
The Stains are available in colors A, B, C, D and A Light, C Light, C Light, D Light, as well as White, Yellow, Orange, Brown, Dark Brown, Blue, Purple, Dark Pink, and Grey. The Fluorescent Paste Glaze is used to achieve an esthetic finishing coat. It also provides fluorescent properties under ultra-violet lighting. The Stain and Glaze Liquid can be mixed with the pastes in order to modify the consistency, and can also be used to clean the brush.
Image /page/1/Picture/4 description: The image shows a collection of small, round containers arranged in a rectangular tray. Each container has a label indicating a different color or shade, such as "Shade A," "Shade B," "yellow," "orange," "purple," and "grey." The containers appear to hold stain paste, as indicated by the text on the labels. There is also a bottle and a small tool in the tray.
{2}------------------------------------------------
# Work flow past production:
The work flow paste production includes melting, milling/sieving, shading and paste preparation, and the process flow is demonstrated below:
Image /page/2/Picture/2 description: The image shows a logo with the words "estetic ceram" written in a simple, sans-serif font. Above the text is a circular design made up of various sized dots, creating a visual element that resembles a stylized flower or burst. The overall design is clean and modern, suggesting a brand identity for a company related to aesthetics or ceramics.
Work flow paste production, glaze/stains/shades ZT
- 1. Melting
- 2. Milling/Sieving
- ကဲ Shading
- 4. Paste preparation
Image /page/2/Figure/8 description: The image shows a flowchart of a process. The process starts with four different steps: Melting, Milling/Sieving, Shading, and Paste Preparation. Each of these steps leads to further sub-steps, such as "Melting in continuous melting furnace" and "Jet milling", eventually leading to the final step, "Finished Product".
Image /page/2/Figure/9 description: The image shows the logo and contact information for estetic ceram ag. The address is Austrasse 15, Postfach 223, 9495 Triesen, Liechtenstein. The phone number is +423 237 48 58, and the fax number is +423 237 48 39. The email address is info@esteticceram.com, and the website is www.esteticceram.com.
{3}------------------------------------------------
### INDICATIONS FOR USE E.
Prismatik's Universal Paste Stains and Glaze are intended to be used in dental applications for coloration and finishing of glass ceramic and zirconia-based restorations.
### SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE NEW F. DEVICE AND THE PREDICATE DEVICES
The following comparison table of the technological characteristics of the new device and the predicate devices outlines and provides the similarities and the substantial equivalency of the Prismatik's" Universal Paste Stains and Glaze and the 3M's Bellus Shading Kit-K090718.
| Elements of<br>Comparison | Prismatik™ Universal Paste Stains<br>and Glaze | 3M's Bellus Shading Kit-<br>K090718 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| General Material | Powder, porcelain | Same |
| Indications | Prismatik's Universal Paste Stains<br>and Glaze are intended to be used in<br>dental applications for coloration and<br>finishing of glass ceramic and<br>zirconia-based restorations. | Color staining and glazing of<br>glass ceramic restorations<br>made from 3M ESPE's Glass<br>Ceramics "Jolly." |
| Biocompatibility | Yes | Same. |
| Sterility | Non-sterile | Same |
| Machining and<br>Sintering | Yes | Same |
| Performance | Simulating the natural tooth dentine | Same |
## Comparison of the Technological Characteristics of the New Device and the Predicate Devices
{4}------------------------------------------------
#### DETERMINITION OF SUBSTANTIAL EQUIVALENCE G.
The above comparison table of the technological characteristics of the new device and the predicate devices was provided for the substantial equivalency of the Prismatik's Universal Paste Stains and Glaze and the 3M's Bellus Shading Kit-K090718. Prismatik believes that the comparative data presented, demonstrate that Prismatik ""Universal Paste Stains and Glaze are essentially the same as currently marketed devices for the same indication for use, and supports our claim of substantial equivalence to predicate Class II devices under the classification of Porcelain powder for clinical use (21 CFR 872.6660) that have previously been found to be substantially equivalent. Both the new and the predicate device consist of general porcelain powder material (Product Code: EIH), that is biocompatible for the same indication for use.
#### SUMMARYOF NON-CLINICAL TESTING H.
Non-clinical test data was used to support the substantial equivalency. To provide evidence for safety, a biocompatibility testing was carried out. The raw materials were tested for cytotoxicity (acc. DEN EN ISO 10993-5) with negative result. From chemical point of view, the porcelains investigated were similar in composition and show similar solubility acc DIN EN ISO 6872.
#### CONCLUSION FROM THE NON-CLINICAL TESTING I.
The results of the above described studies demonstrate that Prismatik's Universal Paste Stains and Glaze is as safe and effective as the cleared predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure with outstretched arms.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### September 12, 2013
Prismatik Dentalcraft, Inc. C/O Mr. Armin Zehtabchi Senior RA Specialist 2212 Dupont Drive, Suite IJK Irvine, CA 92612
Re: K130604
Trade/Device Name: Universal Paste Stains and Glaze Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Codes: EIH Dated: June 26, 2013 Received: June 27, 2013
Dear Mr. Zehtabchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Zehtabchi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note
the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for Prismatik Dentalcraft, Inc. The logo consists of a triangle on the left and the company name on the right. The triangle is filled with a pattern of black and white shapes. The text is in a simple, sans-serif font.
004-Indications for Use Statement
510 (K) Number (if known): K I 30604
Universal Paste Stains and Glaze Device Name:
Indications for Use: Prismatik's Universal Paste Stains and Glaze are intended to be used in dental applications for coloration and finishing of glass ceramic and zirconia-based restorations.
Prescription Use: Yes > No _ (Part 21 CFR 801 Subpart D)
Over-the-Counter Use: Yes [ No Z (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BEL.OW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
# Andrew I. Steen -S 2013.09.12 11:34:27 -04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Respiratory, Infection Control and Dental Devices 510 (k) Number :
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.