K130602 · Biosense Webster, Inc. · DRF · Aug 14, 2013 · Cardiovascular
Device Facts
Record ID
K130602
Device Name
LASSO NAV DUO LOOP CATHETER
Applicant
Biosense Webster, Inc.
Product Code
DRF · Cardiovascular
Decision Date
Aug 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The LASSO® NAV Duo Loop eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart. The LASSO® NAV Duo Loop eco Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.2.)
Device Story
The LASSO® NAV Duo Loop eco Catheter is a diagnostic electrophysiology catheter used to map atrial cardiac structures. It features a fixed spiral loop distal tip with 20 platinum/iridium electrodes for recording intracardiac signals and cardiac stimulation. The device is inserted into the right or left atrium via an 8F guiding sheath. It incorporates a tri-axial magnetic location sensor for integration with the CARTO® 3 EP Navigation System (v3.2 or later) to provide real-time location data. The catheter includes a bi-directional deflection mechanism for maneuverability. Physicians use the device to obtain electrograms, which are transmitted to external recording equipment and the navigation system for display and analysis. This information assists clinicians in identifying and diagnosing heart arrhythmias. The device benefits patients by facilitating precise anatomical and electrical mapping of the heart during electrophysiological procedures.
Clinical Evidence
Bench testing only. Validation included integrity and performance testing after ethylene oxide sterilization, thermal cycling, accelerated aging, and simulated transportation. Compatibility with the CARTO® 3 EP Navigation System was verified. No clinical data provided.
Technological Characteristics
7 Fr diameter catheter with 20 platinum/iridium ring electrodes in a fixed spiral loop configuration. Features tri-axial magnetic location sensor for navigation. Bi-directional deflection mechanism. Connects to CARTO® 3 EP Navigation System and standard recording equipment via accessory cables. Sterilized via ethylene oxide.
Indications for Use
Indicated for multiple electrode electrophysiological mapping (recording or stimulation) of cardiac structures, specifically atrial regions, in patients requiring cardiac arrhythmia diagnosis.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
Predicate Devices
Biosense Webster LASSO® 2515 NAV eco Variable Catheter (K113213)
Reference Devices
Biosense Webster Webster® CS Catheter with EZ Steer Technology (K101345)
Submission Summary (Full Text)
{0}------------------------------------------------
### 2. 510(K) SUMMARY
# K130602
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | Biosense Webster, Inc.<br>3333 Diamond Canyon Road<br>Diamond Bar, CA 91765 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| OFFICIAL<br>CORRESPONDENT | John Jimenez<br>Senior Specialist, Regulatory Affairs<br>Telephone: 909-839-8534<br>Fax: 909-839-8804<br>Email: jjimene7@its.jnj.com<br>AUG 1 4 2013 |
| TRADE NAME | LASSO® NAV Duo Loop eco Catheter |
| COMMON NAME | Electrophysiological Mapping Catheter |
| CLASSIFICATION NAME | Electrode recording catheter or electrode recording<br>probe |
| DEVICE CLASSIFICATION | Class II, 21 CFR §870.1220<br>Product Code: DRF |
| PRODUCT CODES | D-1354-01-S, D-1354-02-S, D-1354-03-S<br>D-1354-04-S, D-1354-05-S, D-1354-06-S |
| PREDICATE DEVICE | Biosense Webster LASSO® 2515 NAV eco Variable<br>Catheter (510(k) K113213) |
| REFERENCE DEVICE | Biosense Webster Webster® CS Catheter with EZ Steer<br>Technology (510(k) K101345) |
#### SUBSTANTIALLY EQUIVALENT TO:
The Biosense Webster LASSO®NA V Duo Loop eco Catheter is substantially equivalent to the Biosense Webster LASSO® 2515 NAV eco Variable Catheter [510(k) K113213, cleared December 5, 2011] and the Biosense Webster® CS Catheter with EZ Steer Technology [510(k) K101345, cleared May 13, 2010]. Like the predicate device, the LASSO®NAV Duo Loop eco Catheter features a tri-axial magnetic location sensor, allowing the ability to provide location information when used with the CARTO® 3 EP Navigation System, Version 3.2. The proposed catheter also has a similar circular loop of 20 electrodes at the distal tip of a 7 Fr diameter catheter, the same patient contact materials, and the same catheter connectivity as the predicate device. The original
{1}------------------------------------------------
intended use of the predicate device for diagnosis of heart arrhythmias remains the same in the proposed device: however, the proposed device will now have the added ability of deflecting in two directions (bi-directional). This latter capability is substantially equivalent to the Reference Device, the Webster® CS Catheter with EZ Steer Technology. The Webster® CS Catheter with EZ Steer Technology went through this same transition with FDA when FDA cleared the update of the uni-directional version to the bi-directional version.
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The LASSO*NAV Duo Loop eco Catheter has been designed to facilitate electrophysiological mapping of the atria of the heart with the CARTO® 3 EP Navigation System and a reference device. The catheter is deployed in the right or left atrium through an 8F guiding sheath. The bi-directional catheter consists of a circular spine on the distal tip, with platinum/iridium electrodes that can be used for stimulation and recording. The purpose of this Premarket Notification is to create a bi-directional fixed multi-loop high density circular diagnostic catheter for recording intracardiac signals and cardiac stimulation. The modifications consist of changing from a single variable loop to a fixed spiral loop, including an insertion tool with atraumatic tip to the catheter, and incorporating a bi-directional deflection in two directions) feature to the distal end of the catheter. The new catheter will retain the extended barrel ("pig tail") connector and will connect to the system cable by way of an accessory cable.
The LASSO® NAV Duo Loop eco Catheter features a fixed spiral loop design available in two diameters, 20mm diameter and 25mm diameter. Each catheter with 20 electrodes spaced along its multiple loops in a uni-polar configuration for the 20mm and either a uni-polar or bi-polar configuration for the 25mm.
The proposed catheters will continue to interface with standard recording equipment via interface cables and appropriate connectors.
# INDICATIONS FOR USE:
The LASSO® NAV Duo Loop eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart.
The LASSO® NAV Duo Loop eco Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO 3 EP Navigation Systems prior to Version 3.2.)
# TECHNICAL CHARACTERISTICS:
The LASSO® NAV Duo Loop eco Catheter is a typical electrophysiological catheter that is unique only in its geometrical arrangement of 20 ring electrodes spaced along its
{2}------------------------------------------------
multiple loops. Otherwise, there are no special technical aspects of the ability of this catheter to detect electrical signals from heart endocardium and transmit this information to the CARTO® 3 EP Navigation System and/or recording equipment for display, analysis, and interpretation in detection of various heart arrhythmias.
### PERFORMANCE DATA:
The safety and performance of the LASSO® NAV Duo Loop eco Catheter have been validated through Bench Testing, Biocompatibility Testing, GLP Animal Study, and System Level Testing. Bench testing evaluated the integrity and performance of the LASSO® NAV Duo Loop eco Catheter following three times ethylene oxide (EtO) sterilization, thermal cycling, accelerated aging (equivalent to one year real-time aging), and simulated transportation conditions. After thorough investigations and mitigations where appropriate, the device passed all intended criteria in accordance with appropriate test criteria and standards. In addition, the proposed catheter was tested with the CARTO® 3 EP Navigation System to demonstrate compatibility. Based on the results of the risk analysis and the bench testing performed, it is concluded that the minor differences in design and use that exist between the predicate device and the LASSO® NAV Duo Loop eco Catheter do not compromise the safety and efficacy of the catheter. This testing demonstrates that the subject device is at least as safe and effective as the legally marketed predicate devices.
#### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
The technological comparison and nonclinical studies demonstrate that the LASSO® NAV Duo Loop eco Catheter is safe and effective for anatomic mapping of the heart and establish equivalence to its predicate and reference devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name arranged in a circular pattern around the perimeter. At the center of the seal is a stylized emblem, consisting of a white abstract design resembling an eagle or bird-like figure, set against a dark background.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
August 14, 2013
Biosense Webster, Inc. C/O John Jimenez 3333 Diamond Canyon Rd Diamond Bar, CA 91765 US
Re: K130602
> Trade/Device Name: Lasso nav duo loop catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Regulatory Class: Class II Product Code: DRF Dated: July 1, 2013 Received: July 2, 2013
Dear John Jimenez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
# Page 2 - John Jimenez
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Owen Farris -S
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 1. INDICATIONS FOR USE STATEMENT
510(k) No (if known): K130602
Device Name: LASSO® NAV Duo Loop eco Catheter
Indication for Use:
The LASSO® NAV Duo Loop eco Catheter is indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. The catheter is designed to obtain electrograms in the atrial regions of the heart.
The LASSO® NAV Duo Loop eco Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.2.)
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by
Owen P. Faris -5
Date: 2013.08.14
08:30:52 -04'00'
Page of of age
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.