POWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)

K130582 · Heze City Green Food Co., Ltd. · LYZ · Feb 4, 2014 · General Hospital

Device Facts

Record IDK130582
Device NamePOWDER FREE VINYL PATIENT EXAMINATION GLOVES, CLEAR (NON-COLORED)
ApplicantHeze City Green Food Co., Ltd.
Product CodeLYZ · General Hospital
Decision DateFeb 4, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Story

Device is a non-sterile, powder-free vinyl patient examination glove. It functions as a physical barrier to body fluids and bloodborne pathogens. The PVC material is water-tight and flexible, conforming to the hand to facilitate medical procedures. Used in clinical settings by healthcare examiners to prevent cross-contamination between patient and examiner. The device is a single-use, disposable product.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical property testing (tensile strength, elongation), freedom from pinholes (AQL 2.5), residual powder content (<2mg/glove), and biocompatibility (skin irritation and dermal sensitization per ISO 10993-10).

Technological Characteristics

Material: Poly (vinyl chloride) (PVC). Surface coating: Polyurethane (PU). Standards: ASTM D5250-06 (physical properties/dimensions), ASTM D5151-06 (pinholes), ASTM D6124-06 (residual powder), ISO 10993-10 (biocompatibility). Non-sterile. Single-use.

Indications for Use

Indicated for use as a non-sterile, disposable patient examination glove worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### Section C 510(k) Summary # 510(K) Summary "This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92." "The assigned 510(k) number is:_______________________________________________________________________________________________________________________________________________ 그 ## Premarket Notification [510(k)] Summary #### 1.0 Submitter: | Submitter's name : | Heze City Green Food Co.,Ltd. | |-------------------------|-------------------------------------------------------------------| | Submitter's address : | Jinzhuang Zone, Suji Caoxian, Heze<br>City,Shandong, 274415 China | | Phone number : | 0086-530-5022111 | | Fax number : | 0086-530-5311988 | | Name of contact person: | Mr.Jerry Zhang | Date of preparation: 2014-01-23 #### Name of the Device 2.0 | Device Name: | Powder Free Vinyl Patient Examination<br>Gloves, Clear (non-colored) | |-------------------------|----------------------------------------------------------------------| | Proprietary/Trade name: | Powder Free Vinyl Patient Examination<br>Gloves, Clear (non-colored) | | Common Name: | Exam gloves | | Classification Name: | Patient examination glove | | Device Classification: | 1 | | Regulation Number: | 21 CFR 880.6250 | | Panel: | General Hospital (80) | | Product Code: | LYZ | ### 3.0 Predicate device | Device Name: | Powder-Free Vinyl Patient Examination Glove | |----------------|---------------------------------------------------| | | (Non-colored) | | Company name: | Zhang Jia Gang Fengyuan Plastic Product Co., Ltd. | | 510(K) Number: | K091663 | ### 4.0 Device Description: - How the device functions: 4.1 PVC films form a barrier to body fluids and bloodborne Pathogens #### 4.2 Scientific concepts that form the basis for the device The PVC rubber is water tight under normal conditions of use. Its tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Section C(rev.03) {1}------------------------------------------------ #### Physical and performance characteristics such as design, materials and 4.3 physical properties: Poly (vinyl chloride) glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements. ## 5.0 Device Intended Use (Indication for use): Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. ### 6.0 Summary of the Technological Characteristics of the Device: The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard. | Characteristics | Standard | Device performance | |--------------------------|--------------------------------------------------------------------------------|---------------------------------------| | Dimension | ASTM standard D<br>5250-06(Reapproved 2011). | Meets | | Physical Properties | ASTM standard D<br>5250-06(Reapproved 2011). | Meets | | Freedom from<br>pinholes | 21 CFR 800.20 | Meets | | Powder Residual | ASTM standard D 5250-06<br>(Reapproved 2011).and<br>D6124-06(Reapproved 2011). | Meets<br><2mg/glove | | Biocompatibility | Primary Skin Irritation in<br>rabbits<br>ISO 10993-10:2002<br>/Amd.1:2006 | Passes<br>Not a Primary Skin Irritant | | | Dermal sensitization in the<br>guinea pig<br>ISO 10993-10: 2002<br>/Amd.1:2006 | Passes<br>Not a Dermal sensitizer | ## 7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), meet per requirements per ASTM D5250-06 (Reapproved 2011)、 ASTM CFR 800.20 10993-10: D6124-06(Reapproved 2011), per 21 and ISO 2002/Amd. 1 :2006. The performance test data of the non-clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above. ### 8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data: Clinical data is not needed for gloves or for most devices cleared by the 510(k) process. Section C(rev.03) {2}------------------------------------------------ | Features & Description | Predicate Device | Medical Guidance Manual(1661) | Glove Subject Device | Result of Comparison | | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | Company | Zhang Jia Gang<br>Fengyuan Plastic<br>Product Co., Ltd.. | | Heze City Green Food<br>Co.,Ltd. | -- | | | 510(K) Number | K091663 | | K130582 | | | | Product name | Powder Free Vinyl<br>Patient Examination<br>Gloves, Clear<br>(non-colored) | | Powder Free Vinyl<br>Patient Examination<br>Gloves, Clear<br>(non-colored) | same | | | Product Code | LYZ | LYZ | LYZ | same | | | Size | Small/Medium/<br>Large/X large | | Small/Medium/<br>Large/X large | same | | | Intend for use | Powder Free Vinyl<br>Patient Examination<br>Gloves, Clear (non-<br>colored) is a disposable<br>device intended for<br>medical purposes that<br>is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | Powder Free<br>Examination Gloves: A<br>powder-free patient<br>examination glove is a<br>disposable device<br>intended for medical<br>purposes that is worn<br>on the examiner's hand<br>or finger to prevent<br>contamination between<br>patient and examiner. | Powder Free Vinyl<br>Patient Examination<br>Gloves, Clear<br>(non-colored) is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hand or<br>finger to prevent<br>contamination between<br>patient and examiner. | Substantially<br>equivalent | | | Device<br>Description and<br>Specifications | Meets ASTM<br>D5250-06<br>(Reapproved 2011) | If vinyl:<br>Do the vinyl<br>examination gloves<br>meet all the current<br>specifications listed<br>under ASTM<br>Specification D5250<br>or an equivalent<br>consensus standard? | Meets ASTM<br>D5250-06<br>(Reapproved 2011) | Substantially<br>equivalent | | | Dimensions<br>--Length | Meets ASTM<br>D5250 -06<br>(Reapproved 2011)<br>≥230mm min | ASTM D5250 | 230mm min for all sizes | Substantially<br>equivalent | | | Dimensions<br>-- Width | Meets ASTM<br>D5250-06<br>(Reapproved 2011) | ASTM D5250 | | Substantially<br>equivalent | | | | Small 80-90 mm<br>Medium 90-100mm<br>Large 100-110mm<br>Xlarge 110-120 mm | | Small 80-85 mm<br>Medium 95-97 mm<br>Large 102-108mm<br>X large 114-118 mm | | | | Dimensions<br>--Thickness | Meets ASTM<br>D5250-06<br>(Reapproved 2011) | | Thickness (mm) min.<br>Finger 0.09-0.12<br>Palm 0.09-0.12 | Substantially<br>equivalent | | | | Finger 0.05mm min.<br>Palm 0.08mm min. | | | | | | Physical<br>Properties | Meets ASTM D<br>5250-06<br>(Reapproved 2011) | ASTM D5250 | | Substantially<br>equivalent | | | | Before aging/after aging<br>Elongation ≥300% | | Before aging/after aging<br>Elongation :330-420% | | | | | Freedom from<br>Pinholes | | | | | | | Meets<br>• 21 CFR 800.20<br>ASTM D5250-06<br>(Reapproved 2011)<br>• ASTM D 5151-06<br>(Reapproved 2011) | 21 CFR 800.20<br>ASTM D5250<br>ASTM D 5151 | Meets ASTM<br>D5151-06<br>(Reapproved 2011)<br><br>Holes at<br>Inspection Level I<br>AQL2.5 | Substantially<br>equivalent | | | | Residual<br>Powder | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)<br><br>below 2mg of residual<br>powder | ASTM D 6124 | Meets ASTM<br>D 6124-06<br>(Reapproved 2011).<br><br>Results generated values<br>below 2mg of residual<br>powder | Substantially<br>equivalent | | | Materials used<br>to fabricate the<br>devices | PVC | If the glove is made<br>of a polymer or other<br>type of material.<br>identify the material. | PVC | Substantially<br>equivalent | | | Dusting or<br>Donning<br>Powder: | PU | If a donning lubricant<br>is used, state the<br>composition<br>and include<br>biocompatibility data<br>for the lubricant in an<br>identified attachment;<br>also state the name,<br>manufacturer, and<br>address below | PU | Substantially<br>equivalent | | | Dusting or<br>Donning<br>Powder: name | PU | Lubricant<br>Generic Name/<br>Lubricant<br>Brand Name | Surface Coating Agent | Substantially<br>equivalent | | | Compare<br>performance<br>data supporting<br>substantial<br>equivalence | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reapproved 2011) | At this time FDA<br>recognizes the<br>following standards:<br>Patient Examination<br>Gloves(PVC)ASTM<br>D5151(Detection of<br>Holes in Medical<br>Gloves)ASTM<br>D6124(Residual<br>Powder on Medical<br>Gloves)ASTM<br>D5250(Poly(vinyl<br>chloride) Gloves) | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reapproved 2011) | Substantially<br>equivalent | | | Single Patient Use | Single Patient Use | Single Patient Use | Single Patient Use | Substantially<br>equivalent | | | Biocompatibility | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10:2002/Amd.1:<br>2006 | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES<br>ISO 10993-10 | The test article was a<br>non-irritant and<br>non-sensitizer.<br><br>SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10:2002<br>/Amd.1:2006 | Substantially<br>equivalent | | | Labeling for the<br>legally marketed | -Powder Free<br>-Patient Examination<br> | Chapter 4 | -Powder Free<br>-Patient Examination | Substantially<br>equivalent | # 9.0 Substantial Equivalence Comparison: . {3}------------------------------------------------ Section C(rev.03) {4}------------------------------------------------ | substantial | -Single Use Only<br>- Manufactured For: | -Single Use Only<br>- Manufactured For: | |----------------------------|-----------------------------------------|-----------------------------------------| | equivalence is<br>claimed. | - Lot | - Lot | #### 10.0 Substantial Equivalence Comparison: It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims. It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is as safe, as effective, and performs as well as the predicate device, Powder-Free Vinyl Patient Examination Glove (Non-colored) Zhang Jia Gang Fengyuan Plastic Product Co., Ltd. K091663. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing health, human services, and the people served by the department. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 4, 2014 Heze City Green Food Company, Limited C/O Mr. Chu Xiaoan Beijing Easy-Link Company Room 1606 Bldg. 1 Jianxiang Yuan #209 Bei Si Huan Zhong Road, Haidian District Beijing 100083 CHINA Re: K130582 Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: December 18, 2013 Received: December 20, 2013 Dear Mr. Xiaoan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 - Mr. Xiaoan Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address Sincerely yours. http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Image /page/6/Picture/7 description: The image contains a combination of text and graphic elements. The text includes "Tejashri Purohit-Sheth, M.D." and "Clinical Deputy Director," suggesting a professional or medical context. The acronym "DAGRID" is also present. There is also a signature and the word "FOR" in the bottom right corner. Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on lest page. ## Indications for Use S10{ik) Number (if known) K130582 #### Device Name Powder Free Viny! Patient Examination Gloves, Clear (non-colored) #### Indications for Use (Describe) Powder Free Viryl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. For FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Elizabeth F. Claverie 2014.02.03 21:51:49.05:00' FORM FDA 3881 (9/13) {8}------------------------------------------------ # This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (9/13)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...