The MASTERtorque LUX 8900 L is intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
Device Story
Air-driven dental handpiece; utilizes dental unit air pressure to rotate turbine up to 400,000 RPM; accepts ISO 1797-1 type 3 burs. Features Direct Stop Technology to reduce turbine/bur stopping time and back suction. Includes fiber optic illumination and internal water filter. Operated by trained dental professionals in clinical settings. Connects to dental units via MULTIflex coupling for air, water, and light supply. Interacts with patient teeth via rotating bur for restorative procedures. Benefits include efficient material removal, reduced stopping time, and improved visibility.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing, including performance testing per international standards for dental air-driven handpieces, biocompatibility studies, and sterilization validation.
Technological Characteristics
Air-driven turbine; 340,000-400,000 RPM; 23W power. Materials: stainless steel, German silver (Ni-Cr coated), PEEK, Fluoride Rubber (Viton). Features: Direct Stop Technology, fiber optic light (25,000 LUX), push-button chuck. Dimensions: 13.0mm head height, 12.5mm head diameter. Connectivity: MULTIflex coupling. Sterilization: steam autoclave. Standards: ISO 7785-1, ISO 7405, ISO 9168.
Indications for Use
Indicated for removal of carious material, reduction of hard tooth structure, cavity/crown preparations, removal of fillings, processing/finishing restorations, and polishing teeth. For use by trained professionals in general dentistry.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for KaVo Dental Excellence. The logo features the letters 'K' and 'V' in a stylized font, with the 'K' appearing larger and more prominent. Below the logo, the words 'KaVo. Dental Excellence.' are printed in a smaller, simpler font.
# Section V - 510(k) Summary
## Submitter:
Kaltenbach & Voigt GmbH Bismarckring 39. 88400 Biberach Stefan Trampler - Contact Person 49 (7351) 56 3515 - Phone number +49 (7351) 56 7 3515 - Facsimile
Date Summary Prepared: August, 28th 2013
SEP 27 2013
Device Name:
- Trade Name MASTERtorque LUX 8900 L .
- Common Name Dental handpiece and accessories �
- Classification Name Dental handpiece and accessories, per 21 CFR § 872.4200 �
# Devices for Which Substantial Equivalence is Claimed:
- ◆ SUPERTORQUE HIGH-SPEED HANDPIECES (K073478)
- HIGH SPEED AIR TURBINE HANDPIECES, PAR-4HEX/4HX SERIES, TWIN Power (K061701)
### Device Description:
The MASTER orque LUX 8900 L air-driven handpieces are dental handpieces for the use by a trained professional in the field of general dentistry. The devices are air-powered handpieces that are reusable and ergonomically shaped, and provided with a fiber optic light system. The devices can be sterilized by the steam autoclave method. Through the MULTIflex coupling connected to a dental unit, the MASTERtorque LUX 8900 L receives the air for high speed turbine, the cooling water and air for cutting treatment through pouring holes and light for illumination the operation area. The MASTERtorque LUX 8900 L is equipped with the Direct Stop Technology which reduces the stopping time of the high speed turbine and so of the bur. The back suction also will be reduced due to the Direct Stop Technology.
Additionally the MASTERtorque LUX 8900 L will be supplied with an exchange filter for the water lines inside of the dental air-driven handpiece. The exchange filter filters the spray water inside the product. The filter is considered to be a spare part. If there is an insufficient amount of cooling water out of the MASTERtorque LUX 8900 L according to the instructions.for use, this spare has to be exchanged.
Further one there is a jet needle supplied with the MASTERtorque LUX 8900 L. By using this part the operator is able to clean the spray holes in the head of the product if there is an insufficient amount of cooling water out of the MASTERtorque LUX 8900 L according to the instructions for use.
Sec 5 - 1
{1}------------------------------------------------
The KaVo MULTIflex connectors (f.e. MULTIflex coupling 465 LED) are accessories to the medical device which will not be delivered together with the MASTERtorque LUX 8900 L. The MULTIflex connectors carry the air for the high speed turbine, the cooling water and air for cutting treatment and light for illumination from the dental treatment unit to the MASTERtorque LUX 8900 L.
The mechanism of action for the MASTERtorque LUX 8900 L is as follows. The dental handpieces / turbine MASTERtorque LUX 8900 L is an air-driven handpiece which will be supplied with water, air and light through the tube and the MULTIflex coupling of a dental treatment unit. Based on the air pressure the turbine rotates up to 400.000 roms. Dental burs (not part of this 510(k)) according to ISO 1797-1 type 3 could be inserted into the chuck system of the turbine. The MASTERtorque LUX 8900 L interacts with the patient through a rotating bur with the patient teeth according to the intended use.
## Intended Use of the Device:
The MASTERtorque LUX 8900 L is intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, removal of fillings, processing and. finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
#### Substantial Equivalence:
The MASTERtorque LUX 8900 L functions in a manner similar to and is intended for the same use as the SUPERTORQUE HIGH-SPEED HANDPIECES marketed by Kaltenbach & Voigt GmbH and to the HIGH SPEED AIR TURBINE HANDPIECES, PAR-4HEX/4HX SERIES, TWIN Power marketed by J. MORITA USA, INC. The MASTERtorque LUX 8900 L is similar to the two (2) predicate devices in that it is a dental air-driven handpiece for the use by a trained professional in the field of general dentistry. The MASTERtorque LUX 8900 L is substantially equivalent in design, application and function to the two (2) predicate devices noted above. All three devices, the MASTERtorque LUX 8900 L and the two (2) predicate devices are reusable and ergonomically shaped. The devices can be sterilized by the steam autoclave method. Based on the run-down behavior the MASTERtorque LUX 8900 L is similar to the HIGH SPEED AIR TURBINE HANDPIECES, PAR-4HEX/4HX SERIES, TWIN Power. Both devices are equipped either with the Direct Stop Technology or with the Quick stop. In addition the MASTERtorque LUX 8900 L and both predicate devices receives the air for their high speed turbines and the cooling water / air for cutting treatment through the tube and the specific coupling of a dental unit.
The MASTERtorque LUX 8900 L.differs from the SUPERTORQUE HIGH-SPEED HANDPIECES and the HIGH SPEED AIR TURBINE HANDPIECES, PAR-4HEX/4HX SERIES, TWIN Power in that the MASTERtorque LUX 8900 L is only available with a fiber optic light system for illumination of the operation area. Furthermore the MASTERtorque LUX 8900 L differs from the SUPERTORQUE HIGH-SPEED HANDPIECES by having the Direct Stop Technology which reduces the stopping time of the high speed turbine and so of the bur. Additionally the back suction will be reduced due to this Direct Stop Technology.
The differences do not render the device NSE because the performance tests demonstrates that the differences in technological characteristics (turbine only with light and the Direct Stop technology) do not raise different questions of safety and effectiveness than the predicate and show that the device is as safe and effective as the predicate.
{2}------------------------------------------------
# Summary of the Technological Characteristics:
.
| Descriptive Information | MASTERtorque LUX 8900 L | SUPERTORQUE HIGH-<br>SPEED HANDPIECES<br>(K073478) | HIGH SPEED AIR TURBINE<br>HANDPIECES, PAR-<br>4HEX/4HX SERIES, TWIN<br>Power (K061701) |
|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The MASTERtorque LUX 8900<br>L is intended for the removal<br>of carious material, reducing<br>of hard tooth structure, cavity<br>and crown preparations,<br>removal of fillings, processing<br>and finishing tooth<br>preparations, restorations,<br>and for polishing teeth. They<br>are designed for use by a<br>trained professional in the<br>field of general dentistry. | Identical | Identical<br>(The PAR-4HEX/ 4 HX series<br>handpiece is for use by<br>authorized persons in the<br>practice of the dentistry.) |
| Principle of operation | Through the tube and the<br>MULTIflex coupling connected<br>to a dental unit, the air-<br>powered handpiece receives<br>the air for high speed turbine,<br>the cooling water and air for<br>cutting treatment through<br>pouring holes and light for<br>illumination the operation<br>area. | Identical | Identical |
| Air / water ports | Up to four (4) | Identical | Three (3) |
| Fiberoptics | With built-in light system | With and without built-in light<br>system | With and without built-in light<br>system |
| Dimensions | Headsize-Height: 13.0 mm<br>Headsize-Diameter: 12,5 mm | Headsize-Height: 15,0 mm<br>Headsize-Diameter: 13,0 mm | Headsize-Height: 13,2 mm<br>Headsize-Diameter: 10,5 mm |
| Type of chuck | Push Button | Identical | Identical |
| Power (approx.) | 23 watts | 18 watts | 22 watts |
| Coupling Dimensions | Length with coupling:<br>Approx. 141 mm | Length with coupling:<br>Approx. 130 mm | Length with coupling:<br>Approx. 122,5 mm |
| Chemical composition of the<br>patient-contacting portions of<br>the device | Stainless steel<br>(See details in Section XV) | German silver and stainless<br>steel | Information not available |
| Chemical composition of the<br>water / air lines | Stainless steel, german silver<br>(nickel - chromium coated),<br>PEEK and Fluoride Rubber<br>Viton<br>(See details in Section XV) | German silver and stainless<br>steel | Information not available |
| Light Intensity | Approx. 25,000 LUX | Identical | Identical |
| Bur retention force | Up to 24 Ncm | Identical | Information not available |
| Operating Pressure | 30 - 61 psi (41 psi<br>recommended) | > 40 psi recommended | 19 - 42 psi (35 psi<br>recommended) |
| Idling Speed | 340,000 - 400,000 rpms | Approx. 350,000 rpms | 340,000 - 400,000 rpms |
| Run-down behavior | Direct Stop | Not applicable | Quick Stop |
.
{3}------------------------------------------------
| Compliance to Standards | Complies with ISO 7785-1.<br>ISO 7405 and ISO 9168 | Identical | Identical |
|-------------------------|----------------------------------------------------|-----------|---------------------------|
| Lubricant | KaVo QUATTROcare<br>(K071288) | Identical | Information not available |
#### Non-Clinical Test Data:
Performance tests according to the international standards for dental air-driven handpieces have been conducted to determine the conformance to the state of the art. Biocompatibility and sterilization studies have been completed which demonstrate that the MASTERtorque LUX 8900 L is safe for his intended use.
## Clinical Test Data:
Clinical testing has not been conducted on this product.
### Conclusion:
Based upon the tests according to the international standards for dental air-driven handpieces, the biocompatibility and sterilization studies and the similar technological / performance characteristics as compared to the predicate devices, the performance of the MASTERtorque LUX 8900 L is deemed to be substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare. The symbol consists of three lines that curve and converge, with a wavy line at the bottom.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 27, 2013
Kaltenbach & Voight GmbH Stefan Trampler Head of Quality Management & Regulatory Affairs Bismarckring 39 BIBERACH GERMANY 88400
Re: K130560
Trade/Device Name: MASTERtorque LUX 8900 L Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFB Dated: August 28, 2013 Received: August 29, 2013
Dear Mr. Trampler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Trampler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
MaryS.Runner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K130560
# Section IV - Indications for Use
510(k) Number (if known):
Device Name: MASTERtorque LUX 8900 L
Indications for Use:
The MASTERtorque LUX 8900 L is intended for the removal of carlous material, reducing of hard tooth structure, cavity and crown preparations, removal of fillings, processing and finishing tooth preparations, restorations, and for polishing teeth. They are designed for use by a trained professional in the field of general dentistry.
Prescription Use X (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
Andrew 1. Stee
2013.09.27
(PI.EASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sec 4 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.