BIOPSY SCIENCES LLC HYDROMARK BREAST BIOPSY SITE MARKER
K130537 · Biopsy Sciences, LLC · NEU · Mar 29, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130537
Device Name
BIOPSY SCIENCES LLC HYDROMARK BREAST BIOPSY SITE MARKER
Applicant
Biopsy Sciences, LLC
Product Code
NEU · General, Plastic Surgery
Decision Date
Mar 29, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The Biopsy Sciences, Inc. HydroMark Breast Biopsy Site Marker is intended to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.
Device Story
HydroMark Breast Biopsy Site Marker consists of a resorbable hydrogel component and a permanent metallic (stainless steel or titanium) coil. Device is pre-loaded in a sterile, disposable applicator. During percutaneous breast biopsy, the delivery system deploys the marker into the biopsy tract. Upon fluid contact, the hydrophilic hydrogel expands to fill the tract, anchoring the marker at the biopsy site and providing temporary ultrasound visibility. Over time, the hydrogel degrades via hydrolysis, leaving the metallic coil for permanent x-ray and MRI visualization. Used by clinicians in clinical settings to facilitate post-biopsy site identification. This submission adds a smaller marker size with a rigid, sharp delivery system for ultrasound-guided deployment, either directly or through a 16-gauge coaxial needle, and updates delivery system compatibility for specific vacuum-assisted biopsy devices.
Clinical Evidence
Bench testing only. Performance validated through visual and dimensional testing, tensile testing of delivery systems, penetration force analysis, usability testing for insertion/deployment, hydration/moisture analysis, and visibility verification under ultrasound, x-ray, and MRI. Biocompatibility confirmed per ISO 10993-1.
Technological Characteristics
Implantable clip (21 CFR 878.4300). Components: Resorbable hydrogel and metallic (stainless steel or titanium) coil. Delivery system: Sterile, disposable, rigid, sharp applicator. Compatibility: 16-gauge coaxial needles and specific vacuum-assisted biopsy systems. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for marking tissue during percutaneous breast biopsy procedures in patients requiring biopsy site localization; visible under ultrasound for at least 6 weeks and permanently visible via x-ray and MRI.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Biopsy Sciences HydroMark Breast Biopsy Site Marker (K060769)
Biopsy Sciences HydroMark Breast Biopsy Site Marker (K083006)
Biopsy Sciences HydroMark Breast Biopsy Site Marker (K090501)
Biopsy Sciences HydroMark Breast Biopsy Site Marker (K121113)
Submission Summary (Full Text)
{0}------------------------------------------------
K130537 Page 1/3
Image /page/0/Picture/1 description: The image shows the logo for Biopsy Sciences. The word "Biopsy" is written in a stylized font, with the "Bio" portion in a black box with white letters. The word "Sciences" is written in a smaller, simpler font below the word "Biopsy."
MAR 2 9 2013
510(K) SUMMARY (21 CFR 807.92)
## HYDROMARK BIOPSY SITE MARKER
510(k) Owner: Biopsy Sciences, Inc. 4900 Creekside Drive, Suite C Clearwater, FL 33760 Tel: 727-290-9825 Fax: 727-290-9828
- Contact Person: Sharon Rockwell Tel: 714-695-9269 E-mail: srockwell@writeme.com
Date Prepared: February, 2013
Trade Name: HydroMark Biopsy Site Marker
Common Name: Implantable clip
Classification Name: Implantable clip per 21 CFR 878.4300, NEU
Predicate Devices: Biopsy Sciences HydroMark Breast Biopsy Site Marker, K060769 Biopsy Sciences HydroMark Breast Biopsy Site Marker, K083006 Biopsy Sciences HydroMark Breast Biopsy Site Marker, K090501 Biopsy Sciences HydroMark Breast Biopsy Site Marker, K121113
Device Description: The HydroMark Breast Biopsy Site Marker contains a resorbable hydrogel component and a metallic component for permanent marking. The hydrogel has features that are unique and highly desirable for breast tissue marking.
> The HydroMark Site Marker is provided pre-loaded in a sterile, disposable applicator that is compatible with specified commercially available biopsy devices. The HydroMark is deployed by the delivery system and is left in the tract created during the biopsy procedure.
> This Special 510(k) addresses the addition of a smaller size HydroMark Breast Biopsy Site Marker which will be provided in a smaller, rigid, sharp delivery system for use under ultrasound
{1}------------------------------------------------
K 130537
deployment either directly or though a commercially available 16 gauge x 9 cm coaxial needle.
This Special 510(k) also describes minor modifications to the delivery system used for the ATEC 9 Ga Biopsy System and the EnCore 10 Ga Directional Vacuum-Assisted Biopsy Devices, and includes compatibility testing to allow the existing Mammotome MR Targeting Set to be labeled for use with other commercial MRI biopsy devices. Associated changes to the IFU and labels are also provided.
Intended Use:
The Biopsy Sciences, Inc. HydroMark Breast Biopsy Site Marker is intended to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.
The indications are identical to those of the predicate device, the Biopsy Sciences HydroMark Biopsy Site Marker, is indicated "to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI."
The hydrogel component expands on fluid contact to fill the track of the biopsy needle anchoring the HydroMark at the exact location of biopsy. Because the hydrogel is hydrophilic, it is clearly distinct from normal breast structure under ultrasound imaging. The hydrogel material degrades via hydrolysis over time leaving the internal stainless steel or titanium coil which provides permanent visibility under x-ray and MRI.
The smaller size marker is a new model in the family of HydroMark Breast Biopsy Site Markers which will be provided with a rigid, sharp delivery system for use under ultrasound deployment either directly or through a 16 gauge x 9 cm coaxial needle. There are no technological differences between this site marker and the others in the family of HydroMark Breast Biopsy Site Markers; therefore the addition of the smaller size marker does not raise new questions of safety or efficacy. All other modifications to the delivery systems, device compatibility and associated IFU and label changes represent minor updates to the products.
Non-Clinical Performance Data:
Non-clinical testing included the following:
Visual and Dimensional testing .
Technological Characteristics:
{2}------------------------------------------------
K130537
- Tensile testing of delivery system ●
- . Penetration force
- Usability testing for ease of insertion and deployment
- . Hydration testing and moisture analysis
- . Visibility under ultrasound, x-ray, and MRI
- . Sterilization validation
The devices performed as intended according to the specifications established for the finished device. The device continues to meet the ISO 10993-1 requirements for biocompatibility.
Conclusions:
The non-clinical bench testing, including simulated use testing, demonstrates that the 18 gauge HydroMark Breast Biopsy Site Marker for use with a 16 gauge x 9 cm coaxial needle performs as intended, accurately marking the biopsy site by fixating in the track of the needle biopsy. The device works in an identical manner to previously cleared 15 ga HydroMark Breast Biopsy Site Markers for use with 13 gauge x 9 cm coaxial needles or for direct puncture. The testing supports a determination of substantial equivalence to predicate HydroMark Breast Biopsy Site Markers cleared by FDA.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter of the circle. The text is in all caps and is evenly spaced around the circle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Letter dated: March 29, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Biopsy Sciences, LLC % Ms. Sharon Rockwell Regulatory Affairs Consultant 5582 Chalon Road Yorba Linda, California 92886
Re: K130537
Trade/Device Name: Hydromark Breast Biopsy Site Marker Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: February 18, 2013 Received: March 1, 2013
Dear Ms. Rockwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Ms. Sharon Rockwell
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark Nijelkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _K130537
Device Name: Biopsy Sciences HydroMark Breast Biopsy Site Marker
Indications for Use:
To mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## David Krause
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K130537
Page __ of ___________________________________________________________________________________________________________________________________________________________________
---
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.