DIPLOMAT

K130534 · Diplomat Dental S.R.O. · EIA · Oct 30, 2013 · Dental

Device Facts

Record IDK130534
Device NameDIPLOMAT
ApplicantDiplomat Dental S.R.O.
Product CodeEIA · Dental
Decision DateOct 30, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6640
Device ClassClass 1

Indications for Use

Diplomat dental operative units are self-contained dental treatment units that contain vacuum system and water supply and separate compressor placed outside the dental unit. These dental operative units are designed to provide air, water, vacuum and electricity to operate various dental hand-pieces, accessories and attachments, and to serve as a base for other dental devices and accessories. Diplomat dental operative units are offered with or without attached dental chairs. When supplied with a dental chair, the chair is intended to properly position a patient to perform different dental procedures. Diplomat dental operative units are to be operated and used by dentists and other legally qualified professionals, and are intended for use to provide general dental restorative care and hygiene procedures in both traditional dental office settings applications by or under the supervision of a licensed dentist or a hygienist if permitted by applicable law.

Device Story

Diplomat dental operative units are AC-powered, self-contained systems providing air, water, vacuum, and electricity to dental hand-pieces and accessories. Units serve as a base for dental instruments (e.g., curing lights, ultrasonic instruments) and may include an integrated patient chair for positioning. Used in dental clinics by dentists, hygienists, and assistants. The device functions as a utility hub for dental procedures; it does not perform diagnostic analysis or automated decision-making. Benefits include centralized power and utility delivery for restorative and hygiene care.

Clinical Evidence

No clinical data. Bench testing only, demonstrating compliance with IEC 60601-1, EN 60601-1-2, and ISO 7494-1.

Technological Characteristics

AC-powered dental treatment unit; provides air, water, vacuum, and electrical power. Available with or without integrated patient chair. Complies with IEC 60601-1 (electrical safety), EN 60601-1-2 (EMC), and ISO 7494-1 (dental unit performance). Class I equipment, Type B applied parts. Professionally installed.

Indications for Use

Indicated for patients requiring general dental restorative care and hygiene procedures in traditional dental office settings, performed by dentists, hygienists, or other legally qualified professionals.

Regulatory Classification

Identification

A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the text "KISOS 34" written in black ink at the top. Below the text is the logo for Emergo Group. The logo consists of the words "EMERGO GROUP" with a stylized globe in the center and a world map above and below the text. OCT 3 0 2013 Image /page/0/Picture/1 description: The image shows the logo for Diplomat Dental Solutions. The logo features the word "DIPLOMAT" in bold, sans-serif font, with the words "DENTAL SOLUTIONS" in a smaller font size underneath. Above the word "DIPLOMAT" is a graphic of a series of rectangles that are decreasing in size from left to right. The rectangles are arranged in a diagonal line. 5. 510(K) Summary Submission Correspondent Company Name: Company Address: Company Contact: Phone: Fax: Submission Date: Website: Submission Sponsor Company Name: Company Address: Country: Phone: Fax: Website: # Device Classification Device Sponsor: Product Classification Name: Product Code: Regulation Number: Classification Panel: Regulatory Class: Emergo Group, Inc. 816 Congress Avenue Suite 1400 Austin, TX 78701 Stuart R. Goldman Senior Consultant project.management@emergogroup.com 512.327.9997 512.327.9998 February 11, 2013 www.emergogroup.com DIPLOMAT DENTAL, s.r.o. Vrbovská cesta 17 921 01 Piešťany Slovak Republic Slovakia 421.33.7954111 421.33.7726326 www.chirana.sk DIPLOMAT DENTAL, s.r.o. Unit, Operative Dental EIA 872.6640 Dental Devices Class I {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Diplomat Dental Solutions. The logo features the word "DIPLOMAT" in bold, sans-serif font, with the words "DENTAL SOLUTIONS" in a smaller font size underneath. Above the word "DIPLOMAT" is a graphic of a series of rectangles that are stacked diagonally, creating a step-like pattern. The logo is black and white. # Predicate Device Sirona C8+Dental Operative Unit (K032543) # Statement of Intended Use Diplomat dental operative units are self-contained dental treatment units that contain vacuum system and water supply and separate compressor placed outside the dental unit. These dental operative units are designed to provide air, water, vacuum and electricity to operate various dental hand-pieces, accessories and attachments, and to serve as a base for other dental devices and accessories. Diplomat dental operative units are offered with or without attached dental chairs. When supplied with a dental chair, the chair is intended to properly position a patient to perform different dental procedures. Diplomat dental operative units are to be operated and used by dentists and other legally qualified professionals, and are intended for use to provide general dental restorative care and hygiene procedures in both traditional dental office settings applications by or under the supervision of a licensed dentist or a hygienist if permitted by applicable law. # Device Description Diplomat dental operative units are modern, ergonomically designed, dental systems offered in a wide range of models depending on the specific end user requirements. Diplomat dental operative units are AC-powered devices that are intended to supply power to, and serve as a base for, other dental devices and accessories - such as hand-pieces, operating lights, curing lights, ultrasonic instruments, hygiene instruments, etc. - and deliver electric, air, water and vacuum power to them. Diplomat dental operative units are offered either with the patient dental chair or as a free-standing unit without the chair for those end-users with existing patient chairs that they want to use. Diplomat dental operative units are offered in the following models and with the following accessories: {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo for Diplomat Dental Solutions. The logo features a stylized graphic of interconnected rectangles, resembling a staircase or a series of steps, positioned above the word "DIPLOMAT" in bold, sans-serif font. Below "DIPLOMAT" are the words "DENTAL SOLUTIONS" in a smaller font size. Diplomat Dental Operative Units: Adept DA 380, Adept DA 370, Adept DA 130, Adept DA 130, Adept DA 110A, Consul DC 350, Consul DC 310, Consul DC 180, Diplomat Consul DC 170, Diplomat Consul DC 170 (Orthodontic), Diplomat Lux DL 320; Diplomat Lux DL 210 Diplomat Dental Accessories: Dental Chair DE 20, Dental Chair DM 20, Dental Stools D 10L, Dental Light Xenos, Dental Light Sirius # Predicate Device Comparison Diplomat dental operative units by DIPLOMAT DENTAL have been shown to be Substantially Equivalent to the C8+Dental Operative Units manufactured by Sirona Dental Systems previously cleared by the FDA under K032543. The applicant and predicate devices are very similar in overall design and technology; principle of operation, intended use, materials, construction, and available accessories offered for them. The differences between the applicant and predicate devices do not raise any new questions of safety or effectiveness. Some select properties and characteristics of the applicant and predicate devices are compared side-by-side in the table below. | Features | Diplomat Dental<br>Operative Units | C8+Dental<br>Operative Unit | Substantially<br>Equivalent | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------------------------| | 510(k) | Pending | K032543 | | | Models | Adept Models: DA 380, DA 370, DA<br>170, DA 130, DA 110;<br>Consul Models: DC 350, DC 310, DC<br>180, DC 170 and DC orthodontic; and<br>Lux Models: DL 320, and DL 210 | Multiple versions | Similar | | Product Code | EIA | EIA | Same | | Regulation<br>Number | 872.6640 | 872.6640 | Same | | Regulation<br>Name | Unit, Operative, Dental | Unit, Operative, Dental | Same | | Place of Use | Dental clinic | Dental clinic | Same | | Intended<br>Users | Dentists; assistants and hygienists | Dentists; assistants and hygienists | Same | | Power Supply | 110V/(230 VAC)<br>50/60 Hz | 110V/115V/127V/(230 VAC)<br>50/60 Hz | Similar | | Utility Supply | Compressed air and water | Compressed air and water | Same | {3}------------------------------------------------ | ABARAGEBAABABABARAHAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | : | | <br>11 | | | | | | 00 | | | | ENTAL SOLUTIONS<br>N 100 | | | | Protection<br>Class | Class 1 equipment | Class 1 equipment | Same | |----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Degree of<br>Protection | Type B of applied parts | Type B of applied parts | Same | | Installation | Professionally installed; available<br>with chair and cart mounted<br>options. | Professionally installed; not available<br>with chair and cart mounted options. | Same | | Air Pressure | $450 kPa/800 kPa$ (min./max.) | $550 kPa/750 kPa$ (min./max.) | Similar | | Water<br>Pressure | $300 kPa/600 kPa$ (min./max.) | $250 kPa/600 kPa$ (min./max.) | Similar | | Electrical<br>Safety | IEC 60601-1 | IEC 60601-1 | Same | | EMC | EN 60601-1-2<br>ISO 7494-1 | IEC 60601-1-2 | Similar | | Performance | Diplomat dental operative units<br>are self-contained dental<br>treatment units that contain<br>vacuum system and water supply<br>and separate compressor placed<br>outside the dental unit. These<br>dental operative units are designed<br>to provide air, water, vacuum and<br>electricity to operate various<br>dental hand-pieces, accessories<br>and attachments, and to serve as a<br>base for other dental devices and<br>accessories. Diplomat dental<br>operative units are offered by with<br>or without attached dental chairs.<br>When supplied with a dental chair,<br>the chair is intended to properly<br>position a patient to perform<br>different dental procedures.<br>Diplomat dental operative<br>units are to be operated and used<br>by dentists and other legally<br>qualified professionals, and are<br>intended for use to provide general<br>dental restorative care and hygiene<br>procedures in both traditional<br>dental office settings applications<br>by or under the supervision of a | The C8 + Dental Operative Unit with<br>accessories are intended to supply<br>power to and serve as a base for dental<br>devices and accessories. This product<br>includes a dental chair. The unit is<br>intended for use in the dental clinic<br>environment and used by trained<br>dentists and/or dental technicians and<br>assistants. The C8 + Dental Operative<br>Unit is offered wilt the optional Sivision<br>3, an intraoral camera system intended<br>to provide the dentist and patient with<br>intraoral video images to view the<br>condition of the teeth and oral cavity. | Similar | | Statement of<br>Intended Use | | | | | of a licensed dentist or a hygienist if permitted by applicable law. | | | | EMERGO O GROUP 21 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Diplomat Dental Solutions. The logo features the word "DIPLOMAT" in large, bold, sans-serif font, with the words "DENTAL SOLUTIONS" in a smaller font below it. Above the word "DIPLOMAT" is a graphic of a series of rectangles that are arranged in a diagonal line. # Non-Clinical Data - Device Bench Testing As part of demonstrating the safety and effectiveness of its line of Diplomat dental operative units, DIPLOMAT DENTAL has submitted these devices for performance testing in accordance with the applicable sections of the following international standards and FDA guidance documents: - . IEC 60601-1:2005, Medical electrical equipment - Part 1, General requirements for basic safety and essential performance - . EN 60601-1-2:2007, Medical electrical equipment – Part 1-2, General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests - . EN 60601-1:2009, Medical electrical equipment - Part 1, General reguirements for basic safety and essential performance - . ISO 7494-1:2004, Dentistry - Dental Units - Part 1: General requirements and test methods # Clinical Data Not applicable to this 510(k) submission. # Substantially Equivalence By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device, or has the same intended use and different technological characteristics, and it can be demonstrated that the device is substantially equivalent to the predicate device, and {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo for "DIPLOMAT DENTAL SOLUTIONS". The word "DIPLOMAT" is in large, bold, sans-serif font, with "DENTAL SOLUTIONS" in a smaller font size underneath. Above the word "DIPLOMAT" is a graphic consisting of a series of connected, progressively smaller, rectangles. The rectangles are arranged in a descending, step-like pattern. that the new device does not raise new questions regarding its safety and effectiveness when compared to the predicate device. It has been shown in this 510(k) submission that the differences between the Diplomat dental operative units by DIPLOMAT DENTAL when compared to the Sirona Dental Systems C8+Dental Operative Units (K032543) are minimal and do not raise any questions regarding their safety and effectiveness. Diplomat dental operative units, as designed and manufactured are therefore determined to be substantially equivalent to the predicate device. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its wings or body. The bird is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002 ### October 30, 2013 Diplomat Dental, S.R.O. C/O Mr. Stuart R. Goldman Senior Consultant Emergo Group, Incorporated 816 Congress Avenue, Suite 1400 Austin, TX 78701 Re: K130534 Trade/Device Name: Dental Operative Unit Regulation Number: 21 CFR 872.6640 Regulation Name: Dental Operative Unit and Accessories Regulatory Class: I Product Code: EIA Dated: September 23, 2013 Received: September 24, 2013 Dear Mr. Goldman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ ### Page 2 - Mr. Goldman Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, MarySTBohner-S Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Devices Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## Indications For Use #### K130534 510(k) Number (if Known): Diplomat dental operative units are self-contained dental treatment units that contain vacuum system and water supply and separate compressor placed outside the dental unit. These dental operative units are designed to provide air, water, vacuum and electricity to operate various dental hand-pieces, accessories and attachments, and to serve as a base for other dental devices and accessories. Diplomat dental operative units are offered with or without attached dental chairs. When supplied with a dental chair, the chair is intended to properly position a patient to perform different dental procedures. Diplomat dental operative units are to be operated and used by dentists and other legally qualified professionals, and are intended for use to provide general dental restorative care and hygiene procedures in both traditional dental office settings applications by or under the supervision of a licensed dentist or a hygienist if permitted by applicable law. Prescription Use X AND/OR Over-The-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Andrew Steens 2013.10.30 21:30:21 -04'00'
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...