ALTAPORE is an implant intended to fill bony voids or gaps of the skeletal system i.e., extremities and pelvis. ALTAPORE can be used in combination with autograft as a bone graft extender in the extremities and pelvis. ALTAPORE can be used in combination with autogenous bone marrow aspirate in the extremities and pelvis. These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process.
Device Story
ALTAPORE is a bioactive, osteoconductive bone void filler composed of silicate-substituted calcium phosphate (Ca, P, O, H, Si). Supplied as 1-2 mm microgranules in an absorbable aqueous gel carrier; does not set in-situ. Implanted by surgeons to fill skeletal voids in extremities and pelvis. Bioactivity demonstrated by apatite-layer formation in simulated body fluid; provides scaffold for bone growth. Osteoconductivity confirmed via animal critical defect models using radiographic, histopathological, and mechanical analyses. Device undergoes cell-mediated remodeling and replacement by natural bone. Used as standalone, autograft extender, or with bone marrow aspirate.
Clinical Evidence
No clinical data. Evidence based on non-clinical benchtop material characterization, dissolution, mechanical testing, in-vitro bioactivity, and in-vivo animal critical size defect models. Biocompatibility established per ISO 10993-1.
Technological Characteristics
Silicate-substituted calcium phosphate; 1-2 mm microgranules; 82.5% nominal porosity; microporous struts. Supplied in absorbable aqueous gel carrier. Sterilization via terminal irradiation. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for filling bony voids or gaps in the skeletal system (extremities and pelvis) that are not intrinsic to the stability of the bony structure. Used as a standalone graft, autograft extender, or with autogenous bone marrow aspirate. Applicable to surgically created or traumatic osseous defects.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
Actifuse® ABX E-Z-fil Putty Bone Graft Substitute (K071206)
{0}------------------------------------------------
Traditional 510(k) Premarket Notification
K130531 - ALTAPORE
.
.
______________________________________________________________________________________________________________________________________________________________________________
Scetion 5, 510(k) Summary
-
・・
| 5. 510(K) SUMMARY | | | APR 08 2 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| DATE SUMMARY<br>PREPARED: | 23 January 2013 | | |
| OWNER: | Baxter Healthcare Corporation<br>One Baxter Way<br>Westlake Village, CA 91362 | | |
| CONTACT PERSON: | Jerzy Wojcik<br>Associate Director, Global Regulatory Affairs<br>Baxter Healthcare Corporation<br>One Baxter Way<br>Westlake Village, CA 91362 | Telephone: | 224-948-3954 |
| | | Fax: | 805-372-3082 |
| | | Email: | jerzy_wojcik@baxter.com |
| DEVICE NAME: | Trade Name: | ALTAPORE | |
| | Common Name: | Bone Void Filler | |
| | Classification: | Resorbable calcium salt bone void filler devices have been<br>classified by the Orthopedics Device Panel as Class II<br>Special Controls per 21CFR888.3045 | |
| | Class: | Class II | |
| | Product Code: | MQV | |
| PREDICATE<br>DEVICES | K071206 | Actifuse® ABX E-Z-fil Putty Bone Graft Substitute | |
| | K081979 | Actifuse® Bone Graft Substitute<br>Actifuse® ABX E-Z-fil Putty Bone Graft Substitute<br>Actifuse® Shape Bone Graft Substitute<br>Actifuse® Flow Bone Graft Substitute | |
| | K082575 | Actifuse® Bone Graft Substitute<br>Actifuse® Microgranules Bone Graft Substitute<br>Actifuse® E-Z Prep<br>Actifuse® ABX E-Z-fil Putty Bone Graft Substitute<br>Actifuse® MIS<br>Actifuse® Shape Bone Graft Substitute | |
ADD . 014
,
{1}------------------------------------------------
## Traditional 510(k) Premarket Notification K130531 - ALTAPORE
# DEVICE DESCRIPTION:
ALTAPORE is a bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler. The interconnected and open porous structure of the silicate-substituted calcium phosphate phase of ALTAPORE is similar to human cancellous bone and is intended to support bone growth with macroand micro- porosity. ALTAPORE is composed solely of elements that exist naturally in normal bone (Ca, P, O, H, Si). ALTAPORE is supplied in a sterile applicator and contains ALTAPORE microgranules, sized 1-2 mm, 80-85% total porosity, suspended in an absorbable aqueous gel carrier. ALTAPORE docs not set in-situ following implantation. ALTAPORE is available in 1.5ml, 2.5ml, 5ml, 10ml, and 20ml configurations.
AI,TAPORE is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary. the product can be mixed with Bone Marrow Aspirate (BMA) or autologous bone at the discretion of the surgeon.
ALTAPORE is bioactive based on in vitro studies that show it forms a surface apatite-layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow allowing complete repair of the defect.
ALTAPORE is osteoconductive based on in vivo animal studies that show it achieves bony healing in a critical defect model as confirmed with radiographic, histolopatholgical, histomorphometric, and mechanical analyses. ALTAPORE undcrgocs ccll-mediated remodeling and is replaced by natural bone.
# STATEMENT OF INTENDED USE:
ALTAPORE is an implant intended to fill bony voids or gaps of the skeletal system i.e., extremities and pelvis. ALTAPORE can be used in combination with autograft as a bone graft extender in the extremities and pelvis. ALTAPORE can be used in combination with autogenous bone marrow aspirate in the extremities and pelvis. These osscous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process.
{2}------------------------------------------------
# TECHNOLOGICAL CHARACTERISTICS:
| ELEMENT OF<br>COMPARISON | ALTAPORE | PREDICATE (ABX) |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Composition | Silicate-substituted<br>calcium phosphate<br>composed solely of<br>elements that exist<br>naturally in normal bone<br>(Ca, P, O, H, Si). | Silicate-substituted<br>calcium phosphate<br>composed solely of<br>elements that exist<br>naturally in normal bone<br>(Ca, P, O, H, Si). |
| Physical<br>Structure | Granules with a porosity<br>similar to cancellous bone | Granules with a porosity<br>similar to cancellous bone |
| Nominal (Total)<br>Porosity | $82.5 \pm 2.5%$ | $80.0 \pm 2.5%$ |
| Strut Porosity | Microporous | Microporous |
| Sterility | Terminal irradiation | Terminal irradiation |
#### ASSESSMENT OF NONCLINICAL DATA:
Testing has shown ALTAPORE to meet the requirements of relevant standards for Calcium Salt Bone Void Fillers. Testing has confirmed ALTAPORE to be safe and effective in providing a scaffold for rapid bone repair via hony infiltration of the porous scaffold. Non-clinical testing included benchtop material characterization, dissolution, and mechanical; as well as in-vitro hioactivity . Biocompatibility of the device has been established in accordance with ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and Testing. Critical size defect implantation in-vivo animal studies have demonstrated that ALTAPORE is efficacious as a standalone bone graft substitute, mixed with Bone Marrow Aspirate (BMA), or mixed with autologous bone.
The conclusions drawn from the non-clinical tests demonstrate that CONCLUSIONS: ALTAPORE is as safe, as cffcctive, and performs as well or better than the predicate devices as a bioactive and osteoconductive bone void filler for osscous defects and is therefore substantially equivalent to the predicate devices. ALTAPORE is efficacious as a standalone bone graft substitute, mixed with Bone Marrow Aspirate (BMA), or mixed with autologous bone. The side-by-side comparative benchtop, in-vitro and in-vivo performance data provided showed no evidence of local or systemic adverse effects related to the device.
)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human figures or waves.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10907 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 8, 2014
Baxter Healthcare Corporation Mr. Jerzy Wojcik Associate Director, Global Regulatory Affairs One Baxter Way Westlake Village, California 91362
Re: K130531
Trade/Device Name: ALTAPORE Regulation Number: 21 CFR 888,3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: March 13, 2014 Received: March 14, 2014
Dear Mr. Wojcik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I D. And Interes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Jerzy Wojcik
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins
#### for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Traditional 510(k) Premarket Notification K130531 - ALTAPORE
## 4. INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
# K130531
Device Name:
### ALTAPORE
#### Indication(s) for Use:
ALTAPORE is an implant intended to fill bony voids or gaps of the skeletal system i.e., extremities and pelvis. ALTAPORE can be used in combination with autograft as a bone graft extender in the extremities and pelvis. ALTAPORE can be used in combination with autogenous bone marrow aspirate in the extremities and pelvis. These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE resorbs and is replaced with bone during the healing process.
তি Prescription Use: 21 CFR 801 Subpart D
Over-the-Counter Use: I 21 CFR 801 Subparı C
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Laurence D. Coyne -S
Page of
(Division Sign-Off) Division of Orthopedia Devices 510(k) Number: K130531
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.