K130503 · Entellus Medical, Inc. · LRC · Jun 11, 2013 · Ear, Nose, Throat
Device Facts
Record ID
K130503
Device Name
PATHASSIST LED LIGHT FIBER
Applicant
Entellus Medical, Inc.
Product Code
LRC · Ear, Nose, Throat
Decision Date
Jun 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Indications for Use
To locate, illuminate within, and transilluminate across nasal and sinus structures in patients aged 18 and over.
Device Story
PathAssist LED Light Fiber is a single-use, sterile, flexible surgical instrument used to illuminate nasal and sinus structures. The device features a flexible illumination fiber, a protective sheath, and an integrated battery-powered LED light source. During ENT procedures, the clinician inserts the device to emit red light from the distal tip, enabling transillumination of sinus structures to assist in anatomical localization. The device is operated by a physician. It provides visual guidance for the surgeon, potentially improving procedural accuracy and patient outcomes during sinus surgery. It is a self-contained unit, eliminating the need for external light source connections.
Clinical Evidence
No clinical data or animal studies were submitted. Evidence consists of bench testing, including EMC and electrical safety (IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2), design verification (functional, mechanical, and compatibility testing), packaging, and shelf-life validation. Simulated use was performed in a cadaver model. Biocompatibility and sterilization (EtO per ISO 11135-1) were validated.
Technological Characteristics
Flexible illumination fiber (27.6cm length, 0.5mm diameter) with integrated battery-powered LED. Biocompatible per ISO 10993-1. Sterilized via Ethylene Oxide (EtO) per ISO 11135-1. Electrical safety/EMC tested per IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2. Single-use, sterile.
Indications for Use
Indicated for locating, illuminating, and transilluminating nasal and sinus structures in patients aged 18 and older.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
{0}------------------------------------------------
Entellus Medical
K13.0503
Image /page/0/Picture/2 description: The image shows the logo for "entellus MEDICAL". To the left of the text is a circular graphic made up of several curved lines. The text "entellus" is in a sans-serif font, and below it is the word "MEDICAL" in a smaller font.
February 26, 2013
3006345872
Karen E. Peterson
(763) 463-7066
Sinus Guidewire
21 CFR 874.4420
Entellus Medical, Inc.
Plymouth, MN 55447
3600 Holly lane North, Suite 40
kpeterson@entellusmedical.com
ENT Manual Surgical Instrument
PathAssist LED Light Fiber
Vice President Clinical, Regulatory and Quality
JUN 1 1 2013
## 510(k) Summary
Date Prepared: Submitter Information:
Establishment Registration:
Contact Information:
Device Information: Trade Name: Common Name: Classification Regulation: Classification Name: Classification Panel: Device Classification: Product Code:
Predicate Devices: PathAssist Light Fiber [K120962] "PLS" Portable Light Source [K091829]
#### Device Description:
The PathAssist LED Light Fiber is a flexible instrument that emits light from the distal end. The LED Light Fiber is provided sterile and is for single use only. The device consists of a flexible illumination fiber, a protective sheath and an integrated battery powered LED light source. When the LED Light Fiber is activated the fiber will emit red light from the distal tip for over 60 minutes. It has a fiber nominal working length of 27.6cm with an outer diameter of 0.5mm (0.020").
ENT
LRC
Class I
Image /page/0/Figure/12 description: This image shows a diagram of a medical device. The device has a housing with an integrated light source and a pull tab. A flexible sheath extends from the housing, and a fiber is connected to the sheath.
PathAssist LED Light Fiber
t
{1}------------------------------------------------
#### Indication for Use
To locate, illuminate within, and transilluminate across nasal and sinus structures in patients aged 18 and over.
The intended use of the subject device is the same as the predicate devices: PathAssist Light Fiber [K120962] and "PLS" Portable Light Source [K091829], and the indications for use of the subject device is the same as the predicate device [K120962].
#### Contraindications:
None
#### Technological Characteristics:
The subject device has the same technological characteristics (i.e., principle of operation, design, function, materials, biocompatibility, shelf life, packaging, sterilization and energy source) as the predicate devices: PathAssist Light Fiber [K120962] and/or "PLS" Portable Light Source [K091829].
Both the subject device and predicate device [K120962] are flexible devices that transmit light from the proximal to distal tip of the device via a light fiber that can be seen via transillumination. Both devices are sterilized using Ethylene Oxide (EtO) and validated per ISO 11135-1, and have a Sterility Assurance Level (SAL) of 10 . Both devices are provided sterile, are for single use only and are biocompatible per ISO 10993-1. The subject device has an internal light source whereas the predicate device [K120962] must be connected to a light source.
Both the subject device and the predicate device [K091829] incorporate an internal battery operated LED light source. Both devices have undergone EMC and electrical safety testing per IEC 60601-1. IEC 60601-2-18 and IEC 60601-1-2. The subject device is provided sterile and is for single use only whereas the predicate device is provided non-sterile and must be cleaned and high level disinfected between each use.
#### Substantial Equivalence:
The intended use of the subject device is the same as the predicate devices [K120962] and [K091829] and the indications for use is the same as the predicate device [K120962]. The technological characteristics of the subject device are the same as the predicate device [K120962] and/or predicate device [K091829], including: principle of operation, design, function, materials, biocompatibility, shelf life, packaging, sterilization and energy source.
#### Performance Data:
Performance testing of the PathAssist LED Light Fiber consisted of EMC and electrical safety, ~ design verification (functional, mechanical, and compatibility testing), packaging, shelf life, and simulated use in a cadaver model. ' Biocompatibility testing and sterilization validation was referenced. Animal and clinical data were not submitted. Performance testing showed that the device meets design specifications and performed as intended.
### Conclusion
In conclusion, the device is substantially equivalent based on a comparison of intended use, indications for use, and technological characteristics. The device is safe and effective for its intended use.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, overlaid with three human figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
June 11, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Entellus Medical, Inc. Karen E. Peterson Vice President Clinical, Regulatory and Quality 3600 Holly lane North, Suite 40 Plymouth, MN 55447
Re: K130503
Trade Name: PathAssist LED Light Fiber Regulation Number: 21 CFR 874.4420 Regulation Name: ENT Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: February 26, 2013 Received: February 28, 2013
Dear Ms. Peterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Issting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Peterson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, MD
-Director -Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Entellus Medical
ר
#### Indications for Use Statement 6.
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: PathAssist LED Light Fiber
# Indications for Use
To locate, illuminate within, and transilluminate across nasal and sinus structures in patients aged 18 and over.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
× Prescription Use OR/AND Over-the-Counter Use
EricAMann-S
Page 16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.