UNITRON TINNITUS MASKER FEATURE

K130494 · Unitron Hearing · KLW · May 7, 2013 · Ear, Nose, Throat

Device Facts

Record IDK130494
Device NameUNITRON TINNITUS MASKER FEATURE
ApplicantUnitron Hearing
Product CodeKLW · Ear, Nose, Throat
Decision DateMay 7, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 874.3400
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

Unitron's tinnitus masker is a broadband sound generator available in Unitron Era hearing aid platforms. It provides a means of sound enrichment therapy that can be used as part of a personalized tinnitus management program to provide temporary relief from tinnitus. Tinnitus masker is intended for adults 18 years of age or older who have both hearing loss and tinnitus. The tinnitus masker within the air conduction hearing instrument is fitted by a hearing health care professional familiar with diagnosis and management of tinnitus.

Device Story

Unitron Tinnitus Masker is a software-based broadband noise generator integrated into Unitron Era hearing aid platforms. It provides sound enrichment therapy to help patients defocus from tinnitus. The device is fitted by a hearing health care professional who adjusts stimulus shaping and output levels. It features amplitude modulation and 'Environment Steering,' which automatically fluctuates noise levels based on the ambient environment. The output is delivered via air conduction hearing instruments. By providing supplementary noise stimulation, the device assists in managing tinnitus when combined with instructional counseling. It is intended for use by adults with both hearing loss and tinnitus.

Clinical Evidence

Bench testing only. Evaluated for safety and performance per ANSI S3.22:2009, IEC 60601-1, and IEC 60601-1-2.

Technological Characteristics

Broadband noise generator integrated into digital hearing aid platforms. Features include amplitude modulation and environment steering for automatic noise level adjustment. Complies with ANSI S3.22:2009, IEC 60601-1, and IEC 60601-1-2 standards.

Indications for Use

Indicated for adults 18+ with both hearing loss and tinnitus. Used as part of a personalized tinnitus management program for temporary relief. Must be fitted by a hearing health care professional.

Regulatory Classification

Identification

A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.

Special Controls

The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ **unitron.** MAY 0 7 2013 ## 5. 510(k) Summary Submitter: Unitron Hearing Inc. 14755 27th Avenue North, Plymouth MN 55447 Phone (763) 744- 3300 Fax (763) 557-8828 Jim Reiter, Contact Person: Quality Assurance Manager Unitron Hearing Inc. 14755 27th Avenue North, Plymouth MN 55447 Phone (763) 744-3463 Fax (763) 557-8828 Date Prepared: 10-January-2013 Device Name: Unitron Tinnitus Masker feature Device Class: Class II Classification Name: Tinnitus Masker Product Code: KLW Regulation Number: 21 CFR §874.3400 Tinnitus Masker Predicate Device(s): K110932, GN ReSound Tinnitus Sound Generator Module Description of Device: Unitron's Tinnitus Masker feature is a noise generator available as an option, in Unitron Era hearing aid platforms providing a sound enrichment source, to provide stimulation to the auditory system, as part of a comprehensive tinnitus management program that works to help distract the wearer's attention from the sound of their tinnitus. Indications for Use: Unitron's tinnitus masker is a broadband sound generator available in Unitron Era hearing aid platforms. It provides a means of sound enrichment therapy that can be used as part of a personalized tinnitus management program to provide temporary relief from tinnitus. Tinnitus masker is intended for adults 18 years of age or older who have both hearing loss and tinnitus The tinnitus masker within the air conduction hearing instrument is fitted by a hearing health care professional familiar with diagnosis and management of tinnitus. Page 9 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "Unitron" in a bold, sans-serif font. The "U" is stylized with a swooping line extending from the bottom left. A period is placed at the end of the word. The underlying principle of sound enrichment is to provide supplementary noise stimulation which can help defocus your attention from your tinnitus and avoid negative reactions. Sound enrichment, coupled with instructional counseling, is an established approach to managing tinnitus. Substantial Equivalence: Comparison to the Predicate Device indicated above shows the Tinnitus Masker feature has the same intended use and does not raise additional safety and effectiveness questions. Technology Characteristics Comparison: Unitron Tinnitus Masker uses the same technology used by the predicate device, sharing the following similarities: - . Similar indications of use, implementation into software, functionality through digital hearing instruments, fitter adjustments to stimulus shaping, maximum output and end user control. The technology characteristics that differ from the predicate device are: - Amplitude Modulation. . - . Environment Steering (automatic fluctuation of the noise level based upon the environment). Performance Data: Unitron Hearing Instruments with the tinnitus masker feature and related software have been evaluated in accordance with ANSI S3.22:2009 American National Standard -Specification of Hearing Aid Characteristics, IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2 Medical electrical equipment Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests. Conclusion: Based upon similar indications of use and technology characteristics it is concluded to be substantially equivalent to K110932, GN ReSound Tinnitus Sound Generator Module 510(k) for Unitron Tinnitus Masker Feature {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 7, 2013 Unitron Hearing, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW Buffalo, MN 55313 Re: K130494 Trade/Device Name: Unitron Tinnitus Masker Feature Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: March 19, 2013 Received: March 20, 2013 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3}------------------------------------------------ ### Page 2 - Mr. Mark Job or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, # Eric Anyann -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "unitron." in a bold, sans-serif font. To the left of the word is a stylized graphic that appears to be a signature or logo. The text is black against a white background, creating a high contrast image. # 4. Indications for Use Indication for Use Form Indications for Use 510(k) Number (if known): K130494 Device Name: Unitron Tinnitus Masker feature Indications for Use: Unitron's tinnitus masker is a broadband sound generator available in Unitron Era hearing aid platforms. It provides a means of sound enrichment therapy that can be used as part of a personalized tinnitus management program to provide temporary relief from tinnitus. Tinnitus masker is intended for adults 18 years of age or older who have both hearing loss and timitus. The tinnitus masker within the air conduction hearing instrument is fitted by a hearing health care professional familiar with diagnosis and management of tinnitus. The underlying principle of sound enrichment is to provide supplementary noise stimulation which can help defocus your attention from your tinnitus and avoid negative reactions. Sound enrichment, coupled with instructional counseling, is an established approach to managing tinnitus. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/4/Picture/15 description: The image shows the name "Srinivas Nandkumar - S" in a clear, sans-serif font. The text is horizontally aligned and appears to be the primary focus of the image. There is some noise or a watermark-like design behind the name, particularly around "Nandkumar", which adds a textured background effect to the text.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...