LIFEPAK 20 DEFIBRILLATOR/MONITOR, LIFEPAK 20E DEFIBRILLATOR/MONITOR
K130454 · Physio-Control, Inc. · MKJ · Aug 21, 2013 · Cardiovascular
Device Facts
Record ID
K130454
Device Name
LIFEPAK 20 DEFIBRILLATOR/MONITOR, LIFEPAK 20E DEFIBRILLATOR/MONITOR
Applicant
Physio-Control, Inc.
Product Code
MKJ · Cardiovascular
Decision Date
Aug 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors are intended for use in the hospital, physician's office, and clinic setting by personnel who are authorized by a physician/medical director. Thev are intended for use on a "crash cart" as well as for portable emergency response throughout a hospital.
Device Story
LIFEPAK 20/20e are hospital/clinic-based defibrillator/monitors for crash carts or portable emergency response. Inputs: ECG (3/5-wire), pulse oximetry, CO2 levels. Operation: Manual/automated defibrillation, noninvasive pacing, synchronized cardioversion, CO2 monitoring, CPR metronome. Powered by AC or internal battery (NiMH for 20, Li-ion for 20e). Output: Visual display of waveforms/vitals, audible/visual CPR metronome, wireless data transmission. Operated by authorized clinical personnel. Output informs clinical decisions regarding resuscitation, pacing, and ventilation. Benefits: Rapid arrhythmia termination, pacing for bradycardia, monitoring of CPR quality and endotracheal tube placement.
Clinical Evidence
No clinical or animal studies submitted. Performance verified via bench testing.
Technological Characteristics
Defibrillator/monitor with manual/AED modes, noninvasive pacing, ECG, pulse oximetry, and CO2 monitoring. Power: AC or internal battery (NiMH or Li-ion). Connectivity: Wireless data transmission. Features: CPR metronome, battery status indicator (20e).
Indications for Use
Indicated for manual defibrillation (ventricular fibrillation, symptomatic ventricular tachycardia, atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia); AED mode for cardiopulmonary arrest (unconscious, pulseless, non-breathing patients >8 years old); noninvasive pacing for symptomatic bradycardia; pulse oximetry for patients at risk of hypoxemia; and EtCO2 monitoring for breathing efficacy/CPR monitoring. Contraindicated for PEA, asystole (defibrillation/pacing), and ventricular fibrillation (pacing).
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Physio-Control. The logo is in white text on a black background. The text is stacked, with "PHYSIO" on top and "CONTROL" on the bottom.
# Submitter:
Physio-Control, Inc. 11811 Willows Road Northeast P.O. Box 97006 Redmond, Washington 98073-9706
**AUG 21 2013**
#### Contact Person:
Michelle Ackermann Principal Regulatory Affairs Specialist Phone: 425-867-4744, Fax: (425) 867-4154
#### Date Summary Prepared:
February 21, 2013
#### Device Trade Name:
LIFEPAK®20 defibrillator/monitor LIFEPAK®20e defibrillator/monitor
#### Device Common Name:
External defibrillator/monitor
### Device Classification:
| Classification Name | Class |
|------------------------------------------------------------------------------------------------|-------|
| Low Energy DC-Defibrillator (Including Paddles), (21CFR<br>870.5300, Product Code LDD) | II |
| Automatic External Defibrillators (Non-Wearable) (21CFR<br>870.5310, Product Code MKJ) | III |
| Cardiac Monitor (Including Cardiotachometer & Rate Alarm)<br>(21CFR870.2300, Product Code DRT) | II |
| External Cardiac Transcutaneous (Non-Invasive)<br>Pacemaker (21CFR870.5550, Product Code DRO) | II |
| Electrocardiograph | II |
| Hydraulic, Pneumatic, or Photoelectric Plethysmographs | II |
| Oximeter (21CFR870.2700, Product Code DQA) | II |
| Carbon-Dioxide Gas Analyzer Gaseous-Phase<br>(21CFR868.1400 Product Code CCK) | II |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Physio-Control. The logo is in a black box with white text. The text is stacked on top of each other with "PHYSIO" on top and "CONTROL" on the bottom.
#### Predicate Devices:
The features and functions of the proposed LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors are substantially equivalent to the previously cleared LIFEPAK 20 defibrillator/monitor, LIFEPAK 20e defibrillator/monitor and the LIFEPAK 15 monitor/defibrillator. 510(k) clearance numbers for the predicate devices are below:
| Predicate Device | 510(k) Number(s) |
|------------------------------------|------------------------------|
| LIFEPAK® 20 defibrillator/monitor | K012274, K033275,<br>K063119 |
| LIFEPAK® 20e defibrillator/monitor | K073089 |
| LIFEPAK® 15 monitor/defibrillator | K082937, K103567 |
#### Description:
The LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors were designed especially for hospitals and clinics for use on "crash carts" as well as for portable emergency response throughout a hospital. Features of the LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors include manual and automated external defibrillation, noninvasive pacing, ECG monitoring (3-wire or 5-wire), pulse oximetry, and synchronized cardioversion. The LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors are powered by AC power and a backup internal battery. The LIFEPAK 20 defibrillator/monitor utilizes a nickel-metal hydride internal battery while the LIFEPAK 20e defibrillator/monitor utilizes a lithium-ion internal battery. The LIFEPAK 20e defibrillator/monitor also has a battery status indicator on the device screen.
In addition to the features described above, the proposed LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors have a new CO2 monitoring feature, CPR metronome feature, wireless data transmission capability and a new optional accessory named the CodeManagement Module™.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Physio-Control. The logo consists of the word "PHYSIO" stacked on top of the word "CONTROL". The text is white and set against a black background. The logo is simple and easy to read.
#### Intended Use:
The LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors are intended for use in the hospital, physician's office, and clinic setting by personnel who are authorized by a physician/medical director. Thev are intended for use on a "crash cart" as well as for portable emergency response throughout a hospital.
Indications for Use (for both the LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors):
#### Manual Defibrillation
Indications: Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.
Contraindications: Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA) such as idioventricular escape rhythms, and in the treatment of asystole.
#### Automated External Defibrillation:
Indications: The AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 20 (and LIFEPAK 20e) defibrillator/monitor is not intended for use on pediatric patients less than 8 years old.
#### Noninvasive Pacing
Indications: Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.
Contraindications: Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Physio-Control. The logo is in white text on a black background. The word "PHYSIO" is on the top line, and the word "CONTROL" is on the bottom line.
## Pulse Oximetry
Indications: The pulse oximeter is indicated for use in any patient who is at risk of developing hypoxemia.
# End-Tidal CO2 monitoring:
# Indications
EtCO2 monitoring is used to detect the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Physio-Control. The logo is in a rectangular box with a black background. The words "PHYSIO" and "CONTROL" are stacked on top of each other in white, sans-serif font.
## Technological characteristics of the proposed and predicate devices:
The main difference between the predicate LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors and the proposed LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors is the new CO2 monitoring indication and CPR metronome feature. Device characteristics such as defibrillation waveform and the ECG analysis system are unchanged from the predicate LIFEPAK 20 and LIFEPAK 20e devices.
The CO2 monitoring and CPR metronome features in the proposed LIFEPAK 20 and LIFEPAK 20e defibrillator/monitor are similar to the CO2 monitoring and CPR metronome features in the predicate LIFEPAK 15 monitor/defibrillator.
# Conclusion of Testing
The information in this 510(k) notification demonstrates that the LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors are substantially equivalent to the predicate LIFEPAK 20 defibrillator/monitor, LIFEPAK 20e defibrillator/monitor and the LIFEPAK 15 monitor/defibrillator with respect to safety, effectiveness, and performance.
No clinical studies or non-clinical (i.e. animal) studies were submitted as part of this 510(k) notification. Performance testing of the proposed devices consisted of bench testing.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circle around the eagle. The eagle is black, and the text is also black.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 21, 2013
Physio-Control, Inc. c/o Ms. Michelle Ackermann Principal Regulatory Affairs Specialist 11811 Willows Road Northeast P.O. Box 97006 Redmond. WA 98073
Re: KI30454
Trade/Device Name: LIFEPAK 20/20e defibrillator/monitor Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ Dated: July 23, 2013 Received: July 24, 2013
Dear Ms. Ackermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Ms. Michelle Ackermann
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Owen P. Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Physio-Control. The logo is a black rectangle with the word "PHYSIO" stacked on top of the word "CONTROL" in white, sans-serif font. The letters are bold and slightly condensed, filling most of the rectangle's space.
Physio-Control Inc. LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 510(k) Premarket Notification
# SECTION D - STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): _ ... . . . . . . . . . . . .
Device Name: LIFEPAK® 20 defibrillator/monitor and LIFEPAK® 20e defibrillator/monitor
Indications For Use:
# Manual Defibrillation
Indications: Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. Delivery of this energy in the synchronized mode is a method for treating atrial fibrillation, atrial flutter, paroxysmal supraventricular tachycardia, and, in relatively stable patients, ventricular tachycardia.
Contraindications: Defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA) such as idioventricular escape rhythms, and in the treatment of asystole.
## Automated External Defibrillation:
Indications: The AED mode is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing normally before using the defibrillator to analyze the patient's ECG rhythm. In AED mode, the LIFEPAK 20 (and LIFEPAK 20e) defibrillator/monitor is not intended for use on pediatric patients less than 8 years old.
र Prescription Use (Part 21 CFR 801 Subpart D) AND/OR AND/OR AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 2
Digitally signed by Owen P. Faris -S Date: 2013.08.21 14:58:43 -04'00'
D-1
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "PHYSIO" stacked on top of the word "CONTROL". The words are in white font and are on a black background. The words are bolded and sans-serif.
Physio-Control Inc. LIFEPAK 20 and LIFEPAK 20e defibrillator/monitors 510(k) Premarket Notification
# SECTION D - STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known):
Device Name: LIFEPAK 20 defibrillator/monitor and LIFEPAK 8 20e defibrillator/monitor
Indications For Use:
# Noninvasive Pacing
Indications: Noninvasive pacing is indicated for symptomatic bradycardia in patients with a pulse.
Contraindications: Noninvasive pacing is contraindicated for the treatment of ventricular fibrillation and asystole.
## Pulse Oximetry
Indications: The pulse oximeter is indicated for use in any patient who is at risk of developing hypoxemia.
## End-Tidal CO2 monitoring:
Indications: EtCO2 monitoring is used to detect the level of expired CO2. It is used for monitoring breathing efficacy and treatment effectiveness in acute cardiopulmonary care, for example, to determine if adequate compressions are being performed during CPR or to rapidly detect whether an endotracheal tube has been placed successfully.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of 2
Digitally signed by
Owen P. Faris -S
Date: 2013.08.21
14:59:01-04'00'
Image /page/8/Picture/21 description: The image contains the text "D-2" in a large, bold, sans-serif font. The "D" is followed by a hyphen and the number "2". The text is black against a white background, and the overall impression is simple and clear.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.