K130451 · Syntermed, Inc. · KPS · May 17, 2013 · Radiology
Device Facts
Record ID
K130451
Device Name
NEUROQ 3.6
Applicant
Syntermed, Inc.
Product Code
KPS · Radiology
Decision Date
May 17, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
1) assist with regional assessment of human brain scans, through automated quantification of mean pixel values lying within standardized regions of interest (S-ROI's), and 2) assist with comparisons of the activity in brain regions of individual scans relative to normal activity values found for brain regions in FDG-PET scans, through quantitative and statistical comparisons of S-ROI's. 3) assist with comparisons of activity in brain regions of individual scans between two studies from the same patient, between symmetric regions of interest within the brain PET study, and to perform an image fusion of the patients PET and CT data 4) NeuroQ 3.6 provides added functionality to provide analysis of amyloid uptake levels in brain regions. The product is intended for use by trained nuclear technicians and nuclear medicine physicians. The clinician remains ultimately responsible for the final interpretation and diagnosis based on standard practices and visual interpretation of all SPECT and PET data.
Device Story
NeuroQ 3.6 is a software program for display and analysis of PET brain studies; operates within IDL environment on nuclear medicine computer systems. Inputs: patient PET brain scans, CT data, and reference database of asymptomatic controls. Processing: automated quantification of mean pixel values in 240 standardized regions of interest (S-ROI); statistical comparison of regional activity against normal database; longitudinal comparison of patient studies; symmetric region comparison; PET/CT image fusion; amyloid uptake analysis. Output: quantitative data and statistical significance maps. Used in clinical settings by nuclear medicine physicians and technicians. Physician integrates quantitative results with visual scan interpretation and clinical history for final diagnosis. Benefits: provides objective, standardized regional quantification to aid diagnostic interpretation; does not replace visual analysis.
Clinical Evidence
No clinical data provided. Effectiveness established through in-house software design, coding, debugging, testing, and validation as documented in the submission. Previous versions (NeuroQ 3.0 and NeuroQ-PET DP) were validated in prior 510(k) submissions (K072307, K041022).
Technological Characteristics
Software-based display and analysis program; operates in IDL (Interactive Data Language) environment; compatible with standard nuclear medicine computer systems. Features automated quantification of pixel values, image co-registration, and statistical comparison against normative databases. No specific hardware materials or energy sources; standalone software application.
Indications for Use
Indicated for use by trained nuclear technicians and physicians to assist in the assessment of human brain scans via automated quantification of mean pixel values in standardized regions of interest (S-ROI). Used for comparing individual FDG-PET brain scans against normal reference databases, comparing longitudinal studies of the same patient, comparing symmetric brain regions, performing PET/CT image fusion, and analyzing amyloid uptake levels.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
## ltem l
MAY 1 7 2013
# 510(k) SUMMARY
## Safety and Effectiveness
### 1. Medical Device Establishment:
Syntermed, Inc. Registration No. 1066019 Owner Operator I.D. 9041128 Device Requlation Number: 892.1200 Product Code: KPS Classification Panel: Radiology Voice: (888) 263-4446 ext 102, FAX: (714) 281-1290 Contact person: Kenneth F. Van Train Address: Syntermed, Inc. 130 Wieuca Road Suite 108 Atlanta, GA 30342 Email: vantrain@syntermed.com
Date Summary Prepared: February 20, 2013
2. Medical Device:
NeuroQ 3.6 - Display and Analysis program for PET Brain studies.
Classification Name - System, Tomography, Computed, Emission
3. Medical Device Equivalence:
NeuroQ™ 3.0, Reference #: K072307 and Siemens Medical Solutions USA, Inc. Scenium 3.0. Reference #:K123528.
#### 4. Device Description:
NeuroQ™ 3.6 has been developed to aid in the assessment of human brain scans through quantification of mean pixel values lying within standardized regions of interest, and to provide quantified comparisons with brain scans derived from FDG-PET studies of defined groups having no identified neuropsychiatric disease or symptoms, i.e., asymptomatic controls (AC). The Program provides automated analysis of brain PET scans, with output that includes quantification of relative activity in 240 different brain regions, as well as measures of the magnitude and statistical significance with which activity in each region
{1}------------------------------------------------
differs from mean activity values of brain regions in the AC database. The program can also be used to compare activity in brain regions of individual scans between two studies from the same patient, between symmetric reqions of interest within the brain PET study, and to perform an image fusion of the patients PET and CT data. The program can also be used to provide analysis of amyloid uptake levels in the brain. This program was developed to run in the IDL operating system environment, which can be executed on any nuclear medicine computer systems which support the IDL software platform. The program processes the studies automatically, however, user verification of output is required and manual processing capability is provided.
### 5. Intended Use and Potential Adverse Effect on Health:
The intended use of this program was to provide the physician with a program which would allow him to co-register and display brain PET scans and compare the patients study to a reference database. The program can be used to compare the activity in brain regions of individual scans between two studies from the same patient, between symmetric regions of interest within the brain PET study, and to perform an image fusion of the patients PET and CT data. The program can also be used to provide analysis of amyloid uptake levels in the brain. This program serves merely as a display and processing program to aid in the diagnostic interpretation of a patient's study. It was not meant to replace or eliminate the standard visual analysis of the PET brain scan. The physician should integrate all of the patients' clinical and diagnostic information, i.e. patients' history, quality control images, visual interpretation of the PET brain scan, and quantitative results, prior to making his final interpretation. This comprehensive processing technique (as with all diagnostic imaging) is not perfect, and will be associated with some false positive and false negative results. The previous validation of the program can be found in the 510(k) submission for NeuroQ™ - PET DP, Ref. 510(k) #: K041022 and NeuroQ™ 3.0 (510(k) #: K072307. The validation for modifications in version 3.6 can be found in Item H. Testing & Validation of this 510(k) and the physician should be aware of the accuracy when integrating the quantitative results for his final interpretation. Therefore, this program has no direct adverse effect on health since the results represent only a part of the information, which the physician will utilize for his final interpretation. The final responsibility for interpretation of the study lies with the physician.
#### 6. Marketing History:
There have been other medical device programs marketed in the past which perform similar functions to those performed by the NeuroQ™ 3.6 program. Most Nuclear Medicine manufacturers have programs that can co-register SPECT/PET data and some of them have programs for comparison of the patient's data to a reference database. NeuroQ™ 3.6 provides a program which executes in the IDL operating system environment and we believe is substantially equivalent to our previous version of NeuroQ™ - PET DP K041022 and NeuroQ™ 3.0 (510(k) #: K072307 and Siemens Medical Solutions USA, Inc. Scenium 3.0, Reference #:K123528. To our knowledge there have been no safety problems with NeuroQ™ - PET DP K041022 which has been in the marketplace since June 2004, NeuroQ™ 3.0 510(k) #: K072307 which has been in the marketplace since March 14, 2008, or Siemens
I2
{2}------------------------------------------------
Medical Solutions USA, Inc. Scenium 3.0, Reference #:K123528 which has been in the marketplace since December 20, 2012.
#### 7. Conclusions:
The safety of this program has been determined through the various stages of software development which included the initial design, coding, debugging, testing, and validation. The effectiveness of the initial program, NeuroQ™ - PET DP and NeuroQ™ 3.0, has been established in in-house testing and clinical validation studies submitted in our previous 510(k) K041022 and 510(k) #: K072307. Specific details and results concerning the validation of the NeuroQ™ 3.6 program are listed in Item H. Testing & Validation. We contend that the method employed for the development and the final in-house validation results of this medical display software program, NeuroQ™ 3.6, have proven its safety and effectiveness. In our opinion, NeuroQ™ 3.6 program is substantially equivalent to our previous version of NeuroQ™ - PET DP and NeuroQ™ 3.0 program and Siemens Medical Solutions USA, Inc. Scenium 3.0 which have been cleared for marketing. NeuroQ™ 3.6 program is intended for the same purpose and raises no new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is a stylized symbol that resembles a caduceus, which is a symbol of medicine. The symbol is made up of three wavy lines that are connected at the top and bottom.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6 Silver Spring, MD 20993-0002
May 17, 2013
Syntermed, Inc. % Mr. Kenneth F. Van Train President 245 Owens Drive ANAHEIM CA 92808
Re: K130451
Trade/Device Name: NeuroQ™ 3.6 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography Regulatory Class: Class II Product Code: KPS Dated: February 20, 2013 Received: February 22, 2013
Dear Mr. Van Train:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Kenneth F. Van Train
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Michael D. O'Hara for
Janine M. Morris Director, Division Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): k130451
Device Name: NeuroQTM 3.6
Indications for Use:
- 1) assist with regional assessment of human brain scans, through automated quantification of mean pixel values lying within standardized regions of interest (S-ROI's), and
- 2) assist with comparisons of the activity in brain regions of individual scans relative to normal activity values found for brain regions in FDG-PET scans, through quantitative and statistical comparisons of S-ROI's.
- 3) assist with comparisons of activity in brain regions of individual scans between two studies from the same patient, between symmetric regions of interest within the brain PET study, and to perform an image fusion of the patients PET and CT data
- 4) NeuroQ 3.6 provides added functionality to provide analysis of amyloid uptake levels in brain regions.
The product is intended for use by trained nuclear technicians and nuclear medicine physicians. The clinician remains ultimately responsible for the final interpretation and diagnosis based on standard practices and visual interpretation of all SPECT and PET data.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. Huse
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) K130451
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.