K130374 · Genesis Digital Imaging, Inc. · LLZ · Aug 29, 2013 · Radiology
Device Facts
Record ID
K130374
Device Name
OMNI-VUE 2 SYSTEM
Applicant
Genesis Digital Imaging, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 29, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Omni-Vue 2™ System is a software device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Device options make possible reading (including mammography), telecommunications; fast demonstration; etc .; and teleconferencing. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that meets technical specifications reviewed and accepted by FDA.
Device Story
Omni-Vue 2 is a PACS software for capturing, storing, distributing, and manipulating medical images. Inputs include DICOM images from modalities (CT, MR, ultrasound, R/F, CR/DR) or digitized film via imaging gateways. System operates on networked workstations; used by radiologists and clinicians for image review, telecommunications, and teleconferencing. Provides tools for image manipulation (W/L, rotation, flipping, zoom, ROI, measurement, annotation). Output is displayed on monitors for clinical decision-making. Enhances workflow by replacing traditional film handling; improves security via FTPS and SSL-based ODBC. Does not contact patients or control life-sustaining equipment.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed that all input/output functions and operational modes met predetermined acceptance criteria. No clinical data provided.
Technological Characteristics
Software-only PACS. Connectivity via DICOM (SCU/SCP, Q/R). Security protocols include FTPS (File Transfer Protocol SSL) and ODBC on SSL. Operates on networked imaging workstations. Does not contact patients. No hardware components.
Indications for Use
Indicated for use by clinicians to receive, store, communicate, process, and display digital medical images (CT, MR, ultrasound, R/F, CR/DR) across networks. Supports mammography interpretation when using FDA-cleared monitors. Contraindicated for primary interpretation of lossy compressed mammographic images and digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K130374
Page 1 of 5
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
#### Date Prepared:
August 25, 2013
#### Submitter's Information: 21 CFR 807.92(a)(1)
Mr. Alain Tabares, Chief Technical Officer Genesis Digital Imaging, Inc. 12921 W. Washington Blvd. Los Angeles. CA 90066
AUG 29 2013
#### Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Trade Name: | Omni-Vue 2™ System |
|------------------------|-----------------------------------------|
| Common Name: | Picture Archiving Communications System |
| Device Classification: | 892.2050 |
| Name: | System, Image Processing |
#### Predicate Device: 21 CFR 807. 92(a)(3)
| Device Classification Name | system, image processing, radiological20 |
|-----------------------------------|------------------------------------------|
| 510(K) Number | K073062 |
| Device Name | OMNI-VUE SYSTEM |
| Applicant | GENESIS DIGITAL IMAGING, INC. |
| Regulation Number | 892.2050 |
| Classification Product Code | LLZ |
| Date Received | 10/30/2007 |
| Decision Date | 03/10/2008 |
| Decision | substantially equivalent (SE) |
| Classification Advisory Committee | Radiology |
| Review Advisory Committee | Radiology |
| Type | Traditional |
#### Device Description: 21 CFR 807 92(a)(4)
Omni-Vue 2™ System makes possible the capturing, storage, distribution, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available, the System can acquire medical images using a DICOM gateway, which generates DICOM-type files. For example, film digitizers obtain images from old film and convert them to meet DICOM standards and stored files are transmitted using a network and can be viewed or manipulated from an imaging workstation.
#### Indications for Use: 21 CFR 807 92(a){5)
Omni-Vue 2™ System is a software device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Device options make possible reading (including mammography), telecommunications; fast demonstration; etc ; and teleconferencing. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic
{1}------------------------------------------------
images may only be interpreted using an FDA cleared monitor that meets technical specifications reviewed and accepted by FDA.
#### Technological Characteristics: 21 CFR 807 92(a)(6)
Omni-Vue 2™ System is a software product that handles and manipulates digital medical images. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed. OmniVue 2 is downsized product of OmniVue related with number of feature, however, OmniVue 2 improves data transfer security, and productivity and brevity of PACS configuration. The reason for data transfer security is improved, is due to FTPS(File Transfer Protocol SSL) and ODBC(Open Database Connectivity) on SSL. Another reason for brevity of PACS configuration is improved, is due to the fact that OmniVue 2 client application provides simple way to get image from CR, DR and digitizer.
The modified device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices. Both systems have been developed to replace traditional film handling in radiology. The 2 devices are substantially equivalent in the areas of design, architecture, general function, application, and intended use.
Any difference between the two devices does not affect safety or efficacy. The predicate device and the new device are compared below:
| Item | OmniVue 2 | OmniVue<br>Predicate | Description |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Login | Yes | Yes | For security purpose, login function is recommended. |
| Patient's list | Yes | Yes | Easy way to find the patient |
| DICOM Storage SCU | Yes | Yes | Routing to another workstation |
| DICOM Storage SCP | Yes | Yes | Getting images from other workstation |
| DICOM Q/R SCU | Yes | Yes | Requests patient from PACS server. |
| DICOM Q/R SCP | Yes | Yes | Response for patient exam request. Neo module has this<br>function. |
| DICOM Print Management | No | Yes | Film printing feature has been removed from the modified<br>device.<br>Difference between the predicate and modified device, but has<br>no impact on safety or efficacy of the modified device and<br>does not raise any new potential safety risks |
| Open DICOM Dir CD | No | Yes | The feature to open patient's CD from other PACS devices has<br>been removed from the modified device.<br>Difference has no impact on safety or efficacy of the modified<br>device and does not raise any new potential safety risks |
| Burn patient's CD | Yes | Yes | |
| CR interface: Direct importing | Yes | Yes | CR: Carestream Vita interface |
| Film digitizer interface | Yes | Yes | Vidar film digitizer interface |
| Multi film scanning | Yes | Yes | Multi films at one time. |
| Manual image import | Yes | Yes | Importing image one by one. |
| Batch import | Yes | Yes | Importing multiple images in a certain folder. |
| Non-DICOM file open and<br>import | No | Yes | The feature to open and import a non-DICOM file such as<br>bmp, jpg and tiff and convert into DICOM format has been<br>removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Paper printing | No | Yes | The feature to print a DICOM image to a paper printer has<br>been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety |
| Item | OmniVue 2 | OmniVue Predicate | Description |
| | | | risks |
| Twain interface | No | Yes | The Interface with twain device feature used to scan images<br>and other items has been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Auto close | No | Yes | The auto feature to close current exam whenever a new exam<br>is opened has been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Chest CT View | No | Yes | The feature to compare images at the same time with<br>difference Window Width/Level has been removed in the<br>modified device.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Image layout change | Yes | Yes | Change layout by user's demand. |
| Series layout change | Yes | Yes | Change series layout by user's demand. |
| Image manipulation tools<br>- W/L<br>- Rotation<br>- Flipping<br>- Zoom in/out<br>- Magnify glass<br>- Auto W/L<br>- Reset modification<br>- Invert | Yes | Yes | Basic tools to manipulate images |
| DICOM header view | Yes | Yes | DICOM header dump for administrator |
| Cine view | No | Yes | The feature for Cine View for a slide show for images has<br>been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks. |
| Hounsfield unit | No | Yes | The feature for using CT exam Hounsfield unit has been<br>removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Scout line | No | Yes | The scout line feature for CT and MR has been removed. The<br>difference has no impact on safety or efficacy of the modified<br>device and does not raise any new potential safety risks |
| Close and Open | Yes | Yes | Close current exam open next exam |
| W/L preset | No | Yes | The Window/Level preset database feature has been<br>removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Sharpen preset | No | Yes | The sharpen preset database has feature been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Look-Up Table preset | Yes | Yes | Use gamma value for image processing |
| Hanging protocol | No | Yes | The limited hanging protocol feature has been removed.<br>The difference has no impact on safety or efficacy of the modified device<br>and does not raise any new potential safety risks |
| Hanging protocol preset | No | Yes | The preset for layouts feature has been removed from<br>modified device.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Item | OmniVue 2 | OmniVue<br>Predicate | Description |
| Thumbnail view | Yes | Yes | Thumbnail image for exam |
| Chiropractic tools<br>Cobb's angle<br>Center mass<br>Atlas plane line<br>New Cobb's angle<br>Extended Cobb's angle<br>Horizontal deflection<br>Chiro ratio<br>Measure from horizontal<br>line | No | Yes | Specially designed tools for chiropractors has been removed<br>from modified device.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| Arbitrary rotation | No | Yes | The rotate image with arbitrary angle feature has been<br>removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| W/L for ROI only | No | Yes | The Window/Level feature in the ROI (region of interest) only<br>has been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks |
| ROI | Yes | Yes | Region of Interest (ROI) |
| Measuring tools<br>Ruler<br>Angle | Yes | Yes | Measure angle or length |
| Customizing DICOM overlay | No | Yes | The feature to insert a customized DICOM overlay on the<br>image has been removed. The difference t has no impact on<br>safety or efficacy of the modified device and does not raise<br>any new potential safety risks. |
| External program interface | Yes | Yes | Launch external program |
| Customizing toolbar set | No | Yes | The feature to customize toolbar icons has been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks. |
| Customizing worklist layout | Yes | Yes | Worklist layout change and column order change |
| Email with attachment | No | Yes | The feature to email an image file has been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks. |
| Old exam list | Yes | Yes | List up old exam for the patient |
| Reporting | Yes | Yes | Create diagnostic report by radiologist |
| L/R mark on image | No | Yes | The feature to mark left/right image is removed because the<br>function resides in a different module that indicates<br>L/R Mark on Image. This feature was a duplicate.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks. |
| Masking | No | Yes | The feature to mask image pixel if the pixel value is same with<br>certain color has been removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks. |
| Image stitching | No | Yes | The feature to stitch multiple images to one has been removed<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks. |
| Annotation | Yes | Yes | Write text on the image |
| All series mode | Yes | Yes | Set multiple series as one series |
| Item | OmniVue 2 | OmniVue<br>Predicate | Description |
| DICOM information edit | Yes | Yes | Other module has it.<br>Only technician can modify DICOM information when they<br>made mistake. |
| Arrow lines | Yes | Yes | Draw arrow lines |
| Overlay show/hide | Yes | Yes | |
| Display original size image | No | Yes | The feature to display the image by real body size has been<br>removed.<br>The difference has no impact on safety or efficacy of the<br>modified device and does not raise any new potential safety<br>risks. |
{2}------------------------------------------------
.
. . . . .
{3}------------------------------------------------
{4}------------------------------------------------
K130374
Page 5 of 5
#### Nonclinical Testing:
The complete system configuration has been assessed and tested at the factory and has passed all in-house testing criteria. The Verification Test Plan was designed to evaluate all input functions, output functions, and actions performed by the Omni-Vue 2™ software in each operational mode and followed the process documented in the Validation Test Plan.
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
If the device is installed Genesis Digital Imaging, Inc., integration and installations verification tests are conducted against acceptance criteria prior to release to the client.
#### Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for Omni-Vue 2™ contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device.
The subject device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The subject and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The modification to the subject device does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Therefore, Omni-Vue 2™ is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling an eagle or bird with three curved lines representing its wings or feathers. The symbol is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2013
Genesis Digital Imaging, Inc. % Mr. Carl Alletto Consultant OTech, Inc. 1600 Manchester Way CORINTH TX 76210
Re: K130374
Trade/Device Name: Omni-Vuc 2™ System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 1, 2013 Received: August 13, 2013
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commence prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it your do roo is enabilional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 l CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Mr. Alletto
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/Medicall.cevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal/roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130374
Device Name: Omni-Vue 2TM
Indications for Use:
Omni-Vue 2™ System is a software device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Device options make possible reading (including mammography), telecommunications; fast demonstration; etc .; and teleconferencing.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor that meets technical specifications reviewed and accepted by FDA
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpari C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
| Sinh for | ) |
|----------|---|
|----------|---|
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) K130374 _______________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.