The Navios Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping. It can be used in conjunction with the following monoclonal antibody reagents and software package: CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 monoclonal antibody reagents. These reagents provide identification and enumeration of CD3+CD4+, CD3+CD8+, CD3+, CD19+ and CD3-CD56+ lymphocyte percentages and absolute counts in peripheral whole blood. Absolute counts may be determined by the Navios flow cytometer using Flow-Count Fluorospheres (single platform technology method) or separate hematology results (dual platform method). These reagents are indicated for use in the immunologic assessment of patients having or suspected of having immune deficiency. Navios tetra Software for automated analysis and results with CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 monoclonal antibody reagents. Navios Software may be installed on an independent computer workstation for off-line analysis of listmode files generated by the Navios Flow Cytometer with the monoclonal antibody reagents and software package listed above. The off-line analysis must be performed in accordance with the product labeling.
Device Story
Navios Flow Cytometer uses hydrodynamic focusing to align cells in suspension, passing them through a 488 nm laser beam. Scattered and fluoresced light is collected by photomultiplier tubes (PMTs), transduced into electronic pulses, and processed by the system computer. The device performs qualitative and quantitative immunophenotyping of lymphocyte subpopulations in whole blood using monoclonal antibody reagents. The Navios tetra Software provides automated gating and analysis of listmode data. The system supports batch or single-tube sampling via an automated carousel loader. Healthcare providers use the resulting cell percentages and absolute counts to assess immune status. The device benefits patients by providing standardized, automated lymphocyte enumeration for diagnosing or monitoring immune deficiencies. It is intended for clinical laboratory use by trained personnel.
Clinical Evidence
Bench testing only. Studies included method comparison (CLSI EP9-A2) against predicate, precision (CLSI EP5-A2), linearity (CLSI EP06-A), carryover (CLSI H26-A2), and limit of detection/quantitation (CLSI EP17-A2). Results showed comparable performance to the predicate device for lymphocyte enumeration. Single vs. dual platform absolute counting methods were compared, and manual vs. automated (PrepPlus 2) sample preparation methods were shown to be comparable.
Technological Characteristics
Flow cytometer with 488 nm laser (22mW). Uses hydrodynamic focusing, photomultiplier tubes (4 standard), and digital integrator circuitry (40 MHz sampling). Materials include polystyrene fluorospheres. Connectivity includes PC-based workstation (Windows Vista). Sterilization/cleaning via FlowClean reagent. Software uses locked algorithms for automated gating. Dimensions/form factor: Benchtop system with 32-tube carousel loader.
Indications for Use
Indicated for immunologic assessment of patients having or suspected of having immune deficiency. Used for identification and enumeration of lymphocyte subpopulations (CD3+CD4+, CD3+CD8+, CD3+, CD19+, CD3-CD56+) in peripheral whole blood.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
Cytomics FC 500 with tetraCXP software (k030828)
Flow-Set Fluorospheres (k944751)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE
A. 510(k) Number:
k130373
B. Purpose for Submission:
New Submission
C. Manufacturer and Instrument Name:
Beckman Coulter, Inc.
Navios Flow Cytometer
D. Type of Test or Tests Performed:
Flow Cytometric Immuno-assay: Cell identification with quantitative cell counts
E. System Descriptions:
1. Device Description:
The Navios Flow Cytometer uses flow cytometric principles to determine qualitative and quantitative measurements of biological and physical properties of cells and other particles. These properties are measured when the cells pass through the laser beam(s) in single file.
The Navios Flow Cytometry System also offers an optional standalone offline software package, Navios software, which enables the user to reanalyze, display, print and export acquired listmode data.
The Navios tetra Software is an optional locked algorithm application plug-in that is designed for the Navios Flow Cytometer. It provides automated analysis and results for tetraCHROME reagents; this application cannot be modified by the user.
Flow-Set Pro Fluorospheres is a suspension of fluorospheres with uniform and stable size and fluorescence intensity. The stability of these product parameters allows for the standardization of light scatter and fluorescence intensity instrument settings. For automated, algorithm-based Beckman Coulter flow cytometry IVD systems such as Navios tetra Software, target value ranges for X-Mode (fluorescent intensity) are provided by Beckman Coulter with each Flow-Set Pro Fluorospheres lot on the Table of Application Target Settings.
The Navios Flow Cytometer system consists of:
- Navios Flow Cytometer
- Navios tetra Software
- Navios Software (off-line analysis tool)
- Flow-Set Pro Fluorospheres
- CYTO-STAT tetraCHROME reagents
- COULTER IMMUNOPREP Reagent System
{1}
- TQ-Prep Workstation (Accessory for Sample Preparation)
- PrepPlus™ 2 Workstation (Accessory for Sample Preparation)
2. Principles of Operation:
The Navios flow cytometer aligns cells in suspension by hydrodynamic focusing and passes them through a laser beam one cell at a time. The scattered and fluoresced light that emanates from these cells as they intersect the laser beam is collected by photomultiplier tubes and transduced to an electronic pulse and counted by the cytometer’s computer. This test depends on the ability of a monoclonal antibody to bind to the surface of cells expressing discrete antigenic determinants. Specific cell staining is accomplished by incubating whole blood with the monoclonal antibody reagent. The CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8- ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5. Monoclonal Antibody Reagents are each a combination of four murine monoclonal antibodies, each conjugated to a specific fluorochrome and specific for a different cell surface antigen. Red blood cells are lysed with the COULTER ImmunoPrep Reagent System. The remaining white blood cells are analyzed on a Navios flow cytometry system with the Navios tetra Software package. Navios Flow Cytometer Software and Navios tetra Software, in conjunction with quality control reagents, provide automated standardization of light scatter and fluorescence intensities and automated adjustment of color compensation settings. The Navios tetra Software, in conjunction with the Navios Flow Cytometer Software, and CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8- ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19- ECD/CD3-PC5 Monoclonal Antibody Reagents, provides automated analysis of lymphocyte subpopulations.
3. Modes of Operation:
Batch or single tube
4. Specimen Identification:
Barcode reader or manual entry
5. Specimen Sampling and Handling:
The automated sample loader for the instrument uses a carousel that holds thirty-two 12 x 75-mm test tubes and includes a barcode reader.
6. Calibration:
Fluorescence intensity is calibrated by adjusting photomultiplier voltage using Flow-Set Pro Fluorospheres.
The absolute count of a population is based on the calibration factor (CAL Factor) and the number of Flow-Count fluorospheres particles within a user defined CAL region.
7. Quality Control:
IMMUNO-TROL Cells and IMMUNO-TROL Low Cells are assayed, lysable, whole-blood quality control products used to verify the activity of the CYTO-STAT tetraCHROME reagents and to verify the methods used for staining targeted cells, lysing
2
{2}
erythrocytes, and analyzing samples with flow cytometry.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR 864.5220 – Automated Differential Cell Counter
2. Classification:
Class II
3. Product code:
OYE- flow cytometric reagents and accessories
PDX- flow cytometry calibrator
4. Panel:
81 Hematology
G. Intended Use:
1. Indication(s) for Use:
Navios Flow Cytometer:
The Navios Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping. It can be used in conjunction with the following monoclonal antibody reagents and software package:
- CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 monoclonal antibody reagents. These reagents provide identification and enumeration of CD3+CD4+, CD3+CD8+, CD3+, CD19+ and CD3-CD56+ lymphocyte percentages and absolute counts in peripheral whole blood. Absolute counts may be determined by the Navios flow cytometer using Flow-Count Fluorospheres (single platform technology method) or separate hematology results (dual platform method). These reagents are indicated for use in the immunologic assessment of patients having or suspected of having immune deficiency.
- Navios tetra Software for automated analysis and results with CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 monoclonal antibody reagents.
Navios tetra Software may be installed on an independent computer workstation for off-line analysis of listmode files generated by the Navios Flow Cytometer with the monoclonal antibody reagents and software package listed above. The
3
{3}
off-line analysis must be performed in accordance with the product labeling.
## Navios tetra Software:
The Navios tetra Software is intended for use as an in vitro diagnostic device for immunophenotyping with CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 monoclonal antibody reagents on the Navios Flow Cytometer. It provides automated analysis and results for the identification and enumeration of CD3+CD4+, CD3+CD8+, CD3+, CD19+ and CD3-CD56+ lymphocyte percentages and absolute counts in peripheral whole blood. Absolute counts may be determined by the Navios flow cytometer using Flow-Count Fluorospheres (single platform technology method) or separate hematology results (dual platform method). It is indicated for use in the immunologic assessment of patients having or suspected of having immune deficiency.
## Flow-Set Pro Fluorospheres
Flow-Set Pro Fluorospheres is a suspension of fluorescent microspheres used as an aid in standardizing forward scatter, side scatter, and fluorescence detectors (FL1-4) on the Cytomics FC 500 and Navios Flow Cytometers.
2. Special Conditions for Use Statement(s):
For prescription use only
Special instrument requirements:
Use with TQ-Prep Workstation (Accessory for Sample Preparation) and PrepPlus™ 2 Workstation (Accessory for Sample Preparation)
## H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
Cytomics FC 500 with tetraCXP software, k030828
Flow-Set Fluorospheres, k944751
2. Comparison with Predicate Device:
Navios Flow Cytometer with Navios tetra Software and Navios Analysis Software
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Navios Flow Cytometer: | Cytomics FC 500 Flow Cytometer with tetraCXP software: |
| | The Navios Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping. It can be used in conjunction with the following monoclonal | The tetraCXP Software for Cytomics FC 500 flow cytometry systems and CYTO-STAT tetraCHROMETM CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROMETM CD45-FITC/CD56-RD1/CD19- |
{4}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | antibody reagents and software package:
• CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 monoclonal antibody reagents. These reagents provide identification and enumeration of CD3+CD4+, CD3+CD8+, CD3+, CD19+ and CD3-CD56+ lymphocyte percentages and absolute counts in peripheral whole blood. Absolute counts may be determined by the Navios Flow Cytometer using Flow-Count Fluorospheres (single platform technology method) or separate hematology results (dual platform method). These reagents are indicated for use in the immunologic assessment of patients having or suspected of having immune deficiency.
•Navios tetra Software for automated analysis and results with CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CYTO-STAT tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 monoclonal antibody reagents.
•Navios Software may be installed on an independent computer workstation for off-line analysis of listmode files | ECD/CD3-PC5 monoclonal antibody reagents combine four-color fluorescent monoclonal antibody reagents, quality control reagents, an optional absolute count reagent, and software for automated analysis of lymphocyte populations in whole blood using Cytomics FC 500 Flow Cytometry systems with CXP Software.
The system with CYTO-STAT tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 is intended "For In Vitro Diagnostic Use" and allows simultaneous identification and enumeration of total CD3+ (T), CD19+ (B), and CD3-/CD56+ (NK) lymphocyte population percentages and absolute counts. This reagent reflects the distribution of the three major subsets comprising the lymphocyte population upon which other lymphocyte enumeration studies are based and provides the total lymphocyte percentage. |
5
{5}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | generated by the Navios Flow Cytometer with the monoclonal antibody reagents and software package listed above. The off-line analysis must be performed in accordance with the product labeling. | |
| Regulation Number and Description, Product Code | 864.5220 –Automated Differential Cell Counter, OYE | 864.5220 –Automated Differential Cell Counter, GKZ |
| Safety Features | Interlocks and mitigation of hazards via software and hardware controls | Same |
| Controlling Software | Same software architecture. | The acquisition software enables the user to acquire data from the FC 500 Flow Cytometry instrument and to analyze, display, print and export acquired listmode data. The embedded software resides in the FC 500 instrument. It controls the instrument functionality including the multi-carousel loader (MCL) for sample introduction.
The embedded software controls the instruments’ lasers, acquisition system and fluidics. The instrument fluidics aspirates the sample, and performs instrument maintenance functions such as startup/shutdown. The embedded software also captures and provides the data to the workstation for processing. |
| System Configuration | Same except PC based workstation running Microsoft Windows Vista application specific software | Bench top
• PC based workstation running Microsoft Windows 2000 application specific software
• Printer |
| tetraCHROME Reagent Assay | Two tube monoclonal antibody reagent assay
• CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5
• CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5
Provides identification and | Same |
{6}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | enumeration of Total CD3+, CD3+CD4+, CD3+CD8+, CD3-CD56+, and CD19+ percentages and absolute count results | |
| tetraCHROME Reagents Specimen and Prepared Sample Stability Claims | Specimen stability – 24 hours room temperature
• Prepared sample stability
• 2 hours room temperature
• 24 hours refrigerated temperature | Same |
| Sample Preparation with Monoclonal Antibodies | Off-board sample preparation following instructions provided with cleared monoclonal antibody reagent | Same |
| Sample Presentation | Prepared sample added to a daughter tube | Same |
| Sample Mixing prior to Acquisition | Prepared sample is vortex mixed | Same |
| Sample Introduction | Tube sampler
• Automated presentation with Multi-tube Carousel Loader (MCL) from 32 test tube capacity carousel
• Manual presentation into a tube location on a MCL via tube access door | Same |
| Aspiration Pathway | Same aspiration pathway used for automated and manual presentation | Same |
| Sample Identification | Bar-code reading of carousel position and labeled sample tube. User may also identify samples based on carousel location with a work list. | Same |
| Sample Analysis | Same except automated gating of tetraCHROME achieved with Navios tetra software | Principle of analysis – Flow Cytometric
• Detection hardware – Lasers, fluidics, optics, electronics
• Sample analysis pathway
• Manual gating of cellular populations per reagent IFU or automated gating of cellular populations for tetraCHROME reagents with tetraCXP software |
| Lasers | IVD configuration – 488 nm | One Laser Standard Configuration |
{7}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | IVD configuration – Standard doubled diode
• 22mW
• 488nm
• Vertical polarization
• Mode 00 | Standard Argon Ion
• 22mW
• 488 nm
• Vertical polarization
• Mode 00 |
| Optics | Same except collection also uses fiber optics | Free space delivery and collection of laser light |
| Electronics | 40 MHz sampling Digital integrator circuitry w/ early stage ADC Yields ~25,500 events/sec | 200 KHz sampling Analog integrator circuitry w/ late stage ADC Yields ~1,500 events/sec |
| Photomultiplier Tubes (PMTs) / Colors | IVD configuration – Standard 4 PMTs | Same |
| Maximum Parameter Detectors | IVD configuration – Six (FS, SS, FL1 – FL4) | Same |
| Off-line Analysis Software | Optional software kits for analysis on off-line workstations | Same |
| Data Handling | Electronics and software required to support data handling | Same |
| Data Reporting | FlowPAGE, Panel Report, Plots, and Statistics printouts | Same |
| Cleaning Cycle Between Samples | Executed with IsoFlow Sheath Fluid ensuring carryover specification is met | Same |
| Cleanse Cycle | Same except uses FlowClean cleaning reagent | Cleaning cycle performed with COULTER Clenz cleaning reagent as part of the daily shutdown process, before and after running samples with vital dyes that stain the tubing, and as part of troubleshooting. |
| Quality Control Techniques | Daily Instrument Checks Commercial Controls
• Inter-laboratory Quality Assurance Program (IQAP) | Same |
| Workstation | Software functionality to allow:
• Patient data management – storage, review, reporting
• Control data management – storage, review, reporting
• System configuration | Same |
8
{8}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | management
• System service test and adjustment procedures. | |
| Laser Optics | Achromatic Crossed Cylinder Lens - 488 spot size = 10 x 80mm | Crossed Cylinder Lens - 488 spot size = 10 x 80mm |
| Fluorescence Collection | Collection lens and light path 1st lens set:
• Gel coupled to flow cell operates at NA1.2
• Forms images at 21x magnification
2nd lens set:
• Located at fiber entry reduces the magnification per channel to 3x
• Light passes through optical fiber to PMT area
• Light leaves the fiber and is collimated
• Collimated light is separated by filters into bands
• PMTs detect light and convert to electrical signals | Collection lens and light path 1st lens series:
• Glued to flow cell operates at NA1.2
• Forms Images at 3x magnification
• Image formed at .020 diameter pinhole
2nd lens series:
• Image at pinhole is transformed into collimated beam
• Collimated light is separated by filters into desired bands
• PMTs detect light and convert to electrical signals |
| Color Separation | Collimated beam is separated into desired components with dichroic filters used in a linear topology | Collimated beam is separated into desired components with dichroic filters used in a tree topology |
| Sample Introduction | • Automated presentation with Multi-tube Carousel Loader (MCL) from 32 test tube capacity carousel
• Manual presentation into a tube location on MCL via tube access door | Same |
| Sample Analysis | Principle of analysis – Flow cytometric
• Detection hardware – Lasers, fluidics, optics, electronics
• Sample analysis pathway
• Manual gating of cellular populations by user or automated gating of | |
9
{9}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | cellular populations with algorithmic software | |
| Automated Algorithm Analysis for tetraCHROME reagents | Navios tetra Software | tetraCXP Software |
| Accessories | Optional software kits for analysis on off-line workstations | Same |
Flow-Set Pro Fluorospheres
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Flow-Set Pro Fluorospheres is a suspension of fluorescent microspheres used as an aid in standardizing forward scatter, side scatter, and fluorescence detectors (FL1-4) on the Cytomics FC 500 and Navios Flow Cytometers. | Flow-Set Fluorospheres are a suspension of Fluorospheres (fluorescent microspheres) used as an aid in optimizing a flow cytometer for quantitative analysis of human leukocytes. |
| Regulation Number and Description, Product code | 864.5220 – Automated Differential Cell Counter, PDX | 864.8625– Hematology Quality Control Mixture, JPK |
| Reagent Components | Same except 3 μm (nominal diameter) polystyrene fluorospheres | 3.6 μm (nominal diameter) polystyrene fluorospheres suspended in an aqueous medium containing surfactants and preservatives at 1 x 106 fluorospheres/mL (nominal concentration) |
| Fluorescence Emission | Ranges from 515-800 nm when excited at 488 nm | Ranges from 525 nm to 700 nm when excited at 488 nm |
| Manufacturing Process | Manufactured by loading a solution of three dyes into a polystyrene particle. | Polymerized in the presence of a single, broad spectrum dye. |
| Sample Preparation | Must be thoroughly mixed prior to use. Samples in test tubes which stand for an extended period of time should be vortex mixed before use. No other sample preparation is necessary. | Same |
{10}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Target Value Ranges | Same as Flow-Set with target value ranges provided for Navios tetra Software | Procedure provided for user to establish target value ranges for each manually gated application. Additionally, target value ranges provided for algorithm-based IVD software applications such as tetraCXP for FC 500 Flow Cytometer |
| Final Product Form | Liquid, ready to use | Same |
| Open Vial Stability | 65 days | Same |
| Closed Vial Stability | 10 months | 18 months |
I. Special Control/Guidance Document Referenced (if applicable):
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline, Second Edition 2004.
CLSI C28-A3, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline.
CLSI EP15-A2 User Verification of Performance for Precision and Trueness; Approved Guideline – Second Edition
NCCLS EP09-A2-IR, Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition (Interim Revision)
CLSI H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Guideline-Second Edition; Section 5.7 – Carryover
CLSI EP17-A2 Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline, Second Edition
CLSI H42-A2, Enumeration of Immunologically Defined Cell Populations by Flow Cytometry; Approved Guideline-Second Edition
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
Method comparison with predicate device:
Several method comparison analyses were performed in order to support the comparability of the automated, algorithm gating methods and manual gating method with tetraCHROME reagents (tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5) and Flow-Count fluorospheres (single platform absolute counting method).
{11}
i. Method Comparison Study 1: Navios Flow Cytometer with Navios tetra Software (algorithm) vs. FC 500 Flow Cytometer with tetraCXP Software (algorithm). Three sites (one in the US and two in Canada) collected a total of 291 specimens (75% abnormal) for the assessment of agreement to the predicate method. Results met all acceptance criteria. Navios Flow Cytometer has demonstrated acceptable clinical performance accuracy as compared to the predicate, FC 500 Flow Cytometer.
| | | Total CD3+ | Total CD3+ | CD3+/ CD4+ | CD3+/ CD4+ | CD3+/ CD8+ | CD3+/ CD8+ |
| --- | --- | --- | --- | --- | --- | --- | --- |
| units | | cells/μL | % | cells/μL | % | cells/μL | % |
| n | | 291 | 291 | 291 | 291 | 291 | 291 |
| Correlation | | 0.990 | 0.990 | 0.994 | 0.997 | 0.993 | 0.998 |
| Mean | tetraCXP | 1366 | 77.96 | 570 | 31.70 | 763 | 44.24 |
| | Navios tetra | 1402 | 77.12 | 601 | 32.27 | 780 | 43.58 |
| Slope (95% Confidence Limits) | | 1.006 (0.994 - 1.018) | 0.992 (0.962 - 1.022) | 1.045 (1.016 - 1.075) | 0.978 (0.960 - 0.995) | 1.019 (1.003 - 1.035) | 0.995 (0.985 - 1.004) |
| Intercept (95% Confidence Limits) | | 29.182 (17.490 - 40.873) | -0.220 (-2.611 - 2.171) | 10.820 (0.8592 - 0.782) | 1.297 (0.837 - 1.758) | 3.995 (-4.812 - 12.802) | -0.377 (-0.749 - -0.005) |
| | | Total CD3+ | Total CD3+ | CD3-/CD56+ | CD3-/CD56+ | CD19+ | CD19+ |
| --- | --- | --- | --- | --- | --- | --- | --- |
| units | | cells/μL | % | cells/μL | % | cells/μL | % |
| n | | 291 | 291 | 291 | 291 | 291 | 291 |
| Correlation | | 0.989 | 0.994 | 0.993 | 0.992 | 0.995 | 0.995 |
| Mean | tetraCXP | 1375 | 78.00 | 137 | 7.87 | 217 | 12.03 |
| | Navios tetra | 1420 | 77.48 | 145 | 8.03 | 228 | 12.15 |
| Slope (95% Confidence Limits) | | 1.027 1.0091.046 | 1.000 0.9861.0 15 | 1.047 1.0271.067 | 1.003 0.9731.033 | 1.036 1.0221.050 | 1.022 0.9951.050 |
| Intercept (95% Confidence Limits) | | 8.779 (-10.940 - 28.498) | -0.553 (-1.726 - 0.620) | 2.350 (0.875 - 3.824) | 0.160 (0.003 - 0.317) | 3.228 (0.474 - 5.983) | -0.010 (-0.157 - 0.137) |
ii. Method Comparison Study 2: Navios Flow Cytometer with Navios tetra Software (algorithm) vs. FC 500 Flow Cytometer with Manual Gating. Results met all acceptance criteria. Navios Flow Cytometer with Navios tetra Software has demonstrated acceptable clinical performance accuracy as compared to the predicate, FC 500 Flow Cytometer with Manual Gating.
iii. Method Comparison Study 3: Navios Flow Cytometer with Manual Gating vs. FC 500 Flow Cytometer with Manual Gating. Results met all acceptance criteria. Navios Flow Cytometer with Manual Gating has demonstrated acceptable clinical performance accuracy as compared to the predicate, FC 500 Flow Cytometer with Manual Gating.
{12}
iv. Method Comparison Study 4: Navios Flow Cytometer with Manual Gating vs. Navios Flow Cytometer with Navios tetra Software (algorithm). Results met all acceptance criteria. Navios Flow Cytometer with Manual Gating has demonstrated acceptable clinical performance accuracy as compared to the Navios Flow Cytometer with Navios tetra Software (algorithm).
# b. Precision/Reproducibility:
To assess variability, three precision studies were conducted:
i. Precision Study 1: CLSI EP05-A2 precision study with control material using Navios tetra software
Precision studies with control materials were conducted in accordance with CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline. Two levels of control material (IMMUNO-TROL Cells and IMMUNO-TROL Low Cells) were evaluated in the study. The controls were prepared with tetraCHROME reagents (CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5) and Flow-Count Fluorospheres. They were run in duplicate twice each day (morning and afternoon) for a minimum of 20 days. The test was executed on three Navios Flow Cytometers with Navios tetra Software at the three independent test sites. All parameters met the precision and reproducibility criteria.
Acceptance Criteria:
| Analyte / Units of Measure | % Positive cells (Total CD3+, CD3+/CD4+, CD3+/CD8+, CD3-/CD56+, CD19+) | | Absolute cell counts (CD3+/CD4+, CD3+/CD8+, Total CD3+, CD3-/CD56+, CD19+) | |
| --- | --- | --- | --- | --- |
| Mean | ≤20% | >20% | <300 cells/μL | ≥300 cells/μL |
| Repeatability Specification (%CV) | ≤10% | ≤5% | ≤10% | ≤5% |
| Reproducibility Specification (%CV) | ≤15% | ≤10% | ≤15% | ≤10% |
ii. Precision Study 2: CLSI EP05-A2 precision study with control materials using the tetraCHROME reagents manual gating method.
To assess the precision of manually gated data on the Navios Flow Cytometer, the listmode data collected at Ohio State University, Columbus, Ohio in Precision Study 1, CLSI EP05-A2 was replayed and manually gated by three qualified operators according to the tetraCHROME reagents and Flow-Count Fluorospheres Instructions for Use (IFU). The acceptability criteria are as stated above for precision study 1. All parameters met the precision and reproducibility criteria.
iii. Precision Study 3: Whole Blood Repeatability
Whole blood specimen prepared with tetraCHROME CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and tetraCHROME CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5
{13}
reagents and Flow Count fluorospheres from three test sites (Ohio State University, Columbus, Ohio, London Health Sciences Centre, London, Ontario, Canada, and Foothills Hospital, Calgary, Canada) were run in duplicate on the Navios Flow Cytometer with Navios tetra Software. The dataset consisted of 336 whole blood specimens including samples representing the CD4 measuring range and analyte concentrations at the medical decision points. Whole blood repeatability followed the approach per CLSI H26-A2 'Validation, Verification and Quality Assurance of Automated Hematology Analyzers' standard, of providing estimates of repeatability using biological samples representing normal and medical decision points. All parameters met the precision and reproducibility criteria.
Acceptance Criteria:
| Analyte / Units of Measure | % Positive cells (Total CD3+, CD3+/CD4+, CD3+/CD8+, CD3-/CD56+, CD19+) | | Absolute cell counts (CD3+CD4+, CD3+/CD8+, Total CD3+, CD3-/CD56+, CD19+) | |
| --- | --- | --- | --- | --- |
| Mean | ≤20% | >20% | <300 cells/μL | ≥300 cells/μL |
| Repeatability Specification (%CV) | ≤10% | ≤5% | ≤10% | ≤5% |
Precision and reproducibility of Flow-Set Pro fluorospheres
Three lots of Flow-Set Pro fluorospheres were tested on a Navios Flow Cytometer over twenty days according to methods defined in CLSI EP5 to assess repeatability and reproducibility. All parameters demonstrated CV's $< 15\%$ and as such met the manufacturer's acceptance criteria. Flow-Set Pro fluorospheres were also tested on an Epics FC500 Flow Cytometer and found to meet the manufacturer's acceptance criteria.
# c. Linearity:
i. Instrument linearity was assessed with two Immunobrite beads of different fluorescent intensities analyzed at 5 different voltages covering the dynamic range of each PMT. The mean channel fluorescent intensity ratios of pairs was calculated at five different voltages levels across three decades on all PMTs. Lower and Upper Limits are calculated from the mean value $+5\%$ . The results of this study demonstrate that the intensity differences between the beads were constant across the range of voltages assessed.
ii. Linearity of the tetraCHROME reagents on the Navios Flow Cytometer with Navios tetra software and the manual gating method was established in accordance with CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. Using the following linearity acceptance limits:
$\mathrm{R}^2\geq 0.95$
0-300 cells/μL: ± 30 cells/μL bias
- >300 cells/μL: ± 10% bias
The analytical system was found to be linear for each parameter as indicated in the following table:
{14}
15
| tetra | Parameters | Navios tetra | Manual Gating |
| --- | --- | --- | --- |
| CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 | CD3+/μL | 50-4653 | 49-4605 |
| | CD3+/CD4+/μL | 33-3147 | 33-3148 |
| | CD3+/CD8+/μL | 15-1408 | 15-1389 |
| CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 | CD3+/μL | 64-5218 | 65-5323 |
| | CD3-/CD56+/μL | 11-969 | 11-994 |
| | CD19+/μL | 11-1081 | 10-996 |
iii. Analytical Measuring Range
Based on the LOD/LOQ studies the following are reported as the analytical measuring range for this device.
| tetra | Parameters | Analytical Measuring Range |
| --- | --- | --- |
| CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 | CD3+/μL | 65-4500 |
| | CD3+/CD4+/μL | 35-3000 |
| | CD3+/CD8+/μL | 15-1300 |
| CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5 | CD3+/μL | 65-4500 |
| | CD3-/CD56+/μL | 35-900 |
| | CD19+/μL | 35-900 |
d. Carryover:
i. A carryover study was performed using Flow Check Pro fluorospheres on each of 4 Navios flow cytometers according to the methodology presented in CLSI Document: H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard – Second Edition, Section 5.7 Carryover was demonstrated to be less than the acceptance limit of 0.1% from one sample to another.
ii. A carryover study was performed to evaluate 1) specimen carryover and 2) reagent carryover using whole blood analyzed with the tetraCHROME system on each of 3 Navios Flow Cytometers according to the methodology presented in CLSI Document: H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard – Second Edition, Section 5.7 Scatter and fluorescence carryover was demonstrated to be less than the acceptance limit of 1% from one specimen to another when the numbers of gated events is 10,000 or greater.
e. Interfering Substances:
Not applicable
2. Other Supportive Instrument Performance Data Not Covered Above:
a. Traceability, Stability, Expected values (controls, calibrators, or methods): Flow-Set Pro Fluorospheres
i. Open vial stability was assessed across three lots of Flow-Set Pro Fluorospheres. Each lot was tested and fluorescence intensity of the associated peaks was assessed on Days 0, 40, 60, 65 and 70. The acceptance criteria are as follows:
{15}
16
| | Specification |
| --- | --- |
| Mean signal intensity and MFI | ± 20% from baseline |
| Monodispersity | • ≥80% for FS and FL1-FL 4
• ≥60% for SS |
ii. Closed vial stability was assessed as above and the drift from time zero to 10 months for all channels on the three lots (TL1, TL2, and TL4) were shown to be within the ±20% allowable drift, indicating that the stability specification are met.
iii. Flow Set Pro fluorospheres meet the open vial claim of 65 days and the closed vial claim of 10 months.
b. Detection limit:
i. Limit of Blank (LoB): Two lots of ImmunoTrol erythrocytes were treated with CytoStat tetraCHROME reagents, Immunoprep reagents and Flow-Count to determine the LoB. The LoB was established for each marker as the upper 95% of the data as recommended in the CLSI EP17-A Guideline.
ii. Limit of Detection (LoD) and Limit of Quantitation (LoQ): Eight dilutions of 2 lots of ImmunoTrol cells were prepared in ImmunoTrol erythrocytes and were treated with 3 lots of CytoStat tetraCHROME reagents, Immunoprep reagents and Flow-Count to measure the LoD and LoQ. LOD and LOQ were determined on two Navios Flow Cytometers by analyzing 2 freshly prepared replicates of each sample over 4 days. The equation provided in section 5.1 of EP17-A was used to calculate the low limit of quantitation (LLoQ) - Spec + 2SDs = Total Error.
| Navios tetra | CD3+ | CD3+/CD4+ | CD3+/CD8+ | CD3-/CD56+ | CD19+ | CD3+ |
| --- | --- | --- | --- | --- | --- | --- |
| LoB (cells/uL) | 1 | 1 | 1 | 0 | 0 | 1 |
| LoD (cells/uL) | 2 | 2 | 2 | 1 | 1 | 2 |
| LLoQ (cells/uL) | 7 | 9 | 6 | 16 | 12 | 6 |
| LLoQ Upper 95% Confidence (cells/uL) | 14 | 15 | 10 | 33 | 15 | 10 |
| Manual | CD3+ | CD3+/CD4+ | CD3+/CD8+ | CD3-/CD56+ | CD19+ | CD3+ |
| --- | --- | --- | --- | --- | --- | --- |
| LoB (cells/uL) | 1 | 1 | 0 | 0 | 0 | 1 |
| LoD (cells/uL) | 2 | 2 | 1 | 1 | 1 | 2 |
| LLOQ (cells/uL) | 4 | 6 | 5 | 13 | 17 | 3 |
| LLoQ Upper 95% Confidence (cells/uL) | 6 | 8 | 8 | 24 | 29 | 12 |
c. Specimen Sampling and Handling:
i. Thirty specimens having CD4 counts across the AMR were prepared within 8 hours and held at room temperature for 24, 48 and 72 hours and prepared again in
{16}
order to verify the expected stability as reported in the literature. Specimens held 24 and 48 hours were refrigerated and tested again after an additional 24 and 48 hours. The acceptance criteria for the reported parameters are as follows:
| Analyte / Units of Measure | Stability Limits |
| --- | --- |
| CD3+/CD4+ count, cells/μL | ±30 cells/μL or ±10%, whichever is greater |
| CD3+/CD4+ % | ±2 percentage points |
| CD3+/CD8+ count, cells/μL | ±45 cells/μL or ±15%, whichever is greater |
| CD3+/CD8+ % | ±3 percentage points |
| Total CD3+ count, cells/μL | ±45 cells/μL or ±15%, whichever is greater |
| Total CD3+ % | ±3 percentage point drift |
| CD3-/CD56+ count, cells/μL | ±45 cells/μL or ±15%, whichever is greater |
| CD3-/CD56+ % | ±2 percentage points |
| CD19+ count, cells/μL | ±45 cells/μL or ±15%, whichever is greater |
| CD19+ % | ±2 percentage points |
All parameters were verified to demonstrate the claimed stability of 24 hours for whole blood stored at room temperature and 24 hours for prepared specimens stored at refrigerated temperature.
ii. Twenty five specimens having CD4 counts across the AMR were prepared within 8 hours and held at room temperature for 24 and 32 hours and prepared again in order to verify the expected stability as reported in the literature. Twenty three specimens prepared as above were also stored an additional 48 hours at refrigerated temperatures and analyzed using the automated and manual methods. The acceptance criteria were as stated above. All parameters were verified to demonstrate the claimed stability of 24 hours for whole blood stored at room temperature and 2 hours for prepared specimens stored at room temperature.
# d. Laser Performance:
Laser performance was assessed on two (2) Navios Flow Cytometers using Flow-Check Pro Fluorospheres and it was demonstrated that integral signal intensity does not vary more than $\pm 5\%$ from the average integral signal channel number obtained over a period of 24 hours for Short-term Laser Performance and 8 days for Long-term Laser.
# e. PrepPlus 2 Specimen Processor:
Studies were performed to assess the accuracy of the PrepPlus 2 specimen processor according to procedures detailed in CLSI EP15. Forty specimens were processed with the PrepPlus 2 and compared to those prepared with manual pipette. The acceptance criteria for the study are as follows:
| Analyte / Units of Measure | Accuracy Specifications |
| --- | --- |
| Cell count (Total CD3+, CD3+/CD4+, CD3+/CD8+, CD3-/CD56+, CD19+) | 0-300 cells/μL: ± 40 bias >300 cells/μL: ± 13% bias |
| Percent (Total CD3+, CD3+/CD4+, CD3+/CD8+, CD3-/CD56+, CD19+) | 0-40%: ± 1.5 percentage point bias >40%: ± 2.5 percentage point bias |
{17}
The results demonstrated that PrepPlus 2 performs comparably to manual pipette for sample preparation when used with the Navios flow cytometer and tetraCHROME reagents and Flow Count for absolute count.
f. Single versus Dual Platform Method Comparison:
Forty one specimens (abnormal and normal) were selected for study. Absolute counts were determined using both Flow-Count beads (single platform) and compared to the absolute counts as calculated as the product of the subset percentage and the lymphocyte count from the LH 780 Hematology Analyzer (K061616) (dual platform). The methods were considered equivalent if the absolute value of the upper or lower 95% confidence limit of the bias is statistically smaller than the allowable difference between the two methods (±10%). The two methods met the acceptance criteria for all reported parameters.
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.