The DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations: - Pulmonary thromboembolism when anticoagulants are contraindicated. - Failure of anticoagulant therapy for thromboembolic disease. - Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced. - Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. DENALI® Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
Device Story
The DENALI® Filter System is a percutaneously placed, retrievable vena cava filter designed to prevent pulmonary embolism. The device consists of a laser-cut Nitinol filter with twelve appendages (six arms, six legs) providing two levels of filtration; features include cranial/caudal anchors for migration resistance and penetration limiters. The delivery system includes an introducer sheath, dilator, and pusher. Used in clinical settings by physicians, the device is deployed via femoral or jugular access under fluoroscopic guidance. Radiopaque markers on the dilator and sheath aid in positioning and measuring IVC diameter. The filter captures emboli while allowing blood flow. Clinical decision-making relies on fluoroscopic visualization during placement and retrieval. The device benefits patients by providing mechanical protection against pulmonary embolism when anticoagulation is not viable.
Clinical Evidence
Prospective, multi-center, single-arm clinical study of 175 patients (mean age 56.7 years). Primary endpoint: Clinical Success Placement (CSP) defined as freedom from PE, embolization, caval occlusion, or procedure-related death. CSP was 96.1% (95% CI: 91.2%). Technical success of placement was 100%. Retrieval success rate was 97.7% (86/88 attempts). No device-related deaths, caval occlusions, or filter fractures reported. Two cases of symptomatic PE occurred. Adverse events were adjudicated by an independent Clinical Events Committee.
Technological Characteristics
Nitinol filter (laser-cut, single-piece with welded snare tip); twelve appendages (six arms, six legs); cranial/caudal anchors; penetration limiters. Delivery system: 8.4 French I.D. introducer sheath, dilator (0.035" guidewire compatible, 800 psi max pressure). Radiopaque marker bands for visualization. Biocompatibility per ISO 10993. Sterilization: standard method (not specified).
Indications for Use
Indicated for prevention of recurrent pulmonary embolism in patients with contraindications to anticoagulants, failure of anticoagulant therapy, or emergency treatment for massive pulmonary embolism. Suitable for placement in the inferior vena cava (diameter ≤ 28 mm).
Regulatory Classification
Identification
A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”
Predicate Devices
ECLIPSE® Filter System – Femoral and Jugular/Subclavian Delivery Kit (K101431)
Submission Summary (Full Text)
{0}------------------------------------------------
Page 2
#### DENALI® Filter System 510(k) Summary 21 CFR 807.92
## MAY 1 5 2013
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 |
|------------|-------------------------------------------------------------------------------|
| Phone: | 480-638-2906 |
| Fax: | 480-449-2546 |
| Contact: | Joni Creal, Sr. Regulatory Affairs Specialist |
| Date: | May 14, 2013 |
#### Subject Device Name
| Device Trade Name: | DENALI® Filter System –<br>Jugular/Subclavian Delivery Kit |
|-----------------------|------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Filter, Intravascular, Cardiovascular |
| Classification: | Class II |
| Classification Panel: | Cardiovascular Devices |
| Product Code: | DTK |
| Predicate Devices: | ECLIPSE® Filter System – Femoral and Jugular/Subclavian<br>Delivery Kit (K101431; Clearance June 25, 2010) |
### Summary of Change
The ECLIPSE® Filter received FDA clearance under K101431 on June 25, 2010. As part of the product improvement life cycle, Bard Peripheral Vascular, Inc. (BPV) has chosen to re-design its vena cava filter platform. The new fitter design is named the DENALL® Filter and incorporates cranial anchors, caudal anchors, penetration limiters and will be terminally electropolished. In addition, minor changes have been made to the IFU.
Bard Peripheral Vascular, Inc.
Image /page/0/Picture/13 description: The image contains the word "BAIRD" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The word appears to be a logo or brand name.
{1}------------------------------------------------
#### Device Description
The DENALI® Filter consists of twelve Nitinol appendages emanating from a central snareable tip. These twelve appendages (six legs and six arms) form two levels of filtration for emboli: the legs provide the lower level of filtration and the arms provide the upper level of filtration. Four out of the six legs have cranial anchors and the remaining two legs have caudal anchors. In addition, all of the legs have penetration limiters. The anchors have been designed to resist cranial and caudal migration, while allowing the filter to be percutaneously removed. The DENALI® Filter is intended to be used in the inferior vena cava with diameters less than or equal to 28 mm.
The DENALI® Delivery Systems consist of an introducer sheath and dilator, and a preloaded DENALI® Filter in a storage tube with a pusher. The dilator accepts a 0.035" guidewire and allows for an 800 psi maximum pressure contrast power injection. Radiopaque marker bands are on the end of the dilator to aid in measuring the maximum indicated IVC diameter. They are spaced at a distance of 28 mm (outer-to-outer). The 55 cm, 8.4 French I.D. introducer sheath contains a radiopaque marker band at the distal tip and hemostasis valve with a side port. The pusher advances the filter through the introducer sheath to the pre-deployment mark and is then used to fix the filter in place while the filter is unsheathed.
#### Indications for Use of Device
The subject device, the DENAL® Filter System – Femoral and Jugular/Subclavian Delivery Kits, is indicated for use in the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulants are contraindicated. .
- . Failure of anticoagulant therapy for thromboembolic disease.
- . Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.
- Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is . contraindicated.
DENALI® Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
Image /page/1/Picture/13 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and prominent appearance. The overall design is simple and eye-catching, making the word easily readable.
{2}------------------------------------------------
#### Technological Comparison to Predicate Devices
The DENALI® Filter System – Femoral and Jugular/Subclavian Delivery Kit has the following similarities to its predicate device, the ECLIPSE® Filter System ~ Femoral and Jugular/Subclavian Delivery Kit (clearance to market via K101431 on June 25, 2010):
- . Same intended use
- Same indications for use .
- . Similar filter and delivery system design
- . Same target population
- . Same operating principle
- . Same fundamental scientific technology
- . Similar packaging configuration and materials
- . Same sterility assurance level and method of sterilization
The differences between the subject device and its predicate device are as follows:
- Laser-cut filter from a Nitinol tube (single piece with welded snare tip) .
- . Caudal anchors
- . Penetration limiters (one on each filter leg)
- Dimensional modifications .
- . Delivery System Modifications
- . Updated IFU
The successful completion of the testing required per the risk assessment demonstrates that the technological characteristics and performance criteria of the DENAL® Filter System - Femoral and Jugular/Subclavian Delivery Kits are comparable to the predicate device and that the subject device can perform in a manner substantially equivalent to devices currently on the market for the same intended use.
#### Performance Testing – In-Vitro Testing
To demonstrate substantial equivalence of the subject device to the predicate device. the technological characteristics and performance criteria were evaluated using the in-vitro testing as outlined below:
- . In-Vitro - Filter
- Fatigue Resistance
Image /page/2/Picture/25 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a three-dimensional appearance. The font is uniform in size and style, and the letters are evenly spaced.
{3}------------------------------------------------
- । Corrosion Resistance
- Cranial Migration Resistance l
- -Caudal Migration Resistance
- l Penetration Resistance (Radial force)
- -Tensile
- -Removal Force
- Clot Trapping -
- Filter Tip Visibility (Radiopacity) ।
- । MRI Compatibility
- In-Vitro Delivery System .
- Deployment Force -
- Deployment Accuracy ।
- -Arm/Leg Entanglement (Configuration)
- Filter Centering (Tilt)
- -Simulated Use
- Delivery System Tensile Strength -
- | Delivery System Torque
- -Visual Inspection (Freedom from surface defects)
- Delivery System Visibility (Radiopacity) l
- Dimensional Verification ー
- Burst Pressure ।
- Biocompatibility, per ISO 10993 Filter .
- Cytotoxicity ।
- Sensitization ।
- -Irritation - Intracutaneous Reactivity
- Acute Systemic Toxicity -
- l Subacute Toxicity and Implantation
- l Genotoxicity
- । Hemocompatibility
- Biocompatibility, per ISO 10993 Delivery System .
- Cytotoxicity ।
- Sensitization ၊
- Irritation Irracutaneous Reactivity |
Image /page/3/Picture/36 description: The image shows the word "BARD" in a stylized, bold font. The letters are outlined in black, giving them a distinct and prominent appearance. The font style is blocky and somewhat geometric, with a unique design for the letter "A" that resembles a triangle.
{4}------------------------------------------------
- । Acute systemic toxicity
- । Hemocompatibility
#### Performance Testing - In-Vivo Testing
Two GLP compliant animal studies were performed in support of the DENALL® Filter System. one to assess the filter and one to assess the delivery systems.
To assess the fitter, an animal study was performed with the primary objective of the study being to evaluate the retrievability of the DENALI® Filter in an ovine animal model following implantation periods of 4 and 12 weeks. Twelve (12) filter retrievals were evaluated for: 1) ease of removal (as assessed by a clinical evaluator) and 2) cava wall damage as assessed by venography, gross evaluation, and histopathology (in two separate subgroups - immediately post-euthanasia (0 week healing) and after an 8 week healing period). The secondary objectives of this study were to assess caval narrowing/stenosis, caval patency, extravasation, filter strut configuration, filter visibility under fluoroscopy, fracture, intimal irregularities, migration, penetration, perforation, thrombus and tilt.
To assess the delivery systems, an acute animal study in an ovine animal model was performed. The acute study validated the following attributes of 12 Femoral and 12 Jugular/Subclavian DENALI® Filter Systems:
- . Dilator Visibility
- Dilator Marker Band Visibility .
- Introducer Sheath Tip (Jugular) and Introducer Marker Band (Femoral) . Visibility
- Dilator/Introducer Trackability .
- Dilator/Introducer Pushability .
- . Aspiration
- Delivery System Trackability .
- . Delivery System Pushability
- . Ease of Deployment (Deployment Force)
- . Deployment Accuracy
- Filter Centering (Tilt) .
- Arm/Leg Entanglement .
4
Image /page/4/Picture/21 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a three-dimensional appearance. The overall design is simple and eye-catching.
{5}------------------------------------------------
- Filter Visibility Under Fluoroscopy .
- Snare Tip Visibility under Fluoroscopy .
- Pusher Assembly Visibility Under Fluoroscopy .
Based upon the performance assessment, all acceptance criteria were met and the DENALL® Filter System was deemed acceptable by clinical evaluators. Specifically to the retreivability animal study, all 12 filters were successfully removed with an acceptable retrieval force. In addition, based on clinical evaluation and confirmed by the pathologist there was no observed caval occlusion/thrombosis, no definitive IVC penetrations/perforations, no contrast extravasation from the VC after filter removal, no significant filter tilting, and no hemodynamically significant caval stenosis. One death occurred following filter placement which the pathologist and attending veterinarian determined not to be device related.
The delivery systems were assessed by a clinical evaluator and the following delivery system attributes were found to be acceptable: dilator visibility, dilator marker band visibility, introducer sheath tip (Jugular), introducer markerband (Femoral) visibility, dilator/introducer trackability, dilator/introducer pushability, aspiration, delivery system (with Filter) trackability, delivery system (with Filter) pushability, ease of deployment (Deployment Force), deployment accuracy, filter centering (Tilt), arm/leg entanglement, filter visibility under fluoroscopy, snare tip visibility under fluoroscopy and pusher assembly visibility under fluoroscopy.
#### Performance Testing - Clinical Testing
A single-arm, prospective, multi-center clinical study was conducted to assess the safety of the DENALI® Filter as both a permanent and retrievable device. Clinical Success Placement (CSP) was defined as freedom from subsequent PE, filter embolization, caval occlusion, filter or procedure related death, insertion adverse events, and technical failure of placement. The preestablished performance goal was that the lower bound of the 95% confidence interval for the CSP was greater than 80%. Technical success of placement (TSP) was defined as deployment of the filter such that the physician judged the location to be suitable to provide sufficient mechanical protection against PE. Additionally, the secondary endpoints of recurrent PE, new or worsening DVT. filter migration. filter fracture and tilt were assessed at the six month visit or the one month post-retrieval visit.
Image /page/5/Picture/11 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, with a white fill. The word is horizontally oriented and appears to be a logo or brand name.
{6}------------------------------------------------
#### Traditional 510/k) DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kit
One hundred seventy five (175) patients (107 males, 68 females) were enrolled at 20 investigational sites across the United States. The mean age was 56.7 ± 15.8 years (range 18 – 89 years). Eighty six (86) patients had their filter successfully retrieved.
Of the 175 patients who underwent DENAL® Filter placement, 95 had active thromboembolic disease (the presence of DVT or PE at the time of filter placement). Of these 95 patients, 63 had a contraindication to anticoagulation, 7 had a complication related to the use of anticoagulant medication, 8 had a failure of anticoagulation, and 17 had a filter placed without contraindication, complication or failure related to anticoagulant medication. Eighty (80) patients without active thromboembolic disease (neither DVT nor PE at the time of filter placement) were enrolled in the study.
Reasons for filter placement were as follows: Surgery (n=86, 49%), Trauma (n=41, 23%), Hypercoagulopathy (n=33, 19%), Cancer (n=7, 4%), Stroke (n=3, 2%) and Other (n=5, 3%).
Sixty nine (69) patients completed a six month visit. Longer term data was available in 21 patients at 12-month follow up and in one patient at 18-month follow up. The study will continue to follow all patients to 24 month post retrieval, whichever comes first. Three (3) patients withdrew their consent, 3 were lost to follow up and 7 died from pre-existing conditions. An independent Clinical Events Committee (CEC) determined that no patient deaths were attributed to the filter device, or implant or retrieval procedures. This clinical experience will be updated in the IFU once the study is complete.
{7}------------------------------------------------
| Eligible<br>for<br>Visit<br>(N) | Visit<br>Completed<br>(N, %) | Retrieved | Reason Visit Not Completed | | | | Events Occurring<br>Before Next Visit | | Pending<br>Visit | |
|---------------------------------|------------------------------|-----------|----------------------------|--------------------------|-----------|-----------------|---------------------------------------|----------|------------------|-----|
| | | | Death1 | Lost to<br>Follow-<br>Up | Withdrawn | Missed<br>Visit | Migration | Fracture | | |
| Baseline/Implant | 175 | 175 | N/A | 0 | 0 | 0 | 0 | 0 | 0 | N/A |
| 3 Months | 128 | 114 (89%) | 28 | 6 | 1 | 0 | 5 | 0 | 0 | 2 |
| 6 Months | 88 | 69 (78%)2 | 32 | 1 | 1 | 3 | 5 | 0 | 0 | 9 |
| 12 Months | 25 | 21 (84%)3 | 23 | 0 | 0 | 0 | 1 | 0 | 0 | 3 |
| 18 Months | 4 | 1 (25%) | 3 | 0 | 0 | 0 | 0 | 0 | 0 | 3 |
| 24 Months | 0 | 0 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | 0 |
| Retrieval | 88 | 88 | 86 | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
| 30 Days Post-<br>Retrieval | 86 | 68 (79%)4 | N/A | N/A | 1 | N/A | N/A | N/A | N/A | 8 |
#### Table 1: Patient Accountability
1 CEC adjudicated, not device related
2 69 patients completed the 6-month visit, 9 patients were retrieved in the 6-month visit window (6-months plus or minus 30 days)
1 patient was retrieved in the 12-month visit window (12-months plus or minus 30 days)
477 patients completed the 30 Day Post-Retrieval Visit, 9 of which occurred during the six month visit window. 68 patients completing the one month post-retrieval visit are reported in the table above to prevent assessment at the 6-month visit and one month post-retrieval visit.
TSP for the DENALI® Filter was 100%. CSP for the DENALI® Filter was 96.1% and the lower bound of the 95% confidence interval was 91.2%. It was concluded that the performance goal was successfully met. Mean placement procedure time was 17.5 minutes.
There were no findings of caval occlusion, filter fracture, cranial migration, filter tilt at placement, or filter tilt at retrieval. There were two (2) cases of symptomatic PE; neither of which caused patient death. There were five (5) cases of asymptomatic penetration; none of which had clinical sequelae. Three (3) cases of penetration were noted at implant and two (2) cases of penetration were noted at retrieval. Twelve (12) patients reported thirteen (13) cases of new or worsening DVT. There were ten (10) cases of new DVT and three (3) cases of worsening DVT. All 10 new DVTs reported were in those patients that had active disease at the time of implant, were considered to be hypercoagulable, or those that had orthopedic procedures on their lower extremities. All site-reported adverse events were adjudicated by the CEC.
Image /page/7/Picture/11 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and appear to be filled with a solid color, likely black or a dark shade. The font style gives the word a strong and prominent appearance.
{8}------------------------------------------------
| Table 2: Complication Rates | |
|---------------------------------|-----------------|
| Recurrent PE | 2 / 1391 (1.4%) |
| Caval Occlusion | 0 / 1372 (0%) |
| New DVT | 10 / 137 (7.3%) |
| Worsening DVT | 3 / 137 (2.2%) |
| Filter Fracture | 0 / 137 (0%) |
| Cranial Migration | 0 / 137 (0%) |
| Caudal Migration | 0 / 137 (0%) |
| Filter Penetration at Placement | 3 / 175 (1.7%) |
| Filter Penetration at Retrieval | 2 / 883 (2.3%) |
| Filter Tilt at Placement | 0 / 175 (0%) |
| Filter Tilt at Retrieval | 0 / 883 (0%) |
The denominator of 139 includes 69 patients who completed the 6-month visit. 68 patients who reached the one month post-retrieval visit and 2 patients who had a reported PE outside of the 6month visit or the one month post-retrieval visit
1All complication rates with a denominator of 137 include 69 patients who completed the 6-month visit and 68 patients who reached the one month post-retrieval visit
388 patients had a retrieval visit with 86 successful retrievals
DENALI® Filter retrieval was attempted in 88 patients and successful in 86 patients (97.7%). In the two (2) unsuccessful retrieval cases, the snare was unable to engage the filter retrieval hook due to anatomical curvature. Mean filter indwell time was 136.2 ± 90.6 days (median 120.0 days, range 5 – 454 days). The right internal jugular vein was used in all retrieval procedures and mean procedure time was 21.9 minutes.
Venacavograms taken before and after the retrieval procedures of the IVC implant site revealed abnormalities that the CEC determined to be related to the device in two patients. One patient had minimal, self limited contrast extravasation post retrieval and another patient experienced intimal injury and caval narrowing of the IVC post retrieval. No clinical sequelae were reported for either patient.
Seventy seven (77) of the 86 patients who had their filter retrieved completed one month followup, one (1) subject was lost to follow-up, and eight (8) were pending. No instances of recurrent PE or new or worsening DVT were reported for any patient completing the one month postretrieval visit.
Bard Peripheral Vascular, Inc.
Image /page/8/Picture/9 description: The image shows the text "Page 10" in a bold, sans-serif font. The text is black against a white background. The words are stacked on top of each other, with "Page" on the top line and "10" on the bottom line.
Image /page/8/Picture/10 description: The image shows the word "Baird" in a bold, sans-serif font. The letters are all capitalized and outlined in black. The word is slightly distorted, with the letters appearing to be stretched horizontally.
{9}------------------------------------------------
| Table 3: DENALI® Filter Retrieval Details | |
|-------------------------------------------|------------|
| Number of Filter Retrieval Attempts | 88 |
| Number of Successful Retrievals | 86 |
| Retrieval Success Rate | 97.7% |
| Mean Indwell Time | 136.2 days |
| Maximum Indwell Time | 454 days |
Image /page/9/Figure/3 description: The image is a histogram titled "Figure 3: Time from Implantation to Retrieval (N=86)". The x-axis is labeled "Time Interval (In Days)" and ranges from 30 to 480 in increments of 30. The y-axis is labeled "Number of Subjects" and ranges from 0 to 15 in increments of 3. The histogram shows the distribution of time intervals, with the highest frequency at 90 and 180 days.
### Conclusion
The DENALI® Filter System – Femoral and Jugular/Subclavian Delivery Kits are substantially equivalent to the legally marketed predicate device, the ECLIPSE® Filter System – Femoral and Jugular/Subclavian Delivery Systems (K101431).
Image /page/9/Picture/8 description: The image shows the word "BARD" in a stylized, outlined font. The letters are bold and sans-serif, with a unique design where the inner parts of the letters are not filled in, creating a hollow effect. The overall appearance is clean and modern.
{10}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/10/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine. The caduceus is made up of three intertwined snakes and a staff.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 15, 2013
Bard Peripheral Vascular, Inc. c/o Ms. Joni Creal Regulatory Affairs Specialist II 1625 West Third Street Tempe, AZ 85281
Re: K130366
Trade/Device Name: Denali Filter System – Femoral Delivery Kit and Jugular Delivery Kit Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular intravascular filter Regulatory Class: Class II Product Code: DTK Dated: February 14, 2013 Received: February 15, 2013
Dear Ms. Creal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability
warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{11}------------------------------------------------
Page 2 - Ms. Joni Creal
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Bram D. Zự chế man -S
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
## Indications for Use
510(k) Number (if known): K130366
Device Name: DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kits
Indications for Use:
The DENALI® Filter System - Femoral and Jugular/Subclavian Delivery Kits are indicated for use in the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulants are contraindicated. .
- . Failure of anticoagulant therapy for thromboembolic disease.
- . Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.
- . Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.
DENALI® Filter may be removed according to the instructions supplied under the section labeled: Optional Procedure for Filter Removal.
Prescription Use X (Part21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bram D. Zut 2013.05.1
Bard Peripheral Vascular, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.