← Product Code [GEX](/productcode/GEX) · K130344

# DORNIER MEDILAS UROPULSE (K130344)

_Dornier Medtech America, Inc. · GEX · Mar 21, 2013 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K130344

## Device Facts

- **Applicant:** Dornier Medtech America, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Mar 21, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Uropulse is intended to be used for cutting, vaporization, and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopes), or for open surgery for contact or non-contact surgery with or without a handpiece for use in incision/excision, vaporization, ablation and coagulation of soft tissue. The Uropulse is indicated for use in medicine and surgery, in the following medical specialties: - Arthroscopy - Urology - Lithotripsy - Pulmonology - Gastroenterology - Gynecology - ENT - General Surgery

## Device Story

Medilas H Uropulse is a solid-state Holmium:YAG laser system (2080 nm wavelength). Device delivers laser energy via sterile fiber optic cable (SMA 905 connector) to target tissue; activated by foot pedal. System features microprocessor control, internal cooling circuit with heat exchanger, and touch-screen graphic control panel for parameter adjustment. Used in clinical settings by physicians for soft tissue incision, excision, vaporization, ablation, and coagulation, as well as lithotripsy. Aiming beam (532 nm) assists targeting. Device provides pulse energy (200-3500mJ) and adjustable frequency (3-20 Hz). Benefits include precise tissue interaction with ~400μm penetration depth, facilitating minimally invasive endoscopic or open surgical procedures.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Solid-state Ho:YAG laser; 2080 nm wavelength; 350 μs pulse duration. Class IV laser (IEC/EN 60825-1). Microprocessor-controlled with internal water cooling/heat exchanger. Fiber optic delivery (SMA 905). Touch-screen interface. Pulse energy 200-3500mJ; frequency 3-20 Hz. 532 nm aiming beam.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Dornier Medilas H Solvo ([K122159](/device/K122159.md))
- Dornier Medilas H20 Laser ([K061455](/device/K061455.md))
- Versa Pulse P20 ([K100228](/device/K100228.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 130344

MAR 2 1 2013

# 510(k) SUMMARY Medilas H Uropulse

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

| Dornier MedTech America | Phone: 770-514-6163              |
|-------------------------|----------------------------------|
| 1155 Roberts Blvd.      | Fax: 770-514-6291                |
| Kennesaw, GA 30144      | Date Prepared: February 21, 2013 |

Contact Person: John Hoffer

Phone: 770-514-6163

## Name of Device and Name/Address of Sponsor

Medilas H Uropulse 1155 Roberts Blvd. Kennesaw, GA 30144

## Common or Usual Name

Holmium: Yttrium Aluminum Garnet (HO:YAG) Laser System

## Classification Name

The General and Plastic Surgery Branch has classified Surgical Powered Laser Instruments (Product Code GEX) as a Class II device pursuant to 21 C.F.R. § 878.4810.

## Predicate Devices

Dornier Medilas H Solvo K122159 Dornier Medilas H20 Laser K061455 Versa Pulse P20 (K100228)

# Purpose of the Special 510(k) Notice

The Uropulse is a modification to the Dornier Medilas H Solvo Laser (K122159).

## Device Description

The Dornier Medilas D Uropulse Laser is a solid-state Holmium YAG Laser. The laser emits laser radiation in the invisible wavelength range of 2080 nm in either a continuous-wave or pulsed mode, which is absorbed primarily by water, with an average penetration depth of approximately 400μm (0.4mm). Due to the application-dependent optimization of the pulse time of 350 microseconds, the Uropulse can be used in various medical applications including breaking up stones, cutting tissue, tissue ablation, coagulation and vaporization.

The laser is microprocessor-controlled utilizing a watchdog-monitored microprocessor and operates with an internal cooling system and a heat exchanger. The bottom of the housing serves as a tank for cooling water and, at the same time, a base for the laser unit. A cooling circuit conducts away heat that is generated by the laser pump lamp. Water heated by the laser pump is conducted out of the cavity and into the stainless steel cooler. Following cooling, it is returned to the tanks.

A graphic control panel regulates and displays the operating parameters, application modes, time functions, system status and messages to the User. The control panel consists of a display with integrated touch screen panel applications to control the functions of the laser.

{1}------------------------------------------------

The laser light emission is transmitted to the application site by a sterile fiber optic delivery system (fiber cable) with a SMA 905 connector. Laser energy is transmitted through the fiber to the application site by depressing a foot pedal. The foot pedals, both the wired and wireless, are water- proof and explosion-proof. Various fiber sizes can be utilized depending on the application.

The laser pulse energy at the distal end of the fiber lies within the range of 200-3500mJ. The pulse frequency can be adjusted in 0.2Hz increments between 3 and 20 HZ. The Uropulse contains an aiming beam at a wavelength of 532nm at the maximum power aperture of 1mW. The device employs a Class IV (4) laser pursuant to 21 CFR 1040 and IEC/EN 60825-1.

## Intended Use

The Uropulse is intended to be used for cutting, vaporization, and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopes), or for open surgery for contact or non-contact surgery with or without a handpiece for use in incision/excision, vaporization, ablation and coagulation of soft tissue. The Uropulse is indicated for use in medicine and surgery, in the following medical specialties:

- Arthroscopy
- Urology
- Lithotripsy
- Pulmonology
- Gastroenterology
- Gynecology
- ENT
- General Surgery

## Substantial Equivalence

From a clinical perspective and comparing design specifications, the Uropulse and the predicate devices are substantially equivalent and have the same intended use. Based on the technological characteristics, design and overall performance of the devices. Dornier MedTech America believes that no significant differences exist between the Uropulse and the Domier Medilas H Solvo Laser (K122159).

Dornier MedTech America, Inc. believes the minor differences do not raise any concerns regarding the overall safety or effectiveness. Thus, the Uropulse is substantially equivalent to its predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Dornier Medtech America, Incorporated % Mr. John Hoffer Vice President of Quality, Regulatory and Clinical 1155 Roberts Boulevard Kennesaw, Georgia 30144

March 21, 2013

Re: K130344 Trade/Device Name: Medilas H Uropulse Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: February 21, 2013 Received: February 25, 2013

Dear Mr. Hoffer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. John Hoffer

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

> FOR Sincerely yours,

# Peter Daggmm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use Statement

K130344 510(k) Number (if known):

Device Name: Medilas H Uropulse

Indications for Use:

The Uropulse is intended to be used for cutting, vaporization, and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes, and colonoscopes), or for open surgery for contact or non-contact surgery with or without a handpiece for use in incision/excision, vaporization, ablation and coagulation of soft tissue. The Uropulse is indicated for use in medicine and surgery, in the following medical specialties:

- Arthroscopy .
- Urology .
- Lithotripsv
- Pulmonology
- Gastroenterology
- Gynecology
- ENT
- General Surgery

Prescription Use × (Per 21 C.F.R. 801.109) AND/OR

Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R Ogden 2013.03.21 09:30:10=04'00'

(Division Sign-Off) for MXM Division of Surgical Devices 510(k) Number K130344

---

**Source:** [https://fda-staging.innolitics.com/device/K130344](https://fda-staging.innolitics.com/device/K130344)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).