K130326 · Tepha, Inc. · OOD · May 7, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130326
Device Name
TEPHAFLEX MELTBLOWN CONSTRUCT
Applicant
Tepha, Inc.
Product Code
OOD · General, Plastic Surgery
Decision Date
May 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TephaFLEX Meltblown Matrix is intended to reinforce soft tissue where weakness exists in patients undergoing surgical procedures that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
Device Story
TephaFLEX Meltblown Matrix is a resorbable, porous, non-woven surgical mesh construct. It is manufactured from poly-4-hydroxybutyrate (P4HB) fibers and provided in single sheets of various dimensions. The device is used by surgeons in clinical settings to reinforce soft tissue during surgical procedures requiring bridging or reinforcement material. It functions as a temporary scaffold that supports tissue during healing before undergoing resorption. The device is provided sterile for single use.
Clinical Evidence
Bench testing only. Evidence includes mechanical characterization (burst strength, suture pull-out, tensile strength, tear resistance) and an in vivo subcutaneous implantation study to characterize strength retention. No human clinical trial data was provided.
Indicated for patients undergoing surgical procedures requiring soft tissue reinforcement or bridging where weakness exists.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
K130326 page 1/3
# tepha
840 Memorial Drive Cambridge. MA 02139 tel: (617) 995-5400 fax: (617) 995-5401
## MAY 0 7 2013
Section IX 510(k) Summary (April 18, 2013)
Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Tepha, Inc. is submitting the following summary of information respecting safety and effectiveness:
| Trade Name: | TephaFLEX® Melt blown Matrix |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Tepha, Inc.<br>99 Hayden Avenue, Suite 360<br>Lexington, MA 02421 |
| Contact Person: | Mary P. LeGraw, V.P., Regulatory Affairs<br>Telephone: 781-357-1709<br>Fax: 781-357-1701<br>E-mail: legraw@tepha.com |
| Device Classification Name: | CFR §878.3300<br>Surgical Mesh - Product Code: OOD |
| Classification: | According to Section 13 of the Federal Food, Drug and Cosmetic Act, the<br>device classification is Class II, Performance Standards. |
| Predicate Devices: | TephaFLEX Surgical Mesh - K113723<br>Cook Biodesign Surgisis Tissue Graft - K062696<br>Gore Bio-A Tissue Reinforcement - K033671<br>MAST Biosurgery Surgi-Wrap - K031995, K050332<br>Please see the attached Substantial Equivalence table comparing the |
| Device Description: | TephaFLEX Meltblown Matrix to the predicate devices.<br>The TephaFLEX Meltblown Matrix is a resorbable construct prepared<br>from poly-4-hydroxybutyrate (P4HB) and is provided either non-dyed or<br>dyed with D&C Violet No. 2. It is a porous, fibrous structure composed of<br>thin P4HB fibers that result in a non-woven mesh like fabric. It is<br>provided in single sheets of varying widths, lengths and shapes ranging<br>from 1x1 to 10x14 inches. |
| Indications for Use: | TephaFLEX Meltblown Matrix is intended to reinforce soft tissue where<br>weakness exists in patients undergoing surgical procedures that require<br>the addition of a reinforcing or bridging material to obtain the desired<br>surgical result. |
| Safety and Performance: | The P4HB material used to manufacture the TephaFLEX melt blown<br>matrix is in compliance with the applicable parts of FDA's Class II Special<br>Controls Guidance Document: Absorbable Poly(hydroxybutyrate)<br>Surgical Suture Produced by Recombinant DNA Technology.<br>Mechanical testing, biocompatibility testing, and <i>in vivo</i> animal testing<br>was performed based on recommendations identified in the FDA surgical<br>mesh guidance document: The Guidance for the Preparation of a Pre- |
{1}------------------------------------------------
K130326 page 2/3
Image /page/1/Picture/1 description: The image shows the word "tepha" in a bold, sans-serif font. A curved line underlines the word, starting from the left side of the 't' and extending to the right side of the 'a'. The text is black against a white background.
840 Memorial Drive Cambridge, MA 02139 tel: (617) 995-5400 fax: (617) 995-5401
market Notification Application for a Surgical Mesh. Specifically, comparative burst strength, suture pull-out strength, tensile strength and tear resistance strength was characterized. In vivo strength retention was characterized via a subcutaneous implantation study. The mechanical and in vivo data collected determined the product to be substantially equivalent to the predicate devices.
#### Conclusion:
Based on the indications for use, technological characteristics, and safety and performance testing, the TephaFLEX Meltblown Matrix has been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act.
{2}------------------------------------------------
Substantial Equivalence Comparison Tabl
-
.
| Characteristic | Indications for Use | Material | Thickness | Dyed, Non-dyed | Size | Performance Results<br>Suture Pullout<br>Tensile Strength | Absorption Profile | Packaging | Sterilization |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------|------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------|---------------------------|
| Tepha, Inc.<br>Proposed TephaFLEX®<br>Melt blown Matrix | The TephaFLEX melt blown<br>matrix is intended to<br>reinforce soft tissue in<br>surgical procedures that<br>require the addition of a<br>reinforcing or bridging<br>material to obtain the desired<br>surgical result. | Poly-4-hydroxybutyrate<br>(P4HB) | ~ 1 mm | Non-dyed & Dyed (D&C<br>Violet #2) | Single sheet sizes of: 1x2<br>through 10 X 12 inches | Substantially<br>Equivalent | Absorption essentially<br>complete within 12-18<br>months | Foil packaging with<br>removable Tyvek header | Ethylene Oxide (EO) |
| Tepha, Inc.<br>Predicate TephaFLEX®<br>Mesh<br>K113723 | The TephaFLEX mesh is<br>intended to reinforce soft<br>tissue where weakness exists<br>in patients requiring soft tissue<br>repair or reinforcement in<br>plastic and reconstructive<br>surgery, or for the repair of<br>hernia or other fascial defects<br>that require the addition of a<br>reinforcing or bridging material<br>to obtain the desire surgical<br>result. | Poly-4-hydroxybutyrate<br>(P4HB) | ~0.6 mm | Non-dyed & dyed (D&C Violet<br>#2) | Single sheet sizes of: 1x2<br>through 12x14 inches | Substantially<br>Equivalent | Absorption essentially<br>complete within 12-18 months | Foil packaging with removable<br>Tyvek header | Ethylene Oxide (EO) |
| Cook Biotech<br>Surgisis<br>K062696 | Surgisis is Intended for<br>implantation to reinforce soft<br>tissue. The device is<br>intended for one-time use. | Porcine small intestinal<br>submucosa | Nominal thickness ranging<br>from 0.04 mm to 0.7 mm | Non-dyed | Single sheet size of: 2x3<br>through 7x10 cm | Substantially Equivalent | Unknown | Inner pouch contained in<br>outer bag composed of<br>polymer film lined paper | Ethylene Oxide |
| Gore<br>Bio-A Tissue<br>Reinforcement<br>K033671 | Bio-A Tissue Reinforcement<br>is intended for use in the<br>reinforcement of soft tissue.<br>Examples of applications<br>where the Bio-A may be<br>used include, but are not<br>limited to, hernia repair (in<br>non-load bearing<br>applications), muscle flap<br>reinforcement, and general<br>tissue reconstruction | Poly(glycolide: trimethylene<br>carbonate) copolymer | ~ 2 mm | Unknown | Available in single sheets<br>and preformed, three-<br>dimensional shapes. | Not tested | Biobsorption process should<br>be complete by the end of<br>six months (labeling) | Unknown, | Ethylene Oxide (EO) |
| MAST Biosurgery, Inc.<br>Surgi-Wrap<br>K031955, K050332 | SurgiWrap MAST Bioresorbable<br>Sheet is to be used wherever<br>temporary wound support is<br>required, to reinforce soft tissues<br>where weakness exists in the<br>urological, gynecological or<br>gastroenterological anatomy, or for<br>the repair of hernia or other fascial<br>defects that require the addition of a<br>reinforcing or bridging material to<br>obtain the desired surgical result.<br>This includes, but is not limited to<br>the following procedures:<br>pubourethral support and bladder<br>support, urethral and vaginal<br>prolapse repair, colon and rectal<br>prolapse repair, reconstruction of<br>the pelvic floor and sacral<br>colposuspension. The absorbable<br>protective film minimizes tissue<br>attachment to the device in case of<br>direct contact with the viscera. | Polylactic acid (PLA) | 20 - 1000 microns | Undyed | Single sheet sizes of<br>25mm x 25mm to 500mm x 500mm | Substantially<br>Equivalent | Absorption complete between 12-18<br>months depending on design. | Tyvek film pouch in individual<br>cardboard box | Electron beam irradiation |
K130326 page
3/3
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Tepha, Inc. % Ms. Mary P. LeGraw Vice President, Regulatory Affairs 99 Hayden Avenue, Suite 360 Lexington, Massachusetts 02421
May 7, 2013
Re: K130326
Trade/Device Name: TephaFLEX® Meltblown Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OOD Dated: April 18, 2013 Received: April 23, 2013
Dear Ms. LeGraw:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Ms. Mary P. LeGraw
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely yours, FOR
Peter D. Rumm-S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): Unknown
Device Name: . TephaFLEX® Melt blown Matrix
Indications for Use: :
TephaFLEX Melt blown Matrix is intended to reinforce soft tissue where weakness exists in patients undergoing surgical procedures that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
Prescription Use: X (21 CFR 801 Subpart D)
AND/OR
Over-The-Counter (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
## David Krajūsēs-S
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K130326
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.