K130311 · Guangdong Transtek Medical Electronics Co., Ltd. · MNW · May 30, 2013 · Cardiovascular
Device Facts
Record ID
K130311
Device Name
TRANSTEK BODY FAT ANALYZER
Applicant
Guangdong Transtek Medical Electronics Co., Ltd.
Product Code
MNW · Cardiovascular
Decision Date
May 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The Transtek Body Fat Analyzer measure weight and uses bioelectrical impedance analysis (BIA) technology to estimate body fat, total body water percentage, muscle mass, and bone mass in generally healthy adults 18 years of age or older. It is intended for use in the home/domestic setting only.
Device Story
Device is a body composition analyzer/scale for home use. Inputs: electrical current passed through body via BIA technology. Principle: measures resistance (impedance) of fat tissue vs. muscle/fluid to estimate body composition metrics. Outputs: weight, body fat, total body water, muscle mass, bone mass displayed on LCD. Includes Bluetooth module for wireless data transmission to external receiving devices. Operated by patient self-use. Data transmission is an independent add-on function; device functions as a conventional scale and analyzer regardless of wireless status. Benefits: provides users with health and fitness tracking data.
Clinical Evidence
Bench testing only. Performance tests, risk management, and design verification were conducted to demonstrate that the addition of wireless communication does not raise new safety or effectiveness questions.
Technological Characteristics
BIA technology; electrical current sensing. Bluetooth module for wireless connectivity. LCD display. Form factor: body composition scale. Standards: IEC60601-1 (electrical safety), IEC60601-1-2 (EMC), ISO14971 (risk management), FCC Part 15 (EMI).
Indications for Use
Indicated for generally healthy adults 18+ to estimate body fat, total body water, muscle mass, and bone mass, and to measure weight. Contraindicated for individuals with pacemakers or other internal medical devices.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Transtek Glass Body Fat Analyzer, Model GBF-950-D (K112932)
Submission Summary (Full Text)
{0}------------------------------------------------
K130311 Page 1 of 3
## TRANSTEK
## Section 6 - 510(k) Summary
Date of Summary Preparation: 1/11/2013
## 1. Submitter's Identifications
Submitter's Name: Guangdong Transtek Medical Electronics Co., Ltd Address: Zone A, 5/F., Investment Building, No. 12 Huizhan East Rd.,
Torch Development District, Zhongshan, Guangdong 528437, China Contact Person: Lisa Li Contact Email Address: lishal@transtek.cn Telephone: 086-760-88282982 ext. 876 Fax: 086-760-85339231
#### 2. Correspondent's Identifications
Correspondent's Name: A03 Lab of BTS Address: No. I Fanghua Street, Hi-tech Zone, Chengdu City, Sichuan, China Contact Person: Leo Wang Contact Email Address: leo.w@hibts.com Telephone: 086-28-86083300 Fax: 086-20-80727399
#### 3. Name of the Device
Device Classification Name: Analyzer, Body Composition (Impedance Plethysmograph) Product Name: Body Fat Analyzer Trade/Proprietary Name: Transtek Body Fat Analyzer Models: GBF-1251-B, BF-1255-B, BF-1256-B, GBF-1257-B Classification Panel: Cardiovascular Common/Usual Name: Body Composition Analyzer/Scales Product Code: MN W Device Classification: Class II
Contraindications: Do not use the Analyzer if you have a pacemaker or other internal medical device.
#### 4. The Predicate Devices
".
Transtek Glass Body Fat Analyzer, Model GBF-950-D, K112932
#### 5. Device Description
Transtek Body Fat Analyzer uses BIA (Bioelectrical Impedance Analysis) technology which passes an electrical current through the body to estimate body far mass, total body water, muscle mass and bone mass. The electrical current is low and may not be felt. The current passes freely through the fluids contained in muscle tissue, but encounters difficulty/resistance when it passes through fat tissue. This
Section 6 - 510(k) Summary
Page 1 of 3
# MAY 3 0 2013
{1}------------------------------------------------
# TRANSTEK
resistance of the fat tissue to the current is termed 'Bioelectrical Impedance', and is accurately measured by Transtek Body Fat Analyzer (GBF-1251-B, BF-1256-B, and GBF-1257-B).
This method simultaneously calculates personal body water, muscle mass, and bone mass, giving a more accurate reading of overall health and fitness. As well as being an analyzer, this device can be used as a conventional weight scale.
The Transtek Body Fat Analyzer embeds a Bluetooth (BT) module that allows it to connect to nearby BT receiving end. The LCD of device displays results. And once measurement is over, the device will start transmission data by BT. Thus users can receive, and display/storage, measurement data of Transtek Body Fat Analyzer unit through their end devices that embedded BT module.
#### 6. Intended Use of Device
The Transtek Body Fat Analyzer measure weight and uses bioelectrical impedance analysis (BIA) technology to estimate body fat, total body water percentage, muscle mass, and bone mass in generally healthy adults 18 years of age or older. It is intended for use in the home/domestic setting only.
#### 7. Design Control Activities and Performance Tests Summary
Design control activities for this modification have been implemented and performance tests of modified devices have been done. These performance tests, risk management, and design verification tests provide demonstration that the differences do not raise any new questions of safety and effectiveness.
Proposed Body Fat Analyzer conforms to (including, but not limited to) the following standards:
IEC60601-1-2, Electromagnetic compatibility
IEC60601-1, Electrical safety
ISO14971, Risk management to medical devices
FCC Part 15, EMI tests of FCC Radiation & RF rules and regulations
#### 8. Summary of Substantial Equivalence
8. I Differences between proposed device and the predicate device
The only significant function difference between the two devices is that the modified device add-on a wireless data communication, what user option, which can transmit measurement results to receiving end equipment.
More modification details are described in this submission.
8.2 Discussion
The Transtek Body Fat Analyzer has identical indication for use, fundamental scientific technology, energy type, and similar performance specifications, dimensional specifications, software/firmware, functions, labeling to the predicate device.
Section 6 - 510(k) Summary
{2}------------------------------------------------
# TRANSTEK
The only function difference between proposed device and the predicate device is that the modified device provides user an optional wireless data transmission. It is an add-on function that is entirely independent to the body compositions analyzer function. Regardless of wireless connection status, the analyzer normal works to carry out a bioelectrical impedance analysis and display its results.
Thus the wireless data transmission function does not affect the safety and effectiveness of the body analyzer function.
#### 9. Conclusions
The Transtek Body Fat Analyzer (GBF-1251-B, BF-1255-B, BF-1256-B, and GBF-1257-B) is substantially equivalent to the predicate device, GBF-950-D, by having the identical indications for use, identical fundamental scientific technologies, and an add-on function which does not impact the safety and effectiveness of the device.
--- End of this section ---
{
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the department's emblem, which is a stylized representation of an eagle or bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 30, 2013
Guandong Transtek Medical Electronics Co., Ltd. % Mr. Leo Wang Senior Consultant A03 Lab of BTS No. 1 Fanghua Street, Hi-tech District, Chengdu Sichuan CHINA 610041
Re: K130311
Trade/Device Name: Transtek Body Fat Analyzer Models: GBF-1251-B, BF-1255-B, BF-1256-B, GBF-1257-B Regulation Number: 21 CFR§ 870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: March 7, 2013 Received: March 7, 2013
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have-determined the device is substantially equivalent-(for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Leo Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## TRANSTEK
## Section 5 - Indications for Use
510(k) Number (if known): K130311
Device Name:
Transtek Body Fat Analyzer Models: GBF-1251-B, BF-1255-B, BF-1256-B, GBF-1257-B
Indications for Use:
The Transtek Body Fat Analyzer measure weight and uses bioelectrical impedance analysis (BIA) technology to estimate body fat, total body water percentage, bone mass, and muscle mass in generally healthy adults 18 years of age or older.
It is intended for use in the home/domestic setting only.
Prescription Use _
AND/OR
X Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Herbert - 2 - S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K130311 510(k) Number
Section 5 - Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.