LINA BIPOLAR LOOP; STANDARD SIZE, LARGE SIZE, EXTRA LARGE SIZE
K130305 · Lina Medical Aps · HIN · Jul 24, 2013 · Obstetrics/Gynecology
Device Facts
Record ID
K130305
Device Name
LINA BIPOLAR LOOP; STANDARD SIZE, LARGE SIZE, EXTRA LARGE SIZE
Applicant
Lina Medical Aps
Product Code
HIN · Obstetrics/Gynecology
Decision Date
Jul 24, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LiNA Bipolar Loop is a 5mm bipolar electrosurgical device intended for amputating the mobilized uterus during laparoscopic supracervical (subtotal) hysterectomy and resection of devasculated subserosal pedunculated myomas. To be used with an electrosurgical generator that provides a bipolar outlet.
Device Story
LiNA Bipolar Loop is 5mm single-use laparoscopic electrosurgical instrument; utilizes bipolar energy for tissue cutting/coagulation. Device consists of gold-plated stainless steel wire loop with zirconia ceramic tip; compatible with standard bipolar electrosurgical generators. Operated by surgeons in clinical/OR settings during gynecological procedures. Surgeon controls energy delivery via generator; loop amputates mobilized uterus or resects pedunculated myomas. Provides precise tissue cutting; benefits include reduced surgical time and effective hemostasis during laparoscopic hysterectomy/myomectomy. Device is sterile, single-use, and disposable.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including efficacy and functionality studies in simulated tissue (pork/beef muscle). Cutting quality and time were evaluated against predicate devices, demonstrating equivalent performance within acceptable limits.
Indicated for patients undergoing laparoscopic supracervical (subtotal) hysterectomy for uterine amputation or resection of devasculated subserosal pedunculated myomas.
Regulatory Classification
Identification
A bipolar endoscopic coagulator-cutter is a device used to perform female sterilization and other operative procedures under endoscopic observation. It destroys tissue with high temperatures by directing a high frequency electrical current through tissue between two electrical contacts of a probe. This generic type of device may include the following accessories: an electrical generator, probes, and electrical cables.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance 2/12/90 (K-90),” and
(iii) “Guidance (‘Guidelines’) for Evaluation of Laproscopic Bipolar and Thermal Coagulators (and Accessories),”
(2) International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety,”
(3) American National Standards Institute/American Association for Medical Instrumentation's HF-18, 1993, “Electrosurgical Devices,”
(4) Labeling:
(i) Indication: For female tubal sterilization, and
(ii) Instructions for use:
(A) Destroy at least 2 centimeters of the fallopian tubes,
(B) Use a cut or undampened sinusoidal waveform,
(C) Use a minimum power of 25 watts, and
(D) For devices with ammeters: continue electrode activation for 5 seconds after the visual endpoint (tissue blanching) is reached or current flow ceases indicating adequate tissue destruction.
{0}------------------------------------------------
LiNA Medical ApS Traditional 510(k) Premarket Submission LiNA Bipolar Loop
# EMERGO GROUP
## Section 5 - 510(k) Summary
#### 1. Submission Sponsor
# JUL 2 4 2013
- LiNA Medical ApS Formervangen 5 2600 Glostrup Denmark Phone: +45 4329 6666 Fax: +45 4329 6699 Contact: Louisa Memborg, Regulatory Affairs Officer
#### 2. Submission Correspondent
Emergo Group 611 West 5th Street, Third Floor Austin, TX 78701 Cell Phone: (406) 579-8124 Office Phone: (512) 327.9997 Fax: (512) 327.9998 . Contact: Richard Gillis, Ph.D., Senior Consultant Email: project.management@emergogroup.com
### 3. Date Prepared
27 JAN 2013
#### Device Identification 4.
| Trade/Proprietary Name: | LiNA Bipolar Loop |
|----------------------------|------------------------------------------------------|
| Common/Usual Name: | Bipolar electrosurgical Loop |
| Classification Name: | Bipolar Endoscopic coagulator-cutter and accessories |
| Classification Regulation: | 21 CFR 884.4150 |
| Product Code: | HIN |
| Device Class: | Class II |
| Classification Panel: | Obstetrics and Gynecology Panel |
#### 5. Predicate Devices
- 1. LiNA Loop (K070315)
- 2. PKS BiLL (K111059)
#### 6. Device Description
The LiNA Bipolar Loop is a 5 mm single use laparoscopic instrument. It is available with three different loop dimensions; 160mm x 80mm, 200mm x 100mm x 120mm x 120mm. The outer 15mm on each side of the loop is not insulated i.e. the bipolar cutting area length
{1}------------------------------------------------
totals 30mm. The device is single use ethylene oxide sterilized and is compatible with most standard electrosurgical generators that provide a bipolar outlet.
### 7. Intended Use
The LiNA Bipolar Loop is a 5mm bipolar electrosurgical device intended for amputating the mobilized uterus during laparoscopic supracervical hysterectomy and resection of devasculated subserosal pedunculated myomas. To be used with an electrosurgical generator that provides a bipolar outlet.
### 8. Comparison of Technological Characteristics
The following table compares the LiNA Bipolar Loop to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | LiNA Medical ApS<br>Formervangen 5<br>2600 Glostrup<br>Denmark | LiNA Medical ApS<br>Formervangen 5<br>2600 Glostrup<br>Denmark | Gyrus ACMI, Inc<br>136 Turnpike Rd.<br>Southborough, MA<br>01772-2104 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | LINA Bipolar Loop | LINA Loop | PKS BILL |
| 510(k) Number | TBD | K070315 | K111059 |
| Product Code | HIN | KNF | HIN |
| Regulation<br>Number | 21 CFR 884.4150 | 21 CFR 884.4160 | 21 CFR 884.4150 |
| Regulation Name | Bipolar endoscopic<br>coagulator-cutter and<br>accessories | Coagulator-cutter,<br>endoscopic, unipolar<br>(and accessories) | Bipolar endoscopic<br>coagulator-cutter and<br>accessories |
| Indications for<br>Use | The LiNA Bipolar Loop is<br>a 5mm bipolar<br>electrosurgical device<br>intended for amputating<br>the mobilized uterus<br>during laparoscopic<br>supracervical (subtotal)<br>hysterectomy and<br>resection of<br>devasculated subserosal<br>pedunculated myomas.<br>To be used with an<br>electrosurgical generator<br>that provides a bipolar<br>outlet. | The LiNA Loop is a<br>5mm monopolar<br>electrosurgical device<br>intended for<br>sectoring the<br>mobilized uterus<br>during Laparoscopic<br>supracervical<br>hysterectomy. To be<br>used with an<br>electrosurgical<br>generator | The PKS BILL is a 5mm<br>bipolar electrosurgical<br>device intended for<br>sectoring the mobilized<br>uterus during<br>Laparoscopic<br>supracervical<br>hysterectomy and the<br>resection of<br>devasculated subserosal<br>pedunculated myomas.<br>Working in conjunction<br>with Gyrus ACMI<br>generator |
| Material | 1. Gold plated stainless<br>steel wire with<br>molded zirconia<br>ceramic tin and | 1. Stainless steel wire | According to PKS BILL<br>510(k), K111059<br>summary "Uses<br>materials that are well |
#### Table 5A - Comparison of Characteristics
{2}------------------------------------------------
# EMERGO GROUP
| Manufacturer | LiNA Medical ApS | LiNA Medical ApS | Gyrus ACMI, Inc |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| | Formervangen 5<br>2600 Glostrup<br>Denmark | Formervangen 5<br>2600 Glostrup<br>Denmark | 136 Turnpike Rd.<br>Southborough, MA<br>01772-2104 |
| Trade Name | LiNA Bipolar Loop | LiNA Loop | PKS BILL<br>established". |
| | polyamide cap.<br>2. Shrink tube: FEP<br>(Teflon)<br>3. Funnel: Black<br>polyamide<br>4. Cannula/Fibre tube:<br>Vinyl ether resin with<br>fiberglass<br>reinforcement<br>5. Inner cannula Tube:<br>Kynar®<br>6. Handle: ABS<br>7. Cable: Thermoplastic<br>polyurethane with<br>brass connector. | 2. Shrink tube:<br>Kynar®(PVDF)<br>3. Funnel: Black<br>polyamide<br>4. Fibre tube: Vinyl<br>ether resin with<br>fiberglass<br>reinforcement<br>5. Inner cannula<br>Tube: Kynar®<br>6. Handle: ABS<br>7. Cable: PVC with<br>brass connector. | |
| Sterile | Sterile packed.<br>Ethylene Oxide Gas<br>SAL 10-6 | Sterile packed.<br>Ethylene Oxide Gas<br>SAL 10-6 | Sterile packed.<br>Gamma Irradiation.<br>SAL 10-6 |
| Single-Use | Intended for single use<br>only. | Intended for single<br>use only. | Intended for single use<br>only. |
| Shelf Life | 3 years | 3 years | Unknown |
| Battery Operated | N/A | N/A | N/A |
| AC Powered | N/A | N/A | N/A |
| Complies with ISO<br>10993-1 | Yes | Yes | Yes |
| Electrical Safety<br>Testing Passed | IEC 60601 | IEC 60601 | IEC 60601 |
#### Non-Clinical Performance Data 9.
The following testing has been performed to support substantial equivalence:
- Efficacy and Functionality Test The LiNA Loop and the LiNA Bipolar Loop had similar . cutting quality in simulated tissue study using pork and beef muscle. Cutting quality was assessed on a scale of 1-9, with 5 being the acceptance criteria. 200mm sized loops were selected as this represented the mid-size loop for both the Bipolar Loop and Loop. The loops were assessed for cutting quality at high and low power settings. Both devices were qualitatively scored as a 5. Furthermore, cutting time was assessed, with an average deviation time between cuts of 0.85 seconds. Cutting time was assessed at the respective high and low power settings using three different generators and found to be within acceptable limits (±4 seconds). Therefore, the cutting time and quality were equivalent between the LiNA Loop and the LiNA Bipolar Loop.
{3}------------------------------------------------
As part of demonstrating the safety and effectiveness of the LINA Bipolar Loop and in showing substantial equivalence the predicate devices that are subject to this 510(k) submission, LiNA Medical ApS completed a number of tests. The LiNA Bipolar Loop meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety confirms that the output meets the design inputs and specifications. The LiNA Bipolar Loop passed all testing stated above as shown by the acceptable results obtained.
The LiNA Bipolar Loop complies with the applicable voluntary standards for biocompatibility and sterilization. The device passed all the testing in accordance with national and international standards.
#### 10. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 11. Statement of Substantial Equivalence
lt has been shown in this 510(k) submission that the difference between the LiNA Bipolar Loop and the predicate devices do not raise any questions regarding its safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the LiNA Bipolar Loop is substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, sterilization, biocompatibility, performance characteristics, and intended use. The LiNA Bipolar Loop, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design featuring three abstract shapes resembling an eagle or bird in flight.
Public Health Service
Food and Drug Administration 10903 New Hamoshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 24, 2013
LiNA Medical ApS % Richard Gillis, Ph.D. Senior Regulatory Consultant Emergo Group, Inc. 611 West 5th Street, Third Floor Austin, TX 78701
Re: K130305
Trade/Device Name: LiNA Bipolar Loop Regulation Number: 21 CFR 884.4150 Regulation Name: Bipolar endoscopic coagulator-cutter and accessories Regulatory Class: Class II Product Code: HIN Dated: June 07, 2013 Received: June 10, 2013
Dear Richard Gillis, Ph.D.,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
## Page 2 - Richard Gillis, Ph.D.
comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its to!!-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
· Enclosure
{6}------------------------------------------------
## Section 4 - Indications for Use Statement
510(k) Number (if known): Not Assigned K130305
Device Name: LiNA Bipolar Loop
Indications for Use:
The LiNA Bipolar Loop is a 5mm bipolar electrosurgical device intended for amputating the mobilized uterus during laparoscopic supracervical (subtotal) hysterectomy and resection of devasculated subserosal pedunculated myomas. To be used with an electrosurgicol generator that provides a bipolar outlet.
Prescription Use_x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Herbert P. Lerner -S
K130305
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.