K130301 · Elimedical Devices (Fujian), Inc. · KGO · Aug 22, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130301
Device Name
LATEX SURGEON'S GLOVES (POWDERED AND POWDER FREE)
Applicant
Elimedical Devices (Fujian), Inc.
Product Code
KGO · General, Plastic Surgery
Decision Date
Aug 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Device Story
Sterilized, disposable natural rubber latex surgical gloves; worn by operating room personnel to protect surgical wounds from contamination. Available in powdered and powder-free versions; creamy white color; various sizes. Provided radiation-sterilized with 3-year shelf life. Device acts as physical barrier between clinician and patient wound site.
Clinical Evidence
Bench testing only. No clinical data. Compliance verified via ASTM D3577 (physical properties), ASTM D5151 (hole detection), ASTM D6124 (residual powder), ASTM D5712/D6499 (protein content), ASTM F1929/F88 (packaging/seal integrity), and ISO 10993-10 (biocompatibility/sensitization). Sterilization validated per ISO 11137-2.
Technological Characteristics
Material: Natural rubber latex (Type 1 per ASTM D3577). Form factor: Disposable surgical gloves, various sizes. Sterilization: Radiation (SAL 10^-6). Standards: ASTM D3577, D5151, D6124, D5712, D6499, F1929, F88, ISO 10993-10, ISO 11137-2.
Indications for Use
Indicated for operating room personnel to protect surgical wounds from contamination. No specific age or gender restrictions.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
## Section III 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
K130301 The assigned 510(k) Number:
- 1. Date of Submission: January 11, 2013
- 2. Sponsor
**AUG 2 2 2013**
Elimedical Devices(Fujian) Inc. E06, Wuping Industrial Park, Wuping, Fujian 364300, China
Contact Person: Mr. He Yangui Position: Manager Tel: +86-10-67617923 Fax: +86-10-67617393 Email: KEVIN@ELIMEDICAL.COM
- 3. Submission Correspondent Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
- Proposed Device Identification 4.
Trade Name: 1). REDLEAF, Elimedical, and 2) Multiple or Customers' Trade Name Device Name: Latex Surgeon's Gloves (Powdered and Powder Free) Common Name: Surgical Gloves
Classification: I Product Code: Surgeon's Glove - 79KGO Classification Name: surgeon's gloves Regulation Number: 21 CFR 878.4460 Review Panel: General & Plastic Surgery
III - 1
{1}------------------------------------------------
Intended Use Statement:
A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
- 5. Predicate Device Identification
510(k) Number: k063757
Product Name: Motex Powder-Free Surgical Gloves & Powdered Latex Surgical Gloves Manufacturer: SHANGHAI MOTEX HEALTHCARE CO., L TD.
- 6. Device Description
The proposed device, Latex Surgeon's Gloves (Powdered and Powder Free) is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.
The proposed device is made of natural rubber latex. per standard ASTM D3577-0961, the rubber surgical gloves classification is:
"Type 1 - gloves compounded primarily from natural rubber latex".
The proposed device include Powder Free Latex Surgeon's Gloves, and variations of different size. All variations share the same colour, creamy white.
The proposed device is provided radiation sterilized to achieve the Sterility Assurance Level (SAL) of 10% and place in a sterility maintenance package to ensure a shelf life of 3 years.
- 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ASTM D3577-0981, Standard Specification for Rubber Surgical Gloves.
ASTM D 5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves.
ASTM D6124-06 (Reaffirmation 2011), Standard Test Method for Residual Powder on Medical Gloves.
ASTM D5712-10. Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method.
ASTM D6499-07, Standard Test Method for the Immunological Measurement of Antigenic Protein in Natural Rubber and Its Products.
ASTM F1929-98 (2004). Standard Test Method for Detecting Seal Leaks in Porous Medical III-2
{2}------------------------------------------------
Packaging by Dye Penetration.
ASTM F88/F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
ISO11137-2: 2006. Sterilization of healthcare products-Radiation-Part2: Establishing the sterilization dose.
- 8. Substantially Equivalent Comparison Conclusion
| ITEM | Proposed Device | Predicate Device (k063757) |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Product Code | KGO | Same |
| Regulation No. | 21 CFR 878.4460 | Same |
| Class | 1 | Same |
| Intended Use | A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. | Similar |
| Powdered or Powered free | Powdered, and Powered free | Same |
| Classification per ASTM D3577 | Type 1 - gloves compounded primarily from natural rubber latex | Same |
| Dimensions | Size 6, 6 $^{1}\!/_{2}$ , 7, 7 $^{1}\!/_{2}$ , 8, 8 $^{1}\!/_{2}$ , 9<br>Dimensions comply with ASTM D3577 | Same |
| Physical Properties | Comply with ASTM D3577 | Same |
| Freedom from Holes | Comply with ASTM D3577 and ASTM D5151 | Same |
| Powder Content | Comply with ASTM D3577 and ASTM D6124 | Same |
| Protein Content | Comply with ASTM D3577, ASTM D5712 and ASTM D6499 | Same |
| Biocompatibility | Comply with ISO 10993-10 | Same |
| Sterilization | Radiation<br>SAL: 10 $^{-6}$ | Same |
| Label and Labeling | Meet FDA's Requirements | Same |
Table III-1 Substantially Equivalent Comparison
Difference in intended use between the proposed and predicate device is discussed in the 510(k) submission documents, it is concluded that the difference will not affectiveness and safety of the proposed device.
The proposed device, Latex Surgeon's Gloves (Powdered and Powder Free), is determined to be Substantially Equivalent (SE) to the predicate device, Motex Powder-Free Surgical Gloves & Powdered Latex Surgical Gloves (k063757), in respect of safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2013
EliMedical Devices (Fujian) Incorporated C/O Ms. Diana Hong General Manager Mid-Link Consulting Company, Limited P.O. Box 237-023, Shanghai, China 200237
Re: K130301
Trade/Device Name: Latex Surgeon's Gloves (Powdered and Powder Free) Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: KGO Product Code: I Dated: July 5, 2013 Received: August 2, 2013
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/4/Picture/6 description: The image shows the name "Mary S. Runner - S" in a bold, sans-serif font. The letters are large and spaced closely together. The word "Runner" is partially obscured by a graphic element that appears to be a stylized logo or design.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Section II Indications for Use
K130301 510(k) Number:
Device Name: Latex Surgeon's Gloves (Powdered and Powder Free)
Indications for Use:
A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
OPRESCRIPTION USE (Part 21 CFR 801 Subpart D) OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sreekanth-Gutala-S
2013.08.22 11:47:44 -04'00'
Page 1 of 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: _13030
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.