K130277 · Hologic, Inc. · KGI · May 31, 2013 · Radiology
Device Facts
Record ID
K130277
Device Name
SINGLE ENERGY (SE) FEMUR EXAMS
Applicant
Hologic, Inc.
Product Code
KGI · Radiology
Decision Date
May 31, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1170
Device Class
Class 2
Indications for Use
Single Energy (SE) Femur Exams are used to visualize focal reaction or thickening along the lateral cortex of the femoral shaft which may be accompanied by a transverse radiolucent line. Clinical correlation is advised as these features may be consistent with atypical femur fractures, a complication associated with long term use of antiresorptive therapy.
Device Story
Hologic Single Energy (SE) Femur Exam is a software option for Discovery bone densitometers. Device utilizes existing IVA-HD (High Definition) scan mode to acquire X-ray images of femoral shaft. System consists of examination table and C-arm with X-ray tube and linear multi-detector array. Patient positioned between source and detector; system performs axial fan beam scan. Output is high-resolution image of femoral shaft displayed on densitometer screen for clinician review. Used in clinical settings to visualize bone deformities, specifically focal reaction or thickening along lateral cortex of femoral shaft, potentially indicating atypical femur fractures. Clinician uses visual assessment of images to correlate with patient history of antiresorptive therapy. No changes to hardware or image acquisition software compared to predicate; only patient positioning differs.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence demonstrated through comparison of technological characteristics, image acquisition parameters, and performance specifications identical to the predicate device.
Technological Characteristics
DXA bone densitometer; single energy fan beam scan; 140 kVp @ 5 mA; 1.4 lp/mm vertical resolution; 0.25 mm x 61 mm aperture; 0.025 mGy entry dose; 15-second acquisition speed. Software-based image visualization.
Indications for Use
Indicated for patients on long-term antiresorptive therapy to visualize focal reaction, thickening, or transverse radiolucent lines along the lateral cortex of the femoral shaft, which may be consistent with atypical femur fractures.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
IVA (MXA-II) software option for the X-Ray bone densitometer (K060111)
{0}------------------------------------------------
K130277
Page 1 of 4
### Section 5 -510(k) Summary
### Date Prepared: January 31, 2013
# MAY 3 1 2013
#### 1.0 Manufacturing Establishment and Contact Information
#### 1.1 Manufacturer Name and Address:
Hologic, Inc. 35 Crosby Drive Bedford, MA 01730
#### 1.2 Establishment Registration Number:
1221300
#### Name, Title, and Telephone Number of Contact: 1.3
Name: Eileen M. Boyle Regulatory Affairs Specialist III Title: Phone: (781) 999-7781 (866) 652-8674 Fax : eileen.boyle@hologic.com Email :
#### 2.0 Device Identification
- 2.1 Device Trade Name: Hologic® Single Energy (SE) Femur Exam
#### 2.2 Common/Usual Name
Software for Bone Densitometers
#### 2.3 Proposed Intended Use:
Single Energy (SE) Femur Exams are used to visualize focal reaction or thickening along the lateral cortex of the femoral shaft which may be accompanied by a transverse radiolucent line. Clinical correlation is advised as these features may be consistent with atypical femur fractures, a complication associated with long term use of antiresorptive therapy.
Page 1 of 4
ﺎ
{1}------------------------------------------------
#### 2.4 Comparison of the Indication for Use:
The proposed indication for use of the SE Femur scan is substantially equivalent to that of the predicate device. Like the proposed device, the predicate device is indicated for the visualization of bone deformities. The predicate device is indicated for the visualization of vertebral bone deformities (as well as for the visualization of abdominal aortic calcifications), whereas the proposed device is for the visualization of bone deformities of the femur.
Since the identical IVA-HD scan is used, the SE Femur scan and IVA scans offer the same fundamental device technology, image acquisition mode, acquisition speed, x-ray technique factors, vertical resolution, x-ray aperture, radiation dose, and image display supported by the bone densitometer. No changes were made to the image acquisition software, report structure, or performance specifications. No new safety and effectiveness questions were raised following the risk assessment for the proposed expansion of the IVA scan's intended use. Therefore, the expanded indication for use statement has no effect on the bone densitometer's safety or effectiveness.
#### 3.0 Device Classification
- 3.1 Classification: Class II
- 3.2 Classification Name and Rule Bone Densitometer: 21 CFR 892-1170
- 3.3 Classification Panel Radiology
- 3.4 Product Code તે જેવી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત-ઉપયોગ વિભાગમાં આવેલું એક ગામના લોકોનો મુખ્યત્વે આવ
- 3.5 Prescription Use The use of the Single Energy (SE) Femur Exam is restricted to prescription use only.
#### 3.6 Predicate Devices
- . 510(k) No: K060111 Trade Name: IVA (MXA-II) software option for the X-Ray bone densitometer SE Date: 4/4/2006 Manufacturer: Hologic, Inc.
{2}------------------------------------------------
#### 4.0 Device Description
The proposed Hologic ® Single Energy (SE) Femur scan allows for the visualization of bone deformities of the femur, specifically focal reaction or thickening along the lateral cortex of the femoral shaft which may be accompanied by a transverse radiolucent line.
Hologic's Discovery (K023398) bone densitometer models: Discovery A, C, W, and SL consist of an examination table and a C-arm. At opposite ends of the C-arm an x-ray tube and a linear multi-detector array are mounted. The patient is positioned on the examination table between the x-ray tube and the detector array. The detector and x-ray source scan the patient axially using an x-ray fan beam perpendicular to the direction of movement. The xray source can be pulsed at a constant voltage to acquire exams as in the case of Hologic's Instant Vertebral Assessment (IVA) exams reviewed and cleared by FDA (K060111). Currently, two types of IVA scans are in use on the Discovery A, C, W, and SL bone densitometers; the IVA and the IVA-HD (High Definition). Both scans are used for the visualization of vertebral bone deformities. The IVA scan has a 10 second scan time, whereas the IVA-HD has a 15 second scan time. IVA-HD scans use a thinner aperture resulting in higher resolution and a lower total dose than IVA scans. The high resolution image of the single energy IVA-HD scan makes it ideal for use as a scan to visualize deformities in the femur.
#### Comparison of Technological Characteristics: 5.0
The Single Energy (SE) Femur Exam is substantially equivalent to that of the predicate device. Like the proposed device, the predicate device is indicated for the visualization of bone deformities. The comparison table summarizes the technological characteristics of the Single Energy Femur Exam and the IVA the predicate device:
| Device Characteristics | Proposed 510(k) Device<br>Single Energy Femur Exam | Hologic, Inc.<br>IVA (K060111) |
|-------------------------------------|----------------------------------------------------|---------------------------------------|
| Method of Deformity<br>Assessment | Visual ) | Same |
| Fundamental Device<br>Technology | DXA Bone Densitometer | Same |
| Exam Site | Femoral shaft | Lumbar and Thoracic Spine |
| Imaging Scan | IVA-HD | IVA-HD |
| Image Acquisition Mode | Single Energy Fan Beam Scan | Same |
| Acquisition Speed for 15.7"<br>Scan | 15 seconds | Same |
| X-ray Technique Factors | 140 kVp @ 5 ma | Same |
| Vertical Resolution | 1.4 lp/mm | Same |
| X-ray Aperture | 0.25 mm (W) x 61 mm (L) | Same |
| Entry Dose | 0.025 mGy | Same |
| Image Display | On screen display with image<br>tools | Same |
| Device Output | Image of femoral shaft | Image of Lumbar and Thoracic<br>Spine |
Comparison table of the Single Energy Femur Exam and IVA:
Page 3 of 4
{3}------------------------------------------------
K130277
Page 4 of 4
#### 6.0 Performance Testing
The SE Femur and IVA-HD scans offer the same fundamental device technology, image acquisition mode, acquisition speed, x-ray technique factors, vertical resolution, x-ray aperture, radiation dose, and image display supported by the bone densitometer. The Single Energy (SE) Femur Exam has the same technological characteristics as the predicate device. The identical IVA-HD scan is used for the acquisition of the lumbar and thoracic spine scan as is used for the proposed device for the acquisition of the SE Femur scan. The sole difference is in the positioning of the patient so that the targeted region (spine or femur) is imaged. No changes have been made to the software and both SE Femur and IVA scans can be viewed on the DXA system's screen and a report can be printed.
The information provided in the 510(k) submission is acceptable for the demonstration of substantial equivalence to the predicate device with respect to physical performance and testing
#### 7.0 Substantial Equivalence
The Single Energy (SE) Femur Exam software is substantially equivalent to commercially IVA software option (K060111) which allows visualization of bone deformities. The predicate device provides substantially equivalent or the same intended use, features and functions as the proposed device.
#### 8.0 Conclusion
The Hologic Single Energy (SE) Femur Exam software allows visualization of bone deformities. The features and functions are substantially equivalent to those of the predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## May 31, 2013
Hologic, Inc. % Ms. Eileen M. Boyle Senior Regulatory Affairs Specialist 35 Crosby Drive BEDFORD MA 01730
Re: K130277
Trade/Device Name: Single Energy (SE) Femur Exam Regulation Number: 21 CFR 892.1170 Regulation Name: Bone densitometer Regulatory Class: II Product Code: KGI Dated: May 15, 2013 Received: May 16, 2013
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general contrôls provisions of the Act include requirements for annual registration, listing of devices, good-manufacturing practice, labeling, and-prohibitions against-misbranding andadulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Boyle
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/Overview/DeviceLabeling/ default.hum. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
· for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Section 4 - Indications for Use
510(k) Number (if known): K130277
## Device Name: Single Energy (SE) Femur Exam
Indications for Use:
Single Energy (SE) Femur Exams are used to visualize focal reaction or thickening along the lateral cortex of the femoral shaft which may be accompanied by a transverse radiolucent line. Clinical correlation is advised as these features may be consistent with atypical femur fractures, a complication associated with long term use of antiresorptive therapy.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
سر
Sm.h.7)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
Office of InVitro Diagnostic Device Evaluation and Safety
K130277 510(k) _______________________________________________________________________________________________________________________________________________________________________
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.