DIMENSION EXL WITH LM SYSTEM
Device Facts
| Record ID | K130276 |
|---|---|
| Device Name | DIMENSION EXL WITH LM SYSTEM |
| Applicant | Siemens Healthcare Diagnostics, Inc. |
| Product Code | JJE · Clinical Chemistry |
| Decision Date | Mar 22, 2013 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 862.2160 |
| Device Class | Class 1 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K130276 · Mar 22, 2013 | DIMENSION EXL WITH LM SYSTEM | Siemens Healthcare Diagnostics, Inc. | Remnant de-identified human serum samples | Remnant patient samples were used to conduct a split-sample method comparison to demonstrate substantial equivalence between the device with a new photomultiplier tube and the predicate device. | Method Comparison; Remnant Samples; Clinical Samples |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Split-sample method comparison | Human serum samples; Sample Size: 45; Number of Sites: 1 (Internal Siemens R&D) | Dimension® EXL™ with LM system with current photomultiplier tube (predicate) | Correlation coefficient, slope, and intercept of FT4L assay results |
Indications for Use
The Dimension® EXL™ with LM system is an in vitro diagnostic device that is intended to measure a variety of analytes in human body fluids. The system utilizes photometric, turbidimetric, chemiluminescence and integrated ion selective multisensory technology for chemical and immunochemical applications for clinical use. The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension EXL™ with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.
Device Story
Dimension® EXL™ with LM is a floor-model, fully automated, microprocessor-controlled clinical chemistry analyzer. It processes human serum and plasma samples using photometric, turbidimetric, chemiluminescent, and ion-selective multisensor technologies. The system utilizes prepackaged Flex® reagent cartridges. This submission specifically addresses a change in the photomultiplier tube (PMT) used for LOCI® chemiluminescent immunoassays. The LOCI® method involves a sequential reaction: sample incubation with biotinylated anti-T4 antibody; addition of T3-coated Chemibeads; and addition of streptavidin-coated Sensibeads. Illumination at 680 nm triggers singlet oxygen diffusion from Sensibeads to Chemibeads, producing a chemiluminescent signal at 612 nm inversely proportional to FT4 concentration. Operated by laboratory technicians in clinical settings, the system provides quantitative results to assist physicians in diagnosing and monitoring thyroid disease.
Clinical Evidence
Bench testing only. Precision evaluated over 20 days using commercial controls and patient pools; results showed comparable performance between new and current PMT. Linearity verified across 0.1–8.0 ng/dL range (R²=1.00). Method comparison against predicate using 45 patient samples showed high correlation (r=0.998, slope=0.97). Limit of Blank (LoB) verified at 0.03 ng/dL and Limit of Detection (LoD) at 0.06 ng/dL.
Technological Characteristics
Floor-model, fully automated, microprocessor-controlled analyzer. Uses LOCI technology (chemiluminescent immunoassay). Reagents: prepackaged Flex cartridges containing synthetic beads and biotinylated mouse monoclonal antibodies. Detection: Photomultiplier tube (enhanced glass single surface tube). Connectivity: LINUX-based system. Temperature control: 37°C for reactions, 2-8°C for reagent storage. Sample volume: 10 μL.
Indications for Use
Indicated for quantitative measurement of Free Thyroxine in human serum and plasma for diagnosis and monitoring of thyroid disease in patients.
Regulatory Classification
Identification
A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.
Predicate Devices
- Dimension® EXL™ with LM system including FT4L assay (k073604)