BIOLIGN ACIF SYSTEMS, BIOLIGN STACC ACIF SYSTEM, BIOLIGN TLIF SYSTEM
K130274 · Elite Surgical Supplies (Pty), Ltd. · MAX · Oct 29, 2013 · Orthopedic
Device Facts
Record ID
K130274
Device Name
BIOLIGN ACIF SYSTEMS, BIOLIGN STACC ACIF SYSTEM, BIOLIGN TLIF SYSTEM
Applicant
Elite Surgical Supplies (Pty), Ltd.
Product Code
MAX · Orthopedic
Decision Date
Oct 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Biolign® Tlif Interbody Fusion Devices are indicated for use with autograft bone graft in skeletally mature patients with degenerative disc disease (DDD) at one level or two contiguous levels of the lumbar spine (L2 - S1). DDD is defined as backpain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six (6) months of non-operative therapy. The Biolign® TLIF Intervertebral fusion devices are used to facilitate fusion in the lumbar spine and are implanted via a postero-lateral approach. The Biolign® intervertebral fusion devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine for example posterior pedicle screw and rod systems. The Biolign® ACIF System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease at one level from C3-C7. Cervical disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six (6) weeks of non-operative treatment. The Biolign® ACIF System is to be used with autograft bone graft and is implanted via an open anterior approach. The Biolign® ACIF System is intended to be used with supplemental fixation systems that have been cleared for in the cervical spine, such as Anterior Cervical plating systems. The Biolign® STACC ACIF System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease at one level from C3-C7. Cervical disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six (6) weeks of non-operative treatment. The Biolign® STACC ACIF System is to be used with autograft bone graft and is implanted via an open anterior approach. The Biolign® STACC ACIF system is intended to be used as a stand-alone system using the bone screws provided for fixation. It may also be used with supplemental fixation system that have been cleared for use in the cervical spine, such as cervical plating systems.
Device Story
Biolign Intervertebral Body Fusion Systems (TLIF, ACIF, STACC) are spinal implants designed to facilitate fusion. TLIF and ACIF cages are manufactured from PEEK Optima LT1 with tantalum markers for radiographic visualization; they feature serrated endplates and central cavities for autograft packing. STACC ACIF is a stand-alone cage including titanium 6Al4V ELI fixation screws. Devices are implanted by surgeons via postero-lateral (TLIF) or anterior (ACIF/STACC) approaches. Cages provide structural support to the disc space during fusion; supplemental fixation (plates or pedicle screws) is typically used, except for the stand-alone STACC configuration. Benefits include stabilization of the spinal segment and promotion of arthrodesis in patients with degenerative disc disease.
Clinical Evidence
No clinical data provided. Substantial equivalence established via bench testing only.
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbar levels (L2-S1) or one cervical level (C3-C7). Patients must have failed 6 weeks (cervical) or 6 months (lumbar) of non-operative therapy. Lumbar patients may have up to Grade I spondylolisthesis. Used with autograft bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
## 510(k) SUMMARY: K130274
[As required by 21CFR807.92]
OCT 2:9 2013
Image /page/0/Picture/4 description: The image shows a series of horizontal lines of varying lengths and thicknesses. The lines are arranged in a somewhat stacked formation, with some lines appearing to overlap or be positioned closely together. The lines are black and are set against a white background.
ELITE SURGICAL SUPPLIES (PTY) LTD
### Contact detail
| | 510(k) Summary - Biolign Intervertebral Cages | |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Prepared | September 11, 2013 | |
| Name of firm | Elite Surgical Supplies (Pty) Ltd.<br>184 Bessemer Road,<br>Pretoria-West Industrial<br>0183<br>South Africa | |
| Postal Address | P.O. Box 26115<br>Arcadia<br>Pretoria<br>0007<br>South Africa | |
| 510(k) Contact | Rudi Verbeek<br>Technical Director and Regulatory Affairs<br>Tel: +27 12 386 0012<br>Facsimile: +27 12 386 2745<br>Email: rudi@elitesurgical.com | |
| USA Agent | Charise de Barros<br>Elite Surgical USA<br>503 Commerce Park Drive, Suite H<br>Marietta<br>Georgia 30060<br>USA<br>Facsimile: 770-590-5152<br>Email: charise@alloutautousa.com | |
Elite Surgical Supplies Biolign® Intervertebral Body Fusion Systems Page 1
{1}------------------------------------------------
## Product detail
| Biolign® Transforaminal Lumbar Intervertebral Fusion (TLIF) device. | |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 888.3080 Intervertebral body fusion device<br>Class II (Special Controls)<br>Orthopaedic and Rehabilitation Devices Panel (87)<br>Product Code MAX (Intervertebral body fusion device with bone graft,<br>Lumbar). |
| Predicates | Spinecraft ORIO-TL - K090887 |
| Device<br>Description | The Biolign TLIF device is a slightly curved cage manufactured<br>from Invibio PEEK Optima LT1 with cavities which can be packed<br>with autograft. The cage has tantalum markers which allow<br>accurate and correct placement of the device. The cage is postero-<br>laterally inserted. The cage incorporates a threaded introducer hole<br>on each end and has serrated inferior and superior endplates. It is<br>available in two footprints (30x13mm, and 36x13mm), two lordotic<br>angles (5° and 10°), and a range of heights ranging from 6 - 12mm<br>with 1mm intervals These sizing options allow selection for<br>individual pathology and anatomical conditions. The device is used<br>with supplemental fixation such as screw and rod systems. The<br>Biolign® TLIF system is provided sterile.. The System is single use<br>only. |
| Indications<br>for<br>use | The Biolign® Tlif Interbody Fusion Devices are indicated for use with<br>autograft bone graft in skeletally mature patients with degenerative<br>disc disease (DDD) at one level or two contiguous levels of the<br>lumbar spine (L2 - S1).<br>DDD is defined as backpain of discogenic origin with degeneration<br>of the disc confirmed by history and radiographic studies. The DDD<br>patients may also have up to Grade I spondylolisthesis at the<br>involved level(s). These patients should be skeletally mature and<br>have six (6) months of non-operative therapy.<br>The Biolign® TLIF Intervertebral fusion devices are used to facilitate<br>fusion in the lumbar spine and are implanted via a postero-lateral<br>approach.<br>The Biolign® intervertebral fusion devices are intended to be used<br>with supplemental fixation systems that have been cleared for use in<br>the lumbosacral spine for example posterior pedicle screw and rod<br>systems. |
| Comparison of<br>the device to<br>the predicate | Elite Surgical Supplies' Biolign® TLIF Spacer is substantially<br>equivalent to the predicates in design, function, performance,<br>material, sterilization and intended use. |
.
.
:
.
:
.
:
Elite Surgical Supplies
Biolign® Intervertebral Body Fusion Systems
Page 2
. .
.
{2}------------------------------------------------
| device(s) | |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>data (Non-<br>Clinical) | Elite Surgical Supplies conducted the following bench testing as<br>recommended by FDA guidances and in accordance with ASTM<br>2077 and ASTM 2267: Static and Dynamic axial compression, Static<br>shear compression, torsion testing and subsidence testing.<br>Results drawn from the testing performed, as well as from an<br>engineering rationale, demonstrate that the Biolign TLIF spacer is<br>substantially equivalent in performance to the predicate device. |
| Performance<br>data (Clinical) | Clinical data and conclusions were not needed for this device. |
| Biolign® Anterior Cervical Interbody Fusion (ACIF) System | |
|-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 888.3080 Intervertebral body fusion device<br>Class II (Special Controls)<br>Orthopaedic and Rehabilitation Devices Panel (87)<br>Product Code ODP (Intervertebral body fusion device with bone graft,<br>Cervical). |
| Predicates | AMT Shell Cervical cage- K080401 |
| Device<br>Description | The Biolign ACIF device is an anatomically correct cage<br>manufactured from Invibio PEEK Optima LT1 with a central cavity<br>which can be packed with autograft. The cage is available in two<br>footprints (15 and 17mm) with a 10° lordotic angle. Cage heights<br>range from 6mm-10mm with 1mm intervals. These sizing options<br>allow selection for individual pathology and anatomical conditions.<br>Tantalum radiographic markers enable clear x-ray control during<br>placement and follow-up. The cage endplates are serrated. The<br>Device is inserted below the anterior surface of the cervical<br>vertebrae. It is used with supplemental cervical plating systems which<br>have prominence from the spine (plate thickness). The Biolign® ACIF<br>system is provided sterile The System is for single use only. |
| Indications for<br>use | The Biolign® ACIF System is indicated for use as an intervertebral<br>body fusion device in skeletally mature patients with degenerative<br>disc disease at one level from C3-C7. Cervical disc disease is defined<br>as discogenic pain with degeneration of the disc confirmed by history<br>and radiographic studies. Patients should have six (6) weeks of non-<br>operative treatment. The Biolign® ACIF System is to be used with<br>autograft bone graft and is implanted via an open anterior approach.<br>The Biolign® ACIF System is intended to be used with supplemental<br>fixation systems that have been cleared for in the cervical spine, such<br>as Anterior Cervical plating systems. |
| Comparison of<br>the device to<br>the predicate | Elite Surgical Supplies' Biolign® ACIF Spacer is substantially<br>equivalent to the predicates in design, function, performance,<br>material, sterilization and intended use. |
Elite Surgical Supplies
Biolign® Intervertebral Body Fusion Systems
Page 3
: .
{3}------------------------------------------------
| device(s) | |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>data (Non-<br>Clinical) | Elite Surgical Supplies conducted the following bench testing as<br>recommended by FDA guidances and in accordance with ASTM<br>2077: Static axial compression, Static shear compression, and static<br>torsion testing.<br>Results drawn from the testing performed, as well as from an<br>engineering rationale, demonstrate that the Biolign ACIF spacer is<br>substantially equivalent in performance to the predicate device. |
| Performance<br>data (Clinical) | Clinical data and conclusions were not needed for this device. |
| Biolign® Stand Alone Cervical Cage (STACC) ACIF System | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 888.3080 Intervertebral body fusion device<br>Class II (Special Controls)<br>Orthopaedic and Rehabilitation Devices Panel (87)<br>Product Code OVE (Intervertebral fusion device with integrated<br>fixation, cervical). |
| Predicates | Surgicraft Ltd Stalif - K072415 |
| Device<br>Description | The Biolign® STACC ACIF Interbody Fusion Device is a stand-<br>alone device and does not rely on additional fixation such as<br>cervical plates. The device is an anatomically shaped ring, with a<br>central cavity which can be packed with autograft. It is<br>manufactured of Invibio Peek Optima LT1. Two Tantalum pins are<br>placed in the device to enable radiographic visualization for<br>placement and follow-up evaluations. The superior, and inferior<br>endplates of the device are serrated. Two fixation screws per cage<br>are used as supplemental fixation. These screws are placed into<br>the cage and purchase into the vertebral endplates. The fixation<br>screws are manufactured of Titanium 6Al4V ELI. The footprint of<br>the cage is available in one size (17 x 14mm), and in heights<br>ranging from 5.5 - 9.5mm with 1mm intervals. The cage is available<br>in a domed (superior endplate is domed, and inferior endplate has<br>3° lordosis angle) and a tapered version (8° lordosis - 4° per<br>endplate) The device is implanted using an anterior approach and<br>after implantation does not stand prominent from the cervical spine.<br>It lies below the anterior surface of the cervical spine. The Biolign®<br>STACC ACIF system is supplied sterile. The System is for single<br>use only. |
| Indications<br>for use | The Biolign® STACC ACIF System is indicated for use as an<br>intervertebral body fusion device in skeletally mature patients with<br>degenerative disc disease at one level from C3-C7. Cervical disc<br>disease is defined as discogenic pain with degeneration of the disc<br>confirmed by history and radiographic studies. Patients should have |
Elite Surgical Supplies
Biolign® Intervertebral Body Fusion Systems
{4}------------------------------------------------
| | six (6) weeks of non-operative treatment. The Biolign® STACC ACIF<br>System is to be used with autograft bone graft and is implanted via<br>an open anterior approach. The Biolign STACC ACIF system is<br>intended to be used as a stand-alone system using the bone screws<br>provided for fixation. It may also be used with supplemental fixation<br>system that have been cleared for use in the cervical spine, such as<br>cervical plating systems. |
|--------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison of<br>the device to<br>the predicate<br>device(s) | Elite Surgical Supplies' Biolign® STACC Spacer is substantially<br>equivalent to the predicates in design, function, performance,<br>material, sterilization and intended use. |
| Performance<br>data (Non-<br>Clinical) | Elite Surgical Supplies conducted the following bench testing as<br>recommended by FDA guidances and in accordance with ASTM<br>2077: Static and dynamic axial compression, Static shear<br>compression, and static and dynamic torsion testing, screw push-out<br>testing, tension offset testing, and impact testing.<br>Results drawn from the testing performed, as well as from an<br>engineering rationale, demonstrate that the Biolign STACC spacer is<br>substantially equivalent in performance to the predicate device. |
| Performance<br>data (Clinical) | Clinical data and conclusions were not needed for this device. |
---
Elite Surgical Supplies
Biolign® Intervertebral Body Fusion Systems
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 29, 2013
Elite Surgical Supplies (Pty) Limited Mr. Rudi Verbeek Technical Director P.O. Box 26115 Arcadia Pretoria 0007 South Africa
Re: K130274
Trade/Device Name: Biolign® ACIF System Biolign® STACC ACIF System Biolign® TLIF Intervertebral Body Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE, MAX, ODP Dated: September 11, 2013 Received: September 25, 2013
Dear Mr. Verbeek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act Control Concession Company Construction of Children ・・・・・・・・・・・・
{6}------------------------------------------------
Page 2 - Mr. Rudi Verbeek
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Erin=FDKéith
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use ·
510(k) Number (if known) K130274
#### Device Name Biolign® ACIF System
#### Indications for Use (Describe)
The Biolign® ACIF System is indicated for use as an intervertebral body fusion device in skeletally mature paients with degenerative disc disease at one level from C3-C7. Cervical discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six (6) weeks of non-operative treatment. The Biolign® ACIF System is to be used with autograft bone graft and is implanted via an open anterior approach. The Biolign® ACIF System is intended to be used with supplemental fixation systems such as Anterior Cervical plating systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
## Anton E. Dmitri Division of Orthopied
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K130274
#### Device Name
#### Biolign® TLIF Intervertebral Body Fusion Device
#### Indications for Use (Describe)
The Biolign® TLIF Intervertebral Body Fusion Devices are indicated for use with autograft bone graft in skeletally mature patients with degenerative disc disease (DDD) at one level or two contiguous levels of the lumbar spine (L2 - S1).
DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radios. The DDd patients may also have up to Grade I spondylolisthesis at the involved level(s). These patients should be skeletally mature and have six {6) months of non-operative therapy.
The Biolign® TLIF Intervertebral Body Fusion Devices are used to facilitate fusion in the lumbar spine and are implanted via a postero-lateral approach.
The Biolign® Intervertebral Body Fusion Devices are intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine, for example posterior pedicle screw and rod systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
[ ] Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Division o
{9}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K130274
#### Device Name Biolign® STACC ACIF System
Diognes STRCC RICH System
#### Indications for Use (Describe)
The Biolign® STACC ACIF System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease at one level from C3-C7. Cervical disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. Patients should have six (6) weeks of non-operative treatment. The Biolign® STACC ACIF System is to be used with autograft bone graft and is implanted via an open anterior approach. The Biolign® STACC ACIF system is intended to be used as a stand-alone screws provided for fixation. It may also be used with supplemental fixation systems that have been cleared for use in the cervical spine, such as cervical plating systems.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
[] Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Anton E. Dmitri Division of Orthopedic
FORM FDA 3881 (9/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.