ESOPHAGUS BOUGIE SET

K130251 · Varian Medical Systems, Inc. · JAQ · Mar 13, 2013 · Radiology

Device Facts

Record IDK130251
Device NameESOPHAGUS BOUGIE SET
ApplicantVarian Medical Systems, Inc.
Product CodeJAQ · Radiology
Decision DateMar 13, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5700
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Esophagus Bougie Set was designed to treat cancer of the esophagus. The Esophagus Bougie Set is an applicator used to facilitate delivery of a radiation source to the esophagus when used in conjunction with a Varian high dose rate afterloader.

Device Story

Esophagus bougie is an HDR radiation applicator; facilitates delivery of radioactive source to esophagus. Used with Varian high dose rate afterloaders. Device consists of bougie tube, catheter, clamping screw, bite protector, threaded ring, fixation strap, and guide wire. Radioactive source is driven from afterloader through transfer guide tubes into applicator positioned in patient. Physician controls source position and dwell time per irradiation treatment prescription. Device is non-electronic; no software. Bougie tube and catheter are disinfected; other components steam sterilized. Fixation strap and guide wire are single-use. Benefits patient by enabling targeted radiation delivery for esophageal cancer treatment.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included design verification of cleaning/disinfection processes per DIN EN ISO 17664 and biocompatibility testing per ISO 10993-1, 10993-5, and 10993-10. Reformulated PVC material was evaluated through design verification to ensure no impact on performance or biocompatibility.

Technological Characteristics

Applicator for HDR radiation therapy. Materials: PVC (reformulated). Components: bougie tube, catheter, clamping screw, bite protector, threaded ring, fixation strap, guide wire. Sterilization: steam sterilization for metal components; high-level disinfection for bougie tube/catheter. No electronics or software. Complies with ISO 10993 and DIN EN ISO 17664.

Indications for Use

Indicated for patients with cancer of the esophagus requiring radiation therapy via high dose rate afterloader.

Regulatory Classification

Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130251 page 1 of 2 # Premarket Notification [510(k)] Summary Esophagus bougie The following information is provided following the format of 21 CFR 807.92. MAR 1 3 2013 | Submitter's Name: | Varian Medical Systems, Inc.<br>3120 Hansen Way C-260<br>Palo Alto, CA 94304 | |----------------------|-----------------------------------------------------------------------------------------------------| | | Contact Name: Peter J. Coronado<br>Phone: 650.424.5731<br>Fax: 650.842.5040<br>Date: December 2012 | | Proprietary Name: | Varian Medical Systems, Inc. Esophagus bougie set | | Classification Name: | Medical charged-particle radiation therapy system<br>21 CFR 892.5700, Class II<br>Product Code: JAQ | | Common/Usual Name: | Varian Medical Systems, Inc. Esophagus bougie set | | Predicate Device: | Varian Esophagus bougie set | The Varian Medical Esophagus Bougie is an HDR applicator designed to Device Description: facilitate delivery of radiation to the Esophagus and has been modified to work with Varian Medical afterloaders. The bougie tube and catheter can be cleaned and disinfected with high level disinfectants and has a maximum insertion time of 24 hours. The clamping screw, bite protector and threaded ring can be steam sterilized. The fixation strap and guide wire are single use. The device does not contain any electronics or software. > A high activity radioactive source is placed within the applicator which has previously been placed for a specified clinical purpose in a patient. The radioactive source (enclosed within the wire/cable) is driven via coupling catheters (Transfer Guide Tubes) from the Afterloader into the applicator within the patient. The length of time and position that the High Dose Rate source spends within the applicator is controlled in accordance with an Irradiation Treatment Prescription. Intended Use Statement The Esophagus Bougie Set was designed to treat cancer of the esophagus. The Esophagus Bougie Set is an applicator used to facilitate delivery of a radiation source to the esophagus when used in conjunction with a Varian high dose rate afterloader. Indications for Use Statement The Esophagus Bougie Set was designed to treat cancer of the esophagus. The Esophagus Bougie Set is an applicator used to facilitate delivery of a radiation source to the esophagus when used in conjunction with a Varian high dose rate afterloader. {1}------------------------------------------------ K130251 Page 2 of 2 # Premarket Notification [510(k)] Summary Esophagus bougie ### Technological Characteristics Significant change to the predicate device is listed below. | Performance | Cleared device | Device with change | |------------------------------------------------|----------------|--------------------| | Sterilization method:<br>Bougie tube, catheter | Sterilized | Disinfected | The clamping screw, bite protector, threaded ring, fixation strap and guide wire sterilization remains the same. ### Discussion of nonclinical testing; The esophagus bougie sterilization efficacy and suitability of the manual cleaning and disinfection process was evaluated through design verification and testing. Non-clinical testing showed the test articles met the pre-defined acceptance criteria, thereby demonstrating the effectiveness of the cleaning and disinfection process plus the cleaning and disinfection agents according to DIN EN ISO 17664 (July 2004). Biocompatibility testing performed in accordance with ISO 10993-1, 10993-5 and 10993-10 show the device is safe for bodily contact. Per the FDA's Guidance document Deciding When to Submit a 510(k) for a Change to an Existing Device, only the formulation of this non-implant PVC material was changed. One of the components used in the composite material is no longer available; however, the composite material remains the same. The reformulated material was evaluated through design verification and testing. Non-clinical testing showed the test articles met the pre-defined acceptance criteria, thereby demonstrating the reformulated material did not impact device performance or biocompatibility. ### Conclusion of nonclinical testing: The change to the esophagus bougie does not raise new questions of safety or effectiveness when compared to the predicate device and, therefore, is substantially equivalent to the predicate esophagus bougie. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings, rendered in a thick, black line. The bird is positioned to the right of a circular emblem. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a sans-serif font. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 13, 2013 Peter J. Coronado Director, Varian Oncology Systems Regulatory Affairs Varian Medical Systems. Inc. 911 Hansen Way, m/s E-110 PALO ALTO CA 94304-1038 Re: K130251 Trade/Device Name: Esophagus bougie Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ Dated: January 30, 2013 Received: February 7, 2013 Dear Mr. Coronado: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2-Mr. Coronado If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, for Janine M. Morris, M.S. Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K130251 Device Name: Esophagus bougie Indications for Use: The Esophagus Bougie Set was designed to treat cancer of the esophagus. The Esophagus Bougie Set is an applicator used to facilitate delivery of a radiation source to the esophagus when used in conjunction with a high dose rate afterloader. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) · Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health 510(k) K130251 Page 1 of
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...