The NeuroStar TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from one prior antidepressant medication at or above the minimal effective dose and duration in the current episode.
Device Story
NeuroStar TMS Therapy System is a computerized, electromechanical device delivering non-invasive, pulsed magnetic fields to induce electrical currents in the cerebral cortex. System components include a mobile console with software-controlled GUI, gantry-supported treatment coil, laser-guided head support system, and SenStar Treatment Link. The SenStar Link provides contact sensing, magnetic field quality control, surface field cancellation, and hygiene barrier. TMS TrakStar (formerly PDMS) is a standalone computer/software system for patient data management and communication between units. Used in inpatient/outpatient clinics and hospitals under physician supervision. Physicians use the system to deliver targeted cortical stimulation; output affects clinical decision-making by providing a non-pharmacological treatment option for MDD patients resistant to initial antidepressant therapy.
Clinical Evidence
No clinical data presented; this submission is a labeling change notification for an existing cleared device.
Technological Characteristics
Computerized electromechanical system; delivers pulsed magnetic fields (~200 µsec duration). Includes mobile console, gantry, treatment coil, laser-guided head positioning, and SenStar contact-sensing link. Software-controlled GUI. Standalone data management system (TMS TrakStar).
Indications for Use
Indicated for adult patients with Major Depressive Disorder (MDD) who have failed to achieve satisfactory improvement from one prior antidepressant medication at or above the minimal effective dose and duration in the current episode.
Regulatory Classification
Identification
A repetitive transcranial magnetic stimulation system is an external device that delivers transcranial repetitive pulsed magnetic fields of sufficient magnitude to induce neural action potentials in the prefrontal cortex to treat the symptoms of major depressive disorder without inducing seizure in patients who have failed at least one antidepressant medication and are currently not on any antidepressant therapy.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's “Class II Special Controls Guidance Document: Repetitive Transcranial Magnetic Stimulation System.” See § 882.1(e) for the availability of this guidance document.
{0}------------------------------------------------
K130233
## NEUR®NETICS
510(k) Summary NeuroStar® TMS Therapy System
APR 3 0 2013
| 510(k) Owner: | Neuronetics, Inc.<br>31 General Warren Blvd<br>Malvern, PA 19355<br>Phone: 610-640-4202<br>Fax: 610-640-4206 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Judy P. Ways, Ph.D.<br>Vice President,<br>Regulatory Affairs and Quality Assurance<br>Neuronetics, Inc.<br>31 General Warren Blvd<br>Malvern, PA 19355<br>Phone: 610-981-4107<br>Fax: 610-640-4206 |
| Date Prepared: | 30 April 2013 |
| Proprietary Name: | NeuroStar® TMS Therapy System |
| Common Name: | Transcranial Magnetic Stimulator |
| Classification Name: | Transcranial Magnetic Stimulator for Treatment of Major<br>Depressive Disorder [21 CFR 882.5805, Product Code OBP] |
| Predicate Device: | NeuroStar TMS Therapy® System [K061053/K083538] |
| Device Description: | The NeuroStar TMS Therapy System is a computerized,<br>electromechanical medical device that produces and delivers non-<br>invasive, magnetic stimulation using brief duration (~200 µsec)<br>rapidly alternating, or pulsed, magnetic fields to induce electrical<br>currents directed at spatially discrete regions of the cerebral<br>cortex. This method of cortical stimulation by application of brief<br>magnetic pulses to the head is known as Transcranial Magnetic<br>Stimulation or TMS.<br>The NeuroStar TMS System is a non-invasive tool for the<br>stimulation of cortical neurons for the treatment of adult patients<br>with Major Depressive Disorder (MDD) who have failed to<br>receive satisfactory improvement from prior antidepressant<br>medication as described under Intended Use. The NeuroStar TMS<br>System is used for patient treatment by prescription only under the<br>supervision of a licensed physician. It can be used in both<br>inpatient and outpatient settings including physician's offices and<br>clinics, and hospitals.<br>The NeuroStar TMS System is an integrated system consisting of<br>a combination of hardware, software, accessories and consumable |
APR 30 2031 General Warren Boulevard / Malven, PA 19355 / 610-640-4202 / 6 10-640-4206 / www.neuronetics.com
·
{1}------------------------------------------------
### NEUR®NETICS
supplies. It includes a Mobile Console which houses the electronics, includes a software controlled graphical user interface, and gantry that supports the Treatment Coil. The ferromagnetic Treatment Coil delivers the TMS TherapyTM. The Head Support System provides accurate positioning of the Treatment Coil using a laser-guided alignment system. A singleuse device, the SenStar® Treatment Link, which is applied to the Treatment Coil, provides contact sensing to monitor contact of the treatment coil with the patient's head throughout a treatment session, quality control by monitoring the magnetic field level prior to patient treatment, surface field cancellation to reduce stimulation of the scalp, and acts as a hygiene barrier from patient to patient. The TMS TrakStar (previously known as Practice Data Management System - PDMS) consists of a stand-alone computer and data management software that facilitates recording and retrieval of patient and treatment information and communication of data among multiple NeuroStar TMS Systems.
NeuroStar TMS Therapy is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from one prior antidepressant medication at or above the minimal effective dose and duration in the current episode.
The subject device, NeuroStar TMS System, has the following similarities to the predicate TMS device (NeuroStar TMS Therapy System, K061053/K083538):
- Indications for use ●
- . Principles of operation
- Design for delivery of Transcranial Magnetic . Stimulation (TMS)
- . Materials
- . Shelf life, and packaged using the same materials and processes.
The proposed change, which is the subject of this 510(k), is a labeling change to the prescribing statement to comply with the FDA's Class II Special Controls Guidance for Repetitive Transcranial Magnetic Stimulation (rTMS) Devices (26 July 2011). No other changes are made to the device or labeling. Therefore, the NeuroStar TMS Therapy System with the change proposed in the 510(k) is substantially equivalent to the predicate · device.
"The NeuroStar @ and NeuroStar TMS Therapy® are registered trademarks of Neuronetics, Inc. TMS Therapy™ is a trademark of Neuronetics, Inc.
1-2 31 General Warren Boulevard / Malvem, PA 19355 / t 610-640-4202 / f 610-640-4206 / www.neuronetics.com
Intended Use
Technological Characteristics and Substantial Equivalence
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the text 'Public Health Service'. The text is in a simple, sans-serif font and is horizontally aligned. The text is likely part of a document or sign, indicating an association with public health services.
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is presented in black and white.
April 30,2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Neuronetics. Inc. c/o Judy P. Ways, Ph.D. Vice President, Regulatory Affairs and Quality Assurance 31 General Warren Boulevard Malvern, PA 19355
Re: K130233 NeuroStar TMS Therapy System Regulation Number: 21 CFR 882.5805 Regulation Name: Repetitive Transcranial Magnetic Stimulator for Treatment of Major Depressive Disorder Regulatory Class: Class II Product Code: OBP Dated: January 29, 2013 Received: January 30, 2013
Dear Dr. Ways:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Judy P. Ways, Ph.D.
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address:
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to:
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address:
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
#### Joyce M. Whang -S
for
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
510(k) Number: K130233
Device Name: NeuroStar TMS Therapy System
Indications For Use:
The NeuroStar TMS Therapy System is indicated for the treatment of Major Depressive Disorder in adult patients who have failed to achieve satisfactory improvement from one prior antidepressant medication at or above the minimal effective dose and duration in the current episode.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M. Whanq -S
(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number K130233
Page 1 of _1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.