BPA-002 Electronic Blood Pressure Monitor for Upper Arm is intended to measure systolic and diastolic blood pressures as well as the determination of heart rate by using the oscillometric method. All values can be read out in one LCD panel. It is suitable for adult arm circumference and can be used by medical departments or for home use. However, it shouldn't be used by people who have severe arrhythmia.
Device Story
BPA-002 Electronic Blood Pressure Monitor for Upper Arm; uses inflatable cuff wrapped around patient upper arm; internal pump inflates cuff; silicon integrated pressure sensor measures pressure via oscillometric method; mechanical valve controls deflation rate; device displays systolic/diastolic blood pressure and heart rate on LCD panel; includes irregular heartbeat detection; automatic shut-off after 60 seconds; maximum pressure safety limit 280mmHg; intended for adult use in clinical or home settings; provides objective measurements to assist healthcare providers or patients in monitoring cardiovascular status.
Clinical Evidence
Bench testing only. Compliance with SP10:2002/(R) 2008 and ANSI/AAMI SP10:2002/A1:2003 for electronic sphygmomanometers. Electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5/10) testing performed.
Indicated for adults to measure systolic/diastolic blood pressure and heart rate via oscillometric method. Suitable for clinical or home use. Contraindicated for patients with severe arrhythmia.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Fully Automatic Digital Blood Pressure Monitor, Model BP-1307 (K120554)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the letters J, X, B, and D. The letters are in a bold, sans-serif font. The letters J, X, and B are solid black, while the letter D is black with white dots inside. There is a small, abstract design to the right of the letter D.
K130232 Jinxinbao Electronic Co., Ltd
## Section 5 510(k) Summary
## (As required by 21 CFR 807.92(a))
### 5.1 Submitter Information
SEP 17 2013
- · Company: Jinxinbao Electronic Co.,Ltd
· Address: No.38, Huanzhenxi Road, Dagang Town, Nansha, Guangzhou, China
· Phone: 086-20-34938449
- · Fax: 086-20-22865301
- · Email: xhoucc@jxb-htb.net
· Contact: Jiacheng Guo, General Manager
· Date: November 15, 2012.
#### 5.2 Device Information
· Trade/Proprietary Name: Electronic Blood Pressure Monitor for Upper Arm, Model
BPA-002
· Common Name: Blood Pressure Monitor
· Classification: Device Class: 2
Review Panel: Cardiovascular
Name: Non-invasive Blood Pressure Measurement System
Regulation Number: 21 CFR 870.1130
Product Code: DXN
· Predicate Device: Fully Automatic Digital Blood Pressure Monitor, Model BP-1307.
#### K120554
· Device Description:
BPA-002 Electronic Blood Pressure Monitor for Upper Arm uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by
an internal pump in the device. The systolic and diastolic blood pressures are
determined by oscillometric method and silicon integrate pressure sensor technology.
The deflation rate is controlled by a preset mechanical valve at a constant rate. The Page 1 of 5
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters J, X, and B in a bold, sans-serif font. The letters are arranged horizontally, with J on the left, X in the middle, and B on the right. The letter B has a design element on the top right that resembles a chain or a series of connected circles. The logo is in black and white.
#### K130232 Jinxinbao Electronic Co., Ltd
pressure of the cuff is completely released automatically at the end of measurement. At the same time, the measurements are displayed on the LCD display for 60 seconds.
There is a maximum pressure safety setting at 280mmHg. The device will not inflate the cuff higher than 280mmHg. It will display an irregular heartbeat symbol if an irregular heartbeat was detected during the measurement process. In addition, it has LCD backlight. After 60 seconds without operation, the blood pressure monitors turn off automatically
· Intended Use:
BPA-002 Electronic Blood Pressure Monitor for Upper Arm is intended to measure systolic and diastolic blood pressures as well as the determination of heart rate by using the oscillometric method. All values can be read out in one LCD panel.
It is suitable for adult arm circumference and can be used by medical departments or for home use. However, it shouldn't be used by people who have severe arrhythmia.
| Item | Subject Device | Predicate Device | |
|----------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------------------------|
| Device Name | Electronic Blood Pressure<br>Monitor for Upper Arm<br>Model BPA-002 | Fully Automatic Digital<br>Blood Pressure Monitor<br>Model BP-1307<br>(K120554) | |
| | Method of Measurement | Oscillometric | Oscillometric |
| | Inflation Method | Automatic Internal Pump | Automatic Internal Pump |
| | Deflation Method | Preset Mechanical Valve at<br>a Constant Rate | Preset Mechanical Valve at<br>a Constant Rate |
| Anatomical Site | | Upper Arm | Upper Arm |
| Range of Measurement | Pressure:<br>0mmHg~280mmHg<br><br>Pulse:<br>30~160 Beats/minute | Pressure:<br>0mmHg~300mmHg<br><br>Pulse:<br>30~180 Beats/minute | |
## 5.3 Comparison of Required Technology Characteristics
{2}------------------------------------------------
| Portugues of | Jinxinbao Electronic Co., Ltd | K130232 |
|------------------------|-------------------------------|--------------------------|
| Accuracy | Pressure: ±3mmHg | Pressure: ±3mmHg |
| | Pulse: ≤5% | Pulse: ±5% |
| Operating Conditions | Temperature: 5℃~40℃ | Temperature: 5°C ~40°C |
| | RH: 15%-85% | RH: <85% |
| Storage Conditions | Temperature: - 10 ℃~60 ℃ | Temperature: - 10°C~55°C |
| | RH: 15%~85% | RH: <95% |
| Power Supply | 4 Batteries Size AA | 4 Batteries Size AA |
| Display | Liquid Crystal Digital | Liquid Crystal Digital |
| | Display | Display |
| Cuff Size | Arm Circumference: | Arm Circumference: |
| | 22~30cm | 22-36cm |
| Unit Weight | About 430g ( Without | About 395g ( Without |
| | Batteries) | Batteries) |
| Sets of Memory | 99 Memories in three | 120 Memories in two |
| | groups | groups |
| Irregular Heart Beat | Yes | Yes |
| Detection | | |
| Low Battery Indicator | Yes | Yes |
| Last 3 Results Average | Yes . | Yes |
#### Brief Summary:
The BPA-002 Electronic Blood Pressure Monitor for Upper Arm incorporates the same fundamental technology characteristics and design with the predicate device, such as measurement method, inflation and deflation method, anatomical site, power supply, display, irregular heart beat detection, low battery indicator and last 3 results average. Though they differ in the concrete parameters of measurement range, accuracy, cuff size and sets of memory, such differences will not influence the intended use of the subject device and will not lead to safety and effectiveness concerns.
{3}------------------------------------------------
#### 5.4 Discussion of Tests Performed
#### · Clinical Tests:
Subject device BPA-002 is compliant to the SP10:2002/(R) 2008&ANSI/AAMI SP10:2002/A1:2003/( Manual, electronic or automated sphygmomanometers. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
#### · Non-Clinical Tests
The subject device was tested to evaluate its safety and effectiveness according to the following standards:
a. Electrical Safety Test according to IEC 60601-1, Medical electrical equipment-Part 1: General requirements for basic safety and essential performance.
b. Electromagnetic Compatibility Test according to IEC 60601-1-2:2007, Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests.
c. Biocompatibility Test according to ISO 10993-5: 2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cytotoxicity and ISO 10993-10: 2010, Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization.
d. Performance Effectiveness Test according to SP10:2002/(R) 2008&ANSI/AAMI SP 10: 2002/A 1: 2003/ (Manual, electronic or automated sphygmomanometers.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters J, X, B, and D. The letters are in a bold, sans-serif font. The letters are black and the background is white. The letters are arranged horizontally, with the J on the left and the D on the right.
5.5 Conclusion:
First, the subject device BPA-002 Electronic Blood Pressure Monitor for Upper Arm enjoys the same intended use and similar technological characteristics with the predicate device. Besides, the performance safety and effectiveness of the subject device has been verified in accordance with the above FDA recognized standards, thus being considered to be as safe and effective as the predicate device. In a word, it is reasonable for us to conclude that the subject device is substantially equivalent to the predicate device according to the above analysis.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is a stylized image of a human figure.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 17, 2013
Jinxinbao Electronic Co., Ltd c/o Ms. Helen Nan, General Manager Wenzhou Cytech Information Service Co., Ltd. RM-404, Bldg. 7, Jinhuichang Homeland, Liuhongqiao Rd. Wenzhou City, 325000, Zhejiang Province. CHINA
K130232 Re:
> Electronic Blood Pressure Monitor for Upper Arm, Model BPA-002 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: Undated Received: August 7, 2013
Dear Ms. Nan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 - Ms. Helen Nan
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours.
# Owen P. Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the letters J, X, and B. The letters are in a bold, sans-serif font and are arranged horizontally. The letters are filled with a textured pattern, giving them a slightly rough appearance. There is a small symbol to the right of the letter B.
Jinxinbao Electronic Co., Ltd
## Section 4 Indications for Use Statement
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
BPA-002 Electronic Blood Pressure Monitor for Upper Arm_ Device Name:
### Indications for Use:
BPA-002 Electronic Blood Pressure Monitor for Upper Arm is intended to measure systolic and diastolic blood pressures as well as the determination of heart rate by using the oscillometric method. All values can be read out in one LCD panel.
It is suitable for adult arm circumference and can be used by medical departments or for home use. However, it shouldn't be used by people who have severe arrhythmia.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
4
Clwez
Digitally signed by Owen P. Faris -S Date: 2013.09.19 14:21:12 -04'00'
Page _1_ of _ 1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.