← Product Code [MHX](/productcode/MHX) · K130208

# BERNOULLI ENTERPRISE SOFTWARE (K130208)

_Cardiopulmonary Corp. · MHX · Dec 20, 2013 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K130208

## Device Facts

- **Applicant:** Cardiopulmonary Corp.
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Dec 20, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

The Bernoulli Management System (MS) Software is a software application intended to be used on a central monitoring station on patients using supported devices in a hospital or hospital-type environment. It is used to provide a secondary display of multi parameter monitor data (ECG, respiratory rate, pulse oximetry, ETCO2, blood pressure, cardiac output, temperature and associated derived outputs), lung ventilator, respiratory gas monitor, IV infusion.(including PCA) pump, feeding pump and related ancillary devices, and to provide remote monitoring and alarm surveillance. This application is also intended to provide primary alarm surveillance for devices where no alarm notification capability is enabled or available in low acuity MedSurg, Cardiac, General ward care areas to monitor patients who are currently not monitored. The Bernoulli MS is intended to supplement and not replace any part of the current device monitoring procedures. The Bernoulli Management System (MS) Software is to be used under license by or on the order of a physician or licensed healthcare professional in the course of his/her professional practice.

## Device Story

Bernoulli Enterprise Software acts as a central monitoring station for hospital-based patient data. It ingests real-time data from bedside devices including multi-parameter monitors, ventilators, respiratory gas monitors, and infusion/feeding pumps. The software aggregates these inputs to provide a secondary display of physiological parameters and device status; it also performs remote alarm surveillance. In low-acuity settings, it serves as a primary alarm surveillance system for otherwise unmonitored patients. Operated by clinicians in hospital environments, the system supplements existing monitoring procedures. By centralizing data and alarms, it enables healthcare providers to observe patient status remotely, facilitating timely clinical intervention and improving patient safety through enhanced alarm awareness.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and software verification/validation.

## Technological Characteristics

Software-based central monitoring station. Operates as a networked system to aggregate data from bedside medical devices. Connectivity via standard hospital network protocols. No physical hardware components; software-only device. Designed for integration with existing hospital infrastructure.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- Bernoulli Management System ([K112655](/device/K112655.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

December 20, 2013

Cardiopulmonary Corp. Urey Kang 200 Cascade Blvd. Milford, CT 06460 US

Re: K130208

Trade/Device Name: Bernoulli Enterprise Software Regulation Number: 21 CFR 870.2300 Regulation Name: Central Monitoring Station Software Regulatory Class: Class II Product Code: PFY, MHX, MSX Dated: November 20, 2013 Received: November 21, 2013

Dear Urey Kang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

Page 2 - Mr. Urey Kang

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Owen P. Faris -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

## 510(k) Number (if known): K130208

## Device Name: Bernoulli® Enterprise Software

#### Indications For Use:

The Bernoulli Management System (MS) Software is a software application intended to be used on a central monitoring station on patients using supported devices in a hospital or hospital-type environment. It is used to provide a secondary display of multi parameter monitor data (ECG, respiratory rate, pulse oximetry, ETCO2, blood pressure, cardiac output, temperature and associated derived outputs), lung ventilator, respiratory gas monitor, IV infusion.(including PCA) pump, feeding pump and related ancillary devices, and to provide remote monitoring and alarm surveillance. This application is also intended to provide primary alarm surveillance for devices where no alarm notification capability is enabled or available in low acuity MedSurg, Cardiac, General ward care areas to monitor patients who are currently not monitored. The Bernoulli MS is intended to supplement and not replace any part of the current device monitoring procedures.

The Bernoulli Management System (MS) Software is to be used under license by or on the order of a physician or licensed healthcare professional in the course of his/her professional practice.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Digitally
Owen F
Date: 20
14:31:5

Digitally signed by Owen P. Faris -5 Date: 2013.12.20 14:31:50 -05'00'

Page 1 of 1

510(k) Submission Page D-18

---

**Source:** [https://fda-staging.innolitics.com/device/K130208](https://fda-staging.innolitics.com/device/K130208)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
